The Expect™ Pulmonary Endobronchial Ultrasound Transbronchial Aspiration Needle is designed to be used with endobronchial ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) of the submucosal and extramural lesions of the tracheobronchial tree. Do not use this instrument for any purpose other than its intended use.
Device Story
Device is an endobronchial ultrasound (EBUS) guided transbronchial aspiration needle; used for fine needle aspiration (FNA) of submucosal and extramural tracheobronchial lesions. System includes needle, adaptor, syringe, and stopcock. Needle passes through EBUS endoscope accessory channel; handle at proximal end actuates needle; sheath and needle lengths are adjustable. Stylet protects sheath during passage and aids sample expulsion. Adaptor locks onto bronchoscope biopsy port; syringe/stopcock provide vacuum suction for aspiration. Used in clinical settings by physicians; output is a tissue sample for diagnostic evaluation. Benefits include minimally invasive tissue acquisition for lesion analysis.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included device flexibility, passability, durability, actuation forces, needle sharpness, tensile strength, and suction capability. Biocompatibility testing performed per ISO 10993-1:2009.
Technological Characteristics
Single-use, sterile (EO) aspiration needle. Features adjustable sheath and needle lengths with locking mechanisms. Includes stylet, adaptor for scope attachment, syringe, and stopcock. Compatible with 2.0mm minimum working channel endoscopes. Materials evaluated per ISO 10993-1. Sterilization validated per ISO 11135-1:2007; residuals conform to ISO 10993-7:2008.
Indications for Use
Indicated for patients requiring ultrasound-guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree using endobronchial ultrasound endoscopes.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
Predicate Devices
Single Use Aspiration Needle NA-201SX-4022 (K050503)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Inside the circle is a symbol that resembles three stylized human profiles facing to the right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 10, 2015
Boston Scientific Corporation Yingying Gao Senior RA Specialist 100 Boston Scientific Way Marlborough, MA 01752
Re: K151315
Trade/Device Name: Expect™ Pulmonary Endobronchial Ultrasound Transbronchial Aspiration Needle Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible or Rigid) and Accessories Regulatory Class: Class II Product Code: EOO Dated: October 1, 2015 Received: October 2, 2015
Dear Ms. Gao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control. and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151315
#### Device Name
Expect™ Pulmonary Endobronchial Ultrasound Transbronchial Aspiration Needle
#### Indications for Use (Describe)
The Expect™ Pulmonary Endobronchial Ultrasound Transbronchial Aspiration Needle is designed to be used with endobronchial ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) of the submucosal and extramural lesions of the tracheobronchial tree. Do no use this instrument for any purpose other than its intended use.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| <div> <span> <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
#### 1. Submitter:
Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01752
Contact: Yingying Gao Sr. Regulatory Affairs Specialist Telephone: 508-683-4356 Fax: 508-683-5939
Secondary Contact: Ashley Santos Regulatory Affairs Manager Telephone: 508-683-4359 Fax: 508-683-5939
Date Prepared: October 29, 2015
# 2. Device:
| Trade Name: | Expect TM Pulmonary Endobronchial Ultrasound<br>Transbronchial Aspiration Needle | |
|----------------------|----------------------------------------------------------------------------------------------|--|
| Device Common Name: | Endobronchial Ultrasound Transbronchial Aspiration<br>Needle/ EBUS-TBNA needle | |
| Classification Name: | Bronchoscope (flexible or rigid) and accessories | |
| Regulation Number: | 874.4680 | |
| Product Code: | EOQ | |
| Classification: | Class II | |
# 3. Predicate Device:
## Predicate device:
| Trade Name: | Single Use Aspiration Needle NA-201SX-4022 |
|----------------------|--------------------------------------------------------------------------------|
| Device Common Name: | Endobronchial Ultrasound Transbronchial Aspiration<br>Needle/ EBUS-TBNA needle |
| Manufacturer: | Olympus Medical Systems Corporation |
| Clearance Number: | K050503 |
| Classification Name: | Gastroenterology-urology biopsy instrument |
{4}------------------------------------------------
| Regulation Number: | 876.1075 |
|----------------------|------------------------------------------------------------------|
| Product Code: | FCG |
| Classification: | Class II |
| And | |
| Reference device: | |
| Trade Name: | Expect™ Slimline (SL) Endoscopic Ultrasound<br>Aspiration needle |
| Device Common Name: | Endoscopic Ultrasound Aspiration needle /<br>EUS-FNA needle |
| Manufacturer: | Boston Scientific Corporation |
| Clearance Number: | K133312 |
| Classification Name: | Endoscope and accessories |
| Regulation Number: | 876.1500 |
| Product Code: | ODG and FCG |
| Classification: | Class II |
### 4. Device Description:
Device Name: Expect™ Pulmonary Endobronchial Ultrasound Transbronchial Aspiration Needle
The ExpectTM Pulmonary device is comprised of the following:
- Expect™ Pulmonary needle ●
- Expect™ Pulmonary adaptor ●
- Syringe .
- Stopcock .
The Expect™ Pulmonary device is an Endobronchial Ultrasound guided Transbronchial Aspiration Needle used for fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree. The device consists of a sheath covered needle which extends into the accessory channel of an endobronchial ultrasound (EBUS) endoscope and is locked into place. A handle on the proximal end of the device is used to actuate the needle in order gather samples. Both the sheath and needle length are adjustable while in the scope. A stylet is in place in order to provide protection to the inside of the sheath during device passage through the scope. The stylet may also be used to expel the sample after the procedure.
Expect™ Pulmonary adaptor is an accessory to be attached and locked onto the biopsy port of the bronchoscope. It allows Expect™ Pulmonary needle to pass through it and to be secured in place with a luer connection.
{5}------------------------------------------------
Syringe and stopcock are accessories to provide and control the vacuum suction to aspirate the sample. They also can be used to expel the samples after the procedure.
### 5. Intended use and Indications for Use:
### Intended Use/Indications for Use:
The Expect™ Pulmonary Endobronchial Ultrasound Transbronchial Aspiration Needle is designed to be used with endobronchial ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) of the submucosal and extramural lesions of the tracheobronchial tree. Do not use this instrument for any purpose other than its intended use.
# 6. Technological Characteristics:
The proposed Expect™ Pulmonary device is designed to acquire sample in the Airway by coupling with an ultrasound bronchoscope. It has the following technological characteristics:
- Adjustable sheath length and sheath locking mechanism ●
- Adjustable needle length and needle locking mechanism ●
- Needle sharpness ●
- . Needle echogenicity in ultrasound image
- Stylet with kink resistance
- Smooth actuation
- Aspiration capability ●
- Passage of device through the scope to the target position ●
- . Secure scope attachment
### Comparison to Predicate:
The proposed Expect Pulmonary device is substantially equivalent to the current marketed predicate, the Olympus Single Use Aspiration Needle NA-201SX-4022 (K050503). The Expect Pulmonary device has similar technological characteristics as the predicate, that is, they operate in the same manner to obtain a tissue biopsy using an ultrasound endoscope. The indication for use of the proposed Expect Pulmonary device is nearly identical to the predicate device.
The comparison of the proposed Expect Pulmonary device and the Olympus Vizishot device is provided in the following Table 6-1.
{6}------------------------------------------------
| Device Characteristics | Predicate Olympus Vizishot device<br>K050503 | Proposed Device BSC Expect™<br>Pulmonary Device |
|-------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | This instrument has been designed to<br>be used with ultrasonic endoscopes<br>for ultrasonically guided fine needle<br>aspiration (FNA) of submucosal and<br>extramural lesions of the<br>tracheobronchial tree and the<br>gastrointestinal tract.<br>Do not use this instrument for any<br>purpose other than its intended use. | The Expect™ Pulmonary Endobronchial<br>Ultrasound Transbronchial Aspiration<br>Needle is designed to be used with<br>endobronchial ultrasound endoscopes for<br>ultrasound guided fine needle aspiration<br>(FNA) of the submucosal and extramural<br>lesions of the tracheobronchial tree.<br>Do not use this instrument for any<br>purpose other than its intended use. |
| Single-Use | Single Use | Identical |
| Sterile | EO | Identical |
| Mechanism of Action<br>for Needle<br>Advancement | Manual | Identical |
| Mechanism for Tissue<br>Sampling | Aspiration | Identical |
| Number of Device<br>Passes during a<br>procedure | Multiple Passes | Identical |
| Needle Gauge Size | 21ga & 22ga | 22ga & 25ga |
| Ability to Visualize<br>with Endoscopic<br>Ultrasound | Echogenic signature on distal end of<br>device | Identical |
| Sheath Length | Adjustable | Identical. |
| Minimum Working<br>Channel<br>Compatibility | 2.0mm (for both 22ga and 21ga) | Identical. 2.0mm (for both 22ga and<br>25ga) |
| Device to scope<br>attachment | Adaptor biopsy valves Attachment for<br>Olympus ultrasound bronchoscopes. | Adaptor Attachment compatible with<br>Olympus and Fujinon ultrasound<br>bronchoscopes. |
Table 6-1 . Comparison of Key Characteristics
{7}------------------------------------------------
The primary difference between the proposed Expect™ Pulmonary device and its predicate devices is the gauge size, patient contact materials and adjustable needle length. With the fully completed biocompatibility and performance bench testing, BSC considers these differences do not raise questions about the safety or effectiveness of the device. Based on the comparison above BSC considers the proposed Expect™ Pulmonary device substantially equivalent to the predicate devices.
The currently cleared Boston Scientific device Expect™ Slimline (SL) Endoscopic Ultrasound Aspiration needle (K133312) is used as a reference device based on its design similarity to the Expect™ Pulmonary device.
#### 7. Performance Data:
#### Biocompatibility Testing Summary:
The proposed Boston Scientific Expect™ Pulmonary devices were evaluated biocompatibility in accordance with ISO 10993-1:2009 Evaluation and Testing. The following tests were performed with acceptable results on the patient contacting portions of the Expect™ Pulmonary device: Cytotoxicity, Sensitization, Irritation, and Systemic Toxicity.
#### Sterilization Testing summary:
The proposed Boston Scientific Expect™ Pulmonary devices meet the requirements of ISO 11135-1:2007 "Sterilization of health care products - Ethylene oxide -- Part 1: Medical devices requirements for development, validation and routine control of a sterilization process for medical devices". This product's Ethylene oxide (EO) sterilization cycle is validated to achieve a minimum sterility assurance level (SAL) of 10 °. And the product Ethylene oxide residual levels conform to ISO 10993-7:2008 (R: 2012) "Biological Evaluation of Medical Devices- Part 7: Ethylene Oxide Sterilization Residuals".
#### Performance Testing Summary:
Non-clinical performance testing was conducted on the Expect™ Pulmonary device which demonstrates the device met the required specifications. The proposed Boston Scientific Expect™ Pulmonary device and packaging were successfully verified after nominal and accelerated age tests, showing the Expect™ Pulmonary device met its shelf life requirements.
The following bench tests were performed:
- 1. Device Flexibility
- 2. Device Passability
- 3. Device Durability (Robustness)
- 4. Needle and Sheath Adjustment Locking Force
- 5. Handle actuation force (Needle extension)
- 6. Needle Sharpness
- 7. Stylet Removal Force
- 8. Handle Home Position
{8}------------------------------------------------
- 9. Needle Extension Length
- 10. Adjustable Working Length (Sheath Extension length)
- 11. Needle to Luer Tensile Strength
- 12. Sheath to Sheath Hub (Actuation Guide)
- 13. Tensile strength; Adaptor to scope tensile
- 14. Device Luer to Adaptor Luer to Scope Tensile
- 15. Adaptor Lock & Unlock Force
- 16. Adaptor Suction
- 17. Needle Extension Length Marking
- 18. Sheath Length Adjustment Markings
- 19. Handle Rotation
- 20. Smooth Actuation
- 21. Packaging
#### 8. Conclusion:
Boston Scientific has demonstrated that the proposed Expect™ Pulmonary Endobronchial Ultrasound Transbronchial Aspiration Needle is substantially equivalent to the currently marketed Olympus Single Use Aspiration Needle NA-201SX-4022 (K050503).
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.