K013111 · Broncus Technologies, Inc. · EOQ · Oct 1, 2001 · Ear, Nose, Throat
Device Facts
Record ID
K013111
Device Name
EXHALE PROBE
Applicant
Broncus Technologies, Inc.
Product Code
EOQ · Ear, Nose, Throat
Decision Date
Oct 1, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4680
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Exhale Probe is intended for blood-flow detection and electrosurgery procedures (i.e. coagulation/cauterization, hemostasis, etc.) in the upper airways and tracheobronchial tree through a bronchoscope.
Device Story
The Exhale Probe is a multi-function catheter designed for use with a standard bronchoscope (2 mm working channel). It integrates an ultrasonic transducer for Doppler blood-flow detection and a coagulation electrode for radiofrequency (RF) electrosurgery. The probe connects to an external Exhale Doppler Processing Unit and a standard RF generator. During bronchoscopy, the clinician inserts the probe to identify pulmonary vessel blood-flow via audio output and perform targeted coagulation or hemostasis in the tracheobronchial tree. The device assists in managing airway procedures by providing real-time vascular feedback and therapeutic energy delivery, potentially reducing bleeding risks during interventions.
Clinical Evidence
Bench testing only. The device was evaluated for compliance with IEC 60601-1 safety standards and FDA 1997 Diagnostic Ultrasound Guidance for Track 1 devices. Biocompatibility of materials was confirmed.
Technological Characteristics
Multi-function catheter featuring a coagulation electrode and an ultrasonic transducer (Model BT 1440). Operates as a Track 1 diagnostic ultrasound system. Connects to external Doppler processing unit and standard RF electrosurgical generator. Designed for insertion through 2 mm bronchoscope working channel. Complies with IEC 60601-1.
Indications for Use
Indicated for blood-flow detection and electrosurgical procedures (coagulation, cauterization, hemostasis) in the upper airways and tracheobronchial tree for patients undergoing bronchoscopy.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the words "Kol 3111" in a bold, handwritten-like font. Below this, the words "Exhale™Probe" are printed in a clear, sans-serif font. The "TM" is a superscript, indicating a trademark.
# OCT - 1 2001
# APPENDIX A. 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
## Name\Address:
Broncus Technologies, Inc. 1400 N. Shoreline Boulevard, Building A, Suite 8 Mountain View, CA 94043
#### Telephone Number:
(650) 428-1600
### Fax Number:
(650) 428-1542
## Contact Person:
Timothy R. Williams Director, Regulatory and Clinical Affairs
Date Summary was Prepared:
June 29, 2001
Trade Name:
Exhale Probe
Classification Number and Name:
Class II, 21 CFR 892.1560, Ultrasound Pulsed Echo Imaging System
Class II, 21 CFR 892.1570, Diagnostic Ultrasound Transducer
Class II 21 CFR 874.4680, Bronchoscope & Accessories
## Device Description:
The Exhale Probe is a multi-function catheter that provides: (1) Doppler audio output in the presence of pulmonary vessel blood-flow; and (2) radiofrequency (RF) energy to a target site within the upper airway or tracheobronchial tree. The probe is inserted into through the 2 mm working channel of a standard bronchoscope and connects to the Exhale Doppler Processing Unit and a standard, commercially available, electrocautery radiofrequency generator.
{1}------------------------------------------------
## Broncus Technologies. Inc.
6013111 Exhale™Probe
#### Substantial Equivalence:
The Exhale Probe is substantially equivalent to the Exhale Doppler Probe (K010649) and the Exhale RF Probe (K011267). with respect to intended use, method of introduction, method of operation, safety, and materials.
#### Safety:
The Exhale Probe consisting of a multi-function catheter contains a coagulation electrode and an ultrasonic transducer and is designed to comply with International Standard IEC 60601-1. The Exhale Probe will meet the requirements of the FDA's 510(k) Diagnostic Ultrasound Guidance for 1997 for a Track 1 device.
#### Non-Clinical Test Results:
#### Biocompatibility
The materials used in the Exhale RF Probe have proven biocompatibility.
# Summary of Substantial Equivalence:
Based on the intended use and product performance information provided in this notification, the subject device has been shown to be substantially equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized, curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT = 1 2001
Broncus Technologies, Inc. C/O Mark Job TÜV Product Services 1775 Old Highway 8 New Brighton, MN 55412-1891
Re: K013111
Trade Name: Exhale Probe Regulation Number: 21 CFR 874.4680; 21 CFR 892.1570; 21 CFR 892.1560 Regulation Name: Bronchoscope and Accessory; Diagnostic Ultrasound Transducer; Ultrasound Pulsed Echo Imaging System Regulatory Class: Class II Product Code: EOQ; ITX; IYO Dated: September 17, 2001 Received: September 18, 2001
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amerodments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the Model BT 1440 transducer intended for use with the Exhale Probe Catheter, as described in your premarket notification.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
Page 2 - Mr. Mark Job
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI I rat 0077; mooning (21 CFR Part 820); and if applicable, the electronic form in the qualis (Sections (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping This device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers In Appendix O, (chelosod) of the Semies tic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and rne special report of the FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nontication. THE PDF Imaning of tastion for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801, please If you desire specific at no issess at (301) 594-4591. Additionally, for questions on the comact the Office of Compilance at (301) 29 + + + > > = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = promotion and advertising or gulation entitled, "Misbranding by reference to premarket 4059. Also, please note the regations on ther general information on your responsibilities under the notification (21 CFR Par 80777). Only Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
{4}------------------------------------------------
Page 3 - Mr. Mark Job
If you have any questions regarding the content of this letter, please contact Karen Baker MSN, RN at (301) 594-2080.
Sincerely yours,
David M. Whipple
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{5}------------------------------------------------
| Broncus Technologies, Inc. | | | |
|----------------------------|--|--|--|
|----------------------------|--|--|--|
| 510(k) Number (if known): | K 013111 |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | The Exhale Probe |
| Indications for Use: | The <i>Exhale Probe</i> is intended for blood-flow detection and electrosurgery procedures (i.e. coagulation/cauterization, hemostasis, etc.) in the upper airways and tracheobronchial tree through a bronchoscope. |
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use_
OR
over-the-counter Use__________________________________________________________________________________________________________________________________________________________
Sonola
(Division Sign-Off) Division of Ophthalmic Devices nor 3111 510(k) Number_
CONFIDENTIAL
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.