K170285 · National Advanced Endoscopy Devices, Inc. · EOB · Aug 1, 2017 · Ear, Nose, Throat
Device Facts
Record ID
K170285
Device Name
AED Sinuscope
Applicant
National Advanced Endoscopy Devices, Inc.
Product Code
EOB · Ear, Nose, Throat
Decision Date
Aug 1, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4760
Device Class
Class 2
Indications for Use
The AED Sinuscope is intended for use in otolaryngology and Head and Neck procedures, including rhinology, and endoscopic plastic and reconstructive surgery.
Device Story
AED Sinuscope is a reusable, rigid rod-lens endoscope used for visualization of nasal passages and sinuses. Device consists of stainless steel shaft, glass rod-lens system, and fiber optic light carrier. Operates by connecting to external light source via light post; images transmitted through lens system to eyepiece. Typically attached to camera/monitor system for physician viewing. Used in clinical settings by otolaryngologists or surgeons to facilitate visualization during endoscopic procedures. Provides direct optical access to anatomical structures, aiding surgical navigation and decision-making. Reusable; requires cleaning and steam sterilization between uses.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included surface temperature measurements per IEC 60601-2-18 and optical parameter verification (field of view, direction of view) per ISO 8600-3. Reprocessing validation performed per AAMI TIR12, AAMI TIR30, and ISO 17665 standards.
Technological Characteristics
Reusable rigid rod-lens endoscope. Materials: surgical grade stainless steel (shaft/light post), glass/acrylic (fiber optics), glass (lens), Ultem (eyepiece), epoxy (lens bonding). Diameters: 2.7mm, 4.0mm. Working lengths: 50mm, 100mm, 175mm. Direction of view: 0°-70°. Field of view: 70°. Illumination: external light source via fiber optics. Connectivity: eyepiece-to-camera/monitor. Sterilization: steam. Standards: IEC 60601-2-18, ISO 8600-3, ISO 14937, ISO 17665.
Indications for Use
Indicated for patients undergoing otolaryngology, Head and Neck, rhinology, and endoscopic plastic and reconstructive surgical procedures.
Regulatory Classification
Identification
A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
# Please wait...
If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.
You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.
For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.
Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.
{1}------------------------------------------------
## 510(K) Summary
#### General Information
| Preparation Date: | June 23, 2017 |
|-------------------|--------------------------------------------------------------------|
| Owner's Name: | National Advanced Endoscopy Devices, Inc.<br>Registration: 9027397 |
| Address: | 22134 Sherman Way |
USA
Telephone Number: Fax Number: Contact Person:
#### Subject Device Name:
Trade Name Common/Usual Name: Classification Name:
| | 21 CFR 874.4760; Class II | |
|-------------------------|-------------------------------------------------------------|--|
| Predicate Device Name: | Schoelly Sinuscope | |
| Trade Name: | Schoelly Sinuscope | |
| Common/Usual Name: | Sinuscope | |
| Classification Name: | EOB- Nasopharyngoscope (flexible or rigid) and accessories | |
| | 21 CFR 874.4760; Class II | |
| Premarket Notification: | K142249, Schoelly Fiberoptic GmbH, SE date January 27, 2015 | |
EOB- Nasopharyngoscope (flexible or rigid) and accessories
Canoga Park, CA 91303
(818) 227-2720
(818) 227-2724
AED Sinuscope
AED Sinuscope
Sinuscope
Gayle Butler
# Premarket Notification:
## Device Description:
## 5700 Series AED Sinuscopes
5700A AED Sinuscope 4.0mm diameter, 0° Angle of View, 175mm working length
5700AXS AED Sinuscope 4.0mm diameter, 0° Angle of View, 50mm working length
5700BAED Sinuscope 4.0mm diameter, 30° Angle of View, 175mm working length
5700CAED Sinuscope 4.0mm diameter, 70° Angle of View, 175mm working length
5700F AED Sinuscope 4.0mm diameter, 45° Angle of View, 175mm working length
{2}------------------------------------------------
### 5727 Series AED Sinuscopes
- 5727A AED Sinuscope 2.7mm diameter, 00 Angle of View, 175mm working length
5727AX AED Sinuscope 2.7mm diameter, 0° Angle of View, 100mm working length
5727AXS AED Sinuscope 2.7mm diameter, 0° Angle of View, 50mm working length
5727BAED Sinuscope 2.7mm diameter, 30° Angle of View, 175mm working length
5727BS AED Sinuscope 2.7mm diameter, 30° Angle of View, 100mm working length
#### 8800 Series AED Sinuscopes
8800A AED Sinuscope 4.0mm diameter, 0° Angle of View, 175mm working length
8800AXS AED Sinuscope 4.0mm diameter, 0° Angle of View, 50mm working length
8800BAED Sinuscope 4.0mm diameter, 30° Angle of View, 175mm working length 8800CAED Sinuscope 4.0mm diameter, 70° Angle of View, 175mm working length
8800F AED Sinuscope 4.0mm diameter. 45° Angle of View. 175mm working length
#### 8827 Series AED Sinuscopes
8827A AED Sinuscope 2.7mm diameter, 0° Angle of View, 175mm working length
8827AX AED Sinuscope 2.7mm diameter, 0° Angle of View, 100mm working length
8827AXS AED Sinuscope 2.7mm diameter, 0° Angle of View, 50mm working length
8827BAED Sinuscope 2.7mm diameter, 30° Angle of View, 175mm working length
{3}------------------------------------------------
8827BS AED Sinuscope 2.7mm diameter, 30° Angle of View, 100mm working length
The AED Sinuscope is a reusable rod lens sinuscope consisting of an eyepiece lens and a light post connection for fiber optic light cables. A shaft composed of surgical grade stainless steel encloses the glass rod-lens system and a built-in fiber optic light carrier made of glass/acrylic. The light post body is also comprised of surgical stainless steel. The eyepiece is made from Ultem.
AED Sinuscopes are manufactured in multiple configurations that differ in insertion tube outer diameter and working length and with respect to optical parameters (direction of view, field of view).
The AED Sinuscope is reusable and provided non-sterile. It must be cleaned and sterilized prior to use.
## Technological Characteristics
To view the sinus and nasal passages the distal end of the sinuscope with a light source attached is passed into a nasal passage. Images are transmitted to the eyepiece which is typically attached to a camera so the images received through the sinuscope can be displayed on a video screen.
Technical parameters of the AED Sinuscope are characterized by the optical view -Direction of View (0° - 70°) and the Field of View (70°). The AED Sinuscopes are available in two different lengths for the insertion tube: 4mm diameter and 2.7mm diameters with working lengths that vary from 50mm, 100mm and 175mm.
The AED Sinuscope differs from the predicate device Schoelly Sinuscope (K142249):
- . The Schoelly Sinuscope comes in a 2.0mm as well as 2.7mm and 4.0mm diameters. The AED Sinuscope comes in 2.7mm and 4.0mm diameters.
- The Schoelly Sinuscope comes in 110mm, 175mm and 187mm working . lengths. The AED Sinuscope comes in 50mm, 100mm and 175mm working lengths.
The different diameters and workings lengths do not change the same basic design or use of the AED Sinuscope. The same basic mechanical design, operating principles and indications for use are the same for both devices.
A detailed comparison of the technological characteristics with the predicate device is shown in the table below:
{4}------------------------------------------------
| Attribute | Proposed<br>Device<br>(current Submission) | Predicate<br>Device<br>K142249 |
|---------------------------------|-------------------------------------------------------------|-------------------------------------------------------------|
| Light<br>Transmission | Fiberoptics | Fiberoptics |
| Light Source | External, connected via light post to light guide connector | External, connected via light post to light guide connector |
| Image<br>Transmission | Rigid Rod Lens | Rigid Rod Lens |
| Diameter | 4mm, 2.7mm | 4mm, 2.7mm, 2.0mm |
| Direction of View | 0° - 70° | 0° - 70° |
| Field of View | 70° | 80° - 100° |
| Image Display | Camera/monitor connected via the eyepiece of the sinuscope | Camera/monitor connected via the eyepiece of the sinuscope |
| Working Length | 50mm, 100mm, 175mm | 110mm, 175mm, 187mm |
| Working<br>Channel | None | None |
| Single Use/Reusable | Reusable | Reusable |
| Reprocessing | Cleaning, sterilization (steam) | Cleaning, sterilization (steam, STERRAD®) |
| Materials | | |
| Insertion Tube (outer surface) | stainless steel | stainless steel |
| Insertion Part (distal surface) | stainless steel, glass | stainless steel, glass |
| Fiberoptics | glass/acrylic | glass/acrylic |
| Lens bonding | epoxy | epoxy |
| Temperature Testing | IEC 60601-2-18 compliant | IEC 60601-2-18 complaint |
| Packaging | Case with foam insert | Case with foam insert |
# Technological Characteristics of the Proposed Device and the Predicate Device
# Indications for Use
The AED Sinuscope is intended for use in otolaryngology and Head and Neck procedures, including rhinology, and endoscopic plastic and reconstructive surgery.
{5}------------------------------------------------
## Non-clinical Performance Testing
Temperature testing: The device was measured for surface temperatures at various locations over time and found to meet the requirements specified in IEC 60601-2-18: Medical Electrical Equipment - Part 2: Particular Requirements For The Basic Safety And Essential Performance Of Endoscopic Equipment. Edition 3.0 2009-08
Optical parameter testing: The device was tested for all relevant optical parameters and found to meet the minimum requirements defined in internal specifications and as specified in the ISO 8600-3:1997 and ISO 8600-3: Amendment 1: 2003 standards (field of view, direction of view ).
# Reprocessina
The AED Sinuscope is the subject of completed reprocessing validations including manual cleaning and steam sterilization.
The cleaning study has been performed in accordance with AAMI TIR12:2010 (Designing, Testing and Labeling Reusable Medical Devices for Reprocessing in Health Care Facilities: A Guide for Device Manufacturers), AAMI TIR30:2011 (A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable devices) and FDA Guideline on General Principles of Process Validation, January 2011, Rev 1.
The sterilization study has been performed in accordance with ISO 14937:2009 (Sterilization of health care products - General requirements for characterization of sterilizing agent and the development, validation and routine control of a sterilization process for medical devices), ANSI/AAMI/ISO 17665-1:2006 (Sterilization of health care products - Moist heat - Requirements for the development, validation and routine control of sterilization process for medical devices),AAMI TIR 12:2010 (Designing, testing and labeling reusable medical devices for reprocessing in health care facilities) and ISO 17665-2:2009 (Sterilization of health care products - Moist heat - Part 2: Guidance of the application of ISO 17665-1).
## Conclusion
The AED Sinuscope meets all the pre-determined acceptance criteria to demonstrate substantial equivalence to the predicate device.
{6}------------------------------------------------
Public Health Service
Image /page/6/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of three human profiles facing right, stacked on top of each other. The profiles are simple and abstract, with flowing lines suggesting hair or clothing. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 1, 2017
National Advanced Endoscopy Devices. Inc. Gayle Butler Compliance Officer 22134 Sherman Way Canoga Park, CA 91303
Re: K170285
Trade/Device Name: AED Sinuscope Regulation Number: 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (Flexible or Rigid) and Accessories Regulatory Class: Class II Product Code: EOB Dated: June 29, 2017 Received: July 3, 2017
Dear Gayle Butler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.