ShenDa® Sinuscope is intended to provide an endoscopic means to view the nasal cavity and nasopharynx.
Device Story
ShenDa® Sinuscope is a reusable, unchanneled rigid endoscope used for visualizing the nasal cavity and nasopharynx. It functions as an optical viewing device, requiring connection to an external light source (e.g., Olympus CLV-180) and light-guide cable to illuminate the target anatomy. The device is available in various configurations, including outer diameters of 2.7mm, 3mm, and 4mm, and multiple viewing angles (0, 12, 30, 45, 70, and 90 degrees). It is intended for use by clinicians in a clinical setting. The device provides direct optical visualization to the physician, aiding in diagnostic or surgical assessment of the nasal and nasopharyngeal regions. It is reprocessed via cleaning and steam sterilization.
Clinical Evidence
No clinical data. Bench testing only. Compliance verified with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, ISO 10993-5, ISO 10993-10, ISO 8600-1, and ISO 8600-5. Performance testing included insertion width, field of view, direction of view, and optical resolution.
Technological Characteristics
Rigid, unchanneled endoscope; optical viewing principle; reusable; steam sterilization. Available in 2.7mm, 3mm, 4mm diameters; 0-90 degree viewing angles. Complies with IEC 60601-1, IEC 60601-2-18, ISO 10993-5/10, and ISO 8600-1/5. No internal software or electronic processing.
Indications for Use
Indicated for visualization of the nasal cavity and nasopharynx in patients requiring endoscopic examination.
Regulatory Classification
Identification
A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three abstract human profiles facing to the right, layered on top of each other.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 25, 2017
Shenyang Shenda Endoscope Co., Ltd % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 China
Re: K161298
Trade/Device Name: Shenda Sinuscope Regulation Number: 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (Flexible or Rigid) and Accessories Regulatory Class: Class II Product Code: EOB Dated: July 26, 2017 Received: July 27, 2017
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K161298
Device Name ShenDa® Sinuscope
Indications for Use (Describe)
ShenDa® Sinuscope is intended to provide an endoscopic means to view the nasal cavity and nasopharynx.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K161298
- 1. Date of Preparation: 08/24/2017
- 2. Sponsor Identification
## Shenyang Shenda Endoscope CO., LTD.
No.123 Hezuo Street, Dadong District, Shenyang, Liaoning, 110044, China
Establishment Registration Number: Not yet registered
Contact Person: Gao, Feng Position: Overseas Sales Manager Tel: +86-24-88093290 Fax: +86-24-88903118 Email: clark_925@163.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Mr. Lee Fu (Alternative Contact Person)
Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 240-238-7587 Email: info@mid-link.net
{4}------------------------------------------------
#### 4. Identification of Proposed Device
Trade Name: ShenDa® Sinuscope Common Name: Sinuscope Model: BD-1 19 Product Codes: J0200G, J0200CG, J1200, J2200, J0212, J0230CG, J1230, J2230, J0245G, J0245E, J0270G, J0270CG, J1270, J2270, J0290E, J0200E, J0230E, and J0270E
Regulatory Information
Classification Name: Nasopharyngoscope (Flexible or Rigid) Classification: Class II Product Code: EOB Regulation Number: 21 CFR part 874.4760 Review Panel: Ear Nose & Throat
Intended Use Statement:
ShenDa® Sinuscope is intended to provide an endoscopic means to view the nasal cavity and nasopharynx.
Device Description
ShenDa® Sinuscope is an unchanneled endoscope intended to provide an endoscopic means to view the nasal cavity and nasopharynx.
The endoscope is available in the various combination of outer diameter (2.7mm, 3mm, 4mm) and vision direction angles (0, 12, 30, 45, 70 and 90 degree), which are listed in Table 1 Specifications of the Proposed Device of this document.
It can be connected to OLYMPUS light source and OLYMPUS light cable; the information of associated OLYMPUS compatible equipments is provided as following:
| Light source | OLYMPUS CLV-180 EVIS EXTRA II XENON LIGHT SOURCE<br>(K100584) |
|----------------------|------------------------------------------------------------------------------|
| Light-guide<br>cable | Size S, plug type, 3m, CF type, identification number: WA03200A<br>(K944072) |
| Connector | Not required |
It is a reusable device subject to cleaning and steam sterilization.
{5}------------------------------------------------
| Table 1 Specifications | | | | | |
|------------------------|---------|-----------------------|-------------------------|----------------------------------|--------------------|
| No. | Code | Direction of view (°) | Field of view field (°) | Diameter of inserting section(Φ) | Working length(mm) |
| 1 | J0200G | 0 | 90 | 4 | 175 |
| 2 | J0200CG | 0 | 110 | 4 | 175 |
| 3 | J1200 | 0 | 60 | 3 | 175 |
| 4 | J2200 | 0 | 60 | 2.7 | 175 |
| 5 | J0212 | 12 | 60 | 4 | 175 |
| 6 | J0230G | 30 | 90 | 4 | 175 |
| 7 | J0230CG | 30 | 110 | 4 | 175 |
| 8 | J1230 | 30 | 60 | 3 | 175 |
| 9 | J2230 | 30 | 60 | 2.7 | 175 |
| 10 | J0245G | 45 | 90 | 4 | 175 |
| 11 | J0245E | 45 | 60 | 4 | 175 |
| 12 | J0270G | 70 | 90 | 4 | 175 |
| 13 | J0270CG | 70 | 110 | 4 | 175 |
| 14 | J1270 | 70 | 60 | 3 | 175 |
| 15 | J2270 | 70 | 60 | 2.7 | 175 |
| 16 | J0290E | 90 | 60 | 4 | 175 |
| 17 | J0200E | 0 | 60 | 4 | 175 |
| 18 | J0230E | 30 | 60 | 4 | 175 |
| 19 | J0270E | 70 | 60 | 4 | 175 |
Table 1 Specifications
- 5. Identification of Predicate Device
510(k) Number: K965233 Device Name: SHARPSITE Ac Manufacturer: Medtronic Xomed Inc.
- 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- A IEC 60601-1-2005, Medical Electrical Equipment- Part 1: General requirements for basic safety and essential performance.
- > IEC 60601-1-2:2007, Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility-Requirements and tests.
- A IEC 60601-2-18:2009, Medical Electrical Equipment- Part 2: Particular requirements for the basic safety and essential performance of endoscopic equipment.
{6}------------------------------------------------
- > ISO 10993-5:2009, Biological Evaluation of Medical Device, Part 5-Tests for Vitro Cytotoxicity;
- > ISO 10993-10:2010, Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization;
- > ISO 8600-1:2015 Endoscopes--Medical endoscopes and Endotherapy Devices--part 1: General Requirements.
- > ISO 8600-5:2005 Optics and photonics-Medical endoscopes and endotherapy devices part 5: Determination of optical resolution of rigid endoscopes with optics
The performance test items include the insertion width, field of view and direction of view, and optical resolution.
- 7. Clinical Test Conclusion
No clinical study is included in this submission.
- Substantially Equivalent (SE) Comparison 8.
| Item | Proposed Device | Predicate Device |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | EOB | K965233 |
| Regulation Number | 21 CFR part 874.4760 | EOB |
| | | 21 CFR part 874.4760 |
| Intended Use | ShenDa® Sinuscope is intended to provide an endoscopic means to view the nasal cavity and nasopharynx. | The Xomed SHARPSITE Ac rigid rod lens endoscopes are intended for use in surgical imaging in otolaryngology and Head and Neck procedures, including rhinology, and endoscopic plastic and reconstructive surgery. |
| Feature | unchanneled rigid endoscope | unchanneled rigid endoscope |
| Mechanism | Viewing Optics | Viewing Optics |
| Working Length | 175mm | 175mm |
| Direction of View | 0, 12, 30, 45, 70, and 90 degree | 0, 30, and 70 degree |
| Field of View | 60, 90 and 110 degress | 100 and 102 degress |
| Shaft Body diameter | 2.7, 3.0, 4.0mm | 4.0mm |
| Depth of Field | 1-50mm | Not Known |
| Safety | Comply with IEC 60601-1, IEC 60601-1-2 and IEC 60601-2-18 | Not Known |
| Biocompatibility | Comply with ISO 10993-5 and ISO 10993-10<br>No adverse reaction demonstrated in cytotoxicity, intracutaneous and skin sensitization | Not Known |
| Reprocessing | Cleaning and Steam Sterilization, | Cleaning and Steam Sterilization, |
### Table 1 SE Comparison
{7}------------------------------------------------
Although the indications for use are expressed differently, the proposed device and predicate device have the similar product structure, working principle and optical performance. This item is considered to be substantially equivalent.
The proposed device provides a more range of field of view and direction of view than the predicate device, additionaly it provides more options of diameters of shaft than the predicate device, to the physicians to fit various patients, these difference it will not affect the safety and effectiveness concerning the SE. Although the depth of filed of the prediate device is unknow, we think the design of the depth of field of the proposed device is appropriate for the clinical use. Although the result of Electrical Safety, EMC and biocompatibility of predicate device is unknown, it is concluded from the relevant testng results of the proposed device, that the proposed device is safe and effective.
- Substantially Equivalent (SE) Conclusion 9.
Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.