K100577 · Acclarent, Inc. · EOB · Jan 6, 2011 · Ear, Nose, Throat
Device Facts
Record ID
K100577
Device Name
ACCLARENT CYCLOPS MULTI-ANGLE ENDOSCOPE
Applicant
Acclarent, Inc.
Product Code
EOB · Ear, Nose, Throat
Decision Date
Jan 6, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4760
Device Class
Class 2
Indications for Use
The Acclarent Cyclops Multi-Angle Endoscope is intended to provide an endoscopic means to view the nasal cavity and nasopharynx in an operating room environment.
Device Story
Acclarent Cyclops Multi-Angle Endoscope is a 4mm rigid endoscope for nasal cavity and nasopharynx visualization. Device features a direction-of-view dial (10° to 100°) and a 320° shaft rotation dial, allowing visualization of structures without rotating the scope body. Rare-earth permanent magnets in the proximal control body drive the direction-of-view change. Device includes a standard eyepiece for camera coupler attachment and a light post compatible with standard medical light sources via included adapters. Used in operating room environments by physicians. Output is a visual image transmitted to a camera system for clinical assessment. Benefits include enhanced visualization capabilities during endoscopic procedures.
Clinical Evidence
Bench testing and a cadaver study were performed. Bench testing evaluated distal shaft diameter, working length, field of view, fixed focus, direction of view, rotation of view, illumination, resolution, dial actuation forces, temperature, magnet field strength, electrical safety, EMC, durability, environmental conditioning, compression, vibration, and shock. No clinical data was necessary.
Technological Characteristics
4mm rigid endoscope; stainless steel construction; glass fiber illumination; rare-earth permanent magnets for direction-of-view control; 10°-100° direction of view; 60° field of view; 5-45mm depth of focus; 175mm working length; compatible with standard camera couplers and medical light sources; reusable; requires manual cleaning and steam sterilization.
Indications for Use
Indicated for visualization of the nasal cavity and nasopharynx in an operating room environment.
Regulatory Classification
Identification
A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
Predicate Devices
OPTIM Inc. ENTity Nasoview Nasopharyngoscope (K080622)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "Acclarent" in a bold, sans-serif font. A curved line extends from the top of the "A" to the top of the "t", creating a visual arc over the word. A small dot or symbol is positioned above the "t", adding a distinctive element to the logo.
JAN 6 2011
:
K100577
# 510(k) SUMMARY
| Sponsor/Submitter: | Acclarent, Inc.<br>1525-B O'Brien Drive<br>Menlo Park, California 94025 |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Keri Yen<br>Regulatory Affairs Manager<br>Phone: (650) 687-5874<br>Fax: (650) 687-4449 |
| Date of Preparation: | January 5, 2011 |
| Device Trade Name/<br>Model Number: | Acclarent Cyclops Multi-Angle Endoscope<br>CYE001 |
| Common Name: | Endoscope |
| Device Classification: | Class II |
| Regulation Number: | 21 CFR 874.4760 |
| Classification Name: | Nasopharyngoscope (Flexible or Rigid) |
| Product Code: | EOB |
| Predicate Devices: | OPTIM Inc. ENTity Nasoview Nasopharyngoscope (K080622)<br>Pollux Endoscopy Inc. Sinuscope (K002214)<br>Optus Sinuscope (K944656)<br>Karl Storz Hopkins Rigid Autoclavable Telescope (K935279)<br>Stryker Endoscopy Arthroscope (K093677) |
| Device Description: | The Acclarent Cyclops Multi-Angle Endoscope is a 4mm rigid<br>endoscope that has the capability of varying direction of view from<br>10° to 100°, which is altered by the direction of view dial. The<br>direction of view is indicated by visible markings on the scope body<br>Cyclops provides a 60° field of view and a depth of focus from 5<br>mm to 40mm. The device shaft can also rotate 320° to allow for<br>visualization of structures without rotating the device; this is<br>controlled by the shaft rotation dial. Small rare-earth permanent<br>magnets are incorporated into the proximal scope control body (≤10<br>gauss at 2cm) and drive the change in the direction of view. A<br>standard eyepiece located on the proximal end of the device is<br>compatible with a standard camera coupler. The light post on the<br>subject device is compatible with an ACMI light source.<br>There are two stainless steel adapters that accompany the Acclarent<br>Cyclops Multi-Angle Endoscope to facilitate connection with Wolf<br>or Storz/Olympus medical light sources. The adapters connect to the |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "Acclarent" in a bold, sans-serif font. A curved line extends from the top of the "l" to the right, creating a visual element above the word. The font is black, and the background is white.
light post. The Acclarent Cyclops Multi-Angle Endoscope is a reusable device and must be cleaned and sterilized according to the user manual prior to every use.
Indications for Use:
The Acclarent Cyclops Multi-Angle Endoscope is intended to provide an endoscopic means to view the nasal cavity and nasopharynx in an operating room environment.
### Technological Characteristics:
| Attribute | Predicate Device<br>(OPTIM Inc ENTity<br>Nasoview<br>Nasopharyngoscope) | Predicate Device<br>(Pollux<br>Endoscopy, Inc<br>Sinuscope) | Predicate Device<br>(Optus Sinuscope) | Subject Device<br>(Acclarent<br>Cyclops Multi-<br>Angle Endoscope) |
|------------------------|-------------------------------------------------------------------------|-------------------------------------------------------------|---------------------------------------|--------------------------------------------------------------------|
| 510(k)<br>number | K080622 | K002214 | K944656 | K100577 |
| Rigidity | Flexible and Steerable | Rigid | Rigid | Same, Rigid |
| Viewing<br>Optics | Lens | Lens | Lens | Same, Lens |
| Depth of<br>View | 5-50mm | 5mm-45mm | Unknown | Same, 5-45 mm |
| Field of View | 70° | 95° | 71° to 83° | 60° |
| Direction of<br>View | 0° to 135° | 0°, 30°, 45°, 70° | 0°, 30°, 70° | 10° to 100° |
| Shaft Body<br>Diameter | 3.6mm | 2.7mm or 4mm | 2.7mm or 4mm | Same, 4mm |
| Working<br>Length | 11.8 inches (30cm) | 9.06 inches<br>(230mm) | 6.89 inches | Same, 6.89 inches<br>(175mm) |
| Illumination<br>Fibers | Glass Fibers | Glass Fibers | Glass Fibers | Same, Glass Fibers |
| Light Source | Integrated LED | Medical light<br>source | Medical light<br>source | Same, Medical<br>light source |
#### Performance Data:
Performance testing of the Acclarent Cyclops Multi-Angle Endoscope consisted of bench testing and a cadaver study. Bench testing met all acceptance criteria for attributes such as distal shaft diameter, working length, field of view, fixed focus, direction of view, rotation of view, illumination, scope resolution, dial actuation forces, temperature testing, field strength testing of magnets, electrical safety, EMC testing, durability testing, environmental conditioning, compression testing, random vibration testing, and shock (free fall drop) testing. Clinical data was not necessary for the subject device. The performance data demonstrates that the Acclarent Cyclops Multi-Angle Endoscope performs as intended.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "Acclarent" in a bold, sans-serif font. A curved line extends from the top of the "r" to the top of the "t", creating a visual element above the word. The font is black and the background is white.
| Validated Reprocessing<br>Methods: | Full manual cleaning with extended enzymatic soak plus general instrument automated washer Pre-vacuum steam sterilization (wrapped) |
|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Summary of Substantial<br>Equivalence: | The Acclarent Cyclops Multi-Angle Endoscope is substantially equivalent to the predicate devices as confirmed through relevant |
bstantially The Acclarent Cyclops Multi-Angle Endoscope is substantially
equivalent to the predicate devices as confirmed through relevant performance tests and attributes.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure with stylized feathers.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
2011 JAN 6
Acclarent, Inc. c/o Ms. Keri Yen Regulatory Affairs Manager 1525-B O'Brien Dr. Menlo Park, CA 94025
Re: K100577
Trade/Device Name: Acclarent Cyclops Multiangle Endoscope Regulation Number: 21 CFR 874.4680 Regulation Name: Nasopharyngoscope, Flexible or Rigid Regulatory Class: II Product Code: EOB Dated: 12/23/2010 Received: 12/27/2010
1. 11. 1
Dear Ms. Yen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Keri Yen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements not the Act or any Federal statutes and regulations administered by other Federal agencies. Your must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Malvina B. Eydelman. M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Acclarent" in a bold, sans-serif font. A curved line extends from the top right of the "t" to above the "e", creating a visual element that resembles a smile or arc over the word. The text is black against a white background.
## INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K100577
Trade Name:
Acclarent Cyclops Multi-Angle Endoscope
Common Name:
Endoscope
Indications For Use:
The Acclarent Cyclops Multi-Angle Endoscope is intended to provide an endoscopic means to view the nasal cavity and nasopharynx in an operating room environment.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
X Prescription Use. (Per 21 CFR 801.109)
_
(Division Sign Off)
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
.10057
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.