K083553 · Schoelly Imaging, Inc. · EOB · Jan 16, 2009 · Ear, Nose, Throat
Device Facts
Record ID
K083553
Device Name
NASO-LARYNGO-PHARYNGOSCOPE
Applicant
Schoelly Imaging, Inc.
Product Code
EOB · Ear, Nose, Throat
Decision Date
Jan 16, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 874.4760
Device Class
Class 2
Indications for Use
The Schoelly Naso-Laryngo-Pharyngoscope is intended to be used by qualified surgeons and physicians to visualize and observe the pharynx and the larynx.
Device Story
Flexible, steerable fiberoptic endoscope for diagnostic visualization of pharynx and larynx. Device captures optical images via fiber bundles; transmits to eyepiece or camera for physician observation. Operated by surgeons/physicians in clinical settings. Provides direct visualization to assist in diagnostic assessment of throat/larynx anatomy. Non-sterile, requires EtO sterilization before use. Benefits include improved diagnostic access to upper airway structures.
Clinical Evidence
Bench testing only. No clinical data provided. Performance validated through comparison of optical and mechanical specifications against predicate device.
Technological Characteristics
Flexible fiberoptic endoscope; 3.2mm distal tip diameter; 300mm working length; 140-degree angulation; 70-degree field of view; 7,000 fiber count. Compliant with IEC 60601-1 and IEC 60601-2-18. Supplied non-sterile; validated for EtO sterilization (6% EtO/94% CO2, 131°F, 120 min exposure). Compatible with chemical disinfectants including Gigasept FF, Lysetol FF, Helipur HplusN, and Cidex.
Indications for Use
Indicated for visualization and observation of the pharynx and larynx by qualified surgeons and physicians.
Regulatory Classification
Identification
A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
## Special 510(k) Summary
| | A) Submitted by Schoelly Imaging, Inc<br>100 Hartwell St<br>West Boylston, MA 01583<br>Registration 8043903 | JAN 1 6 2009 |
|---------|-------------------------------------------------------------------------------------------------------------|--------------|
| Contact | MEDIcept<br>200 Homer Ave<br>Ashland, MA 01721<br>F. David Rothkopf | |
Date of Application November 26, 2008
B) Device Name Naso-Laryngo-Pharyngoscope
508-231-8842 x20 508-231-8861 Fax
Common Name Nasopharyngoscope
Device Class 21 CFR 874 4760 CLASS II
- Product Code EOB
C) Predicate Naso-Laryngo-Pharyngoscope (K991560)
D) Device Description
The Schoelly Naso-Laryngo-Pharyngoscope is a flexible and steerable fiberoptic endoscope for diagnostic purposes
- E) Intended Use
The Schoelly Naso-Laryngo-Pharyngoscope is intended to be used by qualified surgeons and physicians to visualize and observe the pharynx and the larynx
- F) Comparison to Predicate Device(s)
The Schoelly Naso-Laryngoscope has the same intended use, target population, clinical setting, and technology as its predicate device Schoelly Naso-Laryngo-Pharyngoscone (K991560) There is no change to in compatibility with the environment or other devices, go differences in electrical, mechanical or chemical safety There are slight changes in the optical and mechanical specifications and patient contact materials, all of which have non-significant impact on device performance All changes have been made under design control, and validation was performed and conformance assured
Schoelly Imaging, Inc Special 510(k) Naso-Laryngo-Pharyngoscope
{1}------------------------------------------------
| Features | Schoelly Naso-Laryngoscope<br>(the Special 510K application<br>herein) | Schoelly-Naso-<br>Laryngoscope<br>(K991560) |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Visualize and observe the<br>pharynx and the larynx | Visualize and observe<br>the pharynx and the<br>larynx |
| Design | Flexible and steerable fiberoptic<br>endoscope for diagnostic<br>purposes | Flexible and steerable<br>fiberoptic endoscope<br>for diagnostic<br>purposes |
| Parts number | 31 1001s | 31 0001s |
| Specifications | 31 0001s/31 1002s | |
| Tip tubing diameter | $3.2 mm$ | $3.8 mm$ |
| Working Length | $300 mm$ | $300 mm$ |
| Angulation Up | $140^0$ | $150^0$ |
| Angulation Down | $140^0$ | $150^0$ |
| Direction of view | $0^0$ | $0^0$ |
| Field of View | $70^0$ | $85^0$ |
| Length of distal tip<br>deflection | $20 mm$ | $25 mm$ |
| Depth of field | $6 - 1000 mm$ | $3mm - 50mm$ |
| Total Fibers (pixels) | $7,000$ | $10,000$ |
| Fiber size | $8 3 \mu m$ | $8 3 \mu m$ |
| Magnification³ | $7X$ | $2 5 X$ |
| Energy used | IEC601-1 and IEC 60601-2-18 2000 compliant | |
| Labeling<br>Sterilization*/Cleaning | The Instruction Manual states<br>Sterilization | Users Manual states |
| *device is supplied as<br>non-sterile | Section 7 5 Flexible Schoelly<br>endoscopes with air-exhaust valve can<br>also be gas sterilized using ethylene<br>oxide "<br>Specifications<br>Gas mixture 6% EtO, 94% CO₂<br>Temperature 131°F+/-5°F, 55°C +/-2°C<br>Relative air humidity 40-90%<br>Pressure (overpressure), 1 7 bar<br>(170kpa)<br>Exposure time 120 mins | "The Schoelly Naso-<br>Laryngo-Pharyngoscope is<br>provided non-sterile, and<br>must be sterilized using<br>EtO sterilization prior to<br>use<br>"<br>Specifications<br>EO concentration 600 +/-<br>25 mg/L<br>Temperature 54+/- 2°C<br>(58°C) max)<br>Relative humidity 70% +/-<br>5%<br>Pressure 14 +/- 1 PSIG |
| Cleaning | Gas exposure time 4 hours | |
| Section 7 2 states | | |
| "Place the endoscope and the<br>unscrewed parts in a suitable<br>disinfection solution (can also double up<br>as a cleaning solution) The disinfection<br>solution used should be permitted by the<br>manufacturer for such use | | |
| Material compatibility releases exist for<br>the following disinfectants<br>*Gigasept FF, Schulke & Mayr GmbH<br>*Lysetol FF<br>*Helipur HplusN, B Braun Medical AG<br>*Cidex, Johnson & Johnson " | "To dissolve various<br>impurities, use mild<br>cleaning substances" | |
{2}------------------------------------------------
## Naso-Laryngo-Pharyngoscope Special 510(k) Premarket-Notification Submission
" Manufactured in two resolutions, 12,000 and 16,000 pixels
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 1 6 2009
Schoelly Imaging, Inc c/o David McNally 100 Hartwell Street West Boylston, MA 01583
Re K083553
Trade/Device Name Naso-Pharyngoscope Regulation Number 21 CFR 874 4760 Regulation Name Nasopharyngoscope (flexible or rigid) and accessories Regulatory Class Class, II Product Code EOB Dated November 20, 2008 Received December 18, 2008
Dear Mr McNally
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA) You may, therefore, market the device, subject to the general controls provisions of the Act The general controls provisions of the Act include requirements for annual registration, insting of devices. good manufacturing practice. labeling, and prohibitions against misbranding and adulteration
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807), labeling (21 CFR Part 801), good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820), and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act), 21 CFR 1000-1050
ﻧﺴﻴ
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115 Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807 97) For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Drysson of Surveillance Systems at 240-276-3464 You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely vours.
M.B. Egelston, mD
Malvina B Eydelman, M D Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Ko83553
## Indications for Use
510(k) Number (if known) Ko 8 3 53 3
Device Name Naso-Laryngo-Pharyngoscope
Indications for Use
The Schoelly Naso-Laryngo-Pharyngoscope is intended to be used by qualified surgeons and physicians to visualize and observe the pharynx and the larynx.
Prescription Use X 21CFR 801, Subpart D OR Over-the-Counter Use_21CFR 801 109
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NECESSARY)
> ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------.. . ..... . ..... .. .. .. .. ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... .. 16 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
Concurrence of CDRH, Office of Device InVitro Diagnostic Device Evaluation and Safety (OIVD)
Karen H. Baker
Division Sign Off Office of Device InVitro Diagnostic Device Evaluation and Safety
510 (K) K083553
Schoelly Imaging, Inc Special 510(k) Naso-Laryngo-Pharyngoscope
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.