ENDOMEDICAL TECHNOLOGIES QUIK-COVER FLEXIBLE ENDOSCOPE BARRIER SHEATH FOR OTOLARYNGOSCOPES AND BRONCHOSCOPES
Applicant
Rtc Inc.-Memcath Technologies, LLC
Product Code
EOB · Ear, Nose, Throat
Decision Date
Sep 6, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4760
Device Class
Class 2
Indications for Use
The EndoMedical Technologies Quik-Cover™ Flexible Endoscope Barrier Sheath provides a sterile, disposable protective covering for flexible endoscopes during endoscopic examination of the upper airway, vocal chords, nasal passages, esophagus and pulmonary structures,
Device Story
The EndoMedical Technologies Quik-Cover™ is a sterile, single-use polymeric sheath designed to cover the patient-contact surface of flexible endoscopes. It eliminates the need for high-level disinfection between procedures. The device consists of a contiguous sheath with a distal optical window, preloaded on a deployment tube. Some models include a Y-connector and side working channels for air, water, suction, and biopsy instruments. The sheath is applied by sliding it over the endoscope without requiring vacuum or pressure sources. It is intended for use by clinicians during endoscopic examinations of the upper airway, vocal chords, nasal passages, esophagus, and pulmonary structures. By providing a physical barrier, it prevents pathogen transmission between patients and the endoscope.
Clinical Evidence
Bench testing only. Testing included evaluation of optical qualities of the sheath window, mechanical testing of the sheath, and finished device barrier testing.
Technological Characteristics
Polymeric sheath with distal optical window; preloaded on deployment tube; includes optional Y-connector and side working channels for air, water, suction, and biopsy instruments. Single-use, sterile device. No energy source required. Manual application.
Indications for Use
Indicated for use as a sterile, disposable protective covering for endoscopes during endoscopic examination of the upper airway, vocal chords, nasal passages, esophagus, and pulmonary structures.
Regulatory Classification
Identification
A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
Vision-Sciences EndoSheaths for use with Flexible ENT Scopes (K990354)
Vision-Sciences Flexible Fiberoptic Bronchoscope and EndoSheath System (K021344)
Vision-Sciences Flexible Cystoscope with EndoSheath System (K040215)
Vision-Sciences Trans-Nasal Esophagoscope with EndoSheath System (K031786)
SS-F32 EndoSheath for use with the Vision-Sciences Model S-F100 Sigmoidoscope (K963344)
Submission Summary (Full Text)
{0}------------------------------------------------
K061663
Document Date: 5/05/06
- 6 2006
#### 510(k) Summary 5.
May 5, 2006 Date Prepared Revised for ENT/Bronchoscope only May 30, 2006
| New Device | EndoMedical Technologies Quik-Cover™ Flexible Endoscope Barrier Sheath |
|-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Predicate Devices | |
| | K000767 Memcath Urology Introducer Sheath |
| | K042531 UPDATED SLIP Urology Introducer Sheath, Percutaneous Systems,<br>Inc. |
| | ✓ K990354 Vision-Sciences EndoSheaths for use with Flexible ENT Scopes |
| | K021344 Vision-Sciences Flexible Fiberoptic Bronchoscope and EndoSheath<br>System |
| | K040215 Vision-Sciences Flexible Cystoscope with EndoSheath System |
| | ✓ K031786 Vision-Sciences Trans-Nasal Esophagoscope with EndoSheath System |
| | K963344 SS-F32 EndoSheath for use with the Vision-Sciences Model S-F100<br>Sigmoidoscope |
| Contact | Marc Jaker, Vice President<br>RTC, Inc. - Memcath Technologies, LLC<br>1777 Oakdale Ave<br>West St. Paul, MN 55118<br>Phone: 651-450-7400 Fax: 651-450-7555 |
## Intended Use
The EndoMedical Technologies Quik-Cover™ Flexible Endoscope Barrier Sheath provides a sterile, disposable protective covering for flexible endoscopes during endoscopic examination of the upper airway, vocal chords, nasal passages, esophagus and pulmonary structures,
## Device Description
The EndoMedical Technologies Quik-Cover™ Flexible Endoscope Barrier Sheath (EndoMedical Technologies Quik-Cover™) is a sterile, single-use device that covers the entire patient contact surface of flexible endoscopes and eliminates the need for high-level disinfection of the endoscope following each procedure. The device is composed of a contiguous polymeric sheath, with an optical window at the distal end, which is preloaded on a deployment tube. Some models of the EndoMedical Technologies Quik-Cover™ include a Y-connector and side port channel(s) that are intended to allow passage of air, water, suction and biopsy instruments.
{1}------------------------------------------------
## Substantial Equivalence Comparison
The intended use of the EndoMedical Technologies Quik-Cover is similar to the predicate devices in that the sheaths are sterile, disposable protective accessories used to cover endoscopes, catheters or instruments to help prevent the transmission of pathogens.
The EndoMedical Technologies Quik-Cover™ device configuration is similar to the predicate devices in that all devices are comprised of a membrane sheath, a proximal connector and, with respect to the endoscope barrier sheaths, a distal polymeric window. Some models of the EndoMedical Technologies Quik-Cover™ and predicate devices also include a side working channel. The EndoMedical Technologies Quik-Cover application method is similar to the predicate devices in that the sheath slides on and off with no vacuum/ pressure source required.
The EndoMedical Technologies Quik-Cover™ sheath material to the sheath material for the Memcath Urology Introducer Sheath (K000767) and the SLIP Urology Introducer Sheath (K042531) predicate devices. Safety and performance characteristics related to minor design differences have been addressed through the following nonclinical tests:
- Optical qualities of sheath window .
- Sheath mechanical tests .
- Finished device barrier testing .
Nonclinical performance testing demonstrated that the EndoMedical Technologies Quik-Cover™ reliably achieves the desired affect and is safe for its intended use. No new questions of safety or effectiveness for endoscope barrier sheaths were raised during the testing. The EndoMedical Technologies Quil-Cover Flexible Endoscope Barrier Sheath is, therefore, substantially equivalent to currently marketed devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three curved lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP - 6 2006
Memcath Technologies c/o Marc Jaker Vice President RTC, Inc. - Memcath Technologies, LLC 1777 Oakdale Avenue West St., Paul, MN 55118-4031
Re: K061663
K001005
Trade/Device Name: Quik-Cover Protective Barrier Sheath Regulation Number: 21 CFR 874.4760 Regulation Name: Nasopharyngoscope and accessories Regulatory Class: Class II Product Code: EOB Dated: August 16, 2006 Received: August 17, 2006
Dear Mr. Jaker:
We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Section 9 ro(x) promatics is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosule) to regars manetical Device Amendments, or to commerce prior to May 28, 1970, the characted with the provisions of the Federal Food, Drug. devices that have been recure approval of a premarket approval approval application (PMA). and Cosment Act (Act) that do not require approvine controls provisions of the Act. The
You may, therefore, market the device, subject to the general controls provisions of You may, therefore, market the device, basjon to the manual registration, listing of general controls provisions of the free herefore required in the manage inst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), If your device is classified (SCC above) into s. Existing major regulations affecting your device .
It may be subject to such additional controls. Existing major regulations it may be subject to such additional control - Internage - Title 21, Parts 800 to 898. In addition, FDA can be found in the Code of Peceral concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that I Dri- s issuation of a complies with other requirements of the Act that FDA has made a decemination administered by other Federal agencies. You must or any Federal Statutes and regulations administered of registration and listing (21 comply will an the Act s requirements, mercess communities practice requirements as set CFR Part 807), labeling (21 CFR Part 801), government 820); and if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR 1000 1050 forth in the quality systems (QD) regulation (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
# Page 2 -- Mr. Marc Jaker
.. +
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
M.B. EgdahlmsiWD
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### Indications for Use Statement 1.
510(k) Number (if known): K061663
Device Name: EndoMedical Technologies Quik-Cover™ Flexible Endoscope Barrier Sheath
Indications for Use: The EndoMedical Technologies Quik-Cover flexible endoscope barrier sheath provides a sterile, disposable protective covering for endoscopes used during examination of the upper airway, vocal chords, nasal passages, esophagus and pulmonary structures.
Prescription Use XX (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
ion Sign-Off) sion of Ophthalmic
Number K5061163
Prescription Use . (Per 21 CFR 801.109)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.