The EndoSheath® System provides a sterile, disposable protective covering for the scope to be used during flexible endoscopic examination of the upper airway, vocal cords and/or nasal passages. The System may also be used in conjunction with the Pentax AP-4000 Air Pulse Sensory Stimulator to elicit Laryngeal Closure Reflex (Swallow) and to measure the sensory discrimination threshold at which the reflex occurs in the area of the Upper Airway innervated by the Superior Laryngeal Nerve.
Device Story
The Slide-On™ EndoSheath® System is a sterile, single-use protective covering for the VSI ENT-2000 flexible endoscope. It functions as a physical barrier during endoscopic examinations of the upper airway, vocal cords, and nasal passages. When used with the Pentax AP-4000 Air Pulse Sensory Stimulator, the sheathed scope allows clinicians to deliver air pulses to the pyriform sinus and aryepiglottic folds to elicit the laryngeal closure reflex (swallow) and measure sensory discrimination thresholds. The device is used in clinical settings by healthcare providers. It benefits patients by providing a sterile barrier, reducing the need for high-level disinfection of the endoscope between procedures, and enabling diagnostic assessment of laryngeal sensory function in patients with suspected dysphagia.
Clinical Evidence
Bench testing only. Performance testing included sheath burst/leak testing, sheath tensile/elongation testing, sheathed scope articulation testing, and air pulse testing to ensure the sheath does not interfere with the function of the endoscope or the sensory stimulator.
Technological Characteristics
Sterile, single-use thermoplastic elastomer sheath with thermoplastic polymer window and luer connector. 12-inch length, 0.002-inch wall thickness. Designed for use with VSI ENT-2000 scope. Provides microbial barrier. Installation is manual (slide-on/off) without vacuum/pressure requirements. No electronic components; purely mechanical accessory.
Indications for Use
Indicated for adult and pediatric patients with suspected dysphagia requiring flexible endoscopic examination of the upper airway, vocal cords, and nasal passages, and for sensory discrimination threshold testing of the upper airway innervated by the superior laryngeal nerve.
Regulatory Classification
Identification
A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
Predicate Devices
Slide-On EndoSheath® for Flexible ENT Scopes (K990354)
EndoSheath® System for Flexible ENT Scopes (K012543)
Slide-On™ EndoSheath® System for Flexible ENT Scopes (K024095)
{0}------------------------------------------------
K031790
### 2 2003 JUL
Vision-Sciences, Inc. June 9, 2003
510(k) Premarket Notification: Traditional Slide-On™ EndoSheath® System for Sensory Testing
# 510(k) Summary
| Trade Name: | Vision-Sciences Slide-On™ EndoSheath® System for Sensory<br>Testing |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | Vision-Sciences, Inc.<br>9 Strathmore Road<br>Natick, MA 01760<br>Registration #1223490 |
| Device Common<br>Name: | Endoscope and Accessories - 77EOB |
| Classification: | According to Section 513 of the Federal Food, Drug, and<br>Cosmetic Act, the device classification is Class II. |
| Predicate Devices: | K990354 - Slide-On EndoSheath® for Flexible ENT Scopes<br>K012543 - EndoSheath® System for Flexible ENT Scopes<br>K024095 - Slide-On™ EndoSheath® System for Flexible ENT Scopes<br>Manufactured by:<br>Vision-Sciences, Inc.<br>9 Strathmore Road<br>Natick, MA 01760<br><br>K964815 - AP-4000 Air Pulse Sensory Stimulator<br>Manufactured by:<br>Pentax Precision Instrument Corp.<br>3117 Commerce Parkway<br>Miramar, FL 33025 |
Product Description: The device system described in this 510(k) consists of a sterile, single use protective sheath for use with the VSI ENT-2000 scope.
Indications for Use: The EndoSheath® System provides a sterile, disposable protective covering for the scope to be used during flexible endoscopic examination of the upper airway, vocal cords and/or nasal passages. The System may also be used in conjunction with the Pentax AP-4000 Air Pulse Sensory Stimulator to elicit Laryngeal Closure Reflex (Swallow) and to measure the sensory discrimination threshold at which the reflex occurs in the area of the Upper Airway innervated by the Superior Laryngeal Nerve.
Safety and Performance: Substantial equivalence for the new device was based on design characteristics, comparison to legally marketed predicate devices, and performance testing. Performance testing included sheath burst/leak testing, sheath tensile/elongation testing, sheathed scope articulation testing and air pulse testing.
Conclusion: Based on the indications for use, technological characteristics, performance testing and comparison to predicate devices, the proposed VSI Slide-On™ EndoSheath® System for Sensory Testing has been shown to be safe and effective for its intended use.
# 000084
{1}------------------------------------------------
Vision-Sciences, Inc.
June 9, 2003
# /SI Trans-Nasal Esophagoscope with EndoSheath® System
Substantial Equivalence Comparison
| Characteristic | Proposed VSI Slide-OnTM<br>EndoSheath® System for Sensory<br>Testing<br>(Current Submission) | Currently Marketed VSI ENT<br>2000 with EndoSheath® System<br>(K990354, K012534-K024095) | Pentax ENL-Series ENT Scopes for use<br>with AP-4000 Air Pulse Sensory<br>Stimulator<br>(K964815) |
|-----------------------------------------|----------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Sheath material | Same as VSI predicate devices | Thermoplastic elastomer | N/A - No sheath |
| Window material | Same as VSI predicate devices | Thermoplastic polymer | N/A - No sheath |
| Luer connector material | Same as VSI predicate devices | N/A - no luer connector | N/A - No sheath |
| Proximal connector<br>tubing material | Same as VSI predicate devices | Thermoplastic polymer | N/A - No sheath |
| Air channel ID | N/A - no working channel | N/A - no working channel | 1.2 mm |
| Adhesives | Same as VSI predicate devices | UV curable | Unknown |
| Microbial barrier claim | Yes | Yes | N/A - No sheath |
| Sheath installation<br>method | Slides on and off (no vacuum/pressure<br>source required) | Slides on and off (no<br>vacuum/pressure source required) | N/A - No sheath |
| Sheath length | 12" | 12" | N/A - No sheath |
| Minimum sheath wall<br>thickness | .002" | .002" | N/A - No sheath |
| Sheath Packaging | Tyvek/Mylar pouch | Tyvek/Mylar pouch | N/A - No sheath |
| Scope working length<br>(with sheath) | 300 mm | 300 mm | 300 mm |
| Scope insertion tube OD<br>(distal tip) | 4.1 mm (w/sheath) | 3.6 mm | 3.4 mm |
| Articulation (Up/Down) | 90°/90° (sheathed scope) | 135°/135° (sheathed scope) | 130°/130° |
| Angle of View | 75° | 75° | 75° |
| Depth of Field | 3 - 50 mm | 3 - 50 mm | Unknown |
{2}------------------------------------------------
Vision-Sciences, Inc.
June 9, 2003
| Proposed VSI Slide-On™<br>EndoSheath® System for Sensory<br>Testing<br>(Current Submission) | Currently Marketed VSI ENT<br>2000 with EndoSheath® System<br>(K990354, K012534, K024095) | Pentax FNL-Series ENT Scopes for use<br>with AP-4000 Air Pulse Sensory<br>Stimulator<br>(K964815) |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The EndoSheath® System provides a sterile,<br>disposable protective covering for the scope<br>to be used during flexible endoscopic<br>examination of the upper airway, vocal<br>cords and/or nasal passages. The System<br>may also be used in conjunction with the<br>Pentax AP-4000 Air Pulse Sensory<br>Stimulator to elicit Laryngeal Closure<br>Reflex (Swallow) and to measure the<br>sensory discrimination threshold at which<br>the reflex occurs in the area of the Upper<br>Airway innervated by the Superior<br>Laryngeal Nerve. | The EndoSheath® System<br>provides a sterile, disposable<br>protective covering for the scope<br>to be used during flexible<br>endoscopic examination of the<br>upper airway, vocal cords and/or<br>nasal passages. | To elicit Laryngeal Closure Reflex (Swallow)<br>and to measure the sensory discrimination<br>threshold at which the reflex occurs in the<br>area of the Upper Airway innervated by the<br>Superior laryngeal Nerve. The structures<br>being stimulated in the area of the Upper<br>Airway innervated by the Superior Laryngeal<br>Nerve are: the Left and Right Anterior Wall<br>of the Pyriform Sinus and the Left and Right<br>Aryepiglottic Folds. The device is intended<br>to be used with a legally marketed endoscope<br>compatible with the AP-4000, introduced per<br>nasally in Adult and Pediatric patient<br>populations with suspected Dysphagia |
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, topped by a circle. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular fashion around the caduceus symbol.
2 2003 JUL
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Vision-Sciences, Inc. c/o Pamela Papineau Delphi Medical Device Consulting, Inc. 5 Whitcomb Ave. Ayer, MA 01432
Re: K031790
Trade/Device Name: Slide-On101 EndoSheath System® for Sensory Testing Regulation Number: 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (flexible or rigid) and accessories Regulatory Class: Class II Product Code: EOB Dated: June 9, 2003 Received: June 10, 2003
Dear Ms. Papineau:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
## Page 2 - Pamela Papineau
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A. Paigy Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
Page
510(k) Number (if known): K031790
Slide-On™ EndoSheath® System for Sensory Testing Device Name:
Indications for Use:
The EndoSheath® System provides a sterile, disposable protective covering for the scope to be used during flexible endoscopic examination of the upper airway, vocal cords and/or nasal passages. The System may also be used in conjunction with the Pentax AP-4000 Air Pulse Sensory Stimulator to elicit Laryngeal Closure Reflex (Swallow) and to measure the sensory discrimination threshold at which the reflex occurs in the area of the Upper Airway innervated by the Superior Laryngeal Nerve.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
1 Prescription Use (Per 21 CFR 801.109)
OR
Over-the -Counter Use
C00007
Kunt Horken
Division Sign-Off
(Division Sign-Off)
Division of Ophthalmic Ear,
Nose and Throat Devises
510(k) Number __
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.