To be utilized through a flexible endoscope or the superDimension system by physicians who are trained in endoscopic techniques for retrieving specimens from patients with endobronchial lesions, peripheral lung nodules, or lung masses.
Device Story
Endoscopic catheter for tissue biopsy; used in endobronchial lesions, peripheral lung nodules, or lung masses. Device consists of outer ETFE sheath and inner assembly with three needle-tipped brushes at distal end. Physician inserts catheter through bronchoscope or superDimension Extended Working Channel; brushes remain retracted during insertion. Upon reaching target, physician advances proximal thumb ring to extend brushes into tissue; sharpened needle-tips abrade tissue to collect samples. Once sample obtained, brushes retracted into sheath and catheter withdrawn. Used by trained physicians in clinical settings. Benefits include improved tissue collection via multi-brush design compared to single-brush alternatives.
Clinical Evidence
No clinical data. Bench testing included radiographic, tensile, dimensional, simulated use, trackability, shelf life (ASTM F1980-07, F2096-11, F88-09), distribution (ASTM D4169-09), sterilization (ISO 11135-1), and biocompatibility (ISO 10993-1, -5, -7, -10). In vivo porcine model testing confirmed safety and tissue collection capability.
Technological Characteristics
Catheter with ETFE sheath; inner assembly with thumb ring and twisted wire shaft. Distal end features three needle-tipped brushes (10mm or 15mm). Mechanical actuation via proximal thumb ring. Sterilization per ISO 11135-1. Biocompatibility per ISO 10993 series.
Indications for Use
Indicated for patients with endobronchial lesions, peripheral lung nodules, or lung masses requiring tissue specimen retrieval via flexible endoscope or superDimension system.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "COVIDIEN" in bold, sans-serif font. To the left of the word is a black square with a white cross in the center. The cross is formed by two perpendicular white rectangles, creating a medical symbol.
# 510(k) Summary K130357 Covidien IIc, dba superDimension Inc. SuperDimension® Triple Needle Cytology Brush
| Date Prepared: | 11/4/2013 | |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------|--------------|
| 510(k) Applicant: | Covidien llc, formerly registered as superDimension Inc.<br>161 Cheshire Lane Suite 100<br>Plymouth, MN 55441 U.S.A. | |
| Contract Person: | Deborah Fleetham , Manager Regulatory Affairs<br>Phone: 763-210-4091<br>Fax: 763-210-4098<br>Email : deborah.fleetham@covidien.com | NOV 0 6 2013 |
| Name of Device : | | |
| Trade Name : | superDimension® Triple Needle Cytology Brush | |
| Common Name: | Bronchial Biopsy Brush | |
| Classification Name: | Bronchoscope (flexible or rigid) and accessories<br>21 CFR Part 874.4680 | |
| Product code: | BTG | |
#### Equivalent Legally-Marketed Devices:
K834402 Gastrointestinal Sheath Brush (KOG) by Hobbs Medical, Inc. K944650 Wang Bronchial Needle Brush (EOQ) by ConMed
#### Description:
The superDimension® Triple Needle Cytology Brush is designed to obtain tissue samples for biopsy from endobronchial lesions, peripheral lung nodules, or lung masses. The superDimension® Triple Needle Cytology Brush is an endoscopic catheter comprised of an outer Ethylene Tetrafluoroethylene (ETFE) sheathing and an inner catheter assembly. The inner catheter assembly consists of a thumb ring at the proximal end and a twisted wire shaft to connect to the distal end. The distal end terminates in three connected brushes available in two lengths: 10 mm and 15mm. The brushes have sharpened ends, referred to as needle-tipped, that can be used to rough up tissue to obtain a sample of tissue or cells. When the catheter is inserted into a channel such as a bronchoscope or superDimension Extended Working Channel (EWC) with the distal brush in a retracted position inside the outer sheath. When the catheter is in position, the brushes can be extended into the tissue to obtain tissue samples by advancing the proximal thumb ring. When the physician believes that an adequate sample has been taken, the brushes are retracted back into the sheath and then the entire catheter is withdrawn from the channel The superDimension® Triple Needle Cytology Brush is for standard tissue analysis. similar to currently marketed cytology brushes except that it has three smaller brushes in place of one larger brush.
{1}------------------------------------------------
# Intended Use:
To be utilized through a flexible endoscope or the superDimension system by physicians who are trained in endoscopic techniques for retrieving specimens from patients with endobronchial lesions, peripheral lung nodules, or lung masses.
# Summary of Substantial Equivalence:
The superTrax Triple Needle-Tipped Cytology brush is substantially equivalent to the ConMed Needle-Tipped Cytology Brush and the Hobbs Cytology Brush predicate devices in the following attributes:
- Indications for Use .
- Mechanism of action .
- Length .
- . Sterilization technique
- Design .
- Size .
- Materials .
- Packaging
- Performance Characteristics 0
The difference between the predicate devices and the superDimension Triple Needle Cytology Brush is that the predicate devices have one, larger brush on the distal end compared to three smaller, flexible brushes on the distal end of the superDimension Triple Needle Cytology Brush.
# Performance Data:
In-vitro and in-vivo testing has been performed on all components, subassemblies, and /or full devices. The results showed that the device met the required specifications for the completed tests and performed similarly to the predicate devices. Testing included the following:
- In Vitro Testing .
- Radiographic Testing o
- Catheter Tensile Testing o
- Dimensional Testing o
- 0 Simulated Use Testing
- Trackability Testing O
- Shelf Life Testing per ASTM F1980-07, ASTM F2096-11, and ASTM F88-09 0
- Distribution Testing per ASTM D4169-09 o
- Sterilization Testing per ISO 11135-1 o
- 0 Biocompatibility Testing (cytotoxicity, irritation, sensitization) per ISO 10993-1. ISO 10993-5, ISO 10993-7, ISO 10993-10
- In Vivo Testing in a porcine model
- o Tissue Collection
- o Safety Testing
#### Clinical Data:
.
Clinical tests were not required to validate the design of the SuperDimension® Triple Needle Cytology Brush due to the extensive history of similar devices.
{2}------------------------------------------------
# Conclusion :
Based on the intended use, technological characteristics, and results from safety and performance testing, the superDimension Triple Needle Cytology Brush is substantially. equivalent to the legally marketed predicate devices, Gastrointestinal Sheath Brush (KOG) by Hobbs Medical, Inc. K834402, and the Wang Bronchial Needle Brush (EOQ) by ConMed K944650.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
November 6, 2013
Covidien Llc c/o Ms. Deborah Fleetham Manager, Regulatory Affairs 161 Cheshire Lane, Suite 100 Minneapolis, MN 55441
Re: K130357
Trade/Device Name: Superdimension Triple-Needle Cytology Brush Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible or Rigid) and Accessories Regulatory Class: Class II Product Code: BTG Dated: September 27, 2013 Received: September 30, 2013
Dear Ms. Fleetham:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class 11 (Special Controls) or class 111 (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Deborah Fleetham
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/Medica|Devices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K130357
Device Name: superDimension® Triple Needle Cytology Brush
Indications For Use:
To be utilized through a flexible endoscope or the superDimension system by physicians who are trained in endoscopic techniques for retrieving specimens from patients with endobronchial lesions, peripheral lung nodules, or lung masses.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
Sunny Park
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.