The Reza Band® Upper Esophageal Sphincter (UES) Assist Device is a prescription device under 21 CFR Part 801.109 that is indicated for patients 18 years and older to reduce the symptoms of laryngopharyngeal reflux (LPR) disease by reducing the regurgitation of stomach contents from passing through the upper esophageal sphincter. The device is worn by the patient when sleeping.
Device Story
Non-invasive, wearable device applies external pressure (20-30 mmHg) to cricoid cartilage to increase luminal pressure in upper esophageal sphincter (UES). Comprises frame, cushion, comfort band, clasp, and comfort dial. Physician fits device using external manometer connected to single-use pressure sensor; sensor uses air-filled tubing to transducer to display pressure in mmHg. Patient wears device during sleep; removes upon waking. Device provides mechanical resistance to reflux; does not treat underlying disease. Benefits include symptom reduction; risks include skin irritation, discomfort, or potential displacement. If displaced, pressure on neck structures is significantly reduced.
Clinical Evidence
Two clinical studies supported safety and effectiveness. Study 1: Prospective, open-label, 95-patient study (mean age 48.8). Primary endpoint: % reduction in Reflux Symptom Index (RSI) at 4 weeks. RSI scores improved significantly (p<0.0001) from baseline (mean 26.0) to 4 weeks (mean 12.0). 60.7% of subjects achieved RSI <13. Adverse events were mild; 6 subjects discontinued due to discomfort. Study 2: 20-subject safety study evaluated intentional lateral displacement; no significant changes in heart rate, blood pressure, cardiac rhythm, or intraocular pressure compared to baseline.
Technological Characteristics
Non-sterile, reusable frame (semi-flexible) with cushion, comfort band, magnetic clasp, and comfort dial. Pressure sensor (miniature transducer) and external manometer (AAA battery-powered) used for fitting. Materials biocompatible per ISO 10993-1. Device provides 20-30 mmHg pressure. Reusable components intended for 6-month life; pressure sensor is single-use. Mechanical integrity verified via FEA and tensile/fatigue testing.
Indications for Use
Indicated for patients 18+ to reduce LPR symptoms by reducing regurgitation through the upper esophageal sphincter. Contraindicated in patients with neck implants, pacemakers, ICDs, vagus nerve stimulators, glaucoma, neck malignancy/surgery, altered mental status, carotid artery disease, thyroid disease, cerebrovascular disease, or connective tissue disorders (e.g., Marfan's, Ehlers-Danlos). Safety not established for pregnancy, breastfeeding, or CPAP users.
Regulatory Classification
Identification
An external upper esophageal sphincter compression device is a prescription device used to apply external pressure on the cricoid cartilage for the purpose of reducing the symptoms of laryngopharyngeal reflux disease.
Special Controls
In combination with the general controls of the FD&C Act, the External UES Compression Device is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) The patient contacting components must be demonstrated to be biocompatible.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be demonstrated:
(i) Mechanical integrity testing (
*e.g.,* tensile strength testing, fatigue testing) and(ii) Shelf life testing.
(3) The technical specifications must include pressure measurement accuracy to characterize device performance.
(4) Clinical performance testing must document any adverse events observed during clinical use, and demonstrate that the device performs as intended under anticipated conditions of use.
(5) Labeling must include the following:
(i) Appropriate warnings and precautions,
(ii) A detailed summary of the clinical testing pertinent to use of the device including a detailed summary of the device-related complications or adverse events,
(iii) Detailed instructions on how to fit the device to the patient, and
(iv) Instructions for reprocessing of any reusable components.
(6) Patient labeling must be provided and must include:
(i) Relevant warnings, precautions, and adverse effects/complications,
(ii) Information on how to correctly wear the device,
(iii) The potential risks and benefits associated with the use of the device,
(iv) Alternative treatments, and
(v) Reprocessing instructions.
Submission Summary (Full Text)
{0}------------------------------------------------
#### DE NOVO CLASSIFICATION REQUEST FOR REZA BAND® UPPER ESOPHAGEAL SPHINCTER (UES) ASSIST DEVICE
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
External Upper Esophageal Sphincter (UES) Compression Device: An external upper esophageal sphincter (UES) compression device is a prescription device used to apply external pressure on the cricoid cartilage for the purpose of reducing the symptoms of laryngopharyngeal reflux (LPR) disease.
#### NEW REGULATION NUMBER: 874.5900
CLASSIFICATION: CLASS II
PRODUCT CODE: PKA
### BACKGROUND
DEVICE NAME: REZA BAND® UPPER ESOPHAGEAL SPHINCTER (UES) ASSIST DEVICE
SUBMISSION NUMBER: DEN130046
DATE OF DE NOVO: NOVEMBER 27, 2013
### CONTACT: SOMNA THERAPEUTICS, LLC W175 N11081 Stonewood Drive GERMANTOWN, WI 53022
### REQUESTER'S RECOMMENDED CLASSIFICATION: CLASS I, MODIFIED TO CLASS II
#### INDICATIONS FOR USE
The Reza Band® UES Assist Device is indicated for patients 18 years and older to reduce the symptoms of larvngopharyngeal reflux (LPR) disease by reducing the regurgitation of stomach contents from passing through the upper esophageal sphincter. The device is worn by the patient when sleeping.
### LIMITATIONS
Prescription Use only: Federal (USA) law restricts this device to sale by or on the order of a physician.
Limitations on device use are included in the Instructions for Use as Warnings and Precautions.
{1}------------------------------------------------
## Warnings
- The safety and effectiveness has not been demonstrated for the following conditions:
- Patient with implants or implant parts that reside in the area where the Reza o Band is applied.
- Patient with an implanted pacemaker, implanted cardioverter defibrillator o (ICD), vagus nerve stimulator, or other such similar devices implanted in the neck.
- Patient diagnosed with glaucoma. O
- Patient had a malignancy of the neck, including neck surgery. O
- Patients that may have an altered mental status including due to the use of O sedative drugs or narcotics.
- Patients with carotid artery disease, thyroid disease, a history of o cerebrovascular disease, or any disorder of connective tissues (e.g., Marfan's Syndrome or Ehlhers-Danlos Syndrome).
- The Reza Band is packaged and designed for single patient use only.
- The Pressure Sensor is packaged and designed for a one-time use only.
## Precautions
- The Reza Band, the External Manometer or the Pressure Sensor should only be ● used by health care providers trained to prescribe and apply the device.
- The safety of the Reza Band for use during pregnancy, in breastfeeding females or in patients under 18 years of age has not been established.
- The safety of the Reza Band® UES Assist Device in patients currently receiving ● treatment with continuous positive airway pressure (CPAP) has not been established.
- If the patient feels unusual pain, skin irritation or the worsening of symptoms, they should discontinue the use of the Reza Band and contact their health care provider.
- . If the Reza Band does not stay in place, the patient should stop using the Reza Band and contact their health care provider. The patient may need to be re-fitted for the Reza Band.
- The External Manometer is not designed for liquid immersion. The External . Manometer should be cleaned with a damp cloth using warm water or presaturated isopropyl alcohol wipes.
## DEVICE DESCRIPTION
# Reza Band® UES Assist Device
The Reza Band® UES Assist Device (Reza Band) (Figure 1) is a non-invasive, non-sterile device worn by the patient and is designed to provide a set pressure (20-30 mm Hg) on the cricoid cartilage, which increases the luminal pressure within the upper esophageal sphincter (UES). The patient wears the Reza Band after being fit by the physician (Figure 2).
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a headband with a dark blue band and a white plastic frame. The headband appears to be designed for comfort and stability, with the soft band providing cushioning against the head. The white plastic frame likely provides structure and support, possibly housing electronic components or sensors. The headband is likely used for sports, fitness, or medical purposes.
Figure 1: Reza Band® UES Assist Device
Figure 2: Fitting the Reza Band® UES Assist Device to the Patient
The Reza Band has 5 main components:
- Frame Is semi-flexible so that it can apply the prescribed pressure as determined by the ● physician, is comfortable for the patient, and allows limited pressure adjustments by the patient. The Cushion is attached to the Frame. The magnet in the Frame attaches to the Clasp.
- . Cushion - Positioned over the cricoid cartilage and designed to conform to the anatomical structures when being worn.
- Comfort Band - Holds the Reza Band in place.
- Clasp - Attaches to the Reza Band Frame by the use of a magnet. This allows for the patient to easily and quickly take the device on and off during routine use and in emergency situations if needed.
- . Comfort Dial - Allows the patient a limited adjustment to the applied pressure.
## Reza Band® Pressure Sensor and the Reza Band® External Manometer
The Reza Band® External Manometer (External Manometer) is a hand-held device and is connected to the Reza Band® Pressure Sensor (Pressure Sensor). It is used by the treating physician, is powered by a AAA battery and displays the pressure being applied to the cricoid cartilage region by the Reza Band in millimeters of Mercury (mm Hg) as the Reza Band® is being fitted (Figure 3).
The Pressure Sensor connected to the External Manometer is regulated as a miniature pressure transducer (21 CFR Part 890.1615) which is Class I exempt. The Pressure Sensor and the
{3}------------------------------------------------
External Manometer were assessed to ensure that the Reza Band® was accurately applying the specified pressure.
Image /page/3/Picture/1 description: The image shows a medical device with a pressure sensor assembly and an external manometer. The pressure sensor assembly is a blue, bulbous component connected to a clear tube. The tube leads to a handheld external manometer, which has a digital display showing "0.0" and a blue button.
Figure 3: Reza Band® Pressure Sensor Connected to the Reza Band® External Manometer
- . External Manometer – Contains the "On" button and displays the pressure being applied by the Reza Band in mmHg.
- Pressure Sensor – A non-sterile, one-time-use component that is connected to the External Manometer with a luer lock connector. It is positioned between the Reza Band and the patient, at the time of the fitting by the physician. When the Pressure Sensor is compressed during the fitting of the Reza Band, air is forced down the tubing to an internal pressure sensor. This, in turn, provides the input for the display.
### SUMMARY OF NONCLINICAL/BENCH STUDIES
Non-clinical/bench studies conducted on the Reza Band® UES Assist Device to demonstrate a reasonable assurance of safety and effectiveness of the device are summarized in the sections below.
### BIOCOMPATIBILITY/MATERIALS
The Reza Band contacts the patient's intact skin surfaces. The Reza Band is categorized as a permanent contact device as the device is intended to be used more than 30 days. In accordance with ISO 10993-1: Biological Evaluation of Medical Devices, Part 1: Evaluation and Testing within a Risk Management Process, the following biocompatibility testing was conducted on the Reza Band. The biocompatibility assessment was deemed adequate.
Table 1: Reza Band® UES Assist Device Biocompatibility Testing
| Company of Concession Comparis and Comments of Comments of Concession Comer College of Concession Come of Children Come of Children Come of Children Comer Collection Comer Co | C month a mallently and a long of comments of the finance of the finance of the final | | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|-------------------|--|--|
| C<br>ﺎﺭﮦ ﮐﺎ ﺭﺑﻂ ﻭﯾﮑﯽ ﮈﯾﭩﺎ ﺳﮯ ﺑﻠﻨﺪﯼ ﭘﺮ ﻭﺍﻗﻊ ﮨﮯ۔ | Purpose | 1997 3<br>TAUDULU | | |
{4}------------------------------------------------
| Cytotoxicity Evaluation –<br>Fluid Extract/L929 Mouse<br>Fibroblast | To assess the biological activity of L-929<br>mouse fibroblast cells (grown in culture) after<br>exposure to extracts prepared from the<br>completed Reza Band device | Non-cytotoxic |
|-------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|
| Primary Skin Irritation<br>with 0.9% NaCl &<br>Cottonseed Oil Extracts | To assess the irritating potential of extracts of<br>the Reza Band to produce dermal irritation | Negligible |
| Dermal Sensitization<br>Maximization with 0.9%<br>NaCl & Cottonseed Oil | To estimate the potential for sensitization of<br>the Reza Band extract using the guinea pig as<br>the animal model | Non-Sensitizer |
## SHELF LIFE/STERILITY
Reza Band® UES Assist Device is provided non-sterile. The Reza Band and External Manometer components are reusable and intended to be used for up to 6 months. The Pressure Sensor component is for single use. The sponsor provided reprocessing instructions (i.e., cleaning with mild soap and warm water) for the Reza Band and External Manometer for the patient and health care facility users, respectively. The information provided is adequate.
## PERFORMANCE TESTING - BENCH
The sponsor conducted finite element analysis (FEA) of the structural behavior to support the design of the Reza Band UES Assist Device. Bench testing was also conducted to demonstrate the mechanical performance of the Reza Band for the intended conditions of use, and compatibility of use with the External Manometer, and Pressure Sensor. The functionality of the devices was tested over a simulated 6-month shelf life period. All bench testing was conducted on the final, finished devices that met established specifications. Table 2 summarizes each of the bench tests.
| Test | Purpose | Method | Results |
|-----------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Reza Band Neck<br>Ring Lifetime<br>Test | To demonstrate<br>that the Reza<br>Band retains its<br>intended<br>functionality over<br>its intended 6-<br>month use period | (b)(4) | All the established<br>acceptance criteria were<br>met. |
| | | functionality, in accordance<br>with established<br>specifications. | |
| Reza Band<br>Mechanical Test | To demonstrate<br>the mechanical<br>integrity of three<br>key components<br>of the Reza Band<br>for their<br>intended use:<br>tensile strength<br>of the Comfort<br>Band;<br>functionality and<br>attachment of<br>the frame pusher<br>after exposure to<br>pull testing; and<br>attachment of<br>the Comfort<br>Band attachment<br>portions of the<br>frame body after<br>pull testing | Comfort Band Evaluation:<br>The Comfort Band was<br>(b)(4)<br>the<br>length was (b)(4)<br>testing | The length of the Comfort<br>Band was not significantly<br>different after being pulled<br>by the (b)(4)<br>when comparing the<br>length before and after the<br>testing. |
| | | Frame Pusher Evaluation:<br>The Frame Pusher was<br>(b)(4)<br>the<br>Comfort Dial actuation. | The Frame Pusher did not<br>exhibit any damage or<br>failures and the Comfort<br>Dial continued to function<br>as intended (b)(4)<br>was applied<br>to the Frame Pusher. |
| | | Frame Body Evaluation:<br>The Frame Body was<br>applied (b)(4)<br>visually<br>examined for evidence of<br>breakage, damage or<br>deformation. | The Frame Body did not<br>fail and there was no<br>damage noted, after being<br>subjected (b)(4)<br>of force. |
| Reza Band<br>Accelerated<br>Aging Shelf Life<br>Testing | To demonstrate<br>that the Reza<br>Band retains its<br>intended<br>functionality over<br>a simulated 6-<br>month shelf life<br>(storage) period | Devices were subjected to<br>(b)(4)<br>and<br>evaluated for continued<br>functionality and<br>conformance with<br>specifications. | With the exception of the<br>observations regarding the<br>Frame Body magnets<br>following exposure to<br>accelerated aging<br>conditions (b)(4)<br>acceptance criteria were<br>met.<br>A rationale (b)(4) |
| Reza Band,<br>External<br>Manometer and<br>Pressure Sensor<br>Shipping Test | To demonstrate<br>that the packaged<br>Reza Band,<br>External<br>Manometer, and<br>Pressure Sensor<br>withstand<br>damage from | Devices were subjected to<br>simulated shipping<br>conditions including<br>temperature, humidity<br>conditions, compression,<br>vibration, shock, and repeat<br>vibration in accordance with | The Reza Band, External<br>Manometer and Pressure<br>Sensor demonstrated<br>continued functionality and<br>conformance with<br>specifications. |
| | and retain their<br>intended<br>functionality<br>after, simulated<br>shipping<br>conditions | (b)(4) | |
| External<br>Manometer<br>Pressure<br>Measurement<br>Accuracy Testing | To demonstrate<br>that the pressure<br>measurement<br>accuracy of the<br>External<br>Manometer<br>meets design<br>specifications | Known pressures (b)(4)<br>were<br>measured by the External<br>Manometer. | The pressure measurement<br>accuracy of the External<br>Manometer is within (b)(4)<br>mmHg as specified over<br>the (b)(4) |
| External<br>Manometer Life<br>Test | To demonstrate<br>that the External<br>Manometer is<br>capable of<br>maintaining<br>functionality<br>over its intended<br>600-use life time | Pressure measurement<br>accuracy was evaluated at<br>baseline (b)(4)<br>with<br>pressure accuracy assessed<br>at from (b)(4) mmHg. | All established acceptance<br>criteria were met following<br>(b)(4) (in excess of the<br>600-use requirement) |
Table 2. Reza Band® UES Assist Device Bench Testing
{5}------------------------------------------------
{6}------------------------------------------------
## SUMMARY OF CLINICAL INFORMATION
The following is a summary of two clinical studies performed by the sponsor to support a reasonable assurance of safety and effectiveness for the Reza Band® UES Assist Device.
## Study #1: Multi-Center, Non-Randomized, Prospective Study of the Reza Band® Upper Esophageal Sphincter (UES) Assist Device for the Treatment of Esophagopharyngeal Reflux
The sponsor conducted a non-randomized, prospective, open label study of 95 patients treated at 5 investigational sites with the Reza Band. The study enrolled subjects that had been clinically diagnosed with laryngopharyngeal reflux (LPR) (i.e., chronic cough, choking, aspiration, chronic post nasal drip, globus, sore throat, throat clearing) and who met the inclusion and exclusion criteria.
The inclusion criteria were:
- . 18 years of age or older
- . The patient must be willing and able to provide informed consent.
- Understands the clinical study requirements and is able to comply with follow-up schedule. .
- . Clinically diagnosed with esophagopharyngeal reflux with extra-esophageal symptoms (i.e., chronic cough, choking, aspiration, chronic post nasal drip, globus, sore throat, throat clearing) for 6 months or longer.
- . Reflux Symptom Index (RSI) >13
{7}------------------------------------------------
The exclusion criteria were:
- Currently being treated with another investigational medical device and/or drug
- . Currently receiving treatment for sleep apnea with continuous positive airway pressure (CPAP)
- The patient is female and is of child bearing potential and is not using an acceptable method of birth control, or is pregnant or breast feeding.
- Previous head or neck surgery or radiation
- Carotid artery disease, thyroid disease, or history of cerebral vascular disease
- Suspected esophageal cancer
- Nasopharyngeal cancer
- Has either a pacemaker or implanted cardioverter defibrillator (ICD)
- Previously undergone Nissen Fundoplication
The study population consisted of 30.5% males and 69.5% females, (81.1% Caucasian, 8.4% Hispanic, 4.2% African-American, 1.1% Asian, 1.1% other subjects, and 4.2% not reported), with a mean age of 48.8 years, a mean body weight of 160.0 pounds, and a mean Body Mass Index (BMI) of 25.5.
The objective of the study was to assess the safety and effectiveness of the Reza Band when worn by subjects that had been clinically diagnosed with LPR (i.e., chronic cough, choking, aspiration, chronic post nasal drip, globus, sore throat, throat clearing). The subjects applied the Reza Band at bedtime and took it off upon waking.
The safety of the Reza Band was evaluated by assessing the incidence, type, duration and severity of adverse events observed in all patients.
The primary effectiveness endpoint was the percent reduction in the Reflux Symptom Index (RSI) from Baseline to Visit 3 (Week 4). The RSI is a validated nine-item, patient-administered outcome questionnaire designed to document the symptoms and severity of LPR, Patients were instructed to rate how the nine symptoms affected them on a scale of 0 (no problem) to 5 (severe problem), with a maximum total score of 45. Effectiveness was also assessed by the evaluation of patient and physician satisfaction, patient response to the RSI, the SF-36° Health Survey, the Functional Outcomes of Sleep Ouestionnaire and patient diaries.
## Safety Results:
The adverse events that were reported during the study were generally mild, short in duration and the majority of those events were not related to the device. Those events that were reported as related to the device were also generally mild and short in duration. Device-related adverse events did not result in reduced outcomes as assessed by the change of the RSI score from baseline to Visit 3 (Week 4), as they were consistent with the overall population. There were no deaths or unexpected adverse events in the study. Six of ninety five (6/95) subjects discontinued use of the device due to reported symptoms of discomfort, and/or difficulty sleeping with the device in place, and therefore did not complete the study (categorized under Reza Band Problem in table below).
{8}------------------------------------------------
Table 3 below summarizes the incidence of adverse events in all subjects (including the 6 who did not complete the study) along with the reported relationship of the adverse event to the device.
| | Definitely<br>(n/%) | Possibly | Probably Not | Definitely Not |
|----------------------------|---------------------|----------|--------------|----------------|
| Dysphagia | 0 (0.0%) | 0 (0.0%) | 1 (1.1%) | 0 (0.0%) |
| Odynophagia | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) |
| Pain | 5 (5.3%) | 2 (2.1%) | 1 (1.1%) | 0 (0.0%) |
| Globus | 1 (1.1%) | 0 (0.0%) | 1 (1.1%) | 5 (5.3%) |
| Laryngospasm | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 1 (1.1%) |
| Regurgitation | 0 (0.0%) | 0 (0.0%) | 1 (1.1%) | 2 (2.1%) |
| Choking | 5 (5.3%) | 2 (2.1%) | 1 (1.1%) | 2 (2.1%) |
| Nausea | 1 (1.1%) | 0 (0.0%) | 2 (2.1%) | 0 (0.0%) |
| Dyspepsia | 0 (0.0%) | 1 (1.1%) | 1 (1.1%) | 1 (1.1%) |
| Hoarseness | 0 (0.0%) | 2 (2.1%) | 3 (3.2%) | 3 (3.2%) |
| Cough | 0 (0.0%) | 2 (2.1%) | 2 (2.1%) | 7 (7.4%) |
| Difficulty Breathing | 0 (0.0%) | 1 (1.1%) | 3 (3.2%) | 5 (5.3%) |
| Unable to Belch | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) |
| Unable to Vomit | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) |
| Skin Reaction | 5 (5.3%) | 1 (1.1%) | 2 (2.1%) | 1 (1.1%) |
| Reza Band Problem | 14 (14.7%) | 0 (0.0%) | 0 (0.0%) | 1 (1.1%) |
| External Manometer Problem | 1 (1.1%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) |
Table 3. Adverse events relationship to device through 1 month follow-up (N = 95)
Effectiveness Results:
The Primary Effectiveness endpoint was defined as the percent change in the RSI from Baseline (Visit 1) to End of Study (4 weeks). Table 4 below provides the percent change in the RSI from the baseline for Visit 2 (2 weeks) and Visit 3 (4 weeks) for the 89 of 95 (93.7%) subjects that wore the Reza Band® for at least 2 weeks and provided one post-baseline RSI assessment. The six (6) subjects that discontinued use of device were excluded from the analysis as they did not return for Visit 2 (2 week) or Visit 3 (4 week). A worst-case sensitivity analysis demonstrated that the exclusion of the 6 subjects who did not return has no impact on the treatment effect or the outcome of the study.
The change in the RSI from baseline to each post treatment visit, as well as any post-treatment visit, were found to be statistically improved for both Visit 2 and Visit 3 (p<0.0001).
Table 4. Reflux Symptom Index (RSI) percent change in RSI from Visit 1 (N = 89)
{9}------------------------------------------------
| Time Point | Statistic | RSI | % Change from<br>Visit 1 - Baseline |
|--------------------------|--------------------|------|-------------------------------------|
| Visit 1 | n | 89 | |
| | Mean | 26.0 | |
| | Standard Deviation | 6.8 | N/A |
| | Median | 25.0 | |
| | Minimum | 14.0 | |
| | Maximum | 42.0 | |
| Visit 2 | n | 86 | 86 |
| | Mean | 14.6 | -43.8 |
| | Standard Deviation | 9.2 | 31.1 |
| | Median | 12.5 | -50.0 |
| | Minimum | 0.0 | -100.0 |
| | Maximum | 41.0 | 35.7 |
| | p-Value (1) | | <0.0001 |
| | p-Value (2) | | <0.0001 |
| Visit 3 | n | 86 | 86 |
| | Mean | 12.0 | -54.3 |
| | Standard Deviation | 8.6 | -27.1 |
| | Median | 10.5 | -57.8 |
| | Minimum | 0.0 | -100.0 |
| | Maximum | 42.0 | 7.7 |
| | p-Value (1) | | <0.0001 |
| | p-Value (2) | | <0.0001 |
| Last Post-Baseline Visit | n | 89 | 89 |
| | Mean | 12.1 | -54.2 |
| | Standard Deviation | 8.6 | 27.5 |
| | Median | 10.0 | -57.9 |
| | Minimum | 0.0 | -100.0 |
| | Maximum | 42.0 | 7.7 |
| | p-Value (1) | | <0.0001 |
| | p-Value (2) | | <0.0001 |
(1) p-Value based on paired t-test to test hypothesis of no mean percent change.
(2) p-value based on one-sided one sample t-test to test hypothesis of mean change >25%.
All study subjects had an RSI score of greater than 13 at baseline. In the literature, an RSI score of greater than 13 suggests the presence of LPR . In this study, 54 out of 89 subjects (60.7%) reported that their RSI score was below 13 at the last post baseline visit.
1 Belafsky PC, Postma GN, Koufman JA. Validity and reliability of the reflux symptom index (RSI). J Voice. 2002 Jun;16(2):274-7.
Of the 95 subjects that were enrolled in the study, 79 (83.2%) reported taking acid reducing medications with 16 subjects (16.8%) not taking acid reducing medications. Seventy-five of the 89 subjects (84.3%) that completed the study reported taking acid reducing medications at the time of enrollment. Of these 75 subjects, 61 (81.3%) continued taking acid reducing medication during the study and 14 (18.7%) discontinued taking acid reducing medications during the study.
{10}------------------------------------------------
The analysis in Table 5 below shows that the percent change in RSI score from baseline is no different whether acid reducing medications were used during the study. The study was not intended to show nor did the analysis of the data show that the Reza Band® provides a significant reduction in use of acid reducing medication.
| | Enrollment | | 2 Week | | 4 Week | | 4 Week<br>% Change | |
|---------|------------|---------|--------|---------|--------|---------|--------------------|---------|
| | Meds | No Meds | Meds | No Meds | Meds | No Meds | Meds | No Meds |
| Mean | 25.8 | 26.2 | 14.1 | 16.6 | 11.7 | 12.8 | 55.7% | 51.7% |
| Std Dev | 6.6 | 7.7 | 9.1 | 9.3 | 8.5 | 10.1 | 27.6% | 29.8% |
| Median | 25 | 27 | 12 | 15 | 9.5 | 11.5 | 59.4% | 56.2% |
Table 5. RSI scores subjects taking acid reducing medications at enrollment (N = 79) compared to subjects not taking acid reducing medications at enrollment (N = 16)
## Study #2: Safety of an Intentionally Displaced Reza Band® Upper Esophageal Sphincter (UES) Assist Device
## Objective
The safety of the Reza Band® UES Assist Device was evaluated when worn as intended over the cricoid cartilage, as well as when intentionally displaced laterally over each side of the neck, based on changes in heart rate, blood pressure, cardiac rhythm and intraocular pressure (IOP), as compared to when the Reza Band was not being worn. The intentional lateral displacement was studied to evaluate potential safety issues that might occur if the device was unintentionally displaced at night during sleep.
## Methods
Twenty (20) subjects who met the inclusion and exclusion criteria were enrolled in the study, and were representative of the intended Reza Band patient population. The inclusion and exclusion criteria were similar with those specified in the pivotal clinical study (Study #1), but in this study patients with a history of glaucoma were excluded. The study population consisted of 60% males and 40% females (95% Caucasian and 5% African-American subjects), with a mean age of 46.3 years, and a mean Body Mass Index (BMI) of 28.7.
Measurements of heart rate, blood pressure, cardiac rhythm and IOP were taken at baseline, immediately after the Reza Band was placed on the neck in each of the specified scenarios (cricoid, right displacement and left displacement). 5 minutes after placement (except for IOP) and again 15 minutes after placement. The Reza Band was removed and 3 minutes later, IOP was again measured in each placement scenario.
After the fitting of the Reza Band, the Comfort Dial was fully actuated until the Comfort Dial came to the stop, to achieve the maximum pressure. The maximum pressure with the Comfort Dial fully actuated was also measured with the Reza Band displaced laterally over the vascular structures on both sides of the neck.
## Results
The results showed that there was no effect on heart rate, blood pressure, cardiac rhythm or IOP. when the Reza Band was worn as intended as well as when the Reza Band was intentionally
{11}------------------------------------------------
displaced laterally, as compared to the baseline. No adverse events were reported during this study.
Analyses of Variance (One Way) were conducted on heart rate, blood pressure and intraocular pressure to determine whether there was any effect of the Reza Band being placed as intended or laterally displaced, when compared to not wearing the Reza Band. There was no significant difference for any of the measurements at any point in time, when compared to the baseline values.
Additionally, electrocardiograms (ECGs) were evaluated to determine whether there was any impact on cardiac rhythm when the Reza Band was placed as intentionally displaced over the jugular and carotid vasculature. It was determined that there was no change in the ECGs for any of the subjects at any time during the study, when compared to the baseline ECG.
The following summary table (Table 6) shows that when the Reza Band is set to apply its maximal pressure with the Comfort Dial fully actuated, and then displaced laterally, the pressure being applied to the left and right sides of the neck by the Reza Band is significantly reduced (mean of 59.2% (13.3 mm Hg), when displaced laterally to the left and a mean of 55.2% (14.6 mm Hg), when displaced laterally to the right), as compared to the applied pressure over the cricoid (20.9 mm Hg). This further reduces the risk of increased pressure to the sensitive structures of the neck.
| Variable | Cricoid Pressure<br>(Comfort Dial<br>OFF)<br>(mmHg) | Cricoid Pressure<br>(Comfort Dial<br>MAXIMUM)<br>(mmHg) | Displaced to Left<br>(Comfort Dial<br>MAXIMUM)<br>(mmHg) | Displaced to Right<br>(Comfort Dial<br>MAXIMUM)<br>(mmHg) |
|--------------------|-----------------------------------------------------|---------------------------------------------------------|----------------------------------------------------------|-----------------------------------------------------------|
| Mean | 20.9 | 32.6 | 13.3 | 14.6 |
| Standard Deviation | 1.4 | 3.3 | 4.9 | 4.0 |
| Median | 20 | 34 | 12 | 15 |
| Range | 20-24 | 26-39 | 7-22 | 6-22 |
Table 6. Comparison of Reza Band Applied Pressure at Cricoid and Laterally Displaced
### Conclusion
In the event that the Reza Band is unintentionally displaced laterally during sleep, the likelihood of impacting sensitive structures of the neck is low. There was no effect on heart rate, blood pressure, cardiac rhythm, or intraocular pressure, either when the Reza Band is placed as intended over the cricoid cartilage, or when it is laterally displaced over the jugular and carotid vasculature. Based on the results of this study, the Reza Band is safe, when worn as intended and if it were to be unintentionally displaced laterally.
## LABELING:
The sponsor provided labeling information which includes Physician Instructions for Use, Patient Information Handbook, and package labels for the Reza Band, External Manometer, and Pressure Sensor.
{12}------------------------------------------------
The labeling is sufficient and satisfies the requirements of 21 CFR 801.109 Prescription devices. The patient labeling also follows the principles identified in FDA's guidance entitled "Medical Device Patient Labeling" (April 2001).
## RISKS TO HEALTH:
Table 7 below identifies the risks to health that may be associated with use of External UES Compression Device type and the measures necessary to mitigate these risks.
| Identified Risk | Mitigation Method |
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Adverse tissue reaction | - Biocompatibility assessment |
| Risk of over-compression | - Clinical study<br>- Labeling<br>- Technical specifications |
| Device misuse/ incorrect fitting/ malfunctions | - Technical specifications<br>- Clinical study<br>- Labeling<br>- Performance testing<br>(mechanical integrity and shelf<br>life testing) |
Table 7: Identified Risks to Health and Mitigation Methods
## SPECIAL CONTROLS:
In combination with the general controls of the FD&C Act, the External UES Compression Device is subject to the following special controls:
- 1. The patient contacting components must be demonstrated to be biocompatible.
- 2. Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be demonstrated:
- (A) Mechanical integrity testing (e.g., tensile strength testing, fatigue testing)
- (B) Shelf life testing
- 3. The technical specifications must include pressure measurement accuracy to characterize device performance.
- 4. Clinical performance testing must document any adverse events observed during clinical use, and demonstrate that the device performs as intended under anticipated conditions of use.
- 5. Labeling must include the following:
- (A) Appropriate warnings and precautions.
- (B) A detailed summary of the clinical testing pertinent to use of the device including a detailed summary of the device-related complications or adverse events.
{13}------------------------------------------------
- (C) Detailed instructions on how to fit the device to the patient.
- (D) Instructions for reprocessing of any reusable components.
- 6. Patient labeling must be provided and must include:
- (A) Relevant warnings, precautions, and adverse effects/complications.
- (B) Information on how to correctly wear the device.
- (C) The potential risks and benefits associated with the use of the device.
- (D) Alternative treatments.
- (E) Reprocessing instructions.
### BENEFIT/RISK DETERMINATION
The observed risks of the device are based on data collected in the clinical studies described above. There were no device-related serious adverse events. Device movement was the most common adverse event. Device-related, non-serious adverse events also include patient discomfort including feeling hot with the device in place. Some patients reported disliking the sensation of the device around their neck. However, given that this device is externally worn, non-surgical, and prescribed by a physician, there is a high likelihood that patients would typically experience only non-serious adverse events. Duration of most adverse events would only last as long as the device is worn. Most adverse events would be reversible by removal of the device.
The observed benefits of the device are also based on data collected in clinical studies as described above. The benefit of the device is patient satisfaction. From the patient satisfaction survey at least 75% of the subjects reported satisfaction (extremely satisfied, 16.5%; very satisfied, 38.9%; satisfied, 20.0%). As this study was an open-label, single arm study, it is difficult to characterize the device's actual treatment effect. In post-hoc analysis, the patient demographics and characteristics were evaluated against outcomes. The analyses revealed that there was no difference among those sub-groups. Although this study was not statistically powered to assess benefit across the subpopulations, the study revealed that all patient subgroups did well, and therefore, the results appear to be applicable to the overall intended population. However, the current data only support the device's benefit while the device is worn at night. Once the device is removed, then it no longer provides the external pressure to the UES (i.e., there is likely no residual effect).
Additional factors that were considered in determining probable risks and benefits for the Reza Band UES Assist Device include:
- . The sponsor conducted additional studies using their prototype (early version) device to demonstrate the reduction in LPR by extra-esophageal pH, impedance, and direct endoscopic videography to support preliminary effectiveness of this device.
- . In the survey conducted by the sponsor as part of the pivotal study, the following satisfaction scores were noted: 16.5% of the subjects were extremely satisfied, 38.9% were very satisfied, 20.0% were satisfied, and 9.4% were somewhat satisfied; 5.9% of the subjects were dissatisfied, 8.2% were very dissatisfied, and 1.2% were extremely dissatisfied.
{14}------------------------------------------------
- . Considering that the device can be customized, removable, and non-implantable, the patients would tolerate even a smaller benefit given the low potential for risks.
- LPR is a chronic disease/condition. The disease is usually managed by strict adherence to behavioral life style modification, and/or oral medication (PPI twice a day for a minimum of 6 months). If LPR becomes life-threatening, surgical therapy of fundoplication (either open or laparoscopic) is indicated.
- Proper application of the Reza Band to the neck and detailed instructions for the physician as well as for the patients help to mitigate the risks. The physician instruction includes clear indications, warnings and precautions, and an appropriate discussion of the use of this device for patients given condition. The patient instruction includes appropriate use of the device, clear instructions to identify the potential risks and an understanding to report to their respective physician if the device is not functioning properly or intolerable.
- . The quickest intervention is for the patient to remove the device if it is causing any difficulty or adverse effects.
In conclusion, given the available information above, the data support that the probable benefits outweigh the probable risks for the Reza Band Upper Esophageal Sphincter Assist Device for reducing the symptoms of LPR in patients 18 years of age and older. The device provides measurable benefit and the risks can be mitigated by the use of general and the identified special controls.
## CONCLUSION
The de novo for the Reza Band Upper Esophageal Sphincter Assist Device is granted and the device is classified under the following:
Product Code: PKA Device Type: External Upper Esophageal Sphincter Compression Device Class: II Regulation: 21 CFR 874.5900
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.