OPTICAL MEDIASTINOSCOPE WITH PLUG-IN CONNECTION/OCULAR EYEPIECE
K971166 · Richard Wolf Medical Instruments Corp. · EWY · Aug 27, 1997 · Ear, Nose, Throat
Device Facts
Record ID
K971166
Device Name
OPTICAL MEDIASTINOSCOPE WITH PLUG-IN CONNECTION/OCULAR EYEPIECE
Applicant
Richard Wolf Medical Instruments Corp.
Product Code
EWY · Ear, Nose, Throat
Decision Date
Aug 27, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4720
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Mediastinoscope, in connection with various auxiliary instruments, is used for diagnostic and/or surgical interventions in the upper area of the mediastinum. Various auxiliary instruments are used in mediastinoscopy: - The smoke evacuation tube is used to evacuate smoke during HF or laser applications. - The telescope rinsing tube (optic irrigation tube) is used to keep the distal part of the telescope clear by rinsing with irrigation fluid. - The mediastinal puncture tube is used to puncture lymph nodules and aspirate the content for cyto/hystological purposes. - The tumor grasping forceps are used for grasping and manipulating tissue. - The micro grasping forceps with HF are used for coagulating small vessels and grasping. - The HF coagulation suction devices are used for suction and for coagulation while holding tissue ‘out of the way’ during endoscopic operations of the larynx in connection with CO2 lasers or in mediastinoscopy. - The bipolar coagulation suction tube (bipolar aspiration tube) is used for aspiration and for coagulation of small lesions.
Device Story
Optical Mediastinoscope; tubular, conical endoscope with handle and 90° bent telescope; fiber illumination; two spreadable spatulas for operative field distention. Used by physicians in clinical settings for mediastinoscopy. Auxiliary instruments include biopsy/grasping forceps, puncture tubes, suction/coagulation tubes, and irrigation tubes. Provides direct visualization of mediastinal structures; facilitates tissue sampling, node removal, and hemostasis via HF/laser integration. Enhances diagnostic accuracy for mediastinal disease and bronchial carcinoma staging.
Clinical Evidence
No clinical tests performed. Bench testing only, including mechanical pull tests and load tests for forceps/extractors, confirming no permanent deformation or breakage under normal use. Steam sterilization validation performed.
Technological Characteristics
90° bent telescope; 2.78 mm lens diameter; 35° direction of view; 98° angle of view. Spreadable spatula (two-piece); 1 mm wall thickness; distal diameter ~17 mm; working length 160 mm. Fiber optic illumination. Manual operation. Steam sterilizable.
Indications for Use
Indicated for diagnostic and/or surgical interventions in the upper mediastinum, including evaluation of bronchial carcinoma operability, treatment planning for mediastinal carcinomas, biopsy, diagnostic clarification of primary/secondary mediastinal diseases, and removal of nodes/tumors. Contraindications are those described in general medical literature.
Regulatory Classification
Identification
A mediastinoscope and accessories is a tubular tapered electrical endoscopic device with any of a group of accessory devices which attach to the mediastinoscope and is intended to examine or treat tissue in the area separating the lungs. The device is inserted transthoracicly and is used in diagnosis of tumors and lesions and to determine whether excision of certain organs or tissues is indicated. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel. This generic type of device includes the flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
{0}
AUG 27 1997
K971166
| 510(k) Summary of Safety and Effectiveness | | | | |
| --- | --- | --- | --- | --- |
| Submitter: | | | Date of Preparation: March 28, 1997 | |
| Company / Institution name: RICHARD WOLF MEDICAL INSTRUMENTS CORP. | | | FDA establishment registration number: 14 184 79 | |
| Division name (if applicable): N.A. | | | Phone number (include area code): (847) 913-1113 | |
| Street address: 353 Corporate Woods Parkway | | | FAX number (include area code): (847) 913-0924 | |
| City: Vernon Hills | State/Province: Illinois | Country: USA | | ZIP / Postal Code: 60061 |
| Contact name: Mr. Robert L. Casarsa | | | | |
| Contact title: Quality Assurance Manager | | | | |
| Product Information: | | | | |
| Trade name: Optical Mediastinoscope and Accessories | | Model number: 8783.401, 8783.411 and accessories (see section 1: Submitted devices) | | |
| Common name: Mediastinoscope | | Classification name: Mediastinoscope and Accessories | | |
| Information on devices to which substantial equivalence is claimed: | | | | |
| 510(k) Number | Trade or proprietary or model name | | Manufacturer | |
| 1 pre-enact. | 1 Fiber Light Mediastinoscope 8960.01/.03 | | 1 Richard Wolf M.I.C. | |
| 2 K 95 49 10 | 2 Optical Mediastinoscope 10970BV | | 2 Karl Storz | |
| 3 | 3 Mediastinoscope 'CARLENS' 10970A/B and operating instruments | | 3 Karl Storz | |
## 1.0 Description
The Optical Mediastinoscope is a tubular, conical shaped endoscope with a handle and a build-in telescope with fiber illumination which is bent at a 90° angle. The endoscope tube consists of two spreadable spatulas which allow distention of the operative field, and therefore, optimizes vision. The spatulas may be dismounted for easier cleaning.
For efficiency during diagnostic and therapeutic procedures, a variety of accessory instruments are needed, e.g. biopsy and grasping forceps, puncture, suction and coagulation tubes or telescope rinsing tubes.
8 - 1
{1}
8 - 2
# 2.0 Intended Use
The Mediastinoscope, in connection with various auxiliary instruments, is used for diagnostic and/or surgical interventions in the upper area of the mediastinum.
Various auxiliary instruments are used in mediastinoscopy:
- The smoke evacuation tube is used to evacuate smoke during HF or laser applications.
- The telescope rinsing tube (optic irrigation tube) is used to keep the distal part of the telescope clear by rinsing with irrigation fluid.
- The mediastinal puncture tube is used to puncture lymph nodules and aspirate the content for cyto/hystological purposes.
- The tumor grasping forceps are used for grasping and manipulating tissue.
- The micro grasping forceps with HF are used for coagulating small vessels and grasping.
- The HF coagulation suction devices are used for suction and for coagulation while holding tissue ‘out of the way’ during endoscopic operations of the larynx in connection with $\mathrm{CO}_{2}$ lasers or in mediastinoscopy.
- The bipolar coagulation suction tube (bipolar aspiration tube) is used for aspiration and for coagulation of small lesions.
# 3.0 Technological Characteristics
- $90^{\circ}$ bent telescope, diameter of lens $2.78 \mathrm{~mm}$
- direction of view $35^{\circ}$, angle of view $98^{\circ}$
- spreadable spatula, consisting of two half pieces, wall thickness $1 \mathrm{~mm}$
- spatula diameter: distal ca. $17 \mathrm{~mm}$, proximal $28 \times 18.5 \mathrm{~mm}$
- working length $160 \mathrm{~mm}$
# 4.0 Substantial Equivalence
The submitted devices are substantially equivalent to existing pre-enactment devices sold by Richard Wolf and $510(k)$ devices sold by Storz.
# 5.0 Performance Data
- Pull tests of the optical mediastinoscope show that there is no permanent deformation if used normally.
- Mechanical load tests of the forceps and stone extractors show that there is no breakage of the jaw or other parts if used normally.
- Steam sterilization in clinical use and tests performed by Richard Wolf show that the steam sterilization has no influence on the functional performance of the submitted devices when using the fractional method.
# 6.0 Clinical Tests
No clinical tests performed.
{2}
7.0 Conclusions Drawn
These devices are designed and tested to guarantee the safety and effectiveness, when used according to the Instruction manual.
By: Robert L. Casarsa
Quality Assurance Manager
Date: July 14, 1997
8 - 3
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
Mr. Robert L. Casarsa
Manager of Quality Assurance
Richard Wolf Medical Instruments Corporation
353 Corporate Woods Parkway
Vernon Hills, Illinois 60061
Re: K971166
Trade Name: Optical Mediastinoscope and Accessories
Regulatory Class: II
Product Code: EWY
Dated: June 14, 1997
Received: July 16, 1997
AUG 27 1997
Dear Mr. Casarsa:
We have reviewed your Section 510(k) notification of intent to market the devices referenced above and we have determined the devices are substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your devices are classified (see above) into either class II (Special Controls) or class III (Premarket Approval), they may be subject to such additional controls. Existing major regulations affecting your devices can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your devices in the Federal Register. Please note: this response to your premarket notification submissions does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}
Page 2 - Mr. Robert L. Casarsa
This letter will allow you to begin marketing your devices as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your devices to a legally marketed predicate device results in a classification for your devices and thus, permits your devices to proceed to the market.
If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your devices, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Enclosure
{5}
Indications for Use
510(k) Number (if known): K971166
Device Name: Mediastinoscope and Auxiliary Instruments
## Indications for Use:
The Mediastinoscope, in connection with various auxiliary instruments, is used for diagnostic and/or surgical interventions in the upper area of the mediastinum.
Mediastinoscopy is carried out:
- In order to evaluate the operability of patients suffering from bronchial carcinoma which have already been verified, for treatment planning of other carcinomas in the mediastinal area and for biopsies.
- For diagnostic clarification of primary and secondary diseases of the mediastinum, when other examination methods have been unable to provide the nature, and in particular the histological character of the disturbance.
- For the removal of nodes and tumors.
- For the diagnosis and removal of lymphatic nodes.
Revised 7/14/97
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH Office of Device Evaluation (ODE)

Prescription Use ☑ Per 21 CFR 801.109
OR
Over-The Counter
{6}
# Indications Continued:
Various auxiliary instruments are used in mediastinoscopy.
- The smoke evacuation tube is used to evacuate smoke during HF or laser applications.
- The telescope rinsing tube (optic irrigation tube) is used to keep the distal area of the telescope clear by rinsing with irrigation fluid.
- The mediastinal puncture tube is used to puncture lymph nodules and aspirate the content for cyto/histological purposes.
- The tumor grasping forceps are used for grasping and manipulating tissue.
- The micro grasping forceps with HF are used for coagulating small vessels and grasping.
- The HF coagulation suction devices are used for suction and for coagulation while holding tissue ‘out of the way’ during endoscopic operations of the larynx in connection with CO₂ lasers or in mediastinoscopy.
- The bipolar aspiration tube (bipolar coagulation suction tube) is used for aspiration and for coagulation of small lesions.
# Contraindications:
Contraindications for the patient which result from the general findings are described in the relevant literature and must be observed.
Revised 7/14/97
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH Office of Device Evaluation (ODE)

Prescription Use ☑ Per 21 CFR 801.109
OR
Over-The Counter