The TruDi® Navigation system is intended for use during surgical procedures in ENT and ENT skull base surgery to support navigation of instruments to targeted anatomy, where reference to a rigid anatomical structure can be identified relative to a CT or MR based model.
Device Story
TruDi® Navigation System V3 is an image-guided surgical navigation system for ENT and skull base procedures. It uses electromagnetic tracking to display the real-time position of navigated instruments relative to pre-acquired CT or MR images. The system consists of a navigation console, emitter pad, instrument hub, patient tracker, registration probe, and workstation. The emitter pad generates a magnetic field that induces current in a sensor at the instrument tip, allowing position calculation. A patient tracker attached to the forehead compensates for head movement. ENT physicians use the system to navigate instruments to targeted anatomy. The output is a visual display of the tool tip icon overlaid on registered patient images, assisting the surgeon in instrument placement. The system supports data archiving to local disks, flash memory, and hospital PACS networks. It is intended for use in clinical settings by trained physicians.
Clinical Evidence
No clinical data was necessary. Substantial equivalence was established through non-clinical bench testing, including location accuracy tests (compared against a high-accuracy robot), software functional testing (error handling, latency, usability), safety and EMC testing, and simulated use accuracy testing (CT registration and navigation workflow).
Technological Characteristics
Electromagnetic tracking system; PC-based platform running Windows 10 (64-bit). Components include Emitter Pad, Instrument Hub, Patient Tracker, Registration Probe, and Workstation. Connectivity via LAN/PACS. Software-embedded device. Accuracy specification: 2mm RMS. Bench location accuracy: 0.55mm (STD 0.2mm). Non-sterile.
Indications for Use
Indicated for use during ENT and skull base surgical procedures to support navigation of instruments to targeted anatomy using CT or MR-based models.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 21, 2023
Acclarent, Inc. % Dikla Dayan Regulatory Affairs Manager Biosense Webster (Israel) Ltd. 4 Hatnufa St Yokneam, 2066717 Israel
Re: K231862
Trade/Device Name: TruDi® Navigation System V3 Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: PGW Dated: June 23, 2023 Received: June 23, 2023
Dear Dikla Dayan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Joyce C. Lin -S
for Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K231862
Device Name TruDi® Navigation System V3
Indications for Use (Describe)
The TruDi® Navigation system is intended for use during surgical procedures in ENT and ENT skull base surgery to support navigation of instruments to targeted anatomy, where reference to a rigid anatomical structure can be identified relative to a CT or MR based model.
| Type of Use (Select one or both, as applicable) | <table border="0"><tr><td><div style="display:inline-block; vertical-align:middle;"> <img alt="Checked" src="checkbox_checked.png" style="width:16px; height:16px; vertical-align:middle;"/> </div> Registration Use (Part 21 CFR 201 Subpart D) </td><td><div style="display:inline-block; vertical-align:middle;"> <img alt="Unchecked" src="checkbox_unchecked.png" style="width:16px; height:16px; vertical-align:middle;"/> </div> Over-The-Counter Use (21 CFR 201 Subpart C) </td></tr></table> | <div style="display:inline-block; vertical-align:middle;"> <img alt="Checked" src="checkbox_checked.png" style="width:16px; height:16px; vertical-align:middle;"/> </div> Registration Use (Part 21 CFR 201 Subpart D) | <div style="display:inline-block; vertical-align:middle;"> <img alt="Unchecked" src="checkbox_unchecked.png" style="width:16px; height:16px; vertical-align:middle;"/> </div> Over-The-Counter Use (21 CFR 201 Subpart C) |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:inline-block; vertical-align:middle;"> <img alt="Checked" src="checkbox_checked.png" style="width:16px; height:16px; vertical-align:middle;"/> </div> Registration Use (Part 21 CFR 201 Subpart D) | <div style="display:inline-block; vertical-align:middle;"> <img alt="Unchecked" src="checkbox_unchecked.png" style="width:16px; height:16px; vertical-align:middle;"/> </div> Over-The-Counter Use (21 CFR 201 Subpart C) | | |
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Acclarent, a medical technology company. The word "Acclarent" is written in a combination of black and purple font, with a purple arc above the "ent" portion of the name. Below the name, in smaller font, it says "PART OF THE Johnson-Johnson FAMILY OF COMPANIES".
Acclarent, Inc 31 Technology Drive
Main 650.687.5888 Fax 650.687.5889 www.acclarent.com
## 510(k) Summary - K231862
| [807.92(a)(1)] Submitter Information | |
|-----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant: | Biosense Webster, Inc.<br>31 Technology Drive, Suite200<br>Irvine, CA 92618, USA<br>Tel.: 419-233-2611<br>Fax: 949-450-6886 |
| Contact person: | Kamrie Sarnosky<br>Senior Regulatory Specialist<br>Acclarent, Inc. |
| Authored by: | Dikla Dayan<br>RA Manager<br>Biosense Webster (Israel) Ltd |
| Date summary prepared: | June 23, 2023 |
| [807.92(a)(2)] Name of Device | |
| Device Trade Name | TruDi® Navigation System V3 |
| Device Common Name | Image Guided Surgery System |
| Device Classification | Class II, 21 CFR 882.4560<br>Ear, Nose, and Throat Manual Surgical Instrument (21 CFR 882.4560) |
| Product Code | PGW |
| [807.92(a)(3)] Legally Marketed Devices | |
| Predicate Device: | TruDi® Navigation System V2 (K192397) |
| [807.92(a)(4)] Device Description | |
| Device Description: | The TruDi® Navigation System V3 is intended to be used during surgical<br>procedures in ENT and ENT skull base surgery to support navigation of<br>instruments to the targeted anatomy, where reference to a rigid anatomical<br>structure can be identified relative to a CT or MR based model.<br>The TruDi® Navigation System V3 enables ENT physicians to access<br>sphenoid, frontal, and maxillary sinuses, as well as the skull base, by using<br>the systems magnetic tracking technology, which is the same technology<br>used by the predicate device.<br>The system incorporates a Navigation Console, Emitter Pad, Instrument<br>Hub, Patient Tracker, Registration Probe, Workstation and peripherals. A<br>magnetic field generated by the Emitter Pad induces a current in the<br>magnetic sensor embedded in the tip of the navigated tool, which helps to<br>accurately calculate the tool tip position. A CT or MR image is imported<br>and registered to the patient coordinates and the navigated tool tip icon is<br>displayed on top of the registered image, indicating the position of the tool<br>tip in reference to the patient anatomy. A Patient Tracker is fixed to the<br>patient forehead to compensate for the head movement during the surgical<br>procedure. |
| [807.92(a)(5)] Intended Use | |
| Intended use | The TruDi® Navigation System V3 is intended for use during surgical<br>procedures in ENT and skull based surgery to support navigation of<br>instruments to targeted anatomy, where reference to rigid anatomic<br>structure can be identified relative to a CT or MR based model. |
| Difference in Indications<br>from Predicate Device: | There is no difference in intended use between the subject device and<br>predicate device. |
| [807.92(a)(6)] Technical Characteristics | |
| Predicate overview and<br>comparison table: | The substantial equivalence of the TruDi® Navigation System V3 to the<br>predicate is shown by similarity in intended use, indications for use,<br>technology, and performance.<br>The TruDi® Navigation System V3 is an image-guided navigation system<br>designed for use during ENT and skull based surgical procedures. ENT<br>physicians can track and display the real-time location of the tip of<br>navigated instruments relative to pre-acquired CT/MR images. Like the<br>predicate device, the subject system utilizes electromagnetic tracking<br>technology for navigation, uses anatomical reference points on the patient<br>anatomy for intraoperative registration to the image-based model of the<br>anatomy, and uses CT/MR image sets as reference images for the image-<br>based model. The table below provides a comparison of the technological<br>characteristics between the subject device and the predicate device. |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Acclarent, a company that is part of the Johnson & Johnson family of companies. The word "Acclarent" is written in a bold, sans-serif font, with the first part of the word in black and the "ent" in purple. Above the purple part of the word is a purple arc. Below the word "Acclarent" is the phrase "PART OF THE Johnson & Johnson FAMILY OF COMPANIES".
| Attributes | PREDICATE DEVICE:<br>TRUDI NAVIGATION SYSTEM V 2.0 | SUBJECT DEVICE:<br>TRUDI NAVIGATION SYSTEM V 3.0 |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Fundamental Scientific<br>Technology | Electromagnetic location | Same |
| Technological<br>Characteristics | Radiation of low intensity electromagnetic field.<br>Acquisition of magnetic sensor induced<br>voltages. Electromagnetic location in reference<br>to registered CT/MR background of patient head | Same |
| Technology | Software embedded device | Same |
| Platform | PC Based | Same |
| Operating System | Windows 10 (64 bits) | Same |
| Programming Languages | Visual C++, C# and C, MATLAB, HTML | Same |
| Accuracy specification | 2mm RMS | Same |
| Bench Location Accuracy | 0.55mm (standard Deviation 0.7 mm) | 0.55 (STD 0.2) |
| Simulated Use Location<br>Accuracy | 1.1 mm (Standard deviation 0.2 mm) | 1.2 mm (Standard deviation 0.2mm) |
| Attributes | PREDICATE DEVICE:<br>TRUDI NAVIGATION SYSTEM V 2.0 | SUBJECT DEVICE:<br>TRUDI NAVIGATION SYSTEM V 3.0 |
| Main Feature | Displays the instruments position in relation to<br>preoperative scans utilizing electromagnetic<br>tracking technology. | Same |
| Location Update Rate | 10Hz | Same |
| Supported Navigation<br>Instruments | Flexible and Rigid instruments with magnetic<br>sensor on instrument tip | Same |
| Supported Preoperative<br>Images | DICOM CT, MR | Same |
| Supported CT/MR Import<br>Media | CD-ROM, USB flash drive, LAN network<br>(PACs) | Same |
| Patient Data Archiving | Not supported | Supported in local disk, flash memory<br>and hospital network PACS (new<br>Software Function) |
| Packaging | Provided non-sterile | Same |
| HW Main components | | |
| Instrument Hub | Enables the user to connect components and<br>navigated tool to the system | Added connection flexibility.<br>Verification testing has demonstrated<br>that this change does not raise new<br>questions on the safety and<br>effectiveness of the device. |
| Workstation | Dell 5810 or Dell5820<br>Enables data processing | Same |
| Mouse and Keyboard | Standard mouse and keyboarded | Add touchpad mouse |
| Console | Drives the magnetic signals from Emitter Pad to<br>Workstation | Same |
| Emitter Pad | Generates magnetic field | Same |
| Patient tracker | Compensates on patient movement (reusable or<br>disposable) | Same |
| Registration Probe | Enables correlation between patient's face and<br>CT image | Same |
| Isolation Transformer | Adjusting power input to the system | Same |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Acclarent, a company that is part of the Johnson & Johnson family of companies. The word "Acclarent" is written in a bold, sans-serif font, with the first part of the word in black and the last part in purple. Above the purple part of the word is a purple arc. Below the word "Acclarent" is the text "PART OF THE Johnson & Johnson FAMILY OF COMPANIES".
Acclarent, Inc 31 Technology in the line
www.acclarent.com
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Acclarent, a medical technology company. The word "Acclarent" is written in a combination of black and purple font, with a purple arc above the "ent" portion of the word. Below the logo, the text "PART OF THE Johnson & Johnson FAMILY OF COMPANIES" is written in red and gray.
Acclarent, Inc 31 Technology Drive Irvine, CA 92618 Main 650.687.5888 Fax 650.687.5889 www.acclarent.com
Summary of Technological Patient Data Archive Characteristics ● Note: The Patient Data Archive is a new Software Function, although, being a Medical Device Data System (MDDS), it is not regulated by the FDA. ● Instrument Hub improvement ● Support of additional navigated tools Software enhancements . Segmentation manual editing o Path to Target 3D visualization O ● Cables robustness improvement Optional touchpad mouse ● The TruDi® Navigation System V3 was tested to ensure that it functions in Non-clinical Performance accordance with the system design specifications related to substantial Data: equivalence in terms of device safety and effectiveness. The following non-clinical tests were performed: 1. Location Accuracy tests, where the TruDi® Navigation System electromagnetic locations were compared to the locations provided by a highly accurate robot system over its entire navigation volume 2. Software functional tests, covering the complete system functionality, and including error handling, usability and time performance (latency). 3. Safety, and EMC tests were performed by a nationally recognized testing laboratory to verify compliance with safety and EMC standards for medical devices. 4. Simulated use accuracy test, in which a complete CT image registration and instrument navigation workflow was performed, to verify the overall accuracy of the system. The proposed TruDi® Navigation System V3 passed all tests in accordance with specified test criteria and standards, and the modified device did not raise new questions of safety or effectiveness. Clinical data was not necessary to determine that the subject TruDi® Clinical performance data: Navigation System V3 performs as intended. Conclusion: The modified TruDi® Navigation System V3 is substantially equivalent to the currently cleared TruDi® Navigation System V2 based on the completion of non-clinical bench testing as well as similar principles of design, operation and indications for use.
The subject device has the same technological characteristics as the predicate device, with addition of:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.