The devices FlexPointer 1.5 Single Use and FlexTube 3 Single Use are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. The devices are indicated for use with the Fiagon Navigation system using electromagnetic navigation. The devices are indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy. Example procedures include, but are not limited to: ENT Procedures; Transphenoidal access procedures. Intranasal procedures. Sinus procedures, such as Maxillary antrostomies, Ethmoidectomies, Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies. ENT related anterior skull base procedures.
Device Story
Disposable, electromagnetically navigated instruments (FlexPointer 1.5, FlexTube 3) for use with Fiagon Navigation system; sensor embedded in tip tracks within low-energy magnetic field. FlexPointer 1.5 acts as pointer; FlexTube 3 acts as suction device. Navigation software displays instrument position relative to preoperative CT/MRI scans. Used in ENT surgery (sinus, skull base) by surgeons to precisely locate anatomical structures. Provides real-time visual guidance; aids surgical navigation; improves accuracy in complex anatomical regions.
Clinical Evidence
Bench testing only. Accuracy evaluated on anatomical skull phantom using 225 trials (15x15 fiducial markers). Mean target registration error < 1.5 mm achieved for both devices. Testing performed pre- and post-sterilization/accelerated aging to confirm performance stability.
Technological Characteristics
Electromagnetic navigation instruments. Materials: PSU (connector), plastic (handle), stainless steel (tube), heat shrink tubing. Single-use, ETO sterilized. Sensor embedded in tip. Dimensions comparable to predicates. Connectivity: Wired to Fiagon Navigation system. Software: Navigation system-based tracking.
Indications for Use
Indicated for patients requiring stereotactic surgery for ENT procedures (e.g., paranasal sinus, mastoid, skull base) where anatomical structures can be referenced to CT/MR models. Applicable for open or percutaneous procedures.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 10, 2020
Fiagon GmbH Dirk Mucha, CTO Neuendorfstrasse 23b Hennigsdorf, DE 16761 Brandenburg
Re: K200041
Trade/Device Name: FlexPointer 1.5 Single Use, FlexTube 3 Single Use Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: Class II Product Code: PGW Dated: February 13, 2020 Received: February 14, 2020
Dear Dirk Mucha:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Michael J. Ryan Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K200041
### Device Name
Fiagon Navigation - FlexPointer 1.5 Single Use, FlexTube 3 Single Use
### Indications for Use (Describe)
The devices FlexPointer 1.5 Single Use and FlexTube 3 Single Use are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. The devices are indicated for use with the Fiagon Navigation system using electromagnetic navigation.
The devices are indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy.
Example procedures include, but are not limited to:
| ENT Procedures; |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Transphenoidal access procedures. |
| Intranasal procedures. |
| Sinus procedures, such as Maxillary antrostomies, Ethmoidectomies, Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies. |
| ENT related anterior skull base procedures. |
Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# January, 6 2020
#### Submitter Information/ 510(k) Holder 1.
| Submitter: | Fiagon GmbH |
|------------|----------------------------|
| Address: | Neuendorfstrasse 23b |
| | 16761 Hennigsdorf, Germany |
| Telephone: | +49 3302 201 21 10 |
|------------|--------------------|
| Telefax: | +49 3302 201 21 15 |
Mr. Dirk Mucha, CTO Contact:
#### Device Information 2.
| Trade Name: | Fiagon Navigation – FlexPointer 1.5 Single Use, FlexTube 3 Single Use |
|-----------------|-----------------------------------------------------------------------|
| Common Name: | Image guided surgery system |
| Classification: | Class II per 21 CFR 882.4560 |
| Device: | Ear, Nose, and Throat Stereotaxic Instrument |
| Product Code: | PGW |
#### 3. Purpose of Submission
The purpose of this submission is to gain clearance for a modified version of the previously cleared instrument, Fiagon Navigation – FlexPointer 1.5 (K150473) and FlexTube 3mm (K141456)
#### Predicate Device Information 4.
Fiagon Navigation – FlexPointer 1.5 (K150473) and FlexTube 3mm (K141456) (in extended instrument set ENT)
#### 5. Device Description
{4}------------------------------------------------
The Fiagon Navigation - FlexPointer 1.5 Single Use and FlexTube 3 Single Use are disposable instruments intended to be used with the Fiagon Navigation system. The FlexPointer 1.5 Single Use is an electromagnetically navigated pointing device (malleable, sensor within the tip). The FlexTube 3 Single Use is an electromagnetically navigated suction device (malleable, sensor within the tip).
Each device incorporates a sensor device, which is tracked by the navigation system within the low-energy magnetic field of a field generator (part of the navigation system). The navigation software (part of the navigation system) displays the position of the instruments in preoperative scans (e.g., CT, MRI, fluoroscopy).
## 6. Indications for Use
The devices FlexPointer 1.5 Single Use and FlexTube 3 Single Use are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. The devices are indicated for use with the Fiagon Navigation system using electromagnetic navigation.
The devices are indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy.
Example procedures include, but are not limited to:
ENT Procedures; Transphenoidal access procedures. Intranasal procedures. Sinus procedures, such as Maxillary antrostomies, Ethmoidectomies, Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies. ENT related anterior skull base procedures.
# 7. Comparison of Technological Characteristics
The FlexPointer 1.5 Single Use is a modified version of the previously cleared FlexPointer 1.5 (K150473). The FlexTube 3 Single Use is a modified version of the previously cleared FlexTube 3mm (K141456). The reason for this Special 510(k) is to describe a change in material as well as a change from a reprocessed device to one that is provided sterile and is Single Use and disposable.
{5}------------------------------------------------
The new devices, FlexPointer 1.5 Single Use and FlexTube 3 Single Use, have very similar technological characteristics compared to the predicate devices. All instruments are navigation instruments and utilize the same sensors for spatial measurement and work equally well with the Fiagon Navigation System.
The new devices utilize a different connector housing material (PSU). The old material used for the connector housing in the predicate device was PEI. The design of the plug remains identical so that the proper connection to the navigation system is not affected by the change in materials.
For the device FlexPointer 1.5 Single Use the handle is slightly modified and uses different material compared to its predicate.
The design change of the handle consist of some more rounded planes to increase ergonomics making use of the designing possibilities of an injected molding plastic part compared to a machined stainless steel part. The modified material results in a lighter design of the instrument increasing the ergonomics.
The new material compositions have been assessed to biocompatibility concerns with identical endpoint tests to the predicate and can be seen substantially equivalent to the predicate.
For the FlexTube 3 Single Use the navigation sensor is supported within an additional heat shrink tubing on the outside of a stainless steal tube where the predicate uses an additional tube as sensor carrier which is fixed inside the stainless steel tube. Therefore, even though the inner diameter in the modified instrument is comparable in dimensions to the unmodified instrument, there is no additional obstruction inside the inner tube, thereby facilitating the ease of flow of fluids within the instruments. Further, the performance testing demonstrates that the instruments perform as expected. Thus, the modified device can be considered substantially equivalent with respect to technological characteristics.
The handle is slightly modified and uses different material. The design change of the handle consist of some more rounded planes to increase ergonomics making use of the designing possibilities of an injected molding plastic part compared to a machined stainless steel part. The modified material results in a lighter design of the instrument increasing the ergonomics. The new material compositions (incl. new handle, heat shrink and the unmodified suction tube) have been assessed to biocompatibility concerns with identical endpoint tests to the predicate and can be seen substantially equivalent to the predicate. The material change does not raise new concerns of safety or effectiveness.
The change of material in the plug and handle as well as the modified sensor fixation with additional heat shrink tubing are motivated by the change of the sterilization method from
{6}------------------------------------------------
steam sterilization (132° C) to ETO sterilization and from 10 times use to single use, which is lowering the requirements for the temperature stability of the materials.
Under the modified sterilization and life time specifications the material modifications do not raise new concerns of safety or effectiveness.
Sterilization validation tests reports are provided to demonstrate that these differences do not raise new issues of safety and effectiveness
Biocompatibility tests have been performed to demonstrate substantial equivalent to the predicates and bench testing of performance have been done using same test protocols for navigation accuracy with the Fiagon Navigation system on a anatomical phantom. Mean accuracy value < 1.5 mm is achieved for both new devices which compares to the predicates.
## 8. Performance Data
Bench testing was performed in order to determine device accuracy of the modified devices with the Fiagon Navigation system.
Also testing was performed to ensure sterilization of the modified disposable instruments does not alter the performance characteristics of the device. Following test were performed:
- -Bench test: Accuracy on anatomical phantom with the Fiagon navigation system Result: mean accuracy < 1.5 mm as mean target registration error on fiducial markers. Protocol: total 15 x 15 fiducial markers are touched on anatomical skull phantom and measured/displayed navigation position by device is compared to known position of marker. Difference of all 225 trial is computed as mean accuracy.
- Bench test: Post sterilization accuracy testing: same protocol as above was used post । sterilization and accelerated aging.
Result: mean accuracy < 1.5 mm as mean target registration error on fiducial markers.
# 9. Conclusion
Based on the indications for use, technological characteristics, performance testing, and comparison to the predicates, the devices FlexPointer1.5 Single Use and FlexTube 3 Single Use have been shown to be substantially equivalent to the predicate devices FlexPointer 1.5 and FlexTube 3mm and the modified devices do not present any new issues of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.