← Product Code [PNZ](/submissions/EN/subpart-e%E2%80%94surgical-devices/PNZ) · K220027

# Audion ET dilation system (K220027)

_Entellus Medical, Inc. · PNZ · Apr 12, 2022 · Ear, Nose, Throat · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/PNZ/K220027

## Device Facts

- **Applicant:** Entellus Medical, Inc.
- **Product Code:** [PNZ](/submissions/EN/subpart-e%E2%80%94surgical-devices/PNZ.md)
- **Decision Date:** Apr 12, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 874.4180
- **Device Class:** Class 2
- **Review Panel:** Ear, Nose, Throat
- **Attributes:** Therapeutic

## Intended Use

To dilate the cartilaginous portion of the Eustachian tube for treating tube dysfunction in patients 18 years and older using a transnasal approach.

## Device Story

Audion ET dilation system is a single-use, sterile device for treating Eustachian tube dysfunction. It consists of a dilation device with a non-malleable shaft, a fixed distal balloon, and an atraumatic polymer ball tip; an inflation device (syringe and extension line); and an optional inflation lock. Operated by physicians in a clinical setting, the device is positioned under endoscopic visualization. The ball tip is seated in the Eustachian tube, and the balloon is advanced and inflated to 12 atm to dilate the cartilaginous portion. The system provides mechanical dilation to improve Eustachian tube function. It does not support guidewires, suction, irrigation, or light fibers.

## Clinical Evidence

Bench testing only. Testing included biocompatibility, design verification (dimensional, functional, strength, HFE/UE), packaging, shelf life, and design validation with ENT physicians. Results demonstrated the device meets design specifications and performs as intended.

## Technological Characteristics

Non-malleable shaft; 1.26 mm OD; 1.91 mm atraumatic polymer ball tip; 6 mm diameter balloon (at 12 atm); 20 mm balloon body length. Materials include polymer. Sterilization via ethylene oxide. Single-patient use. Manual operation via inflation syringe and optional inflation lock. No software or electronic components.

## Regulatory Identification

A Eustachian tube balloon dilation system is a prescription device that includes a flexible catheter attached to an inflatable balloon. The system is intended for use in dilating the cartilaginous portion of the Eustachian tube for treating persistent Eustachian tube dysfunction.

## Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be evaluated:
(i) Mechanical testing, including tensile and flexural testing of catheter joints and materials.
(ii) Durability testing, including fatigue and burst pressure testing of the balloon materials and components.
(iii) Inflation and deflation characterization testing, including time and pressure measurements, and leak testing of the balloon.
(iv) Verification testing of safety features built into the device must be performed, including the characterization of catheter geometries and distal tip insertion limitation mechanisms.
(2) Simulated use testing in a clinically relevant model must demonstrate the reliability of the device to remain mechanically functional throughout the anticipated conditions of use, and validate that the design features limit access to only the cartilaginous portion of the Eustachian tube.
(3) The patient-contacting components of the device must be demonstrated to be biocompatible.
(4) Performance data must demonstrate the sterility of the device.
(5) Performance data must support shelf life by demonstrating continued sterility of the device, package integrity, and device functionality over the identified shelf life.
(6) Training must include simulated use on cadavers to ensure users can follow the instructions for use to allow safe use of the device.
(7) Labeling must include:
(i) Detailed instructions for use.
(ii) A detailed summary of the device technical parameters, including maximum allowed inflation pressure, allowable catheter geometries, and available balloon sizes.
(iii) A shelf life.

In combination with the general controls of the FD&C Act, the Acclarent Aera™ Eustachian Tube Balloon Dilation System is subject to the following special controls:

- 1. Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be evaluated:
	- a. Mechanical testing, including tensile and flexural testing of catheter joints and materials.
- b. Durability testing, including fatigue and burst pressure testing of the balloon materials and components.
- Inflation and deflation characterization testing, including time and pressure C. measurements, and leak testing of the balloon.
- d. Verification testing of safety features built into the device must be performed. including the characterization of catheter geometries and distal tip insertion limitation mechanisms.
- 2. Simulated use testing in a clinically relevant model must demonstrate the reliability of the device to remain mechanically functional throughout the anticipated conditions of use, and validate that the design features limit access to only the cartilaginous portion of the Eustachian tube.
- 3. The patient-contacting components of the device must be demonstrated to be biocompatible.
- 4. Performance data must demonstrate the sterility of the device.
- 5. Performance data must support shelf life by demonstrating continued sterility of the device, package integrity and device functionality over the identified shelf life.
- 6. Training must include simulated use on cadavers to ensure users can follow the instructions for use to allow safe use of the device.
- 7. Labeling must include:
	- a. Detailed instructions for use.
	- b. A detailed summary of the device technical parameters, including maximum allowed inflation pressure, allowable catheter geometries, and available balloon sizes.
	- c. A shelf life.

## Predicate Devices

- XprESS ENT Dilation System ([K163509](/device/K163509.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 12, 2022

Entellus Medical, Inc. Katie Wilson Sr. Staff, Regulatory Affairs Specialist 3600 Holly Lane North, Suite 40 Plymouth, Minnesota 55447

Re: K220027

Trade/Device Name: Audion ET dilation system Regulation Number: 21 CFR 874.4180 Regulation Name: Eustachian Tube Balloon Dilation System Regulatory Class: Class II Product Code: PNZ Dated: March16, 2022 Received: January 5, 2022

Dear Katie Wilson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K220027

Device Name Audion ET dilation system

#### Indications for Use (Describe)

To dilate the cartilaginous portion of the Eustachian tube for treating tube dysfunction in patients 18 years and older using a transnasal approach.

| Type of Use (Select one or both, as applicable)                                                                               |                                                                        |
|-------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(K) SUMMARY

| Date Prepared:                                                             | December 31, 2021                                                                                               |
|----------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Submitter Information:                                                     | Entellus Medical, Inc.<br>3600 Holly Lane North, Suite 40<br>Plymouth, MN 55447                                 |
| Establishment Registration:                                                | 3006345872                                                                                                      |
| Contact Information:                                                       | Katie Wilson<br>Sr. Staff, Regulatory Affairs Specialist<br>(269) 888-0315<br>katie.wilson@stryker.com          |
| Alternate Contact:                                                         | Garrett Ahlborg<br>Regulatory Affairs Manager<br>(763) 463-7089<br>garrett.ahlborg@stryker.com                  |
| Device Information:<br>Trade Name:<br>Common Name:<br>Classification Name: | Audion ET dilation system<br>Eustachian Tube Balloon Dilation Device<br>Eustachian Tube Balloon Dilation Device |
| Product Code:                                                              | PNZ                                                                                                             |
| Regulation Number:                                                         | Class II 21 CFR 874.4180                                                                                        |
| Predicate Devices:                                                         | K163509 XprESS ENT Dilation System                                                                              |

#### Device Description:

The Audion ET dilation system (ET is an abbreviation for Eustachian tube) is intended to dilate the Eustachian tube through use of balloon dilation and a transnasal approach. It contains a dilation device, inflation device and inflation lock. The dilation device has a 1.26 mm outer diameter (OD) curved non-malleable shaft with a balloon fixed at the distal end and a 1.91 mm atraumatic polymer ball tip. The dilation device is positioned under endoscopic visualization with the balloon fully retracted. Once the portion of the device from the ball tip to the base of the curve is seated in the Eustachian tube, the balloon is advanced and then inflated. The deployed balloon angle of the dilation device is 45° for optimal treatment of the Eustachian tubes. The distal leg length with balloon extended is 18.5 mm. When the balloon is inflated to 12 atm, the balloon diameter is 6mm and the body length is 20 mm.

{4}------------------------------------------------

The inflation device is an accessory that consists of an inflation syringe and an extension line. The inflation device is designed to deliver a pressure of 12 atm and is used to inflate the balloon on the dilation device.

The inflation lock is an optional accessory that is intended to interface with the Audion dilation device, inflation syringe and extension line connections to hold or release pressure during balloon dilation.

The Audion ET dilation system is provided sterile and is for single patient use only. The dilation device, inflation device and inflation lock are sterilized using ethylene oxide.

The Audion ET dilation system has been tested to withstand multiple inflations in a surgical case

## Indications for Use:

To dilate the cartilaginous portion of the Eustachian Tube for treating persistent Eustachian Tube dysfunction in patients 18 years and older using a transnasal approach.

## Contraindications:

There are not any known contraindications that directly refer to the product. The physician is responsible for deciding if the general condition of the patient allows the intended application.

## Technological Characteristics:

The technological characteristics of the Audion ET dilation system are similar to the predicate device [K163509], including Principles of Operation, design, function, materials, biocompatibility, and sterility.

The Audion ET dilation system and its predicate [K163509] both:

- . have the same intended use and indications for use with respect to Eustachian tube dilation
- use balloon technology with 12 atm of pressure to dilate Eustachian tube anatomy .
- extend no more than 19.5 mm into the cartilaginous portion of the Eustachian tube and ● feature similar distal ball tip dimensions
- include the same inflation syringe and extension line ●
- are sterilized using ethylene oxide
- provided for single patient use only, and
- are for use by healthcare professionals (physicians) only .

There are some minor design differences between the subject and predicate device:

- . The subject Audion dilation device features a non-malleable shaft that cannot be reshaped and is only intended for Eustachian Tube dilation, while the predicate device [K163509] has a malleable shaft that may be reshaped and is intended to dilate both Eustachian tube and sinus anatomy.

{5}------------------------------------------------

- The subject device has a formed, atraumatic polymer ball tip while the predicate device ● [K163509] features an atraumatic stainless-steel ball tip. Unlike the predicate device [K163509], the Audion ET dilation system is not designed to accommodate the use of guidewires nor is it designed to allow for suction, irrigation or use of the LED Light Fiber.
- Audion ET dilation system includes an inflation lock for optional use .

### Substantial Equivalence:

The Audion ET dilation system has been shown to be substantially equivalent to the predicate device [K163509], based on Intended Use, Indications for Use, Principles of Operations and Technological Characteristics. The Intended Use and Indications for Use are identical to the predicate device's [K163509] Intended Use and Indications for Use with respect to Eustachian Tube balloon dilation. The main differences between the devices are that the predicate [K163509] has a malleable shaft that allows the predicate [K163509] to be configured for use in sinus anatomy and in the Eustachian tube while the subject device has a non-malleable shaft and is for use only in the Eustachian tube. Additionally, the predicate [K163509] has a stainless-steel ball tip while the subject device has a polymer ball tip.

While there are some minor design differences between the subject and predicate [K163509] devices, these differences do not raise different questions of safety or efficacy.

#### Performance Data:

Bench testing was performed to demonstrate that the Audion ET dilation system performs as intended and meets the design specification. The testing included biocompatibility, design verification (dimensional, functional, strength, HFE/UE verification testing), packaging, shelf life and design validation with ENT physicians (HFE/UE). Performance testing showed that the device meets design specifications and performs as intended.

## Conclusion:

The Audion ET dilation system has been shown to be substantially equivalent to the predicate device based on Indication for Use, Intended Use, Principles of Operation and technological characteristics. The minor design differences between the subject device and the predicate device (e.g., update to a non-malleable shaft with polymer ball tip) do not raise different questions of safety or efficacy and performance testing has been conducted to demonstrate the Audion ET dilation system performs as intended and is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/PNZ/K220027](https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/PNZ/K220027)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).