K181696 · Gyrus Acmi, Inc. · LYA · Nov 30, 2018 · Ear, Nose, Throat
Device Facts
Record ID
K181696
Device Name
ChitoZolve
Applicant
Gyrus Acmi, Inc.
Product Code
LYA · Ear, Nose, Throat
Decision Date
Nov 30, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4780
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
Indicated for use in patients undergoing nasal/sinus surgery as a space occupying hemostat/splint to: - Separate tissue or structures compromised by surgical trauma - Separate and prevent adhesions between mucosal surfaces during mesothelial cell regeneration in the nasal cavity - Help control minimal bleeding following surgery or trauma - Help control minimal bleeding following surgery or nasal trauma by tamponade effect blood absorption and platelet aggregation - Act as an adjunct to aid in the natural healing process Indicated for use as a nasal hemostat to treat epistaxis. Intended for use under the direction of a licensed healthcare provider.
Device Story
ChitoZolve is a dissolvable, porous, flexible nasal splint/hemostat. Input: saline solution for reconstitution. Operation: surgeon inserts dry splint into nasal cavity; splint absorbs saline, swells to exert pressure (tamponade) on bleeding sites, and acts as a physical spacer to prevent mucosal adhesions. Output: physical barrier/hemostat. Used in clinical settings by physicians following nasal/sinus surgery or for epistaxis. Benefits: controls bleeding, prevents adhesions, supports natural healing; dissolves within 2-14 days, eliminating need for manual removal.
Clinical Evidence
No clinical or animal testing conducted. Evidence based on non-clinical mechanical/functional bench testing (reconstitution, expansion, dissolution, platelet adhesion, coagulation, firmness, pliability, thickness, pH) and GLP-compliant biocompatibility testing (cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity).
Technological Characteristics
Materials: Carboxymethyl Chitosan, Methyl Cellulose, Hydroxyethyl Cellulose. Form: Absorbent porous sponge. Principle: Swellable, dissolvable nasal spacer/hemostat. Sterilization: Gamma radiation (ISO 11137). Standards: ISO 10993-1, 5, 10 (biocompatibility); ISO 14971 (risk); ISO 11137-1/2 (sterilization); ISO 11607-1/2 (packaging).
Indications for Use
Indicated for patients undergoing nasal/sinus surgery or experiencing epistaxis. Used as a space-occupying hemostat/splint to separate tissues, prevent adhesions, control minimal bleeding, and aid healing.
Regulatory Classification
Identification
An intranasal splint is intended to minimize bleeding and edema and to prevent adhesions between the septum and the nasal cavity. It is placed in the nasal cavity after surgery or trauma. The intranasal splint is constructed from plastic, silicone, or absorbent material.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services seal on the left and the FDA acronym and name on the right. The FDA part of the logo is in blue, with the acronym in a square and the name "U.S. Food & Drug Administration" written out in full.
November 30, 2018
Gyrus ACMI, Inc Dolan Mills Sr. Specialist, Regulatory Affairs, ENT 136 Turnpike Road Southborough, MA 01772
Re: K181696
Trade/Device Name: ChitoZolve Regulation Number: 21 CFR 874.4780 Regulation Name: Intranasal Splint Regulatory Class: Class I Product Code: LYA, EMX Dated: October 31, 2018 Received: November 1, 2018
Dear Dolan Mills:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device. please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Srinivas Nandkumar -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181696
Device Name ChitoZolve
Indications for Use (Describe)
Indicated for use in patients undergoing nasal/sinus surgery as a space occupying hemostat/splint to:
- Separate tissue or structures compromised by surgical trauma
- Separate and prevent adhesions between mucosal surfaces during mesothelial cell regeneration in the nasal cavity
- Help control minimal bleeding following surgery or trauma
- Help control minimal bleeding following surgery or nasal trauma by tamponade effect blood absorption and platelet aggregation
- Act as an adjunct to aid in the natural healing process
Indicated for use as a nasal hemostat to treat epistaxis.
Intended for use under the direction of a licensed healthcare provider.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Traditional 510(k) Notification K181696
## 510(k) Summary
## General Information
| Manufacturer and 510(k) Submitter: | Gyrus ACMI, Inc., an Olympus company<br>136 Turnpike Road<br>Southborough, MA 01772-2104<br>Phone: 1-800-262-3540<br>Fax: 1-901-373-0220 |
|-----------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment Registration Number: | 3003790304 |
| Contact Person: | Dolan Mills<br>Senior Specialist, Regulatory Affairs |
| Date Prepared: | Oct 31, 2018 |
| <b>Device Description</b> | |
| Classification Name: | Intranasal Splint |
| Regulatory Class<br>Regulation Number<br>Review Panel<br>Product Code | Class 1<br>21 CFR 874.4780 / 874.4100<br>Ear, Nose, & Throat Panel<br>LYA / EMX |
| Project Name: | Chitosan Splint (14-006) |
| Trade Name(s): | ChitoZolve, Chitosan IntraNasal Pack |
| Generic/Common Name: | Intranasal Splint |
| Model Number: | CNS155 |
{4}------------------------------------------------
#### Predicate Device
Hemostasis PosiSep / PosiSep X:
K122494
#### Product Description
Intranasal splints are utilized to stop nasal hemorrhage and for the prevention of tissue adhesion in the nasal and/or sinus cavities. It is often used after FESS, nasal/sinus surgery, or surgical trauma in the event of excessive epistaxis. Splints are gently inserted into the nasal cavity. The device is designed to swell within the nasal cavity in order to exert pressure on the bleeding sites.
The splint material is dissolvable. Splints can also be used following sinus surgery in order to hold cartilage and membranes in their proper places during healing following surgerv.
The Chitosan Intranasal Splint is made from Carboxymethyl Chitosan, mixed with Methyl Cellulose, and Hydroxyethyl Cellulose to make an absorbent porous sponge that is flexible and dissolvable. The device acts as a nasal splint and temporary spacer.
#### Technological Characteristics
Chitosan is a polysaccharide derivative of chitin which can be obtained from a number of natural sources (ie. shellfish chitin) and has well-documented medical and healing properties.
The device is used in the same way as the predicate. At the point of use, the splint is inserted by the surgeon into the patient's nasal cavity and reconstituted fully with a saline solution to expand and act as a moisture barrier and tissue spacer during healing. The nasal splints dissolve between 2-14 days after insertion.
The technological differences between the subject and predicate device are: dry weight % of Chitosan, and the use of Methyl Cellulose instead of Carboxymethylcellulose.
The performance of the device was compared against the known performance characteristics of the predicate device. Testing demonstrated that the performance requirements were met, and that the subject device meets all applicable user requirements. The technological differences do not impact the safety or efficacy of the device.
## Material
The device uses the same or similar patient-contacting materials that are utilized in the predicate device. The Chitosan Intranasal Splint is made from Carboxymethyl Chitosan, mixed with Methyl Cellulose, and Hydroxyethyl Cellulose to make an absorbent porous sponge that is flexible and dissolvable.
{5}------------------------------------------------
#### Intended Use / Indications
Indicated for use in patients undergoing nasal/sinus surgery as a space occupying hemostat/splint to:
- Separate tissue or structures compromised by surgical trauma
- Separate and prevent adhesions between mucosal surfaces during mesothelial cell regeneration in the nasal cavity
- Help control minimal bleeding following surgery or trauma
- Help control minimal bleeding following surgery or nasal trauma by tamponade effect blood absorption and platelet aggregation
- Act as an adjunct to aid in the natural healing process
Indicated for use as a nasal hemostat to treat epistaxis.
Intended for use under the direction of a licensed healthcare provider.
#### Compliance to Standards
The design of the device complies with the following standards:
ISO 10993-1, 5, 10, Biological Evaluation of Medical Devices ISO 14971, Risk Analysis ISO 15223-1:2016, Medical Devices - Symbols to be used ISO 11137-1:2006, Sterilization of medical devices, Radiation ISO 11137-2:2013, Sterilization of medical devices, Radiation ISO 11607-1:2006, Packaging for Terminally Sterilized Medical Devices ISO 11607-2:2006, Packaging for Terminally Sterilized Medical Devices ISO 11737-1:2006, Sterilization of Medical Devices - Microbiological Methods
Risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971. Design verification tests were identified and performed as a result of risk analysis assessment.
## Summary of Sterilization and Shelf Life Discussion
The device is provided sterile, single-use.
The device is sterilized with Gamma, using a cycle validated in accordance with ISO 11137 to provide a sterility assurance level of 10-6.
{6}------------------------------------------------
The Shelf Life period for the device was determined via testing and through an analysis of the shelf-life stability of the materials used in the design of the device, as well as an analysis of the packaging materials and processes used with other Gyrus ACMI devices. Shelf-life studies are on file to support the labeled shelf life.
#### Summary of Performance Testing
Performance tests were executed to ensure that the device functions as intended and meets design specifications. The following non-clinical and preclinical tests and usability studies were conducted:
## Non-Clinical / Preclinical Performance
Evidence of safety and effectiveness was obtained from two primary areas:
- 1) non-clinical (mechanical, functional) performance testing
- 2) preclinical (bench) evaluations and testing
Non-clinical: Basic safety and performance testing was performed. In addition, verification and comparison bench studies were conducted to evaluate the mechanical and functional performance.
#### Overview of Product Performance and Mechanical / Functional Testing
- a. Reconstitution (volume)
- b. Expansion (dimensional height)
- c. Dissolution time (In vitro)
- d. Platelet adhesion, whole blood coagulation
- e. Firmness
- f. Pliability
- g. Thickness
- h. pH
Biocompatibility testing: The splints are classified in accordance with ISO 10993-1, as a Surface Device, Breached or Compromised Surfaces, for Prolonged exposure (<14 days.). ISO 10993-1 and FDA Blue Book memo #G95-1 guidelines recommend that these devices have supporting data for cytotoxicity, sensitization, and irritation. Full GLP biocompatibility testing (Cytotoxicity, sensitization, irritation, acute systemic toxicity, and material-mediated pyrogenicity) is on file for these devices.
{7}------------------------------------------------
Stability: Production equivalent samples were subjected to accelerated and real time aging to confirm that the devices maintain functionality and continue to meet specifications over time. The results of the age testing demonstrate that the device will be stable for the stated shelf-life. Samples were also subjected to environmental conditioning and ship testing.
Preclinical: Evidence obtained from preclinical bench studies demonstrate that the device performs substantially equivalent to the predicate device in relevant aspects associated with performance.
Testing demonstrated that the device performs as well as or better than the predicate device.
No animal or clinical testing was conducted. The use of the device type has been documented in published literature and indicates safe and effective use for the target procedures and expected patient populations.
## Substantial Equivalence
The device is equivalent to the predicate device both physically and with respect to performance.
The device has the same intended use / indications for use as the predicate.
The device uses similar or same patient-contacting materials in similar quantities that are utilized in the predicate device.
## Conclusion
The performance of the device was compared against the known performance characteristics of the predicate device. Testing demonstrated that the performance requirements were met, and that the device exhibited comparable performance characteristics to the predicate.
In summary, the device is substantially equivalent to the predicate device and presents no new questions of safety or efficacy.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.