ADVACOAT SINUS GEL, MODEL 9100, ADVA SINUS STENT, MODEL 9200
Applicant
Carbylan Biosurgery
Product Code
LYA · Ear, Nose, Throat
Decision Date
Dec 4, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4780
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
AdvaCoat™ Sinus Gel and Stent are indicated for use in patients undergoing nasal/sinus surgery as a space-occupying dressing and/or stent intended to prevent adhesions in the nasal cavity, separate mucosal surfaces compromised by surgery or trauma, minimize edema and help control minimal bleeding following surgery or trauma, and act as an adjunct to aid in the natural healing process.
Device Story
AdvaCoat™ Sinus Gel and Stent are sterile, bioresorbable, viscoelastic materials composed of cross-linked derivatized hyaluronan. Provided as a gel (syringe-delivered) or dry stent, the device acts as a space-occupying dressing. Upon contact with fluids, it forms a gelatinous, hygroscopic mass that adheres to mucosal surfaces, preventing migration. Used in clinical settings by physicians following nasal/sinus surgery or trauma to separate mucosal surfaces, minimize edema, and control minimal bleeding. The device is naturally eliminated or removed via gentle irrigation; any residual material remaining at 14 days may be removed by the surgeon. Benefits include adhesion prevention and support of the natural healing process.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing, biocompatibility data (ISO 10993-1), and comparison of technological characteristics to legally marketed predicate devices.
Technological Characteristics
Material: Cross-linked derivatized hyaluronan. Form: Gel (syringe) or dry stent. Principle: Hygroscopic, forms gelatinous mass in contact with fluids. Biocompatibility: ISO 10993-1. Sterilization: E-beam radiation (SAL 10^-6). Resorption: Bioresorbable; residual material removable at 14 days. No electronic components or software.
Indications for Use
Indicated for patients undergoing nasal/sinus surgery or trauma; used as space-occupying dressing/stent to prevent adhesions, separate mucosal surfaces, minimize edema, control minimal bleeding, and aid natural healing.
Regulatory Classification
Identification
An intranasal splint is intended to minimize bleeding and edema and to prevent adhesions between the septum and the nasal cavity. It is placed in the nasal cavity after surgery or trauma. The intranasal splint is constructed from plastic, silicone, or absorbent material.
{0}------------------------------------------------
K 063308
### Premarket Notification 510(k) Summary
Tab 3
DEC - 4 2006
#### Date Prepared:
September 1, 2006
### Submitter (Contact):
Roxanne Dubois, VP, Regulatory Affairs Carbylan BioSurgery, Inc. Address: 3181 Porter Drive, Palo Alto, CA 94304 Telephone: 650-855-6775 Fax: 650-855-9119 Email: rdubois@carbylan.com
#### Device Name and Classification:
Proprietary Name: AdvaCoat™ Sinus Gel and Stent Common Name: Sinus Packing, Gel or Stent Classification Name: Splint, Intranasal Splint Classification: Class f, Per 21 CFR 874.4780 Product Code Number: LYA
#### Performance Standards:
No FDA performance standards exist for this product.
#### Predicate Devices:
Hyalsine Hylan B Gel (K012532) MeroPack (K041381) Merogel Nasal Dressing And Sinus Stent (K982731)
#### Device Description:
AdvaCoat™ Sinus Gel and Stent (AdvaCoat™) is a sterile, non-pyrogenic, non-swelling, viscoelastic, bioresorbable material composed of cross-linked polymers of a derivatized hyaluronan. AdvaCoat™ is produced from a nonanimal, non-pathogenic source using a highly purified hyaluronan. AdvaCoat™ Sinus Gel and Stent and its break down products are biocompatible, non-immunoqenic and non-toxic.
AdvaCoat™ Sinus Gel and Stent are provided in two formats, a gel and a stent. The gel is delivered using a syringe and the stent is applied dry as a packing material.
Due to its tissue adhesive properties, AdvaCoat™ will not migrate or be dislodged from the application site. AdvaCoat™ may be used as a space occupying dressing and/or stent to minimize bleeding. Upon application,
{1}------------------------------------------------
K063308
AdvaCoat™ forms a viscous and transparent gel conforming to mucosal surfaces.
The final packaged product is sterilized with E-beam radiation to a Sterility Assurance Level (SAL) of 10th and intended for single use only. The sterilization information for AdvaCoat™ is substantially equivalent to the data for the predicate devices.
#### Significant Performance Characteristics
There are no new performance characteristics for AdvaCoat™ compared to the substantially equivalent predicate devices marketed for sale. These devices are indicated for use to prevent adhesions and minimize edema and bleeding following nasal/sinus surgery or trauma.
#### Indication For Use:
AdvaCoat™ Sinus Gel and Stent are indicated for use in patients undergoing nasal/sinus surqery as a space-occupying dressing and/or stent intended to prevent adhesions in the nasal cavity, separate mucosal surfaces compromised by surgery or trauma, minimize edema and help control minimal bleeding following surgery or trauma, and act as an adjunct to aid in the natural healing process.
#### Technological Characteristics:
There are no significant differences in the technological characteristics of this device compared to the predicate devices which adversely affect safety or effectiveness. Provided below is a table summarizing and comparing the technological characteristics of AdvaCoat™ and the predicate devices.
| Attribute | Proposed Device<br>AdvaCoat™, Gel<br>and Stent | Hyalsine Hylan B<br>Gel (K012532) | MeroPack<br>(K041381) | Merogel Nasal<br>Dressing And<br>Sinus Stent<br>(K982731) |
|------------------|--------------------------------------------------|-----------------------------------|-----------------------|-----------------------------------------------------------|
| Company<br>Name: | Carbylan<br>BioSurgery, Inc. | Genzyme | Medtronic<br>Xomed | Xomed, Inc. |
| Date Cleared: | N/A | 10/30/01 | 9/10/04 | 2/2/99 |
| Code/Class: | LYA, Class I | SAME | SAME | EMX, Class I |
| 21 CFR No. | 874.4780 | SAME | SAME | SAME |
| Device Name | splint, intranasal<br>septal | SAME | SAME | balloon, epistaxis |
| Intended Use | nasal / sinus<br>surgery | SAME | SAME | SAME |
| Indications | Post-op, help<br>control minimal<br>bleeding and | SAME | SAME | SAME |
#### Technological Characteristics of AdvaCoat™ and Predicate Devices Table 1.1
{2}------------------------------------------------
K063308
| | separate mucosal surfaces / adhesion prevention | | | |
|----------------------------|------------------------------------------------------------------------------------------------------------------------|---------|--------------------------------------------------|--------------------------------------------------|
| Material /<br>Construction | Derivative<br>hyaluronic acid | SAME | Derivative<br>hyaluronic acid and<br>collagen | SAME |
| Absorbant<br>Qualities | In excess of 10<br>times weight of the<br>device | unknown | In excess of 10<br>times weight of the<br>device | In excess of 10<br>times weight of the<br>device |
| Sterility | E-beam irradiation | unknown | Gamma irradiation | Gamma irradiation |
| Resorption<br>Time | If the material is still<br>seen at 14 days<br>follow-up, any<br>residual material<br>may be removed by<br>the surgeon | SAME | SAME | SAME |
| Biocompatibility | ISO 10993-1 | SAME | SAME | SAME |
| Method of<br>Action | Hygroscopic, forms<br>gelatinous mass in<br>contact with fluids | SAME | SAME | SAME |
| Method of<br>Removal | Natural elimination<br>or gentle irrigation<br>of residues | SAME | SAME | SAME |
| Bioreabsorbable | Yes | SAME | SAME | SAME |
#### Summary of Substantial Equivalence:
The information submitted in this Premarket Notification supports a determination that AdvaCoat™ Gel and Stent are substantially equivalent to the predicate devices. The design, technological characteristics and intended use (safety and effectiveness) of AdvaCoat™ are substantially equivalent to those for the predicate devices (Hyalsine Hylan B Gel, K012532; MeroPack, K041381; and, Merogel Nasal Dressing And Sinus Stent, K982731). The risks and benefits to the patient are the same as those normally attributed to the use of an intranasal splint. The technological characteristics, design and intended use are well known, low risk, and are substantially equivalent to the previously cleared predicate devices. In addition, the biocompatibility data support the safety of the material.
#### Conclusions
The device meets all the biocompatibility test requirements and is substantially equivalent in design, intended use and technological characteristics to the predicate devices. Therefore, a determination can be made that AdvaCoat™ Gel and Stent are considered substantially equivalent to the predicate devices.
{3}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Carbylan Biosurgery c/o Roxanne J. Dubois Vice President, Regulatory Affairs 3181 Porter Drive Palo Alto, CA 94304
DEC - 4 2006
Re: K063308
Trade/Device Name: AdvaCoat™ Sinus Gel and Stent Regulation Number: 21 CFR 874.4780 Regulation Name: Intranasal splint Regulatory Class: Class I Product Code: LYA Dated: November 1, 2006 Received: November 2, 2006
Dear Ms. Dubois:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
#### Page 2 - Roxanne J. Dubois
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
M.B. Eichelman SMD
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: AdvaCoat™ Sinus Stent
Indications for Use:
AdvaCoat™ Sinus Stent is intended for use in patients undergoing nasal/sinus surgery as a space-occupying dressing and/or stent intended to prevent adhesions in the nasal cavity, separate mucosal surfaces compromised by surgery or trauma, minimize edema and help control minimal bleeding following surgery or trauma, and act as an adjunct to aid in the natural healing process.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Kuu A. Bble
Division Sian-Off Division in Ophmalmic Far. Nose and Throat Devises
× Prescription Use (Per 21 CFR 801.109)
510(k) Number.
000023
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name: AdvaCoat™ Sinus Gel
Indications for Use:
AdvaCoat™ Sinus Gel is intended for use in patients undergoing nasal/sinus surgery as a space-occupying dressing and/or stent intended to prevent adhesions in the nasal cavity, separate mucosal surfaces compromised by surgery or trauma, minimize edema and help control minimal bleeding following surgery or trauma, and act as an adjunct to aid in the natural healing process.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
:
Karen A. Baker
n of Onthalmic Ear, e and Tribat Devises
510(k) Number
000022
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.