K181838 · 3nt Medical , Ltd. · LRC · Dec 20, 2018 · Ear, Nose, Throat
Device Facts
Record ID
K181838
Device Name
Sinusway Dilation System
Applicant
3nt Medical , Ltd.
Product Code
LRC · Ear, Nose, Throat
Decision Date
Dec 20, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4420
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The Sinusway Dilation System is intended to access and treat the frontal, maxillary and sphenoid sinuses in sinus procedures in adults using a trans-nasal approach, by dilation and displacement of the anatomic structures along the sinus drainage pathways.
Device Story
Sinusway Dilation System is a manual surgical instrument for sinus procedures; used by surgeons in clinical settings. System comprises a single-use sterile Dilation Kit and a re-processible, autoclavable Holder. Dilation Kit includes an inflatable balloon device, three interchangeable pre-shaped seekers (stylets) to adjust distal end geometry, and an inflation device. Operation involves advancing the device into nasal/sinus anatomy under direct endoscopic visualization; seeker provides guidance to sinus drainage pathways; balloon is manually inflated with saline to dilate and displace anatomic structures. Inflation device provides visual and tactile pressure feedback. Device facilitates sinus drainage; benefits patient by restoring sinus patency.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including balloon fatigue, inflation/deflation timing, dimensional attributes, compliance (diameter vs. pressure), rated burst pressure, bond strength, inflation mechanism pressure feedback, and simulated use functionality. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, irritation, sensitization, acute systemic toxicity). Sterilization validated per ISO 11135-1.
Technological Characteristics
Manual surgical instrument; balloon dilation technology. Components: single-use sterile dilation device with inflatable balloon, interchangeable pre-shaped seekers, and syringe-based inflation device. Holder is re-processible/autoclavable. Biocompatibility per ISO 10993-1. Sterilization via EtO (single-use components). No software or electronic components.
Indications for Use
Indicated for adults requiring access and treatment of frontal, maxillary, and sphenoid sinuses via trans-nasal approach for dilation and displacement of sinus drainage pathway structures.
Regulatory Classification
Identification
An ear, nose, and throat manual surgical instrument is one of a variety of devices intended for use in surgical procedures to examine or treat the bronchus, esophagus, trachea, larynx, pharynx, nasal and paranasal sinus, or ear. This generic type of device includes the esophageal dilator; tracheal bistour (a long, narrow surgical knife); tracheal dilator; tracheal hook; laryngeal injection set; laryngeal knife; laryngeal saw; laryngeal trocar; laryngectomy tube; adenoid curette; adenotome; metal tongue depressor; mouth gag; oral screw; salpingeal curette; tonsillectome; tonsil guillotine; tonsil screw; tonsil snare; tonsil suction tube; tonsil suturing hook; antom reforator; ethmoid curette; frontal sinus-rasp; nasal curette; nasal rasp; nasal rongeur; nasal saw; nasal scissors; nasal snare; sinus irrigator; sinus trephine; ear curette; ear excavator; ear rasp; ear scissor, ear snare; ear spoon; ear suction tube; malleous ripper; mastoid gauge; microsurgical ear chisel; myringotomy tube inserter; ossici holding clamp; sacculotomy tack inserter; vein press; wire ear loop; microrule; mirror; mobilizer; ear, nose, and throat punch; ear, nose and throat knife; and ear, nose, and throat trocar.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue.
December 20, 2018
3NT Medical Ltd. % Orly Maor Company Regulatory Consultant 25 Sirkin Street Kfar Saba 44421, Israel
Re: K181838
Trade/Device Name: Sinusway Dilation System Regulation Number: 21 CFR 874.4420 Regulation Name: Ear, Nose, And Throat Manual Surgical Instrument Regulatory Class: Class I Product Code: LRC Dated: November 28, 2018 Received: November 28, 2018
Dear Orly Maor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device. please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
## Srinivas Nandkumar -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K181838
Device Name Sinusway Dilation System
Indications for Use (Describe)
The Sinusway Dilation System is intended to access and treat the frontal, maxillary and sphenoid sinuses in sinus procedures in adults using a trans-nasal approach, by dilation and displacement of the anatomic structures along the sinus drainage pathways.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10pt;"> ☑ Prescription Use (Part 21 CFR 201 Subpart D) </span> | <span style="font-size: 10pt;"> ☐ Over-The-Counter Use (21 CFR 201 Subpart C) </span> |
|-------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for 3NT MEDICAL. The logo is composed of the number 3, the letters NT, and the word MEDICAL. The number 3 and the letters NT are in blue, while the word MEDICAL is in gray.
Sinusway Dilation System K181838
### Traditional Premarket Notification Submission - 510(k) Sinusway Dilation System 510(k) Number K181838
Date Prepared: December 13, 2018
#### I. SUBMITTER
3NT Medical Ltd. 22 Hamelacha Street, PO Box 11384, Rosh Ha'ayin 4809169, Israel Tel: +972.73.7154057 Fax: +972.73.7154058
#### Contact Person
Orly Maor 25A Sirkin Street Kfar Saba 44421, Israel Tel: +972-9-7453607 Fax: +972-153-9-7453607 oram.ma@gmail.com
#### II. DEVICE
Name of Device: Sinusway Dilation System Common or Usual Name: Sinusway Dilation System Classification Name: Ear, nose, and throat manual surgical instrument (21 CFR 874.4420) Regulatory Class: I (510(k) exempt) Product Code: LRC
#### III. PREDICATE DEVICE
3NT Medical Ltd. believes that the Sinusway Dilation System is substantially equivalent to the following predicate device cleared under multiple clearances:
- Entellus Medical Inc. XprESS Multi-Sinus Dilation Tool cleared under K102003, K112506 and K121174, Product code LRC.
{4}------------------------------------------------
## BNTMEDICAL
#### IV. DEVICE DESCRIPTION
Sinusway Dilation System combines the tissue expansion effect of balloon dilation with the features of a curved sinus seeker, and by that allows the user to track the dilation device into the sinus drainage pathways. The shape of the distal end of the dilation system can be changed by inserting pre-shaped seekers into it. The system includes:
- A Dilation Kit (Single use, provided sterile) consists of the following 1. components:
- a. A Dilation Device includes an inflatable balloon at its distal end and connects to the inflation device at its proximal end.
- b. Three interchangeable pre-shaped Seekers (also referred to as stylets).
- c. An Inflation Device connects to the dilation device and inflates it while providing visual and tactile pressure indication.
- All Dilation Kit components are disposed of at the end of the procedure.
- 2. A Holder (re-processible and auto-clavable) - holds the components of the Dilation Kit in place and locks them, allowing the user to operate the dilation tool.
#### V. INDICATIONS FOR USE
Sinusway Dilation System is intended to access and treat the frontal, maxillary and sphenoid sinuses in sinus procedures in adults using the trans-nasal approach, by dilation and displacement of the anatomic structures along the sinus drainage pathways.
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Sinusway Dilation System has the same intended use and indications for use as the Entellus predicate device. Sinusway Dilation System, same as the predicate device, use balloon dilation technology to access and treat the target anatomy. Both balloons are positioned over a rigid pre-shaped seeker and are manually inflated with saline until the desired result is achieved. The Sinusway Dilation System and Entellus predicate device principles of operation are identical. Both devices are advanced into the nasal and sinus anatomy under direct endoscopic visualization and treat the anatomy by dilating and displacing anatomic structures along the sinus drainage pathways. Both devices are used through a trans-nasal approach.
Similar tests and tests methods performed in accordance with the same standards were used in both Sinusway Dilation System and the predicate device to validate the design. The testing results showed that the minor differences in device characteristics and principles of operation between the subject device and predicate device do not raise any new questions of safety or effectiveness.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for 3NT Medical. The "3NT" is in a dark teal color, while "MEDICAL" is in gray. The logo is simple and modern.
### Sinusway Dilation System K181838
The Sinusway Dilation System has the same technological characteristics as the predicate devices as demonstrated in the table below:
| Name | 3NT Medical<br>Sinusway Dilation<br>System | Entellus Medical<br>XprESS Multi-Sinus<br>Dilation Tool | SE Justification |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| 510(k) number | K181838 | K102003 and subsequent<br>K112506 and K121174 | |
| Product Code | LRC | LRC | Same |
| CFR | Class I, 21 CFR 874.4420 | Class I, 21 CFR 874.4420 | Same |
| Indications for<br>Use | Intended to access and treat<br>the frontal, maxillary and<br>sphenoid sinuses in sinus<br>procedures in adults, by<br>dilation and displacement<br>of the anatomic structures<br>along the sinus drainage<br>pathways. | Intended to access and treat<br>the frontal recesses.<br>sphenoid sinus ostia,<br>maxillary ostia/ethmoid<br>infundibula in adults using<br>a trans-nasal approach. The<br>bony sinus outflow tracts<br>are remodeled by balloon<br>displacement of adjacent<br>bone and paranasal sinuses<br>structure. | Same |
| Balloon Diameter | 6mm | 5mm, 6mm, 7mm | Same<br>The 3NT size is<br>included within the<br>predicate range of<br>sizes |
| Balloon Length | 10mm | 8mm, 18mm | SE<br>The 3NT size is<br>included within the<br>predicate range of<br>sizes |
| Distal Ball Tip<br>Diameter | 1.5mm | 2mm | SE |
| Dilation Device<br>Angles | Similar to sinus seekers,<br>pre-set by replaceable pre-<br>bent seekers | Similar to sinus seekers,<br>pre-set by bending a<br>shapeable seeker | SE |
| Guided by Direct<br>Visualization? | Yes | Yes | Same |
| Trans-Nasal<br>Approach? | Yes | Yes | Same |
| Inflation device | Syringe barrel and plunger<br>based | Syringe barrel and plunger<br>based | Same |
| Name | 3NT Medical<br>Sinusway Dilation<br>System | Entellus Medical<br>XprESS Multi-Sinus<br>Dilation Tool | SE Justification |
| Balloon inflation<br>substance | Saline | Saline | Same |
| Pressure<br>Indication | Visual and tactile | Visual only | SE |
| Rated Burst<br>Pressure | 17.4 atm | 19-22 atm (6mm balloon) | Similar, both are<br>above the maximal<br>inflation pressure |
| Dilation Tool<br>Sterilization | EtO (single-use) | EtO (single-use) | Same |
| Biocompatibility | All surfaces which come in<br>contact with the patient's<br>mucosal surfaces were<br>successfully tested for<br>biocompatibility in<br>accordance with ISO<br>10993-1. | All surfaces which come in<br>contact with the patient's<br>mucosal surfaces were<br>successfully tested for<br>biocompatibility in<br>accordance with ISO<br>10993-1. | Same |
| Prescription Use | The device should be used<br>only by trained surgeon<br>under a physician order. | The device should be used<br>only by trained surgeon<br>under a physician order. | Same |
{6}------------------------------------------------
# 3NTMEDICAL
#### Sinusway Dilation System K181838
Based on the above analysis, 3NT Medical Ltd. believes that the Sinusway Dilation System is substantially equivalent to the legally marketed predicate devices
#### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination:
- Risk analysis per ISO 14971:2012 -
#### -Biocompatibility testing
An evaluation of biocompatibility was performed in compliance with ISO 10993-1. Biocompatibility evaluation included cytotoxicity, irritation, sensitization and acute systemic cytotoxicity testing. All tests were completed with passing results.
#### Sterilization, Packaging and Shelf Life Testing -
Sterilization validation testing of the endoscope was performed to demonstrate compliance with ISO 11135-1. The handle was validated for cleaning and re-use by autoclave. In addition, shelf life and packaging testing were performed to support the labeled shelf life. All tests were successfully completed.
{7}------------------------------------------------
# ЗПТМЕДІСАЦ
Sinusway Dilation System K181838
#### Bench Testing -
Bench testing included the following:
- Balloon Dilation Tool Fatigue ●
- Inflation/Deflation time test protocol ●
- Visual and demonstration Report
- Dimensional Attribute ●
- Compliance (Diameter vs. Pressure) Report ●
- Balloon Rated Burst Pressure Report
- Bond Strength Report
- Inflation Mechanism Pressure feedback range testing Report ●
- Functionality and simulated use ●
All tests met the predefined acceptance criteria.
#### VIII. CONCLUSIONS
The Sinusway Dilation System has the same intended use as the predicate device. The principal features of the device that were described, as well as the testing provided, show that the minor differences in device characteristics between the subject device and predicate device do not raise any new questions of safety or effectiveness. Performance data has been provided, establishing that the Sinusway Dilation System performs as intended and in a manner that is substantially equivalent to the predicate. Therefore, the device may be found substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.