K181546 · Intuit Medical Products, LLC · LRC · Jan 3, 2020 · Ear, Nose, Throat
Device Facts
Record ID
K181546
Device Name
Dillard Nasal Balloon Catheter
Applicant
Intuit Medical Products, LLC
Product Code
LRC · Ear, Nose, Throat
Decision Date
Jan 3, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4420
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The Dillard Nasal Balloon (DNB) Catheter is an instrument intended to provide increased intransal space to facilitate access for endonasal and transnasal procedures and/or temporarily address nasal obstruction by displacing the inferior turbinate and lower nasal septum. The Dillard Nasal Balloon (DNB) Catheter is intended for use in ages 17 years or older.
Device Story
Dillard Nasal Balloon (DNB) Catheter is a single-use, sterile, manual surgical instrument. Device consists of a catheter with a distal inflatable balloon, a plastic hub with a luer connector for inflation, and a secondary lumen for guidewire use. Available in lengths of 25cm or 100cm with balloon diameters from 5mm to 12mm. Radiopaque markers on the balloon assist with radiographic positioning. Operated by a physician in a clinical setting; the device is inserted into the nasal cavity. Hydraulic pressure is applied via an external inflation device to expand the balloon, which mechanically dilates internal passages by displacing the inferior turbinate and lower nasal septum. This provides increased space for procedures or temporary relief of obstruction. The device is identical to previously cleared Dillard catheters (K143738, K153484) except for the indication for use.
Clinical Evidence
No clinical data was required. Substantial equivalence was established through bench testing, including dimensional specifications, joint separation strength, balloon burst pressure, inflation/deflation times, balloon cycle fatigue in simulated use, cadaveric simulated use, biocompatibility (ISO 10993), sterilization validation (ISO 11135), packaging integrity (ISO 11607/ISTA 2A), and shelf-life testing (ASTM F1980).
Technological Characteristics
Materials: medical grade polymers and metal alloys. Principle: hydraulic pressurization via inflatable balloon for mechanical dilation. Dimensions: 25cm/100cm length; 5-12mm diameter. Connectivity: none. Sterilization: 100% Ethylene Oxide gas. Single-use. Compatible with standard inflation devices and guidewires.
Indications for Use
Indicated for patients 17 years or older requiring increased intranasal space for endonasal/transnasal procedures or temporary relief of nasal obstruction via displacement of the inferior turbinate and lower nasal septum.
Regulatory Classification
Identification
An ear, nose, and throat manual surgical instrument is one of a variety of devices intended for use in surgical procedures to examine or treat the bronchus, esophagus, trachea, larynx, pharynx, nasal and paranasal sinus, or ear. This generic type of device includes the esophageal dilator; tracheal bistour (a long, narrow surgical knife); tracheal dilator; tracheal hook; laryngeal injection set; laryngeal knife; laryngeal saw; laryngeal trocar; laryngectomy tube; adenoid curette; adenotome; metal tongue depressor; mouth gag; oral screw; salpingeal curette; tonsillectome; tonsil guillotine; tonsil screw; tonsil snare; tonsil suction tube; tonsil suturing hook; antom reforator; ethmoid curette; frontal sinus-rasp; nasal curette; nasal rasp; nasal rongeur; nasal saw; nasal scissors; nasal snare; sinus irrigator; sinus trephine; ear curette; ear excavator; ear rasp; ear scissor, ear snare; ear spoon; ear suction tube; malleous ripper; mastoid gauge; microsurgical ear chisel; myringotomy tube inserter; ossici holding clamp; sacculotomy tack inserter; vein press; wire ear loop; microrule; mirror; mobilizer; ear, nose, and throat punch; ear, nose and throat knife; and ear, nose, and throat trocar.
Predicate Devices
Acclarant RELIEVA TRACT™ Balloon Dilation System (K183090)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below.
January 3, 2020
Intuit Medical Products, LLC Jack Griffis Vice President, R&D 6018 Eagles Rest Trail Sugar Hill, Georgia 30518
Re: K181546
Trade/Device Name: Dillard Nasal Balloon Catheter Regulation Number: 21 CFR 874.4420 Regulation Name: Ear, Nose, And Throat Manual Surgical Instrument Regulatory Class: Class I, reserved Product Code: QGK Dated: October 7, 2019 Received: October 8, 2019
Dear Jack Griffis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Michael J. Ryan Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K181546
Device Name Dillard Nasal Balloon (DNB) Catheter
#### Indications for Use (Describe)
The Dillard Nasal Balloon (DNB) Catheter is an instrument intended to provide increased intransal space to facilitate access for endonasal and transnasal procedures and/or temporarily address nasal obstruction by displacing the inferior turbinate and lower nasal septum.
The Dillard Nasal Balloon (DNB) Catheter is intended for use in ages 17 years or older.
Type of Use (Select one or both, as applicable)
| <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|-------------------------------------------------------------|------------------------------------------------------------|
|-------------------------------------------------------------|------------------------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a logo for Intuit Medical Products, LLC. On the left side of the logo is a blue illustration of a human head and neck in profile, showing the nasal passages and throat. To the right of the illustration is a light blue rectangle containing the letters "IMP" in large, white font, with the words "Intuit Medical Products, LLC" written in a smaller, white font below the letters.
# 510(k) Summary
510(k) Number: K181546
December 31st, 2019 Date Revised:
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
- A. Submitter: Intuit Medical Products (IMP), LLC 6018 Eagles Rest Trail Sugar Hill, Georgia 30518
- B. Company Contact: Jack Griffis Vice President, Research & Development (404) 583-6889 (direct) jgriffis@intuitmedicalproducts.com
- C. Device Information: Trade Name: Dillard Nasal Balloon Catheter (DNB) Common Name: Balloon, Nasal Airway
- D. Classification: Ear, nose, and throat manual surgical instruments QGK, 21 CFR 874.4420 Class I
- E. Predicate Devices: Acclarant RELIEVA TRACT™ Balloon Dilation System, K183090
Additional Reference Devices (for technological comparison ONLY): ECONOM™ speculum for ENT applications, K925754 Dillard Sinuplasty Balloon Catheter (DSB), K143738 Dillard Airway Dilatation System (DAS), K153484
- F. Physical Description:
The Dillard Nasal Balloon (DNB) Catheter has a total length of either 25cm or 100cm and is available in balloon diameters from 5mm to 12mm and balloon lengths of 20mm and 40mm. The balloon component of the catheter has radiopaque markers to assist with radiographic positioning (as applicable). The proximal end of the device is a common catheter design consisting of a plastic hub and strain relief. The hub is used to inflate the balloon and the luer connector integrated into the hub is compatible with standard inflation devices. A second lumen within the catheter, intended for guidewire use, extends through the central lumen back to the proximal hub and through the distal tip. All catheters and accessories are supplied sterile and intended for single use.
NOTE: the proposed device is identical to those previously cleared under K143738 and K153484 with the exception of the indication for use.
{4}------------------------------------------------
- G. Indications for Use:
The Dillard Nasal Balloon Catheter is an instrument intended to provide increased intranasal space to facilitate access for endonasal and transnasal procedures and/or temporarily address nasal obstruction by displacing the inferior turbinate and lower nasal septum.
The Dillard Nasal Balloon Catheter is intended for use in ages 17 years or older.
- H. Comparison of Characteristics / Performance Testing / Substantial Equivalence: The DNB Catheter is substantially equivalent to the predicate devices in intended use, indications for use or fundamental scientific technology and important performance specifications. Refer to the Table below for a summary:
| Characteristic | Proposed Device<br>(Dillard Nasal Balloon Catheter,<br>K181546) | Predicate Device<br>(Acclarent TRACT™ Balloon<br>Catheter, K183090) | Reference Device<br>(Single use speculum for<br>ENT applications, K925754) |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use | "…an instrument intended to<br>provide increased intranasal<br>space to facilitate access for<br>endonasal and transnasal<br>procedures and/or temporarily<br>address nasal obstruction by<br>displacing the inferior turbinate<br>and lower nasal septum" | "…an instrument intended to<br>provide increased intranasal<br>space to facilitate access for<br>endonasal and transnasal<br>procedures and/or temporarily<br>address nasal obstruction by<br>displacing the inferior turbinate<br>and lower nasal septum" | Speculum (ENT) – Procode<br>EPY: ...a device intended to<br>be inserted into a body<br>cavity to aid observation. It<br>is either nonilluminated or<br>illuminated and may have<br>various accessories |
| Classification Name | Ear, nose, and throat manual surgical instrument | | Speculum, ENT |
| Product Code | QGK | | EPY |
| Classification Section | 21 CFR 874.4420 | | 21 CFR 878.1800 |
| Principle of Operation | Operates on the principle of hydraulic pressurization applied<br>through an inflatable balloon attached to the distal end to<br>mechanically dilate internal passages | | Operates on the principle of<br>mechanical leverage<br>applied through a rigid<br>body and a fulcrum for<br>leverage to expand internal<br>passages |
| Materials | Various medical grade polymers and metal alloys - refer to Table A,<br>page 38 of the submission | | Various medical grade<br>polymers and metal alloys |
| Single Patient Use | Yes | Yes | Yes |
| Direct Patient Contact | Yes | Yes | Yes |
| Balloon Diameter and<br>Length | 5mm X 20mm & 40mm<br>6mm X 20mm & 40mm<br>7mm X 20mm & 40mm<br>8mm X 20mm & 40mm<br>9mm X 20mm & 40mm<br>10mm X 20mm & 40mm<br>11mm X 20mm & 40mm<br>12mm X 20mm & 40mm | 5mm x 24mm<br>7mm x 24mm<br>8.5mm x 24mm<br>10mm x 40mm<br>12mm x 40mm<br>14mm x 40mm<br>16mm x 40mm | N/A |
Image /page/4/Picture/6 description: The image shows a diagram of the human head and neck with the nasal passages and throat highlighted in blue and purple. To the right of the diagram is a blue square with the text "IMP Intuit Medical Products, LLC" in white. The text is stacked, with "IMP" on the top line, "Intuit Medical" on the second line, and "Products, LLC" on the third line.
{5}------------------------------------------------
| Characteristic | Proposed Device<br>(Dillard Nasal Balloon Catheter,<br>K181546) | Predicate Device<br>(Acclarent TRACT™ Balloon<br>Catheter, K183090) | Reference Device<br>(Single use speculum for<br>ENT applications, K925754) |
|-------------------------------|-----------------------------------------------------------------|---------------------------------------------------------------------|----------------------------------------------------------------------------|
| Maximum Inflation<br>Pressure | 10 – 12ATM | 8 – 16ATM | N/A |
| Flexible Shaft | Yes | Yes | No - rigid body |
| Catheter Length | 25CM and 100CM | 45CM | N/A |
| Sterilization Method | 100% Ethylene Oxide gas | | |
| Packaging | Pouch, Protective Tube, Stylet | | Pouch |
In addition, the device was subjected to the following performance tests to support the assertion of substantial equivalence:
- Dimensional specifications
- Joint separation strength
- 트 Compatibility with (standard) accessories
- l Balloon burst pressure
- I Inflation and deflation times
- . Balloon cycle fatigue in simulated use (simulated nasal septum and surrounding support structures)
- . Simulated use in cadavers
- . Biocompatibility Testing in compliance with the ISO 10993
- Sterilization Validation in compliance with the ISO 11135
- . Packaging Integrity and Transportation Validation in compliance with the ISO 11607 and ISTA Part 2A
- Shelf life testing to 3 years real-time equivalent in compliance with the ASTM F1980
- NOTE: no clinical testing was required to establish safety and efficacy
#### l. Conclusion from Clinical and Non-Clinical Testing:
No new questions of safety or effectiveness were identified for the Dillard Nasal Balloon Catheter compared to the predicate. Performance testing demonstrated the DNB performs as intended and is as safe and as effective as the predicate device. Therefore, the DNB is considered substantially equivalent to the predicate device.
Jack Griffis Vice President, Research & Development
Image /page/5/Picture/19 description: The image shows a diagram of the human respiratory system next to a logo for Intuit Medical Products, LLC. The diagram shows the nasal cavity, pharynx, larynx, and trachea. The logo is a blue rectangle with the letters "IMP" in large, white font. Below the letters, the words "Intuit Medical Products, LLC" are written in a smaller, white font.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.