← Product Code [LRC](/submissions/EN/subpart-e%E2%80%94surgical-devices/LRC) · K152435

# PassAssist LED Light Fiber (K152435)

_Entellus Medical, Inc. · LRC · Nov 20, 2015 · Ear, Nose, Throat · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/LRC/K152435

## Device Facts

- **Applicant:** Entellus Medical, Inc.
- **Product Code:** [LRC](/submissions/EN/subpart-e%E2%80%94surgical-devices/LRC.md)
- **Decision Date:** Nov 20, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 874.4420
- **Device Class:** Class 1
- **Review Panel:** Ear, Nose, Throat
- **Attributes:** Pediatric

## Intended Use

To locate, illuminate within, and transilluminate across nasal and sinus structures.

## Device Story

PathAssist LED Light Fiber is a sterile, single-use, flexible instrument for ENT procedures. Device comprises flexible illumination fiber, protective sheath, and integrated battery-powered LED light source. Upon activation, distal tip emits red light for >60 minutes. Used by physicians to locate and transilluminate nasal and sinus structures. Provides visual guidance during sinus procedures; assists in confirming anatomical positioning. Benefits include improved visualization of sinus structures during minimally invasive interventions.

## Clinical Evidence

Prospective, multicenter, single-arm study (IDE G140080) in 50 subjects aged 2-21 years. Evaluated 129 sinus illumination/transillumination attempts. Primary endpoints: technical success rate and complication rate. Results: 98.4% technical success rate (127/129 attempts); 0% complication rate; no device-related adverse events reported.

## Technological Characteristics

Flexible illumination fiber with 27.6cm working length and 0.375mm outer diameter. Integrated battery-powered LED light source. Single-use, sterile. Principle of operation: fiber-optic light transmission for transillumination. Identical materials and design to predicate.

## Regulatory Identification

An ear, nose, and throat manual surgical instrument is one of a variety of devices intended for use in surgical procedures to examine or treat the bronchus, esophagus, trachea, larynx, pharynx, nasal and paranasal sinus, or ear. This generic type of device includes the esophageal dilator; tracheal bistour (a long, narrow surgical knife); tracheal dilator; tracheal hook; laryngeal injection set; laryngeal knife; laryngeal saw; laryngeal trocar; laryngectomy tube; adenoid curette; adenotome; metal tongue depressor; mouth gag; oral screw; salpingeal curette; tonsillectome; tonsil guillotine; tonsil screw; tonsil snare; tonsil suction tube; tonsil suturing hook; antom reforator; ethmoid curette; frontal sinus-rasp; nasal curette; nasal rasp; nasal rongeur; nasal saw; nasal scissors; nasal snare; sinus irrigator; sinus trephine; ear curette; ear excavator; ear rasp; ear scissor, ear snare; ear spoon; ear suction tube; malleous ripper; mastoid gauge; microsurgical ear chisel; myringotomy tube inserter; ossici holding clamp; sacculotomy tack inserter; vein press; wire ear loop; microrule; mirror; mobilizer; ear, nose, and throat punch; ear, nose and throat knife; and ear, nose, and throat trocar.

## Predicate Devices

- PathAssist LED Light Fiber ([K141916](/device/K141916.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 20, 2015

Entellus Medical, Inc. Ms. Karen Peterson Vice President, Clinical, Regulatory And Quality 3600 Holly Lane North, Suite 40 Plymouth, MN 55447

Re: K152435

Trade/Device Name: PathAssist LED Light Fiber Regulation Number: 21 CFR 874.4420 Regulation Name: Ear, Nose, and Throat Manual Surgical Instrument Regulatory Class: Class I Product Code: LRC Dated: October 29, 2015 Received: October 30, 2015

Dear Karen Peterson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm_for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Eric A. Mann -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K152435

Device Name PathAssist LED Light Fiber

Indications for Use (Describe)

To locate, illuminate within, and transilluminate across nasal and sinus structures.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for "entellus MEDICAL". On the left side of the logo is a circular design made up of curved lines in black and blue. To the right of the design is the word "entellus" in a sans-serif font. Below the word "entellus" is the word "MEDICAL" in a smaller, sans-serif font with a trademark symbol.

510(k) Summary

Date Prepared: Submitter Information:

Establishment Registration:

Contact Information:

August 26, 2015 Entellus Medical, Inc. 3600 Holly Lane North, Suite 40 Plymouth, MN 55447

3006345872

Karen E. Peterson Vice President Clinical, Regulatory and Quality (763) 463-7066 kpeterson@entellusmedical.com

Device Information: Trade Name: Common Name: Classification Name: Product Code: Regulation Number:

PathAssist LED Light Fiber Sinus Guidewire ENT Manual Surgical Instrument LRC Class I, 21 CFR 874.4420

PathAssist LED Light Fiber [K141916]

# Device Description:

Predicate Device:

The PathAssist LED Light Fiber is a flexible instrument that emits light from the distal end. The LED Light Fiber is provided sterile and is for single use only. The device consists of a flexible illumination fiber, a protective sheath and an integrated battery powered LED light source. When the LED Light Fiber is activated the fiber will emit red light from the distal tip for over 60 minutes. It has a fiber nominal working length of 27.6cm with an outer diameter of 0.375mm (0.015").

The PathAssist LED Light Fiber is packaged alone or may be packaged with XprESS (LoProfile or Ultra).

Image /page/3/Figure/14 description: This image shows a diagram of a medical device. The device has a housing, an integrated light source, a flexible sheath, and a fiber. There is also a pull tab on the top of the device. 

PathAssist LED Light Fiber

# Indication for Use:

To locate, illuminate within, and transilluminate across nasal and sinus structures.

# Contraindications:

None known.

#### Technological Characteristics:

The LED Light Fiber has expanded indications for use, identical intended use, and identical scientific technology (i.e., principle of operation, design, function, materials, biocompatibility, packaging, shelf life, and sterilization) as the predicate device.

{4}------------------------------------------------

### Substantial Equivalence:

The LED Light Fiber has expanded indications for use, identical intended use, and identical fundamental scientific technology as the predicate device. The LED Light Fiber is substantially equivalent to the predicate device.

### Performance Data:

Performance testing of the LED Light Fiber consisted of clinical testing to support the expanded indications for use. A prospective, multicenter, single-arm study investigating the use of LED Light Fiber in patients aged 2-21 years was conducted in the United States under IDE G140080. The primary outcome measures were procedural technical success rate and complication rate. A total of 129 sinus illuminations/transilluminations were attempted in 50 subjects. A total of 127 attempts were successfully completed for an overall LED Light Fiber technical success rate of 98.4%. No complications (0%) were reported during the study. In addition, no adverse events related to the device procedure were reported during the study. Performance testing showed that the device performed as intended.

### Conclusion:

In conclusion, the technological characteristics are identical the predicate device. Performance testing has demonstrated that the device is safe and effective and that its performance is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/LRC/K152435](https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/LRC/K152435)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
