The DEFENDO Disposable Suction Valve is intended to be used to control the suction function of a compatible bronchoscope during a pulmonary procedure.
Device Story
Sterile, single-use disposable valve; attaches to bronchoscope suction port. User depresses valve to activate suction; controls flow of fluids from patient to suction pump/waste canister. Used in clinical pulmonary procedures. Provides mechanical control of suction; prevents leakage. Benefits patient by ensuring proper suctioning during bronchoscopy.
Clinical Evidence
Bench testing only. Performance data includes design verification (valve connection, depression force, suction flow), shelf-life validation (functional, dye penetration, peel strength, visual seal inspection), and sterilization validation.
Technological Characteristics
Sterile, single-use, disposable mechanical valve. Dimensions/geometry designed for compatibility with specific bronchoscope suction ports. No energy source; purely mechanical operation. Sterilized.
Indications for Use
Indicated for use by clinicians to control suction function of compatible bronchoscopes during pulmonary procedures.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized human figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 17, 2015
Medivators, Inc. % Ms. Kinnari Shah Sr. Regulatory Affairs Specialist 14605 28th Ave N Minneapolis, MN 55447
Re: K151522
Trade/Device Name: DEFENDO Disposable Suction Valve Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible or Rigid) and Accessories Regulatory Class: Class II Product Code: KTI Dated: "August 21, 2015" Received: August 20, 2015
Dear Ms. Kinnari Shah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control. and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151522
Device Name DEFENDO Disposable Suction Valve
#### Indications for Use (Describe)
The DEFENDO Disposable Suction Valve is intended to be used to control the suction function of a compatible bronchoscope during a pulmonary procedure.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Medivators, a Cantel Medical Company. The logo consists of a blue circular graphic on the left, followed by the word "MEDIVATORS" in blue capital letters. Below the company name is the text "A Cantel Medical Company" in a smaller font size, also in blue. The logo is clean and professional, with a focus on the company name and affiliation.
# Indications for Use
510(k) number (if known): K151522
Device Name: DEFENDO Disposable Suction Valve
#### Indications for Use:
The DEFENDO Disposable Suction Valve is intended to control the suction function of a compatible bronchoscope during a pulmonary procedure.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Medivators, a Cantel Medical Company. The logo consists of a blue circular graphic on the left, which has several curved lines inside of it. To the right of the graphic is the word "MEDIVATORS" in a large, sans-serif, blue font. Below the company name is the text "A Cantel Medical Company" in a smaller, sans-serif, blue font.
### 510(k) Summary
| Manufacturer: | Medivators Inc., a Cantel Medical Company |
|----------------------|--------------------------------------------------------------------|
| Address: | 3150 Pollock Drive,<br>Conroe, TX 77303<br>(800) 328-3345 |
| Official Contact: | Kinnari Shah<br>Sr. Regulatory Affairs Specialist, Medivators Inc. |
| Date Prepared: | August 21, 2015 |
| Trade Name: | <b>DEFENDO Disposable Suction Valve</b> |
| Common Name: | Bronchoscope Accessory |
| Classification Name: | Bronchoscope (flexible or rigid) and accessories |
| Product Code: | KTI |
| Device Class: | II |
| Regulation No: | 874.4680 |
Medivators Inc. has supplied the following information to support substantial equivalence of the DEFENDO Disposable Suction Valve to other Suction Valves currently cleared for sale in the United States of America.
#### 1. Intended Use
The DEFENDO Disposable Suction Valve is intended to be used to control the suction function of a compatible bronchoscope during a pulmonary procedure.
#### 2. Device Description
The subject device is a sterile, single use, disposable medical device. It is designed to be attached to the suction port of a bronchoscope during a pulmonary procedure to help the user engage in the suction function of the bronchoscope by depressing/activating the valve. The activation of the suction valve allows the user to control the suction flow of fluids from the patient to a suction pump/waste canister.
#### 3. Comparison to Other Devices in Commercial Distribution Within the United States
DEFENDO Disposable Suction Valve is equivalent in function, intended use and scientific technology to its predicate devices - Olympus Disposable Suction Valve cleared under 510(k) – K920025 and Defendo Disposable Suction Valve cleared under 510(k) – K102581.
Similarities between Subject and Predicate Devices - Subject and predicate devices have similar intended use, principle of operation and scientific technology. They are provided sterile and must be disposed after a single use. The subject device and the predicate device –
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for "MEDIVATORS A Cantel Medical Company". The logo consists of a blue circular icon with several curved lines inside, followed by the word "MEDIVATORS" in blue, and the text "A Cantel Medical Company" in a smaller font size below it. The logo is simple and professional, and the blue color gives it a clean and modern look.
Olympus Disposable Suction Valve have the same valve body (subcomponent of valve that mates with the endoscope) geometry and thus are compatible with the same bronchoscope models.
Difference between Subject and Predicate Devices - The only critical difference between the subject device and the predicate device - Olympus Disposable Suction Valve is the material of construction. However subject device has the same material of composition as its other predicate - Defendo Disposable Suction Valve cleared under 510(k) - K102581. When used as indicated, the subject device will not contact the patient either directly.
A device comparison table which supports substantial equivalence of the subject device is provided below:
| Device<br>Parameters | Subject Device –<br>DEFENDO Disposable<br>Suction Valve | Predicate Device –<br>Olympus Disposable Suction<br>Valve (K920025) | Predicate Device –<br>DEFENDO Disposable<br>Valve (K102581) |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | DEFENDO Disposable<br>Suction Valve | Olympus Disposable Suction<br>Valve | DEFENDO Disposable<br>Suction Valve |
| Regulation<br>Number | 874.4680 | 874.4680 | 876.1500 |
| Device Class | Class II | Class II | Class II |
| Certification<br>Panel | Ear, Nose & Throat | Ear, Nose & Throat | Gastroenterology/Urology |
| Product Code | KTI | EOQ | ODC |
| Indication of<br>Use | The DEFENDO Disposable<br>Suction Valve is intended to<br>be used to control the suction<br>function of a compatible<br>bronchoscope during a<br>pulmonary procedure. | The OLYMPUS<br>DISPOSABLE SUCTION<br>VALVE is an attachment for<br>Olympus bronchoscopes. The<br>device is designed to fit into<br>the suction port of the<br>endoscope to insure the<br>control and proper flow of<br>liquids (body fluids) through<br>the scope without leakage. | The DEFENDO Disposable<br>Suction Valve is intended<br>to be used to control the<br>suction function of an<br>endoscope during a GI<br>endoscopic procedure. |
| Compatible<br>Endoscopes | Compatible bronchoscopes | OLYMPUS® bronchoscopes | GI Endoscopes |
| Single Use | Yes | Yes | Yes |
| Supplied<br>sterile | Yes | Yes | Yes |
### Table 2 – Device Comparison Table
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Medivators, a Cantel Medical Company. The logo consists of a blue circular graphic with several curved lines inside, followed by the word "MEDIVATORS" in blue, block letters. Below the company name is the text "A Cantel Medical Company" in a smaller, sans-serif font, also in blue.
## 4. Summary of Non-Clinical Performance Data
Medivators has conducted the following testing to demonstrate substantial equivalence of DEFENDO Disposable Suction Valve-
- Design Verification Comparison Tests to Predicate Device ●
- o Valve Connection Test
- Valve Depression Force Test o
- o Suction Flow Test
- Shelf-life validation ●
- o Functional Testing
- Dye Penetration Testing o
- Peel Strength Testing о
- O Visual Seal Inspection
- Sterilization validation ●
### 5. Conclusion
Medivators Inc. has provided appropriate premarket notification information in the form of this 510(k) to support the substantial equivalence of the subject device to legally marketed predicate devices. The information and performance data provided indicates that the subject device is as safe and as effective as its predicate devices for its intended use when used in accordance with the device labeling.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.