MONTGOMERY RADIOPAQUE SALIVARY BYPASS TUBE

K962688 · Boston Medical Products, Inc. · KCF · Jul 25, 1996 · Ear, Nose, Throat

Device Facts

Record IDK962688
Device NameMONTGOMERY RADIOPAQUE SALIVARY BYPASS TUBE
ApplicantBoston Medical Products, Inc.
Product CodeKCF · Ear, Nose, Throat
Decision DateJul 25, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.4420
Device ClassClass 1
AttributesTherapeutic

Indications for Use

1. Treatment of pharyngocutaneous fistulas following laryngectomy or laryngoesophagectomy. 2. Management of cervical esophageal and hypopharyngeal strictures and fistulas resulting from a variety of causes including surgery, radiation, advanced malignancy, trauma, and ingestion of caustic agents.

Device Story

Silicone tube with funnel-shaped end; used for esophageal strictures and fistulas. Inserted by clinicians to bypass damaged areas. Provides mechanical support/conduit for saliva/food passage. Available in 8mm-20mm diameters. Radiopaque for visualization.

Clinical Evidence

Bench testing only. Finished device samples passed cytotoxicity testing (Elution Test, USP 23) and were determined to be non-reactive and non-cytotoxic.

Technological Characteristics

Silicone tube with funnel-shaped end; radiopaque. Available in 8mm, 10mm, 12mm, 14mm, 16mm, 18mm, and 20mm outside diameters. Biocompatibility verified via USP 23 Elution Test.

Indications for Use

Indicated for patients with pharyngocutaneous fistulas post-laryngectomy/laryngoesophagectomy or cervical esophageal/hypopharyngeal strictures/fistulas caused by surgery, radiation, malignancy, trauma, or caustic ingestion.

Regulatory Classification

Identification

An ear, nose, and throat manual surgical instrument is one of a variety of devices intended for use in surgical procedures to examine or treat the bronchus, esophagus, trachea, larynx, pharynx, nasal and paranasal sinus, or ear. This generic type of device includes the esophageal dilator; tracheal bistour (a long, narrow surgical knife); tracheal dilator; tracheal hook; laryngeal injection set; laryngeal knife; laryngeal saw; laryngeal trocar; laryngectomy tube; adenoid curette; adenotome; metal tongue depressor; mouth gag; oral screw; salpingeal curette; tonsillectome; tonsil guillotine; tonsil screw; tonsil snare; tonsil suction tube; tonsil suturing hook; antom reforator; ethmoid curette; frontal sinus-rasp; nasal curette; nasal rasp; nasal rongeur; nasal saw; nasal scissors; nasal snare; sinus irrigator; sinus trephine; ear curette; ear excavator; ear rasp; ear scissor, ear snare; ear spoon; ear suction tube; malleous ripper; mastoid gauge; microsurgical ear chisel; myringotomy tube inserter; ossici holding clamp; sacculotomy tack inserter; vein press; wire ear loop; microrule; mirror; mobilizer; ear, nose, and throat punch; ear, nose and throat knife; and ear, nose, and throat trocar.

Predicate Devices

Submission Summary (Full Text)

{0} JUL 25 1996 K962688 # SMDA Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. **Product:** Montgomery Radiopaque Salivary Bypass Tube **Description:** The Montgomery Salivary Bypass Tube is a silicone tube with a funnel-shaped end for use in treating esophageal strictures and fistulas. The tube is available is sizes 8mm, 10mm, 12mm, 14mm, 16mm, 18mm, and 20mm (outside diameter). **Indication For Use:** 1. Treatment of pharyngocutaneous fistulas following laryngectomy or laryngoesophagectomy. 2. Management of cervical esophageal and hypopharyngeal strictures and fistulas resulting from a variety of causes including surgery, radiation, advanced malignancy, trauma, and ingestion of caustic agents. **Predicate Device:** Montgomery Salivary Bypass Tube. **Testing:** Finished device samples passed cytotoxicity testing and were determined to be non-reactive and non-cytotoxic based on Elution Test, USP 23. Submitted by: Stuart K. Montgomery, President Date: 7/10/96
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...