K180063 · The Anspach Effort, Inc. · ERL · Mar 27, 2018 · Ear, Nose, Throat
Device Facts
Record ID
K180063
Device Name
OCM-G1 Attachment
Applicant
The Anspach Effort, Inc.
Product Code
ERL · Ear, Nose, Throat
Decision Date
Mar 27, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 874.4250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
When used with the ANSPACH Systems, the OCM-G1 Attachment and Burr Support Sleeves are intended for cutting and shaping bone primarily in otology procedures such as cocleostomies.
Device Story
OCM-G1 Attachment is a one-piece straight surgical accessory for G1 electric drill systems; used for cutting/shaping bone in otology procedures (e.g., cochleostomies). Device connects to distal end of handpiece via Diamond Flat locking mechanism; accepts OCM curved burr support sleeve assembly. Operated by surgeons in clinical settings. User selects rotation speed on drill system to drive burr. Knurl sleeve on proximal end facilitates manual connection/disconnection without tools. Device is non-sterile, reusable, and steam autoclavable. Output is mechanical rotation of burr for bone resection. Benefits include precise bone shaping during delicate ear surgeries.
Clinical Evidence
No clinical testing was performed. Substantial equivalence supported by bench testing, including locking feature compatibility, temperature monitoring, system functionality, and simulated use studies demonstrating performance in cutting and shaping bone.
Technological Characteristics
Materials: 17-4PH and 303 stainless steel (ASTM A564, ASTM A582). Form factor: One-piece straight attachment, 2.5mm length. Mechanism: Diamond Flat locking configuration. Connectivity: Attaches to G1 electric handpiece. Sterilization: Steam autoclavable (reusable).
Indications for Use
Indicated for cutting and shaping bone in otology procedures, such as cochleostomies, when used with ANSPACH systems.
Regulatory Classification
Identification
An ear, nose, and throat electric or pneumatic surgical drill is a rotating drilling device, including the handpiece, that is intended to drive various accessories, such as an ear, nose, and throat bur (§ 874.4140), for the controlled incision or removal of bone in the ear, nose, and throat area.
Predicate Devices
Anspach XMax Pneumatic and eMax 2 Plus Electric Systems with Otologic Attachment System (K131053)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below.
March 27, 2018
The Anspach Effort, Inc. Tamara West Sr. Regulatory Affairs Specialist 4500 Riverside Drive Palm Beach Gardens, FL 33410
Re: K180063
Trade/Device Name: OCM-G1 Attachment Regulation Number: 21 CFR 874.4250 Regulation Name: Ear, nose, and throat electric or pneumatic surgical drill Regulatory Class: Class II Product Code: ERL Dated: February 27, 2018 Received: February 28, 2018
Dear Tamara West:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration 10903 New Hampshire Avenue
Silver Spring, MD 20993 www.fda.qov
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely yours,
Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180063
Device Name Anspach OCM-G1 Attachment
Indications for Use (Describe)
When used with the ANSPACH® Systems, the OCM-G1 Attachment and Burr Support Sleeves are intended for cutting and shaping bone primarily in otology procedures such as cocleostomies.
Type of Use (Select one or both, as applicable)
[ Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
## I. SUBMITTER
The Anspach Effort, Inc. 4500 Riverside Drive Palm Beach Gardens, Florida 33410 USA
Registration Number: 1045834
Contact: Tamara West Phone: (561) 494-3712 Fax: (561) 624-8383 Email: twest1@its.jnj.com
Alternate Contact: Jeannette G. Dailey Phone: (561) 494-3710 Fax: (561) 624-8383 Email: jdailey5@its.jnj.com
Date Prepared: January 02, 2018
### II. DEVICE
Trade Name: OCM-G1 Attachment
Common Name: Surgical Cutting Tools
Classification Information: 874.4250 Drill, Surgical, ENT (Electric or Pneumatic) Including Handpiece
Regulatory Class: II
Product Code: ERL
Classification panel: Ear, Nose and Throat
### III. PREDICATE DEVICE
K131053 Anspach XMax Pneumatic and eMax 2 Plus Electric Systems with Otologic Attachment System (Otologic Attachment and Curved Burrs)
> Anspach Companies 4500 Riverside Drive, Palm Beach Gardens, Florida 334
Phone (561) 627-1080 · Fax (561) 627-312 Toll Free Phone (800) 327-6887 · Toll Free Fax (800) 327-6661
{4}------------------------------------------------
# IV. DEVICE DESCRIPTION
The OCM-G1 (Otologic Curved Micro) Attachment is an accessory to the G1 electric drill system. It is a one-piece straight attachment that is manufactured from 17-4PH and 303 stainless steels (ASTM A564 and ASTM A582). The OCM-G1 Attachment can be connected and disconnected to the distal end of the handpiece via a locking mechanism by the user.
After the OCM-G1 Attachment is secured to the handpiece, it will accept and hold an OCM curved burr support sleeve assembly (previously cleared). The OCM curved burr support sleeve assembly will rotate at a speed selected by the user for cutting and shaping bone, primarily in otology procedures such as cochleostomies.
The OCM-G1 Attachment is provided non-sterile and is reusable.
# V. INDICATIONS FOR USE
When used with the ANSPACH Systems, the OCM-G1 Attachment is intended for culting and shaping bone primarily in otology procedures such as cochleostomies.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| | OCM-G1 Attachment | eMax 2 Plus Electric Systems with<br>Otologic Attachment System (Otologic<br>Attachment and Curved Burrs) |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)<br>Number | K180063 | K131053 |
| Indications<br>for Use | When used with the ANSPACH Systems,<br>the OCM-G1 Attachment is intended for<br>cutting and shaping bone primarily in<br>otology procedures such as<br>cochleostomies. | When used with the ANSPACH Systems, the<br>OCM Attachment and OCM Burr Support<br>Sleeves are intended for cutting and<br>shaping bone primarily in otology<br>procedures such as cochleostomies. |
| Sterilization | Non-sterile | Non-sterile |
| Product Code | ERL | ERL, EQJ |
| Dissection Tools - NA | | |
| Console - NA | | |
| Handpiece - NA | | |
| Foot Control - NA | | |
| Attachment | | |
| Description | Same | Straight Attachment |
| Locking<br>Mechanism | Diamond Flat configuration | 560 Flat configuration |
| Material | Same | Stainless Steel |
| Length | 2.5mm | 2.6mm |
| Knurl sleeve | Yes | No |
| Connection to | Same | Quick Disconnect |
Anspach Companies
4500 Riverside Drive, Palm Beach Gardens, Florida 33
Phone (561) 627-1080 · Fax (561) 627-3120 Toll Free Phone (800) 327-6887 · Toll Free Fax (800) 327-6661
{5}------------------------------------------------
| handpiece | | | |
|-----------|------|-------------|--|
| Labeling | Same | Non-sterile | |
The predicate OCM Attachment (K131053) is an accessory to an electric/pneumatic handpiece that utilizes the 560 Flat configuration locking mechanism. A minor design change was made to the locking mechanism of the predicate attachment (K131053) modifying it from connecting to a handpiece with a 560 Flat locking configuration to connecting to a handpiece with a Diamond Flat configuration. To allow for ease of connection/disconnection with the handpiece, a knurl sleeve was added to the proximal end of the proposed OCM-G1 Attachment unlike the predicate (K131053) where the knurl sleeve was located on the handpiece.
The proposed OCM-G1 Attachment remains similar to the predicate attachment (K131053) in terms of operating principles, where it holds and secures a curved burr support sleeve assembly intended for cutting and shaping bone primarily in otology procedures such as cochleostomies. The indications for use, material and manufacturing remain similar to the predicate attachment.
In addition, the OCM-G1 Attachment will utilize the same OCM Curved Burr Support Sleeves as the predicate.
The technological characteristics of the proposed OCM-G1 Attachment are similar to those of the predicate (K131053) device in the following ways:
- Both attachments use a patented locking mechanism (input shafts) to secure to . handpiece
- Both attachments are compatible with the same OCM curved burr support sleeve 0 assembly currently on the market
- . Both attachments by design have a straight outer metal cover with internal components
- Both attachments are non-sterile reusable, steam autoclavable devices .
The following technological differences between the proposed and predicate (K131053) device were considered in relation to the substantial equivalence determination:
- . Use of a Diamond Flat configuration on the proposed OCM-G1 Attachment for locking mechanism opposed to 560 Flat configuration. The Diamond Flat confiquration allows the proposed OCM-G1 Attachment to be compatible with the G1 handpiece only. The predicate (K131053) is compatible with handpieces compatible with the 560 Flat configuration.
- The proposed OCM-G1 Attachment is compatible with an electric powered drill . system while the predicate (K131053) is compatible with various electric and pneumatic powered drill systems. There is no change to the performance of the attachment as compared to the predicate (K131053).
- . A knurl sleeve was incorporated on the proximal end of the OCM-G1 Attachment for ease of use and which allows the attachment to be removed from the handpiece
Anspach Companies 4500 Riverside Drive, Palm Beach Gardens, Florida 33410
{6}------------------------------------------------
Image /page/6/Picture/21 description: The image shows the word "ANSPACH" in blue, with a registered trademark symbol to the right of the word. To the left of the word is a blue square with a white bird-like figure inside. The text is in a bold, sans-serif font. The overall image has a clean and professional look.
using one less step than the predicate. For the proposed OCM-G1 Attachment, the act of twisting the attachment locks in the dissection tools or burr on handpiece, whereas on the predicate (K131053) retracting (and releasing) the disconnect sleeve allows the dissection tool or burr to be installed/removed.
- . Total length of the proposed OCM-G1 Attachment is 2.5mm opposed to 2.6mm of predicate (K131053).
The proposed changes to the OCM-G1 Attachment have no impact on the safety and effectiveness of the device or its ability to perform as intended.
# VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
### Biocompatibility Testing:
The OCM-G1 Attachment does not have direct or indirect patient contact, same as the predicate. A biocompatibility evaluation was conducted for the OCM-G1 Attachment per the FDA quidance document Use of International Standard ISO 10993-1. "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
The OCM-G1 Attachment is manufactured and packaged in the same manner with identical material as the predicate device (K131053). No new or additional processing steps were introduced; therefore, biocompatibility requirements have been met.
### Design Control Activities:
The following verification tests were performed to demonstrate that the device meets the performance requirements under its indications for use conditions.
- Locking feature compatibility .
- Temperature .
- Disconnect without tools e
- System functionality .
The above tests demonstrate proper functionality, compatibility, safe external temperatures, and the ability to connect and disconnect without the use of tools, the same as the predicate (K131053). A simulated use study demonstrated the OCM-G1 Attachment and curved cutting burrs perform as intended in cutting and shaping bone properly in the same manner as the predicate (K131053).
### Cleaning and Reprocessing:
The OCM-G1 Attachment was assessed to determine whether or not current validations for clinical processing and steam sterilization are capable of producing a sterility assurance
### Ansnach Comnanie
4500 Riverside Drive, Palm Beach Gardens, Florida 334
Phone (561) 627-1080 · Fax (561) 627-3120 Toll Free Phone (800) 327-6887 · Toll Free Fax (800) 327-6661
{7}------------------------------------------------
# ANSPACH
level (SAL) of 10-6. Overall, the proposed OCM-G1 Attachment device characteristics would be equal or less of a challenge to sterilize than the currently validated "worst case" with the ability to be cleaned and sterilized effectively without adverse effects, same as the predicate (K131053).
### Clinical Testing:
No clinical testing was necessary for a determination of substantial equivalence.
### VII. CONCLUSIONS
The OCM-G1 Attachment described in this submission has the same intended use, general design, material, technological characteristics, and principles of operation as the predicate device (K131053). The introduction of the attachment does not raise any new issues of safety or efficacy and was found to be substantially equivalent to the predicate (K131053) device.
# 4500 Riverside Drive. Palm Beach Gardens. Florida
Phone (561) 627-1080 · Fax (561) 627 Toll Free Phone (800) 327-6887 · Toll Free Fax (800) 327-6661
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.