Indicated for the excision of polyps in the nasal passageways and in sinuses for which access has been previously established.
Device Story
PolypVac Microdebrider is a sterile, single-use, disposable surgical instrument for cutting and aspirating polyp tissue during nasal and sinus procedures. System components include a hand-held cutting device, tissue filter, and stylet. Operated by a surgeon under visual guidance using a user-supplied endoscope (≤5mm diameter) and vacuum pump. Principle of operation: vacuum suction draws polyp tissue into a side-facing cutting window; a reciprocating cutter excises the tissue. The device features a rotatable shaver blade (360°) and an ergonomic handle with a trigger to activate/halt cutting. Irrigant is routed to the distal tip through the shaver blade. Output is the physical removal of sinonasal polyps, facilitating surgical clearance of nasal passageways. Benefits include effective tissue resection and improved ergonomic control for the surgeon.
Clinical Evidence
No clinical data was performed or necessary. Substantial equivalence is supported by non-clinical bench testing, including functionality, resection rate, and simulated use testing, alongside biocompatibility, shelf-life, packaging, and sterilization validation.
Technological Characteristics
Hand-held microdebrider; 3.3mm and 4.0mm blade diameters; reciprocating cutter mechanism; vacuum-powered aspiration. Materials evaluated per ISO 10993. Packaging meets ASTM F1980-07 and D4169-14. Sterilization via E-beam radiation (SAL 10^-6). Requires external vacuum pump and endoscope (≤5mm).
Indications for Use
Indicated for the excision of polyps in the nasal passageways and in sinuses for which access has been previously established.
Regulatory Classification
Identification
An ear, nose, and throat electric or pneumatic surgical drill is a rotating drilling device, including the handpiece, that is intended to drive various accessories, such as an ear, nose, and throat bur (§ 874.4140), for the controlled incision or removal of bone in the ear, nose, and throat area.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three human profiles or faces stacked on top of each other, with flowing lines suggesting movement or connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 22, 2016
Laurimed LLC Mr. Brian R. Dubois Vice President of Engineering 500 Arguello St., Suite 100 Redwood City, CA 94063
Re: K161101
Trade/Device Name: Polypvac Microdebrider (3.3mm And 4.0mm) Regulation Number: 21 CFR 874.4250 Regulation Name: Ear, Nose, And Throat Electric Or Pneumatic Surgical Drill Regulatory Class: Class II Product Code: ERL Dated: May 19, 2016 Received: May 23, 2016
Dear Mr. Dubois:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) к 161101
Device Name
PolypVac Microdebrider
Indications for Use (Describe)
Indicated for the excision of polyps in the nasal passageways and in sinuses for which access has been previously established.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### l. General Information [807.92(a)(1)]
## Name, Address, Phone and Fax Number of Applicant
Laurimed LLC 500 Arguello Street, Suite 100 Redwood City, CA 94063 Phone: (650) 587-5296 Fax: (650) 587-3823
# Contact Person
Alex Gordon Engineering and Manufacturing Manager Laurimed LLC 500 Arguello Street, Suite 100 Redwood City, CA 94063 Phone: (650) 587-5296 Fax: (650) 587-3823
### Date Prepared
April 18, 2016
#### II. Device Information [§807.92(a)(2)]
Trade Name: Laurimed PolypVac Microdebrider Common Name: Microdebrider Classification Name: Ear, nose, and throat electric or pneumatic drill Classification: 21 CFR §874.4250 Product Code: ERL Device Class: Class II
#### lll. Predicate Device [§807.92(a)(3)]
The Laurimed PolypVac Microdebrider is substantially equivalent to the Laurimed PolypVac Microdebrider (K133133)
{4}------------------------------------------------
#### IV. Device Description [§807.92(a)(4)]
The Laurimed PolypVac Microdebrider is intended for cutting and aspirating polyp tissue during sinus and nasal procedures. The system consists of a hand-held cutting device, tissue filter, and stylet. The PolypVac Microdebrider is supplied sterile and is single-use and disposable.
#### V. Indications for Use [§807.92(a)(5)]
Indicated for the excision of polyps in the nasal passageways and in sinuses for which access has been previously established.
#### VI. Comparison of Technological Characteristics with the Predicate Device [§807.92(a)(6)]
The technological characteristics and principals of operation of the Laurimed PolypVac Microdebrider are substantially equivalent to the Laurimed PolypVac predicate device (K133133).
| Model Name | PolypVac Microdebrider | PolypVac Microdebrider | Rationale for<br>Substantial<br>Equivalence |
|----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K133133 | TBD | N/A (Same). |
| Product Code | ERL | ERL | N/A (Same). |
| Indications for Use | Indicated for the excision of<br>polyps in the nasal<br>passageways and in sinuses<br>for which access has been<br>previously established. | Indicated for the excision of<br>polyps in the nasal<br>passageways and in sinuses<br>for which access has been<br>previously established. | N/A (Same). |
| Contraindications | None | None | N/A (Same). |
| Anatomical Sites | Nose (including accessible<br>sinuses) | Nose (including accessible<br>sinuses) | N/A (Same). |
| Target Tissue | Sinonasal Polyps | Sinonasal Polyps | N/A (Same). |
| Equipment /<br>Materials Required | User-supplied endoscope (≤5<br>mm diameter), vacuum pump,<br>and sterile irrigant syringe /<br>bag. | User-supplied endoscope (≤5<br>mm diameter), vacuum pump,<br>and sterile irrigant syringe /<br>bag. | N/A (Same). |
| Method of<br>Introduction | Manual placement under<br>visual guidance. | Manual placement under<br>visual guidance. | N/A (Same). |
| Mechanism of<br>Action | Suction from user-supplied<br>vacuum pump draws polyp<br>tissue into side-facing cutting<br>window, tissue excised by<br>reciprocating cutter. | Suction from user-supplied<br>vacuum pump draws polyp<br>tissue into side-facing cutting<br>window, tissue excised by<br>reciprocating cutter. | N/A (Same). |
| Model Name | PolypVac Microdebrider | PolypVac Microdebrider | Rationale for<br>Substantial<br>Equivalence |
| 510(k) Number | K133133 | TBD | |
| User Interface | The handle body fits in the<br>palm of the surgeon's hand<br>and the shaver blade extends<br>approximately perpendicular<br>to the handle body. | Same.<br>The handle shape has been<br>optimized to reduce size and<br>improve ergonomics. | The modifications to the<br>User Interface do not<br>affect the principles of<br>operation or the<br>performance of the<br>device and no additional<br>risks or hazards have<br>been identified with these<br>changes. |
| | Rotatable shaver blade<br>(+/- 90°). | Rotatable shaver blade (360°)<br>with knob. | |
| | Trigger located on side of<br>handle activates and halts<br>reciprocating motion of cutter. | Same.<br>The trigger shape has been<br>optimized. | |
| | Connection for vacuum<br>(nipple) located at base of<br>handpiece. | Connection for vacuum located<br>at back of barrel. | |
| | PVC Tubing with a luer<br>connects to the base of the<br>handpiece. | Smaller diameter tubing with a<br>luer connects at the back of the<br>barrel. | |
| | Filter integrated into<br>handpiece. | Filter is separate from<br>handpiece. | |
| Physical /<br>Dimensional<br>Characteristics | Blade diameter: 3.5 mm. | Blade diameter: 3.3 mm &<br>4.0 mm. | The devices utilize the<br>same technology and |
| | Rounded atraumatic plastic<br>tip. | Rounded atraumatic metal tip. | principles of operation<br>and the rounded tips are<br>equivalent. No additional<br>risks or hazards have<br>been identified with<br>these changes. |
| Irrigation | Irrigant is routed to the distal<br>tip through a lumen in the<br>shaver blade's plastic sheath. | Irrigant is routed to the distal<br>tip within the shaver blade<br>(between the evacuation tube<br>and the stainless steel tube). | The modified device<br>meets all fluid flow<br>requirements. |
| Materials | All contact materials have<br>been evaluated per ISO<br>10993. | All contact materials have<br>been evaluated per ISO<br>10993. | Finish product was tested<br>and is biocompatible in<br>accordance with ISO<br>10993. |
| Packaging | Package meets ASTM<br>F1980-07 and D4169-14. | Package meets ASTM<br>F1980-07 and D4169-14. | The modified packaging<br>configuration has passed<br>all packaging validation<br>tests, showing that it<br>equivalently protects the<br>device and maintains<br>sterility. |
| Sterilization | E-beam radiation with<br>SAL 10-6 | E-beam radiation with<br>SAL 10-6. | N/A (Same). |
Table 1 Summary of Technological Characteristics
{5}------------------------------------------------
{6}------------------------------------------------
#### VII. Performance Data [§807.92(b)]
Results of non-clinical testing demonstrated that the Laurimed PolypVac Microdebrider is safe and effective for its intended use and substantially equivalent to the predicate.
# Summary of Nonclinical, Bench Testing [§807.92(b)(1)]
The Laurimed PolypVac Microdebrider has been carefully compared to a legally marketed device with respect to intended use and technological characteristics. In addition, non-clinical testing was conducted to validate the performance of the device and ensure the Laurimed PolypVac Microdebrider functions as intended and meets design specifications. The comparison and nonclinical results demonstrate that the Laurimed PolypVac Microdebrider is substantially equivalent to the predicate device for its intended use and does not raise new issues of safety or effectiveness.
| Testing Type | Test Description | Results Supporting<br>Substantial Equivalence |
|--------------------------|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Bench Testing | Functionality Testing | The modified PolypVac<br>Microdebrider passed all<br>functional testing and met all<br>product specification<br>requirements demonstrating<br>equivalent performance to that<br>of the predicate device. |
| | Resection Rate | |
| | Simulated Use Testing | |
| Biocompatibility Testing | Cytotoxicity | The modified PolypVac<br>Microdebrider and the<br>predicate device are<br>biocompatible in accordance<br>with ISO 10993. |
| | Sensitization | |
| | Irritation | |
Table 2. Performance Testing and Substantial Equivalence Support
In addition to the performance testing listed in Table 2, Laurimed conducted Shelf-life, Packaging, and Sterilization Validation testing on the Polypvac Microdebrider.
# Summary of Clinical Data [§807.92(b)(2)]
No clinical testing was performed or necessary in support of this premarket notification.
#### VIII. Conclusion [§807.92(b)(3)]
The Laurimed PolypVac Microdebrider does not raise any new issues of safety or effectiveness, as both the modified and the predicate devices have the same intended use, and utilize similar performance specifications and comparable technological features to achieve the same mechanism of action: therefore, the Laurimed PolypVac Microdebrider is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.