XPS Nexus System, XPS Nexus Foot Control, XPS Nexus IV Pole
K150728 · Medtronic Xomed, Inc. · ERL · Aug 14, 2015 · Ear, Nose, Throat
Device Facts
Record ID
K150728
Device Name
XPS Nexus System, XPS Nexus Foot Control, XPS Nexus IV Pole
Applicant
Medtronic Xomed, Inc.
Product Code
ERL · Ear, Nose, Throat
Decision Date
Aug 14, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The XPS Nexus™ System is indicated for the incision/cutting, removal and drilling of soft and hard tissue and bone in head and neck/ENT, oral/maxillofacial and plastic/reconstructive/aesthetic surgical procedures.
Device Story
XPS Nexus System is a powered microdebrider and drill system for surgical tissue/bone removal. System components include a power control console, foot control unit, and various hand-pieces (StraightShot M2/M4 Microdebriders, Indigo Otology Drill) to drive burs, blades, drills, rasps, and cannulas. Console features integrated irrigation pump for blades/burs. Used in OR settings by surgeons. Device transforms electrical power into mechanical motion for cutting/drilling. Healthcare providers operate via foot control and console touch pad to manage speed and irrigation. Output is mechanical action on tissue; assists in precise surgical resection. Benefits include efficient tissue removal in ENT and maxillofacial procedures.
Clinical Evidence
No clinical data. Bench testing only, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), software validation (IEC 62304), and usability engineering (IEC 62366/60601-1-6).
Technological Characteristics
Powered surgical drill/microdebrider system. Materials: stainless steel and medical polymer. Components: power console, foot control, hand-pieces, irrigation pump. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 62304 (software), IEC 62366 (usability). Connectivity: standalone console with foot control interface.
Indications for Use
Indicated for incision, cutting, removal, and drilling of soft/hard tissue and bone in head and neck/ENT, oral/maxillofacial, and plastic/reconstructive/aesthetic surgical procedures.
Regulatory Classification
Identification
An ear, nose, and throat electric or pneumatic surgical drill is a rotating drilling device, including the handpiece, that is intended to drive various accessories, such as an ear, nose, and throat bur (§ 874.4140), for the controlled incision or removal of bone in the ear, nose, and throat area.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The seal is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 14, 2015
Medtronic Xomed, Inc. Ms. Gabriela Anchondo Principal Regulatory Affairs Specialist 6743 Southpoint Dr. North Jacksonville, FL 32216
Re: K150728
Trade/Device Name: XPS Nexus System, XPS Nexus Foot Control, XPS Nexus IV Pole Regulation Number: 21 CFR 874.4250 Regulation Name: Ear, Nose, and Throat Electric or Pneumatic Surgical Drill Regulatory Class: Class II Product Code: ERL Dated: July 7, 2015 Received: July 9, 2015
Dear Ms. Anchondo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Kesia Y. Alexander -A
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K150728
Device Name XPS Nexus™ System
Indications for Use (Describe)
The XPS Nexus™ System is indicated for the incision/cutting, removal and drilling of soft and hard tissue and bone in head and neck/ENT, oral/maxillofacial and plastic/reconstructive/aesthetic surgical procedures.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5.0 510(k) Summary
| A. 510(k) Owner | Medtronic Xomed, Inc<br>6743 Southpoint Drive North<br>Jacksonville, Florida 32216-0980 USA<br>904-296-9600<br>904-296-2386 (FAX) |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| B. Contact Information | Gabriela Anchondo<br>Regulatory Affairs Manager<br>Medtronic Xomed, Inc<br>gabriela.anchondo@medtronic.com |
| C. Date Summary Prepared | March 19, 2015 |
| D. Proprietary Name | XPS Nexus™ System |
| E. Device Name | |
| Trade name: | XPS Nexus™ |
|----------------------|-------------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Drill, Surgical, ENT (electric or pneumatic) including<br>hand piece (ERL) |
| Classification Name: | Ear, nose and throat electric or pneumatic surgical drill<br>(21 CFR 874.4250, Product Code: ERL, Class II) |
### F. Predicate Devices:
| Trade Name: | XPS 3000 System |
|-------------------------|-------------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Drill, Surgical, ENT (electric or pneumatic) including<br>hand piece (ERL) |
| Classification Name: | Ear, nose and throat electric or pneumatic surgical drill<br>(21 CFR 874.4250, Product Code: ERL, Class II) |
| Premarket Notification: | K041413 |
| Trade Name: | IPC® Integrated Power Console |
| Common/Usual Name: | Drill, Surgical, ENT (electric or pneumatic) including<br>hand piece (ERL) |
| Classification Name: | Ear, nose and throat electric or pneumatic surgical drill<br>(21 CFR 874.4250, Product Code: ERL, Class II) |
| Premarket Notification: | K081277 |
### G. Device Description
The XPS Nexus™ System is a powered microdebrider and drill system that removes soft tissue, hard tissue and bone during surgical procedures. The system consists of a power control console, a foot control unit and assorted hand-pieces to drive various
{4}------------------------------------------------
burs, blades, drills, rasps and cannulas. Hand-piece options include the StraightShot® M2 Microdebrider, the StraightShot® M4 Microdebrider and the Indigo® Otology Drill.
The power control console also includes integrated irrigation pump for irrigation of blades and burs.
Optional accessories include an IV pole to hold a bag(s) of irrigation solution as well as a carrying case.
### H. Intended Use/Indications for Use:
The XPS Nexus™ System is indicated for the incision/cutting, removal and drilling of soft and hard tissue and bone in head and neck / ENT, oral / maxillofacial and plastic / reconstructive / aesthetic surgical procedures.
{5}------------------------------------------------
## I. Substantial Equivalence
## Summary of Technological Characteristics
| | Subject Device | Predicate Device(s) | |
|---------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name | XPS Nexus™ System | IPC® Integrated Power Console<br>K081277 | XPS 3000 System<br>K041413 |
| Indications for Use | The XPS Nexus™ System is indicated for the<br>incision/cutting, removal and drilling of soft<br>and hard tissue and bone in head and neck /<br>ENT, oral / maxillofacial and plastic /<br>reconstructive / aesthetic surgical procedures. | The Electric Drill System is indicated<br>for the incision / cutting, removal,<br>drilling, and sawing of soft and hard<br>tissue and bone in Head & Neck / ENT<br>(Otologic, Neurologic, Neurotologic,<br>Sinus, Rhinologic, Nasopharyngeal /<br>Laryngeal), Oral / Maxillofacial, and<br>Plastic / Reconstructive / Aesthetic<br>Surgical Procedures. | The XPS 3000 is intended for the incision and removal of soft and hard<br>tissue or bone in general otorhinolaryngology, head and neck, and<br>otoneurological surgery.<br>Otology/ neurotology indications include mastoidectomy, mastoidotomy,<br>and acoustic neuroma.<br>Sinus indications include septoplasty, removal of septal spurs,<br>polypectomy, antrostomy, ethmoidectomy/sphenoethmoidectomy, frontal<br>sinus trephination and irrigation, frontal sinus drill out, endoscopic DCR,<br>trans-spheniodal procedures, maxillary sinus polypectomy, circumferential<br>maxillary antrostomy, choanal atresia, sphenoidotomy, and medial, lateral,<br>and posterior frontal sinusotomy<br>Nasopharyngeal / laryngeal indications include adenoidectomy, tracheal<br>procedures, laryngeal polypectomy, laryngeal lesion debulking,<br>tonsillectomy, tonsillotomy for obstructive tonsillar disease, removal of<br>endobronchial lesions, and the surgical management of recurrent<br>respiratory papillomatosis (RRP).<br>Head and neck (ENT) indications include soft tissue shaving, rhinoplasty<br>(narrowing of the bony vault and revision of the bony pyramid), removal<br>and shaping of bone during rhinoplasty procedures, removal of adipose<br>tissue (lipo debridement) in the maxillary and mandibular regions of the<br>face, removal of acoustic neuroma, and incision and removal of soft tissue<br>during plastic, reconstructive, and/or aesthetic surgery.<br>The XPS 3000 system using the PowerSculpt handpiece and reciprocating<br>cutting blades / rasps is indicated to cut hard and soft tissue or bone in<br>otorhinolaryngology and head and neck surgery.<br>The XPS 3000 system with reciprocating adapter and suction cannula is<br>intended for the removal of soft tissue and fluid during general surgical<br>procedures including suction lipoplasty for aesthetic body contouring.<br>The XPS 3000 system is indicated for use in orthopedic surgical<br>procedures where the cutting and removal of soft and hard tissue or bone is<br>required. These include spinal and small and large joint arthroscopic<br>procedures.<br>An integral pump is provided for irrigation, and a second integral pump<br>maybe provided for handpiece cooling. |
| Microdebriders<br>StraightShot® Magnum II<br>StraightShot® M4 | Yes<br>Yes | Yes<br>Yes | Yes<br>Yes |
| Drills<br>Indigo™<br>Visao® | Yes<br>No | -<br>Yes<br>Yes | -<br>No<br>Yes |
| Basic Features | Hand piece w/suction<br>Control unit w/Footswitch<br>1 Irrigation Pump | Hand piece w/suction<br>Control unit w/Footswitch<br>2 Irrigation Pumps | Hand piece w/suction<br>Control unit w/Footswitch<br>1 Irrigation Pumps, 1 optional handpiece cooling pump |
| User interface | Touch pad | Touch screen | Control buttons, displays |
| Device name | Subject Device | Predicate Device(s) | |
| | XPS Nexus™ System | IPC® Integrated Power Console<br>K081277 | XPS 3000 System<br>K041413 |
| Provided sterile | - | - | - |
| Console/handpiece | No | No | No |
| Foot Control | No | No | No |
| Blade/bur/drill bit | Yes | Yes | Yes |
| Tubing sets | Yes | Yes | Yes |
| Patient contact | - | - | - |
| Console/handpiece | No | No | No |
| Foot control | No | No | No |
| Tubing set (fluid path) | Yes | Yes | Yes |
| Blade/bur/drill bit | Yes | Yes | Yes |
| Body contacting<br>materials | Stainless steel and medical polymer | Stainless steel and medical polymer | Stainless steel and medical polymer |
{6}------------------------------------------------
{7}------------------------------------------------
### J. Testing
The verification and validation testing of the XPS Nexus™ System included the following testing:
### Electrical Safety Testing
Electrical Safety compliance is demonstrated through testing in accordance with:
| FDA<br>Recognition<br>Number | Standard<br>Developing<br>Organization | Recognition<br>List Number | Standard<br>Designation<br>Number and<br>Date | Title of Standard | Effective<br>Date | Category |
|------------------------------|----------------------------------------|----------------------------|-----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|------------------------|
| 19-4 | AAMI ANSI | 036 | ES 60601-1:<br>2005/(R)2012<br>and A1:2012, | C1:2009/(R)2012 and<br>A2:2010/(R)2012<br>(Consolidated Text)<br>Medical electrical<br>equipment - Part 1:<br>General requirements<br>for basic safety and<br>essential performance<br>(IEC 60601-1:2005,<br>MOD) | 07/09/2014 | General II<br>(ES/EMC) |
Test results indicated that XPS Nexus System complies with the applicable standards.
#### Electromagnetic Compatibility Testing
Electromagnetic Compatibility compliance is demonstrated through testing in accordance with:
| FDA<br>Recognition<br>Number | Standard<br>Developing<br>Organization | Recognition<br>List Number | Standard<br>Designation<br>Number and<br>Date | Title of Standard | FR<br>Publication<br>Date | Category |
|------------------------------|----------------------------------------|----------------------------|-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|------------------------|
| 19-1 | IEC | 037 | 60601-1-2<br>Edition 3:<br>2007-03 | Medical electrical<br>equipment - Part 1-2:<br>General requirements<br>for basic safety and<br>essential performance -<br>Collateral standard:<br>Electromagnetic<br>compatibility -<br>Requirements and tests | 10/17/2014 | General II<br>(ES/EMC) |
Test results indicated that XPS Nexus System complies with the applicable standards.
{8}------------------------------------------------
### Software Testing
Software testing was performed in compliance with the following guidance and standards:
- FDA Guidance: The content of premarket submissions for software contained in medical devices, May 11, 2005
- FDA Guidance: General principles of software validation; Final guidance for industry ● and FDA staff, January 11, 2002
- IEC 60601-1-4 Medical Electrical Equipment, Part 1-4: Collateral Standard: ● Programmable Electrical Medical Systems; IEC 60601-1-4:1996 (First Ed.) + Am. 1:1999 (Consolidated 1.1 Ed.) for use with IEC 60601-1 (1988). Amts. 1 (1991) and 2 (1995)
| FDA<br>Recognition<br>Number | Standard<br>Developing<br>Organization | Recognition<br>List Number | Standard<br>Designation<br>Number and<br>Date | Title of Standard | FR<br>Publication<br>Date | Category |
|------------------------------|----------------------------------------|----------------------------|-----------------------------------------------|---------------------------------------------------------------|---------------------------|---------------------------------|
| 13-8 | IEC | 029 | 62304: 2006-<br>05 | Medical device<br>software - Software life<br>cycle processes | 08/20/2012 | (Software/<br>Informatics<br>s) |
Test results indicated that XPS Nexus System complies with the applicable standards.
### Performance Testing
Performance testing was performed in compliance with the following standards:
| FDA<br>Recognition<br>Number | Standard<br>Developing<br>Organization | Recognition<br>List Number | Standard<br>Designation<br>Number and<br>Date | Title of Standard | FR<br>Publication<br>Date | Category |
|------------------------------|----------------------------------------|----------------------------|-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|-----------------------|
| 5-85 | IEC | 036 | 60601-1-6<br>Edition 3.0<br>2010-01 | Medical electrical<br>equipment -- Part 1-6:<br>General requirements<br>for basic safety and<br>essential performance —<br>Collaterial Standard:<br>Usability | 07/09/2014 | (General I<br>(QS/RM) |
| 5-67 | AAMI ANSI<br>IEC | 036 | 62366:2007/(R<br>)2013 | Medical devices -<br>Application of usability<br>engineering to medical<br>devices | 07/09/2014 | (General I<br>(QS/RM) |
Test results indicated that XPS Nexus System complies with the applicable standards.
General performance verification and validation testing of the subject XPS Nexus System was also performed to verify the performance and output characteristics.
No clinical testing was deemed necessary for this Premarket Notification.
{9}------------------------------------------------
### K. Conclusion
The data presented in this Premarket Notification support that the subject device is safe and effective and performs in the same manner as the predicate device when used in accordance with the labeled directions for use and for the specified indication(s).
The risks of the subject device, as well as the benefits to the patient, are the same as those attributed to the use of the predicate devices. No new risks have been identified.
The scope of intended use of XPS Nexus™ System is limited as compared with predicate devices; it has the same conditions of use and the same key technological characteristics as predicate devices and does not raise new issues of safety or effectiveness. The XPS Nexus™ System is therefore substantially equivalent to the IPC® Integrated Power Console cleared by K081277 and XPS 3000 System cleared by K041413.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.