ANSPACH XMAX, EMAX2 AND EMAX2 PLUS SYSTEM WITH OTOLOGIC ATTACHMENT SYSTEM
K131053 · The Anspach Effort, Inc. · ERL · Dec 31, 2013 · Ear, Nose, Throat
Device Facts
Record ID
K131053
Device Name
ANSPACH XMAX, EMAX2 AND EMAX2 PLUS SYSTEM WITH OTOLOGIC ATTACHMENT SYSTEM
Applicant
The Anspach Effort, Inc.
Product Code
ERL · Ear, Nose, Throat
Decision Date
Dec 31, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ANSPACH® XMax Pneumatic and eMax 2 and eMax2 Plus Electric Systems are intended for cutting and shaping bone including spine and cranium. When used with the ANSPACH® Systems, the OCM Attachment and OCM Burr Support Sleeves are intended for cutting and shaping bone primarily in otology procedures such as cochleostomies.
Device Story
System comprises pneumatic or electric surgical drill handpieces (XMax, eMax 2, eMax2 Plus) paired with Otologic Curved Micro (OCM) attachment and OCM Burr Support Sleeves. OCM attachment connects to handpieces via twist-lock; burr support sleeves snap into attachment. Sleeves feature 20° curve and small diameter for improved visibility; non-rotating sleeve protects soft tissue from rotating burr. Operated by surgeons in clinical/OR settings for bone dissection. System provides mechanical power to cutting burrs at speeds of 10,000–80,000 rpm. Benefits include precise bone shaping and enhanced access in delicate otologic procedures.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including biocompatibility testing per GLP (21 CFR Part 58), functional performance testing (repeated use, temperature safety, cleaning/sterilization validation), and a simulated use study confirming performance in bone cutting and shaping.
Technological Characteristics
Pneumatic and electric surgical drill systems. OCM attachment uses standard twist-lock connection; OCM Burr Support Sleeves use snap-in connection. Materials are biocompatible, consistent with currently marketed attachments. Operating speed: 10,000–80,000 rpm. OCM sleeves available in 5.6 cm and 7.3 cm lengths. Non-rotating sleeve design for soft tissue protection. Reusable attachment; single-use sterile burr support sleeves.
Indications for Use
Indicated for patients requiring bone cutting and shaping during surgical procedures, including spine, cranium, and otology procedures such as cochleostomies.
Regulatory Classification
Identification
An ear, nose, and throat electric or pneumatic surgical drill is a rotating drilling device, including the handpiece, that is intended to drive various accessories, such as an ear, nose, and throat bur (§ 874.4140), for the controlled incision or removal of bone in the ear, nose, and throat area.
{0}------------------------------------------------
DEC 3 1 2013
510(k) PREMARKET NOTIFICATION ANSPACH® Systems with Otologic Curved Micro (OCM) Attachment and OCM Burr Support Sleeves
## 510(k) Summary
| Submitter<br>Information | Company: | The Anspach Effort, Inc.<br>4500 Riverside Drive<br>Palm Beach Gardens, FL 33410 |
|---------------------------------|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Contact Person: | Jeannette G. Dailey<br>Regulatory Affairs Manager |
| | Email: | Dailey.Jeannette@synthes.com |
| | Telephone: | 561-494-3710 |
| | FAX: | 561-625-9110 |
| | Date Prepared: | November 8, 2013 |
| Identification of<br>the Device | Common Names: | Pneumatic and Electric Surgical Instruments<br>and Attachments |
| | Device Name: | ANSPACH® XMax Pneumatic and eMax 2 and<br>eMax2 Plus Electric Systems with the<br>ANSPACH® Otologic Curved Micro (OCM)<br>Attachment and OCM Burr Support Sleeves |
| | Classification Names: | • Ear, nose and throat electric or pneumatic<br>surgical drill<br>• Ear, nose and throat bur |
| | Device Classification: | • Class II, 21 CFR 874.5250 (drill)<br>• Class 1, 510(k) exempt, 21 CFR 874.4140 (bur) |
Predicate Devices(s) The subject devices are substantially equivalent to the following predicate devices:
The electric and pneumatic surgical drills are identical to the currently marketed electric and pneumatic surgical drills:
- ANSPACH® XMax Pneumatic System cleared via K965080 .
- ANSPACH® eMax 2 and eMax2 Plus Electric Systems cleared via . K011444
The Otologic Curved Micro (OCM) Attachment and OCM Burr Support Sleeves are substantially equivalent to the handpiece and burrs associated with the:
- . Medtronic. Inc., XPS® 3000 System cleared most recently via K073255 on March 24, 2008
Continued on next page
{1}------------------------------------------------
## 510(k) Summary, Continued
| Description of Device | The ANSPACH® XMax Pneumatic and eMax 2 and eMax2 Plus Electric Systems are surgical drills that handle a complete range of surgical procedures ranging from power demanding applications to the most delicate dissection. |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The ANSPACH® Otologic Curved Micro (OCM) Attachment consists of the attachment which connects directly to the ANSPACH® XMax Pneumatic and eMax 2 and eMax2 Plus Electric Systems handpieces with the standard twist lock mechanism, and the OCM Burr Support Sleeves. The OCM is provided non-sterile and is reusable. Cleaning and sterilization instructions are provided in the ANSPACH® OCM Attachment package insert. |
| | The ANSPACH® OCM Burr Support Sleeves with integral cutting burrs mount to the OCM Attachment with a simple snap-in connection. The 20° curve and small diameter of the OCM Burr Support Sleeves provides excellent visibility. The non-rotating support sleeve protects soft tissue from the rotating cutting burr shaft. The OCM Burr Support Sleeves can be used at speeds from 10,000 to 80,000 rpm. They are available in two lengths (5.6 and 7.3 cm), each with a wide variety of different cutting burr styles and diameters to accommodate varying surgeon preferences. The OCM Burr Support Sleeves are provided sterile and for single use only. |
| Indications for Use | The ANSPACH® XMax Pneumatic and eMax 2 and eMax2 Plus Electric Systems are intended for cutting and shaping bone including spine and cranium. |
| | When used with the ANSPACH® Systems, the OCM Attachment and OCM Burr Support Sleeves are intended for cutting and shaping bone primarily in otology procedures such as cochleostomies. |
Continued on next page
.
.
{2}------------------------------------------------
## 510(k) Summary, Continued
| Technological<br>Characteristics | There are no new technological characteristics associated with the<br>ANSPACH® Pneumatic and Electric Systems, the OCM Attachment<br>and the OCM Burr Support Sleeves. The ANSPACH® XMax<br>Pneumatic and eMax 2 and eMax2 Plus Electric Surgical Drills are<br>identical to the currently marketed electric and pneumatic surgical drills.<br>The new ANSPACH® Otologic Curved Micro (OCM) Attachment and<br>OCM Burr Support Sleeves utilize the same biocompatible materials as<br>currently marketed attachments and burrs. The packaging,<br>manufacturing method, and cleaning and sterilization methods of the<br>OCM Attachment are the same as currently marketed ANSPACH® |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | attachments. The OCM Burr Support Sleeves packaging,<br>manufacturing method and sterilization method are also the same as<br>currently marketed ANSPACH® burrs.<br>The new indications for use are consistent with numerous other<br>currently marketed surgical drill systems and accessories. |
| Non-Clinical<br>Testing | A series of biocompatibility tests demonstrated the safety of the OCM<br>Burr Support Sleeves patient contacting materials. All tests were<br>conducted in accordance with the GLP regulation (21 CFR Part 58).<br>Several performance tests demonstrated the safety and effectiveness of<br>the OCM Attachment and OCM Burr Support Sleeves. Specifically,<br>OCM Attachment testing demonstrated proper functionality after<br>repeated use, safe external temperatures, and the ability to clean and<br>sterilize effectively without adverse effects on functionality or<br>performance. OCM Burr Support Sleeves testing demonstrated proper<br>functionality and safe external temperatures. In addition, a simulated<br>use study demonstrated both the OCM Attachment and the OCM Burr<br>Support Sleeves perform as intended in cutting and shaping bone<br>properly. |
| Conclusions | Scientifically valid data demonstrates that the ANSPACH® XMax<br>Pneumatic and eMax 2 and eMax2 Plus Electric Systems with the<br>ANSPACH® Otologic Curved Micro (OCM) Attachment and OCM Burr<br>Support Sleeves is as safe and effective as the predicate devices when<br>used in accordance with the labeled directions for use and for the<br>requested indication. |
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three curved lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Public Health Service
December 31, 2013
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
The ANSPACH Effort, Inc. % Ms. Annette M. Hillring 3012 St. Charles Drive Tampa, FL 33618
Re: K131053
Trade/Device Name: ANSPACH® XMax Pneumatic and eMax 2 and eMax2 Plus Electrical Systems with Otologic Attachment System (Otologic Attachment and Curved Burrs) Regulation Number: 21 CFR 874.4250 Regulation Name: Ear, Nose and Throat Electric or Pneumatic Surgical Drill Regulatory Class: Class II Product Code: ERL; EQJ Dated: November 26, 2013 Received: December 2, 2013
Dear Ms. Hillring:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register,
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{4}------------------------------------------------
Page 2 - Ms. Annette M. Hillring
You must comply with all the Act's requirements, including, but not limited to: Registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.goy/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Bradley S. Cunningham -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear,
Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): _K131053_
Device Name: ANSPACH® XMax Pneumatic and eMax2 and eMax2 Plus Electrical Systems with Otologic Attachment System
The ANSPACH® XMax Pneumatic and eMax 2 and eMax2 Plus Indications for Use: Electric Systems are intended for cutting and shaping bone including spine and cranium.
When used with the ANSPACH® Systems, the OCM Attachment and OCM Burr Support Sleeves are intended for cutting and shaping bone primarily in otology procedures such as cochleostomies.
Prescription Use V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sageev George -S 2013.12.27 17:13:53 -05'00'
Page of of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.