Ambu aScope 5 Broncho HD 5.6/2.8 Sampler Set, Ambu aScope 5 Broncho HD 5.0/2.2 Sampler Set
Applicant
Ambu A/S
Product Code
EOQ · Ear, Nose, Throat
Decision Date
Jan 12, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 874.4680
Device Class
Class 2
Indications for Use
Ambu® aScope™ 5 Broncho HD Sampler Set is intended for endoscopic procedures and examination within the airways and tracheobronchial tree. Ambu® aScope™ 5 Broncho HD Sampler Set is intended to provide visualization via a compatible Ambu displaying unit. and to allow passing of endotherapy instruments via its working channel. Ambu® aScope™ 5 Broncho HD Sampler Set enables aspiration and collection of fluid samples.
Device Story
Single-use, flexible bronchoscope system; includes aScope 5 Broncho HD endoscope, BronchoSampler 60 SC container, Bronchoscope Attachment Part (BAP), Suction Connection Tube (SCT), and luer lock adapters. Device inputs: visual signals from distal tip camera/LED; fluid samples via suction pathway. Operation: endoscope tip maneuvered by clinician; BronchoSampler inserted into suction pathway between endoscope and vacuum source; BAP maintains suction pathway connections. Output: real-time visualization on compatible Ambu display unit; collected fluid samples in container. Used in clinical settings for airway examination and sampling. Clinician uses visual output to guide endotherapy instruments and perform diagnostic aspiration. Benefits: sterile, single-use design eliminates reprocessing needs; integrated sampling capability allows multiple samples during one procedure.
Clinical Evidence
Bench testing only. No clinical data provided. Verification included transportation, packaging, suction, connectivity, functional testing of the sampler, biological evaluation, and sterilization validation.
Technological Characteristics
Flexible, single-use endoscope with distal camera and LED light source. Materials sterilized via Ethylene Oxide. Includes suction pathway accessories (sampler container, BAP, SCT). Connectivity: wired to compatible Ambu display unit. Design verification performed per ISO 14971.
Indications for Use
Indicated for endoscopic procedures and examination within the airways and tracheobronchial tree, including visualization, passage of endotherapy instruments, and aspiration/collection of fluid samples.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 12, 2023
Ambu A/S % Sanjay Parikh Director, QA/RA Ambu Inc. 6721 Columbia Gateway Drive, suite 200 Columbia, Maryland 21046
Re: K223782
Trade/Device Name: Ambu aScope 5 Broncho HD 5.6/2.8 Sampler Set, Ambu aScope 5 Broncho HD 5.0/2.2 Sampler Set Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: EOQ Dated: December 12, 2022 Received: December 16, 2022
Dear Sanjay Parikh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation -emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory. ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K223782
Device Name
Ambu® aScope™ 5 Broncho HD 5.6/2.8 Sampler Set Ambu® aScope™ 5 Broncho HD 5.0/2.2 Sampler Set
## Indications for Use (Describe)
Ambu® aScope™ 5 Broncho HD Sampler Set is intended for endoscopic procedures and examination within the airways and tracheobronchial tree.
Ambu® aScope™ 5 Broncho HD Sampler Set is intended to provide visualization via a compatible Ambu displaying unit. and to allow passing of endotherapy instruments via its working channel.
Ambu® aScope™ 5 Broncho HD Sampler Set enables aspiration and collection of fluid samples.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| <span> <span style="text-decoration: underline;"></span>Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
This Special 510(k) summary has been prepared in accordance with 21 CFR 807.87(h) and the content and format of the 510(k) summary has been prepared in accordance with 21 CFR 807.92.
| Submitter | Ambu A/S<br>Baltorpbakken 13<br>2750 Ballerup<br>Denmark<br>Tel.: +45 7225 2000<br>Fax.: +45 7225 2050 | | | |
|--------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|---------------------------------------------|-----------------------------|
| Contact Person | Name: Kristian Moltved<br>Job Title: Associate Regulatory Affairs Professional<br>Address: Ambu A/S, Baltorpbakken 13, 2750 Ballerup<br>Telephone number: +45 7225 2116<br>Fax number: +45 7225 2050 | | | |
| Date Summary<br>Prepared | December 12, 2022 | | | |
| Device Trade<br>Name | Ambu® aScope™ 5 Broncho HD 5.6/2.8 Sampler Set<br>Ambu® aScope™ 5 Broncho HD 5.0/2.2 Sampler Set | | | |
| Device Common<br>Name | Flexible Endoscope and Specimen Sampling System - Single Use | | | |
| Device<br>Classification | Ambu® aScope™ 5 Broncho HD Sampler Set:<br>Bronchoscope (flexible or rigid) and accessories<br>Product Codes: EOQ<br>21 CFR 874.4680<br>Class II | | | |
| Legally Marketed<br>devices to which<br>the device is<br>substantially<br>equivalent | Predicate | Manufacturer<br>Ambu A/S | Trade Name<br>Ambu® aScope™<br>5 Broncho HD | 510(k)<br>number<br>K220606 |
{4}------------------------------------------------
| Description of<br>the Device | The Ambu® aScope™ 5 Broncho HD Sampler Set consists of:<br>Ambu® aScope™ 5 Broncho HD Two Sample Containers (aScope BronchoSampler™ 60 SC) Suction Connection Tube (SCT) Bronchoscope Attachment Part (BAP) 2 Luer lock adapters (introducers) |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Ambu® aScope™ 5 Broncho HD is currently being sold on its own as a sterile, single use, flexible endoscope intended for endoscopic procedures and examination within the airways and tracheobronchial tree. |
| | The Ambu® BronchoSampler™ 60 SC is a sterile, single-use medical device designed as an accessory to aScope 5 Broncho HD single- use endoscopes. It enables aspiration and collection of fluid samples. |
| | The Ambu® BronchoSampler™ 60 SC will be sold alone as an add-on to aScope 5 Broncho HD and as a part in the Ambu® aScope™ 5 Broncho HD Sampler Set. |
| | The Ambu® BronchoSampler™ 60 SC achieves its function by being inserted into the suction pathway between the endoscope and the vacuum source. |
| | The Sample Container Ambu® BronchoSampler™ 60 SC can be removed and replaced by the user to support multiple samples being taken during the same procedure. |
| | The Bronchoscope Attachment Part (BAP) is inserted into the suction pathway between the Ambu® aScope™ 5 Broncho HD endoscope and the vacuum source. It functions to maintain the connections and the controls for the suction pathway and the Ambu® BronchoSampler™ 60 SC, where the aspirated sample is stored. |
{5}------------------------------------------------
The Suction Connection Tube (SCT) facilitates connection to suction tubing having end connectors of the male type, instead of the more typical female type.
The Ambu® BronchoSampler™ 60 SC plus an additional Sampler Container, BAP, SCT and two introducers will be packaged and sterilized together with a aScope 5 Broncho HD endoscope to form a self-contained set, the aScope 5 Broncho HD Sampler Set.
There will be two sets, corresponding to the two sizes of aScope 5 Broncho HD:
- a. Ambu® aScope 5 Broncho HD 5.6/2.8 Sampler Set
- b. Ambu® aScope 5 Broncho HD 5.0/2.2 Sampler Set
Ambu® aScope™ 5 Broncho HD Sampler Set has the following physical and performance characteristics:
- Maneuverable Ambu® aScope 5 Broncho HD endoscope tip . controlled by the user
- Flexible Ambu® aScope 5 Broncho HD endoscope insertion . cord
- . Camera and LED light source at the distal tip of the Ambu® aScope 5 Broncho HD endoscope
- . Sterilized by Ethylene Oxide
- For single use
- Enables aspiration and sample collection ●
The following characteristic of the endoscope varies between sizes:
- Distal end outer diameter
- Insertion tube outer diameter
- Working channel inner diameter ●
- . Angulation range
{6}------------------------------------------------
| Indications for<br>Use | Ambu® aScope™ 5 Broncho HD Sampler Set is intended for<br>endoscopic procedures and examination within the airways and<br>tracheobronchial tree.<br><br>Ambu® aScope™ 5 Broncho HD Sampler Set is intended to<br>provide visualization via a compatible Ambu displaying unit and to<br>allow passing of endotherapy instruments via its working channel.<br><br>Ambu® aScope™ 5 Broncho HD Sampler Set enables aspiration<br>and collection of fluid samples. |
|---------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary of the<br>technological<br>characteristics<br>in comparison to<br>the predicate<br>device | The Ambu® aScope™ 5 Broncho HD endoscope in the Ambu®<br>aScope™ 5 HD Broncho Sampler Set is identical to the predicate. |
| | Ambu® aScope™ 5 Broncho HD Sampler Set differs from the<br>predicate in the following areas:<br>Ambu® aScope™ BronchoSampler 60 SC, BAP, SCT and<br>two introducers are added to the Ambu® aScope™ 5<br>Broncho HD endoscope. Ambu® aScope™ BronchoSampler 60 SC, BAP, SCT and<br>two introducers are packaged and sterilized together with<br>aScope 5 Broncho HD endoscopes to form a self-contained<br>set, the Ambu® aScope™ 5 Broncho HD Sampler Set. |
| Performance<br>Data -Bench | Performance tests related to the modifications was performed to<br>document the following properties of the modification to Ambu®<br>aScope™ 5 Broncho HD.<br><br>In addition to the non-clinical performance testing documented for<br>the Ambu® aScope™ 5 Broncho HD (K220606), additional bench<br>testing has been performed in accordance with ISO 14971 as<br>Design Verification of those Design Input requirements that are<br>specified as risk control measures for those risks arising from the<br>addition of sampling functionality to the endoscope. A table of this<br>special 510(k) cross references these additional risks, documented<br>in the Product Risk Evaluation for the Ambu® aScope™ 5 Broncho<br>Sampler Set, with the risk control measures and the subsequent<br>verification results. |
| | A summary of the methods to verify the additional requirements<br>and identified risks arising from the risk control measures are as<br>follows:<br>• Transportation testing<br>• Packaging testing<br>• Suction Testing<br>• Connectivity testing<br>• Tests for basic function of BronchoSampler 60<br>• Biological evaluation<br>• Sterilization Validation<br>Result: All requirements are verified successfully. |
| Performance<br>Data - Clinical | Not applicable. |
| Conclusion | The Ambu® aScope™ 5 Broncho Sampler Set has the same<br>intended use and indications for use, and similar technological<br>characteristics and principles of operation as the predicate device. |
| | It is concluded that Ambu® aScope™ 5 Broncho Sampler Set is<br>substantial equivalent to its predicate device. |
{7}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.