← Product Code [EOQ](/submissions/EN/subpart-e%E2%80%94surgical-devices/EOQ) · K222014

# EXALT Model B Single-Use Bronchoscope (K222014)

_Boston Scientific Corporation · EOQ · Jul 28, 2022 · Ear, Nose, Throat · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/EOQ/K222014

## Device Facts

- **Applicant:** Boston Scientific Corporation
- **Product Code:** [EOQ](/submissions/EN/subpart-e%E2%80%94surgical-devices/EOQ.md)
- **Decision Date:** Jul 28, 2022
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 874.4680
- **Device Class:** Class 2
- **Review Panel:** Ear, Nose, Throat

## Intended Use

The EXALT Model B Single-Use Bronchoscope is intended for use with the EXALT Monitor for endoscopic procedures within the airways and tracheobronchial tree.

## Device Story

EXALT Model B is a sterile, single-use flexible bronchoscope available in Slim, Regular, and Large sizes. It functions by capturing live video of the airways and trachea, which is transmitted to the EXALT Monitor for clinician visualization. Regular and Large sizes support the delivery of endoscopic accessories. The device is used in clinical settings for airway examination. It is operated by a physician or qualified clinician. The output allows real-time visualization of the tracheobronchial tree, facilitating diagnostic or therapeutic interventions. The single-use nature eliminates the need for reprocessing, reducing cross-contamination risks.

## Clinical Evidence

No clinical data. Bench testing was conducted on samples of the original predicate device (Ambu aScope 4 Broncho, K173727) to establish insertion tube stiffness specifications.

## Technological Characteristics

Single-use, flexible bronchoscope available in three sizes (Slim, Regular, Large). Features include an insertion tube, video imaging capability, and accessory channels (Regular/Large). Connectivity is via the EXALT Monitor. Sterilized for single-use. Technological characteristics are identical to the predicate device, with updated insertion tube stiffness specifications.

## Regulatory Identification

A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.

## Predicate Devices

- EXALT Model B Single-Use Bronchoscope ([K211030](/device/K211030.md))

## Reference Devices

- Ambu aScope 4 Broncho ([K173727](/device/K173727.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

July 28, 2022

Boston Scientific Corporation Carter Navarro Fellow, Regulatory Affairs 100 Boston Scientific Way Marlborough, Massachusetts 01752

## Re: K222014

Trade/Device Name: EXALT Model B Single-Use Bronchoscope Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: EOQ Dated: July 7, 2022 Received: July 8, 2022

## Dear Carter Navarro:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K222014

Device Name EXALT Model B Single-Use Bronchoscope

Indications for Use (Describe)

The EXALT Model B Single-Use Bronchoscope is intended for use with the EXALT Monitor for endoscopic procedures within the airways and tracheobronchial tree.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Boston Scientific. The logo is in blue and features the words "Boston Scientific" stacked on top of each other. The font is a serif typeface, and the overall design is clean and professional.

Boston Scientific Corporation Endoscopy Division 100 Boston Scientific Way Marlborough, MA 01752 (508) 683-4000 www.bostonscientific.com

## 510(k) Summary for EXALT Model B Single-Use Bronchoscope (K222014)

#### 1. Submitter

Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01752

Carter Navarro Contact: Fellow, Regulatory Affairs Phone: (508) 382-0356 E-mail: carter.navarro@bsci.com

Date Prepared: July 7, 2022

#### 2. Device

| Trade Name:                | EXALT Model B Single-Use Bronchoscope (Slim);         |
|----------------------------|-------------------------------------------------------|
|                            | EXALT Model B Single-Use Bronchoscope (Regular);      |
|                            | EXALT Model B Single-Use Bronchoscope (Large);        |
|                            | Common Name:                                          |
| Product Code:              | EOQ                                                   |
| Device Class:              | Class II                                              |
| Device Panel:              | Ear Nose & Throat                                     |
| Classification Regulation: | 21 CFR 874.4680, Bronchoscope (flexible or rigid) and |
|                            | accessories                                           |

#### 3. Predicate Device

| Trade Name:                | EXALT Model B Single-Use Bronchoscope (Slim);                     |
|----------------------------|-------------------------------------------------------------------|
|                            | EXALT Model B Single-Use Bronchoscope (Regular);                  |
|                            | EXALT Model B Single-Use Bronchoscope (Large);                    |
| Clearance Number:          | K211030                                                           |
| Common Name:               | Bronchoscope (flexible or rigid)                                  |
| Product Code:              | EOQ                                                               |
| Device Class:              | Class II                                                          |
| Device Panel:              | Ear Nose & Throat                                                 |
| Classification Regulation: | 21 CFR 874.4680, Bronchoscope (flexible or rigid) and accessories |

{4}------------------------------------------------

#### 4. Device Description

The EXALT Model B Single-Use Bronchoscope is a sterile, single-use bronchoscope available in three sizes (Slim, Regular, and Large). The proposed device facilitates examination of the airways and trachea. delivery of accessories (Regular and Large sizes only), and delivery of live video when connected to an EXALT Monitor.

#### 5. Indications for Use

The EXALT Model B Single-Use Bronchoscope is intended for use with the EXALT Monitor for endoscopic procedures within the airways and tracheobronchial tree.

#### 6. Technological Characteristics

The proposed EXALT Model B Single-Use Bronchoscope is identical to the predicate device.

#### 7. Substantial Equivalence

The proposed EXALT Model B Single-Use Bronchoscope is identical to the predicate device. The purpose of this Special 510(k) was to revise the specifications for insertion tube stiffness for the Slim, Regular, and Large sizes of the proposed device. For all three sizes, the upper end of the two-sided specification was increased. The new specifications were set based on test results from new, unopened samples of the original predicate device (Ambu aScope 4 Broncho, K173727) to which the EXALT Model B Single-Use Bronchoscope was shown to be substantially equivalent in the original premarket notification (K211030).

#### 8. Performance Data

Performance testing was not required to establish a finding of substantial equivalence between the proposed device and the predicate device, since the device is unchanged, Performance testing was conducted on new, unopened samples of the original predicate device (Ambu aScope 4 Broncho, K173727) to establish the insertion tube stiffness seen in commercial use.

#### 9. Conclusion

Boston Scientific has demonstrated that the proposed EXALT Model B Single-Use Bronchoscope is substantially equivalent to the currently marketed predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/EOQ/K222014](https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/EOQ/K222014)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
