The iNod Ultrasound Guided Biopsy Needle is intended for use through a flexible bronchoscope for intraluminal sonographic imaging in the tracheobronchial tree and retrieval of specimens from patients with endobronchial lesions, peripheral lung nodules, or lung masses. The iNod Ultrasound Guidance Console is intended to be used with iNod Ultrasound Guided Biopsy Needle, to observe and to store real-time ultrasound images of endobronchial lesions, peripheral lung nodules, or lung masses located in airways and tracheobronchial tree.
Device Story
iNod Ultrasound Guidance System combines radial ultrasound probe and biopsy needle into single-use sterile device (SUD); used with Motor Drive Unit (MDU) and console. MDU rotates transducer to generate 360° ultrasound images; console interprets signals and displays real-time images on touchscreen tablet. Needle includes stainless-steel indicator strip for ultrasound visualization; needle exits at 11° angle for biopsy. Nitinol stylet expels samples. System used in clinical settings by physicians; provides real-time visualization of pulmonary lesions to guide biopsy; aids in specimen retrieval for diagnosis. Console allows image storage, annotation, and export. Benefits include simultaneous imaging and biopsy capability.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing only, including device passability, rotation, needle actuation, acoustic output, pulse-echo performance, and software functionality.
Technological Characteristics
Single-use radial ultrasound transducer and biopsy needle; stainless steel needle; Nitinol stylet. MDU provides electromechanical rotation. Console is software-based tablet with acquisition PC. Sterilization: E-beam (ISO 11137). Biocompatibility: ISO 10993-1. Electrical safety: IEC 60601-1, 60601-1-2, 60601-1-6, 60601-2-18, 60601-2-37. Connectivity: DICOM archiving.
Indications for Use
Indicated for patients with endobronchial lesions, peripheral lung nodules, or lung masses requiring intraluminal sonographic imaging and specimen retrieval via flexible bronchoscope.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 2, 2023
Boston Scientific Corporation Inderdeep Tiwana Principal, Regulatory Affairs 100 Boston Scientific Wav Marlborough, Massachusetts 01752
Re: K221340
Trade/Device Name: iNod Ultrasound Guided Biopsy Needle (UPN: M00502060), iNod Ultrasound Guidance Console (UPN: M00503210) Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: EOO, ITX, IYO, ODG Dated: February 8, 2023 Received: February 9, 2023
Dear Inderdeep Tiwana:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Ethan L. Nyberg -S
for James Lee, Ph.D Division Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221430
Device Name iNod Ultrasound Guidance Console
#### Indications for Use (Describe)
The iNod Ultrasound Guidance Console is intended to be used with iNod Ultrasound Guided Biopsy Needle, to observe and to store real-time ultrasound images of endobronchial lesions, peripheral lung nodules, or lung masses located in airways and tracheobronchial tree.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M14.9998 0H1.00024C0.44873 0 0 0.448486 0 1.00006V15C0 15.5515 0.44873 16 1.00024 16H14.9998C15.5513 16 16 15.5515 16 15V1.00006C16 0.448486 15.5513 0 14.9998 0ZM14.9998 15H1.00024V1H14.9998V15ZM12.7068 4.29297L6.20684 10.793L3.29285 7.87891L2.58589 8.58594L6.20684 12.207L13.4138 5.00006L12.7068 4.29297Z" fill="black"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <rect height="15" rx="1.5" stroke="black" width="15" x="0.5" y="0.5"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221430
Device Name iNod Ultrasound Guided Biopsy Needle
#### Indications for Use (Describe)
The iNod Ultrasound Guided Biopsy Needle is intended for use through a flexible bronchoscope for intraluminal sonographic imaging in the tracheobronchial tree and retrieval of specimens with endobronchial lesions, peripheral lung nodules, or lung masses.
Type of Use (Select one or both, as applicable)
| <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) Summary for iNod Ultrasound Guidance System
# 1. Submitter
Boston Scientific Corporation Endoscopy Division 100 Boston Scientific Way Marlborough, MA 01752
# Contact:
| Primary Contact: | Inderdeep Tiwana<br>Principal Regulatory Specialist<br>Phone: (508) 683-4005<br>E-mail: Inderdeep.tiwana@bsci.com |
|--------------------|--------------------------------------------------------------------------------------------------------------------------|
| Secondary Contact: | Ashley Santos<br>Senior Director, Regulatory Affairs-Endoscopy<br>Phone: (508) 382-0284<br>Email: Ashley.santos@bsci.com |
Date Prepared: May 06, 2022
{6}------------------------------------------------
# 2. Device
| Trade Name: | iNod Ultrasound Guided Biopsy Needle |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Diagnostic Ultrasound Transducer<br>Bronchoscope (flexible or rigid) and accessories |
| Product Code: | ITX and EOQ |
| Device Class: | Class II |
| Device Panel | Ear, Nose & Throat |
| Classification Regulation: | 21 CFR.874.4680 Bronchoscope (flexible or rigid) and<br>accessories |
| Trade Name: | iNod Ultrasound Guidance Console |
| Common Name: | IYO: System, imaging, pulsed echo, ultrasonic<br>ITX: Diagnostic Ultrasound Transducer<br>ODG: Endoscopic ultrasound system, gastroenterology-<br>urology |
| Product Code: | IYO, ITX and ODG |
| Device Class: | Class II |
| Device Panel | Radiology and Gastroenterology and Urology |
| Classification Regulation: | 21 CFR.892.1560 System, imaging, pulsed echo, ultrasonic |
# 3. Predicate Devices
| Proposed<br>Device | Component | Predicate Devices |
|---------------------------------------|--------------------------------------------|----------------------------------------------------------------------------------------------------|
| iNod Ultrasound<br>Guidance<br>System | iNod Ultrasound<br>Guided Biopsy<br>Needle | Probe:<br>• Olympus Ultrasound Probe (K982323)<br>Needle:<br>• Arcpoint Pulmonary Needle (K163537) |
| | iNod Ultrasound<br>Guidance<br>Console | • EVIS EUS Endoscopic Ultrasound Center<br>(K121564)<br>• Avvigo Ultrasound System II (K212490) |
| | iNod Motor<br>Drive Unit | • MDU5 Plus – 510K cleared with Avvigo Ultrasound<br>System II (K212490) |
{7}------------------------------------------------
## 4. Device Description
The proposed iNod Ultrasound Guidance System (iNod System), comprises: the sterile, single use iNod Ultrasound Guided Biopsy Needle (also referred to as iNod Single Use Device (SUD), iNod Ultrasound Guidance Console (iNod Console) and the iNod Motor Drive Unit (iNod MDU). The proposed iNod Ultrasound Guidance System is intended to perform pulmonary needle biopsy under real-time visualization using Radial Endobronchial Ultrasound (R-EBUS).
The iNod single-use-device (iNod SUD) is a single use, sterile device that has combined functionality of a radial ultrasound probe and a biopsy needle. The iNod SUD allows real-time visualization of pulmonary lesions with radial ultrasound, while simultaneously allowing biopsy of the pulmonary lesions. The three main parts of iNod SUD are a handle, a radial ultrasound transducer, and biopsy needle. The biopsy needle is actuated at an angle that allows biopsy of eccentric and concentric nodules. A stainless-steel needle indicator strip is housed in the imaging window at distal tip, which provides an ultrasound signature, indicating the orientation of the needle to the user in the ultrasound image. The biopsy needle will exit the iNod SUD distal tip at 180 degrees relative to the needle indicator position visible in the ultrasound image. Once the needle is positioned to biopsy a nodule, the needle is unlocked by user to the exit ramp at 11° angle. Finally, when a sample is collected, the Nitinol stylet can be passed down the needle lumen to expel the sample.
The iNod Console is an electronic device that consists of:
- The iNod MDU, which rotates the ultrasound transducer in the iNod SUD to generate 360° ● images as well as provides patient isolation from the rest of the Console components.
- A software-based touchscreen tablet with battery, which can be mounted to a mobile pole yia . the pole docking station.
- A software-based Acquisition PC that interprets the received ultrasound signals from the ● iNod SUD and generates the image that is displayed on the tablet.
- Functionality that displays ultrasonic image received from iNod SUD through iNod MDU ● and Acquisition PC on the iNod tablet and
- . Ability to save and export procedure images and recordings.
The iNod MDU provides rotation of the iNod SUD's ultrasound transducer, required for generating a 360° ultrasound image. An electromechanical connector interface at the proximal end of the iNod SUD makes the connection to the MDU-catheter interface consists of an integrated mechanical drive hub and electrical connection. The iNod Console interfaces with the iNod SUD through the iNod Motor Drive Unit (iNod MDU), which provides the electromechanics for the rotating parts of the imaging catheter, and the interface between the iNod SUD and the iNod console. The iNod MDU is nearly identical in design to its predicate MDU5 Plus, which is currently used with Avvigo Guidance System II (K212490).
{8}------------------------------------------------
## 5. Indications for Use
### iNod Ultrasound Guided Biopsy Needle:
The iNod Ultrasound Guided Biopsy Needle is intended for use through a flexible bronchoscope for intraluminal sonographic imaging in the tracheobronchial tree and retrieval of specimens from patients with endobronchial lesions, peripheral lung nodules, or lung masses. .
### iNod Ultrasound Guidance Console:
The iNod Ultrasound Guidance Console is intended to be used with iNod Ultrasound Guided Biopsy Needle, to observe and to store real-time ultrasound images of endobronchial lesions, peripheral lung nodules, or lung masses located in airways and tracheobronchial tree.
## 6. Technological Characteristics
The iNod Ultrasound Guided Biopsy Needle merges two devices: an ultrasound probe, and a lung biopsy needle, into one device. The vast majority of design characteristics are identical to the predicate devices, however, there are a few key technological differences between iNod SUD and its predicates. Performance testing has been executed to evaluate all design characteristics, as well as those in which there were differences in technological characteristics as compared to the predicates. The results of the bench-top testing were passing and demonstrates the proposed iNod SUD performs substantially equivalent to the predicate devices.
The iNod Ultrasound Guidance Console and its elements including the iNod MDU are nearly identical in design and physical attributes to its predicate AVVIGO Guidance System II (K212490). The main difference is the brand name, intended use as the AVVIGO Guidance System II is intended for use with catheters having cardiovascular indications for use, and removal of cardiovascular feature (pull-back feature) from the iNod MDU. All components and accessories of the proposed iNod Console are identical to the AVVIGO Guidance System II (K212490). In addition, the indications of use of the proposed iNod Console are identical to those of the primary predicate device, EVIS EUS Endoscopic Ultrasound (K203128) and there are no significant design differences between iNod Console and EVIS EUS Endoscopic Ultrasound (K203128).
{9}------------------------------------------------
## 7. Substantial Equivalence
The iNod Ultrasound Guided Biopsy Needle, iNod Console and iNod MDU devices are substantially equivalent to the following predicate devices.
| Proposed<br>Device | Component | Predicate Devices |
|---------------------------------------|--------------------------------------------------|----------------------------------------------------------------------------------------------------|
| iNod Ultrasound<br>Guidance<br>System | iNod<br>Ultrasound<br>Guided<br>Biopsy<br>Needle | Probe:<br>• Olympus Ultrasound Probe (K982323)<br>Needle:<br>• Arcpoint Pulmonary Needle (K163537) |
| | iNod<br>Console | • EVIS EUS Endoscopic Ultrasound Center (K121564)<br>• Avvigo Ultrasound System II (K212490) |
| | iNod MDU | • MDU5 Plus – 510K cleared with Avvigo Ultrasound<br>System II (K212490) |
Table 5-1 - Predicate Devices and 510(k) numbers
## 8. Performance Data
Non-clinical (bench) testing was successfully performed on the proposed iNod Ultrasound Guided Biopsy Needle, iNod Ultrasound Guidance Console and the iNod System altogether. Performance testing (bench) that represents device's clinical use was successfully completed. In addition, all necessary software tests were performed for the iNod Software. The following bench testing was conducted to evaluate the changes on the proposed device. All Performance testing (bench) was successfully completed. The results of performance (bench) testing demonstrate that the proposed iNod Ultrasound Guided Biopsy Needle, iNod Ultrasound Guidance Console and iNod Motor Drive Unit (MDU) are considered substantially equivalent to the predicate devices.
| Component | Product Specification | Results (Pass/Fail) |
|--------------|-----------------------------------------------------------------------------------------------------------------|---------------------|
| iNod SUD | Device Passability | Pass |
| iNod SUD | Device Rotation | Pass |
| iNod SUD | Needle Actuation Force | Pass |
| iNod SUD | Needle Extension Length | Pass |
| iNod SUD | Needle Angle | Pass |
| iNod SUD | Needle Outer Diameter | Pass |
| iNod SUD | Needle Inner Diameter | Pass |
| iNod SUD | Needle Sharpness | Pass |
| iNod SUD | Needle Aspiration | Pass |
| iNod SUD | Stylet Removal Force | Pass |
| iNod SUD | Stylet Kink Resistance | Pass |
| iNod SUD | Catheter Working Length | Pass |
| Component | Product Specification | Results (Pass/Fail) |
| iNod SUD | Handle to Catheter Tensile | Pass |
| iNod SUD | Drive Cable to Handle Tensile | Pass |
| iNod SUD | Drive Cable to MDU Connector Tensile | Pass |
| iNod SUD | Device Reliability | Pass |
| iNod SUD | Pulse Echo Sensitivity | Pass |
| iNod SUD | Pulse Echo Bandwidth | Pass |
| iNod SUD | Pulse Echo Center Frequency | Pass |
| iNod SUD | Pulse Echo Pulse Length | Pass |
| iNod SUD | Acoustic Output | Pass |
| iNod SUD | NURD (Non-Uniform Rotation Distortion) | Pass |
| iNod SUD | Needle Lock Override Force | Pass |
| iNod SUD | ISO 80369<br>Small-bore connectors for liquid and gases in<br>healthcare applications | Pass |
| iNod SUD | ISO 9626 Stainless Steel Needle Tubing for<br>Manufacture of Medical Devices - Requirements and<br>test methods | Pass |
| iNod SUD | Catheter Drive Cable Length | Pass |
| iNod SUD | Device Bronchoscopic Visibility | Pass |
| iNod SUD | Device Radiopacity | Pass |
| iNod SUD | Needle Indicator | Pass |
| iNod SUD | Needle Lock Engage/Disengage Force | Pass |
| iNod SUD | Handle Grip Area | Pass |
| iNod SUD | Actuation Mechanism Location | Pass |
| iNod SUD | Needle Locking Mechanism Location | Pass |
| iNod SUD | Handle Rotation | Pass |
| iNod SUD | Product can be held with scope | Pass |
| iNod SUD | Catheter Outer Diameter | Pass |
| iNod SUD | Imaging System Compatibility | Pass |
| iNod SUD | Motor Drive Compatibility - Torque | Pass |
| iNod SUD | Motor Drive Compatibility – Load | Pass |
| iNod SUD | Needle Exit Location | Pass |
| iNod Console | Archive Case Studies | Pass |
| iNod Console | Create Case Study | Pass |
| iNod Console | Annotate | Pass |
| iNod Console | Patient Display | Pass |
| iNod Console | Add Snapshots | Pass |
| iNod Console | Select and Review Case Studies | Pass |
| iNod Console | Export Event Logs | Pass |
| iNod Console | Export Recordings and Screenshots | Pass |
| iNod Console | Import/Export Case studies | Pass |
| iNod Console | Patient List Query | Pass |
| iNod Console | Select Patient | Pass |
| iNod Console | Mobile Configuration | Pass |
| iNod Console | Record Management | Pass |
| iNod Console | Default Modality | Pass |
| iNod Console | Power-On Self Test Status | Pass |
| iNod Console | Operating System | Pass |
| iNod Console | Power Up Duration | Pass |
| iNod Console | DICOM Archiving Performance | Pass |
| iNod Console | Expected Service Life | Pass |
| Component | Product Specification | Results (Pass/Fail) |
| iNod Console | System Access to health delivery organization<br>network | Pass |
| iNod Console | User Control | Pass |
| iNod Console | System Displays | Pass |
| iNod Console | Image Artifact Removal | Pass |
| iNod Console | Start Recording | Pass |
| iNod Console | Stop Recording | Pass |
| iNod Console | Linear Measurement on Ultrasound Frame | Pass |
| iNod Console | Imaging Depth | Pass |
| iNod Console | Distance Accuracy on Ultrasound Frame | Pass |
| iNod Console | Image Frame Rate: Ultrasound catheters | Pass |
| iNod Console | Gap detection | Pass |
| iNod Console | Catheter compatibility and Bandwidth | Pass |
| iNod Console | SUD Retention Force | Pass |
| iNod MDU | Receive Path Spectral Response | Pass |
| iNod MDU | Bi-Polar Tx - P-P Output Voltage Amplitude | Pass |
| iNod MDU | Pulse Uniformity – Bi-Polar Pulses | Pass |
| iNod MDU | Bi-Polar Output Bandwidth | Pass |
| iNod MDU | Receive Path Common Mode Gain | Pass |
| iNod MDU | RF Receive Path Noise Figure | Pass |
| iNod MDU | Receive Path Spurious Noise | Pass |
| iNod MDU | Pull Force and Side Force | Pass |
| iNod MDU | Rotational Speed | Pass |
Table 5-2 Bench/ Performance Testing Summary
{10}------------------------------------------------
{11}------------------------------------------------
iNod Ultrasound Guided Biopsy Needle is sterilized using E-beam sterilization method and E-Beam Radiation cycle is validated per Method VDmax27.5 to a sterility assurance level of 100 in accordance with EN ISO 11137-1: 2006, EN ISO 11137-2: 2013, and ISO TS13004:2014. Validation included dose mapping, bioburden assessment, and dose audit / dose verification testing to validate an SAL of 106 in routine sterilization processing. All relevant aspects of the applicable standards were met. In addition, the iNod Ultrasound Guided Biopsy Needle is undergoing routine bacterial endotoxins test (BET), also known as the Limulus amebocyte lysate (LAL) test, in accordance with standard AAMI/ANSI ST72: Bacterial endotoxin - test methods, routine monitoring and alternatives to batch resting (Sterility).
Biocompatibility of the proposed iNod Ultrasound Guided Biopsy Needle was evaluated in accordance with ISO 10993-1. Based on the biocompatibility classification of the device the following biocompatibility tests were conducted: Cytotoxicity, Sensitization, Toxicity, and Pyrogen Testing.
Lastly, iNod Ultrasound Guidance System was evaluated for electrical safety under electoral safety standards ISO 60601-1, ISO 60601-1-2, ISO 60601-1-6, 60601-2-18 and ISO 60601-2-37. All tests met the predefined requirements and further support the safety of the proposed devices.
# 9. Conclusion
Boston Scientific has demonstrated that the proposed iNod Ultrasound Guidance System is substantially equivalent to the currently marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.