aScope 4 Broncho Regular Sampler Set 5.0/2.2, aScope 4 Broncho Large Sampler Set 5.8/2.8
Applicant
Ambu A/S
Product Code
EOQ · Ear, Nose, Throat
Decision Date
Jun 17, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 874.4680
Device Class
Class 2
Indications for Use
aScope 4 Broncho Sampler Set consists of sterile, single use, flexible endoscope with sample containers (aScope BronchoSampler), intended for endoscopic procedures and examination within the airways and tracheobronchial tree. aScope BronchoSampler is designed as an add-on to aScope 4 Bronchial Alveolar Lavage (BAL) or Bronchial Wash (BW) procedure which enables aspiration of fluid sample(s) from the bronchial or alveolar part of the lung. It is designed for use in adults and intended for use in a hospital environment. It is intended to provide visualization via Ambu Displaying Unit.
Device Story
Sterile, single-use flexible endoscope with integrated sample container (aScope BronchoSampler); used for airway/tracheobronchial tree examination and fluid aspiration (BAL/BW). Device features maneuverable tip, insertion cord, distal camera, and LED light source. Operated by clinicians in hospital settings; connects to Ambu Displaying Unit for real-time visualization. BronchoSampler attaches to endoscope handle to facilitate fluid collection. Provides direct visualization to guide procedures and enables diagnostic sample acquisition. Benefits include reduced cross-contamination risk due to single-use design and streamlined workflow for specimen collection.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including ISO 8600-1/3/4 (endoscope performance), EN 14254 (receptacles), IEC 60601-1/2-18 (electrical safety), and IEC 60601-1-2 (EMC).
Technological Characteristics
Single-use flexible video endoscope; distal camera and LED illumination; suction functionality. Sterilized via Ethylene Oxide. Complies with ISO 8600-1, ISO 8600-3, ISO 8600-4, IEC 60601-1, IEC 60601-2-18, and IEC 60601-1-2. Connects to external Ambu Displaying Unit.
Indications for Use
Indicated for endoscopic procedures and examination of the airways and tracheobronchial tree in adults in a hospital environment, including aspiration of fluid samples during BAL or BW procedures.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
Predicate Devices
Ambu® aScope™ 4 Broncho Regular and Large (K173727)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
June 17, 2021
Ambu A/S % Sanjay Parikh Director, QA/RA Ambu Inc. 6230 Old Dobbin Lane, Suite 250 Columbia, Maryland 21045
Re: K210883
Trade/Device Name: aScope 4 Broncho Regular Sampler Set 5.0/2.2, aScope 4 Broncho Large Sampler Set 5.8/2.8 Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: EOQ Dated: May 17, 2021 Received: May 18, 2021
Dear Sanjay Parikh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Shu-Chen Peng Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K210883
#### Device Name
Ambu® aScope™ 4 Broncho Regular Sampler Set 5.0/2.2 Ambu® aScope™ 4 Broncho Large Sampler Set 5.8/2.8
#### Indications for Use (Describe)
aScope 4 Broncho Sampler Set consists of sterile, single use, flexible endoscope with sample containers (aScope BronchoSampler), intended for endoscopic procedures and examination within the airways and tracheobronchial tree. aScope BronchoSampler is designed as an add-on to aScope 4 Bronchial Alveolar Lavage (BAL) or Bronchial Wash (BW) procedure which enables aspiration of fluid sample(s) from the bronchial or alveolar part of the lung.
It is designed for use in adults and intended for use in a hospital environment.
It is intended to provide visualization via Ambu Displaying Unit.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
This Special 510(k) summary has been prepared in accordance with 21 CFR 807.87(h) and the content and format of the 510(k) summery has been prepared in accordance with 21 CFR 807.92.
| Submitter | Ambu A/S<br>Baltorpbakken 13<br>DK-2750 Ballerup<br>Denmark<br>Tel.: +45 7225 2000<br>Fax.: +45 7225 2050 | | | |
|-----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|-------------------------------------------------|-------------|
| Contact Person | Name: Gurpreet Kaur Rehal<br>Job Title: Regulatory Affairs Professional<br>Address: Ambu A/S, Baltorpbakken 13, DK-2750 Ballerup<br>Telephone number: +45 7225 2116<br>Fax number: +45 7225 2050 | | | |
| Date Summary<br>Prepared | March, 23, 2021 | | | |
| 510(k) no. | K210883 | | | |
| Device Trade<br>Name | Ambu® aScope™ 4 Broncho Regular Sampler Set 5.0/2.2<br>Ambu® aScope™ 4 Broncho Large Sampler Set 5.8/2.8 | | | |
| Device Common<br>Name | Flexible Endoscope and Specimen Sampling System - Single Use | | | |
| Device<br>Classification | Bronchoscope (flexible or rigid) and accessories<br>Product Codes: EOQ<br>21 CFR 874.4680<br>Class II | | | |
| Legally<br>Marketed<br>devices to which<br>the device is<br>substantially<br>equivalent | Predicate<br>A: | Manufacturer | Trade Name | 510k number |
| | | Ambu A/S | Ambu® aScope™ 4<br>Broncho Regular and<br>Large | K173727 |
{5}------------------------------------------------
| Description of the Device | The Ambu® aScope™ 4 Broncho Sampler Set consists of:<br>• aScope 4 Regular/Large endoscope<br>• aScope BronchoSampler ™<br><br>Ambu® aScope™ 4 Broncho Sampler Set consists of sterile, single use, flexible endoscope with sample containers (aScope BronchoSampler), intended for endoscopic procedures and examination within the airways and tracheobronchial tree. aScope BronchoSampler is designed as an add-on to aScope 4 Broncho during Bronchial Alveolar Lavage (BAL) or Bronchial Wash (BW) procedure which enables aspiration and collection of fluid sample(s) from the bronchial or alveolar part of the lung. Both of the aScope 4 Broncho and BronchoSampler are already marketed separately.<br><br>Ambu® aScope™ 4 Broncho Sampler Set has the following physical and performance characteristics:<br>• Maneuverable tip controlled by the user<br>• Flexible insertion cord<br>• Camera and LED light source at the distal tip<br>• Sterilized by Ethylene Oxide<br>• For single use<br>• Enables aspiration and sample collection in BAL and BW procedures<br><br>The differences between the endoscope sizes are as follows:<br>• Distal end outer diameter<br>• Insertion tube outer diameter<br>• Working channel inner diameter<br>• Angulation range |
|-------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Ambu® aScope™ 4 Broncho Sampler Set consists of sterile, single use, flexible endoscope with sample containers (aScope BronchoSampler), intended for endoscopic procedures and examination within the airways and tracheobronchial tree. aScope BronchoSampler is designed as an add-on to aScope 4 Broncho during Bronchial Alveolar Lavage (BAL) or Bronchial Wash (BW) procedure which enables aspiration and collection of fluid sample(s) from the bronchial or alveolar part of the lung.<br><br>It is designed for use in adults and intended for use in a hospital environment.<br><br>It is intended to provide visualization via Ambu Displaying Unit. |
| Summary of the technological characteristics in comparison to the predicate devices | The endoscope of Ambu® aScope™ 4 Broncho Sampler Set is similar to the predicate device A in the following areas:<br>• They are all single-use devices delivered sterile.<br>• They are all flexible endoscopes with a maneuverable tip.<br>• They are all video endoscopes with a camera located in the distal tip to provide an image on a separate displaying unit.<br>• They all provide illumination from the distal tip.<br>• They all have suction functionality<br>• They all have the same insertion tube working length |
| | They all have equivalent inner and outer diameters in their<br>corresponding sizes.<br>The endoscope of Ambu® aScope™ 4 Broncho Sampler Set differs from<br>the predicate device A in the following areas:<br>Bronchosampler is added in Ambu® aScope™ 4 Broncho Sampler<br>Set which can be attached to the handle of the endoscope. |
| Performance<br>Data -Bench | The following data are described for the product line extension Ambu®<br>aScope™ 4 Broncho Sampler Set in the premarket notification: |
| | Declaration of Conformity with the product specific standards ISO<br>8600-1, ISO 8600-3 and ISO 8600-4Aging performance testSterile Packaging IntegrityElectromagnetic Compatibility according to IEC 60601-1-2Electrical Safety according to IEC 60601-1 and IEC 60601-2-18.Performance test for Single-use receptacles according to EN<br>14254 |
| | Result: All tests were passed. |
| Performance<br>Data - Clinical | Not applicable. |
| Conclusion | Based on the indication for use, technological characteristics,<br>performance data and comparison to predicate device it has been<br>concluded that the functionality and intended use of Ambu® aScope™ 4<br>Broncho Sampler Set is substantially equivalent to the predicate device.<br>It is concluded that Ambu® aScope™ 4 Broncho Sampler Set is as safe<br>and as effective and performs as well as the predicate device. |
{6}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.