LungVision Tool is an instrument designed as a working channel intended to be used with standard bronchoscopes, endotherapy accessories and ultrasound probe to guide the endotherapy accessories or ultrasound probe to the target area, specifically within the respiratory system.
Device Story
LungVision Tool functions as a working channel for standard bronchoscopes; enables guidance of endotherapy accessories or ultrasound probes to target areas within respiratory system. Operated by pulmonologists or thoracic surgeons in hospitals, sub-acute care, or physician offices. Device consists of handle, adaptor, sheath, and wire/stylet. Physician connects tool to bronchoscope instrument port; inserts accessories through sheath. Provides mechanical pathway for instruments; facilitates navigation to target anatomy. Benefits patient by enabling targeted diagnostic or therapeutic access during bronchoscopy. Single-use, sterile device.
Clinical Evidence
No clinical data. Bench testing only. Testing included biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity), mechanical performance (tensile strength, sheath insertion/withdrawal, twisting, drop tests), and sterilization validation (ETO, SAL 10^-6).
Technological Characteristics
Working channel for bronchoscopes; includes handle, adaptor, sheath, and wire. Sheath max OD 2.72 mm, min ID 2.08 mm. Materials biocompatible per ISO 10993. Manual attachment mechanism. Single-use. Sterile (Ethylene Oxide).
Indications for Use
Indicated for patients undergoing endoscopic procedures in the respiratory system, serving as a working channel to guide endotherapy accessories or ultrasound probes to target areas.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
Boston Scientific Expect™ Pulmonary Endobronchial Ultrasound Transbronchial Aspiration Needle (K151315)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA acronym is in a blue square, and the full name "U.S. Food & Drug Administration" is in blue text.
April 26, 2018
Body Vision Medical Ltd. % Paul Dryden Consultant ProMedic, LLC 131 Bay Point Dr. NE St. Petersburg, FL 33704
Re: K172955
Trade/Device Name: LungVision Tool Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible or Rigid) and Accessories Regulatory Class: Class II Product Code: EOQ Dated: March 22, 2018 Received: March 23, 2018
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely yours,
Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172955
Device Name LungVision Tool
Indications for Use (Describe)
Lung Vision Tool is an instrument designed as a working channel intended to be used with standard bronchoscopes, endotherapy accessories and ultrasound probe to guide the endotherapy accessories or ultrasound probe to the target area, specifically within the respiratory system.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| <span style="padding-right: 5px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="padding-right: 5px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| Official Contact: | Dorian Averbuch<br>Body Vision Medical Ltd.<br>34 Sokolov St.<br>Ramat Hasharon Israel<br>Tel: 646- 863-7848 (US) |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | April 19, 2018 |
| Proprietary or Trade Name: | LungVision Tool |
| Common/Usual Name: | Bronchoscope (flexible or rigid) and accessories |
| Classification Name: | Bronchoscope (flexible or rigid) and accessories<br>EOQ, Class II, CFR 874.4680 |
| Predicate Device:<br>Reference Device: | K060243 - Olympus Sheath Guide<br>K151315 - Boston Scientific Expect™ Pulmonary Endobronchial<br>Ultrasound Transbronchial Aspiration Needle ("EPEUTAN") |
#### Device Description:
Lung Vision Tool is an instrument designed as a working channel intended to be used with standard bronchoscopes, endo-therapy accessories and ultrasound probe to guide the endo-therapy accessories or ultrasound probe to the target area, specifically within the respiratory system.
LungVision tool is designed to be used by pulmonologists or thoracic surgeons during pulmonary procedures.
The LungVision Tool is designed to be connected to a standard Bronchoscope Instrument Port.
The Lung Vision Bronchoscope Adaptor is used to connect the LungVision Handle to the Bronchoscope instrument port / working channel entrance, and to allow the physician to release the tool from the bronchoscope. The current adaptor is for the Olympus Bronchoscope Model BF-1T160.
More types of Bronchoscope Adaptors for other bronchoscopy brands and models.
Lung Vision Tool Sheath can accommodate endo-therapy accessories and ultrasound probes with an outer diameter up to 1.9 mm.
#### Indications for Use:
Lung Vision Tool is an instrument designed as a working channel intended to be used with standard bronchoscopes, endotherapy accessories and ultrasound probe to guide the endotherapy accessories or ultrasound probe to the target area, specifically within the respiratory system.
#### Patient Population
Patients undergoing endoscopic procedures.
#### Contraindications
Contraindications are limited to those of the use of a bronchoscope of the endotherapy instruments that the clinician may desire to use.
{5}------------------------------------------------
#### Environments of use
Anywhere a patient may undergo a bronchoscopy procedure, e.g., hospital, sub-acute care or physician office settings.
#### Device Comparison
The following tables compare the subject device to the predicate and reference devices.
| Feature | Predicate<br>K060243<br>Olympus Guide Sheath | Reference<br>K151315<br>Boston EPEUTAN | Subject device<br>K172955<br>LungVision Tool |
|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | To be used with<br>bronchoscopes and endo-<br>therapy accessories | To be used with<br>bronchoscopes and endo-<br>therapy accessories | To be used with<br>bronchoscopes and endo-<br>therapy accessories |
| Anatomical location | Respiratory organs | Respiratory organs | Respiratory organs |
| Components | | | |
| Handle | Handle-like | Yes | Yes |
| Connector / Adaptor to<br>attach to working channel | Stopper | Yes | Yes |
| Sheath | Yes | Yes | Yes |
| Wire / stylet to stiffen or<br>protect the sheath | No | Yes | Yes |
| Works with endo-therapy<br>instruments | Yes | Yes | Yes |
| Feature | Predicate<br>K060243<br>Olympus Guide Sheath | Reference<br>K151315<br>Boston EPEUTAN | Subject Device<br>K172955<br>LungVision Tool |
| Indications for Use | This instrument has been designed to be used<br>with Olympus bronchoscopes, endo-therapy<br>accessories and ultrasound probe to guide<br>the endo-therapy accessories or ultrasound<br>probe to the target area within the respiratory<br>organs. | The Expect™ Pulmonary Endobronchial<br>Ultrasound Transbronchial Aspiration Needle<br>is designed to be used with endobronchial<br>ultrasound endoscopes for ultrasound guided<br>fine needle aspiration (FNA) of the<br>submucosal and extramural lesions of the<br>tracheobronchial tree. Do not use this<br>instrument for any purpose other than its<br>intended use. | LungVision Tool is an instrument designed<br>as a working channel intended to be used<br>with standard bronchoscopes, endotherapy<br>accessories and ultrasound probe to guide<br>the endotherapy accessories or ultrasound<br>probe to the target area, specifically within<br>the respiratory system. |
| Classification | Bronchoscope (flexible or rigid) and<br>accessories.<br>21CFR 874.4680<br>Procut Code: EOQ | Bronchoscope (flexible or rigid) and<br>accessories.<br>21CFR 874.4680<br>Procut Code: EOQ | Bronchoscope (flexible or rigid) and<br>accessories.<br>21CFR 874.4680<br>Procut Code: EOQ |
| Target anatomy | Respiratory Organs | Respiratory Organs | Respiratory Organs |
| Anatomy access | Bronchial airways | Bronchial airways | Bronchial airways |
| Patient Population | Not specified but for patients undergoing<br>endoscopic procedures | Not specified but for patients undergoing<br>endoscopic procedures | Patients undergoing endoscopic procedures |
| Environment of use | Not specified but where endoscopic<br>procedures are performed: hospitals, sub-<br>acute and physician office settings | Not specified but where endoscopic procedures<br>are performed: hospitals, sub-acute and<br>physician office settings | Not specified but where endoscopic<br>procedures are performed: hospitals, sub-<br>acute and physician office settings |
| Single Use | Yes | Yes | Yes |
| Sterile | Yes | Yes | Yes |
| Mechanism of action | Manual attachment | Manual attachment | Manual attachment |
| Number of device<br>passes during a<br>procedure | Multiple passes | Multiple passes | Multiple passes |
| Echogenic | No | Yes | No |
| X-ray detection | Yes | No | Yes |
| Components | Handle-like<br>Stopper (Connector)<br>Sheath | Handle<br>Adaptor / Connector<br>Sheath<br>Wire / Stylet | Handle<br>Adaptor / Connector<br>Sheath<br>Wire |
| Sheath Max Outer<br>Diameter | 2.7 mm | N/A | 2.72 mm |
| Sheath Min Inner<br>Diameter | 2.1 mm | N/A | 2.08 mm |
| Feature | Predicate<br>K060243<br>Olympus Guide Sheath | Reference<br>K151315<br>Boston EPEUTAN | Subject Device<br>K172955<br>LungVision Tool |
| Working Length | 900 mm | N/A | 1000 mm |
| Length extended<br>beyond the scope | 27.5 mm | N/A | 17 mm |
| Sheath length<br>adjustable | Yes | Yes | Yes |
| Recommended<br>Channel Size | 2.8 mm | 2.0 mm (this device is inserted into a sheath | 2.8mm |
| Curved distal tip | No, but endotherapy instruments do | N/A | Yes |
| Performance - Non-clinical | | | |
| Biocompatibility | Surface Contact<br>Skin and Breached<br>Limited duration (<24 hours) | Surface Contact<br>Skin and Breached<br>Limited duration (<24 hours) | Surface Contact<br>Skin and Breached<br>Limited duration (<24 hours) |
| Testing | | | Cytotoxicity<br>Sensitization<br>Irritation or Intracutaneous Reactivity<br>Acute Systemic Toxicity<br>Material Mediated Pyrogenicity – no claim<br>of non-pyrogenic |
| Bench testing | Not listed | Passability<br>Tensile strength - Stylet (wire)<br>Durability<br>Removal Force<br>Rotation | Bronchoscope Compatibility Test<br>Simulated Use<br>Fluoroscopy<br>Tensile Test for Sheath and Wire<br>Tensile Test for Wire Distal End<br>Repeatability Use<br>Sheath Insertion/Withdrawal<br>Twisting Test<br>Drop Test<br>Accelerated Aging Verification Test<br>Package integrity Test<br>Package Stability Test |
As can be seen the predicate, reference and subject device have similar intended uses, conditions of use, and components.
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
### Substantial Equivalence Discussion
#### Indications for Use / Patient Population / Environment of Use:
As in comparison of Indications For Use above, we can conclude that the indications for use for the LungVision Tool and the predicate are substantially equivalent.
Discussion - The differences in proposed indications for use relate to associating the specific sheath to that bronchoscope manufacturer's own instrument of scope. The subject device is intended for use with the compatible bronchoscopes with equivalent working channel and instrument diameters.
This difference does not raise new risk or safety concerns, and the subject device can be found substantially equivalent.
#### Prescriptive:
Both the LungVision Tool and predicate are prescription devices. Discussion -There are no differences.
#### Design and Technology:
The LungVision Tool is constructed of similar materials and components to both the predicate and reference.
Discussion -There are no differences in design, technology, or principle of operation which would raise new safety or effectiveness concerns thus the subject device can be found substantially equivalent.
#### Performance and Specifications:
We performed the equivalent tests and used the predicate for reference of acceptance criteria. Discussion -There are no differences which would raise safety or effectiveness concerns compared to the predicate, thus the subject device can be found substantially equivalent.
#### Performance Testing:
#### Nonclinical / Bench
We have performed bench tests and found that the Lung Vision Tool met all requirements specifications and can be found to be substantially equivalent to the predicate.
- Bronchoscope Compatibility Test
- Simulated Use ●
- Fluoroscopy ●
- Tensile Test for Sheath and Wire ●
- Tensile Test for Wire Distal End ●
- Repeatability Use
- Sheath Insertion/Withdrawal ●
- Twisting Test ●
- Drop Test ●
- Accelerated Aging Verification Test ●
{9}------------------------------------------------
## Biocompatibility
- We assessed the patient contact type and performed the applicable ISO 10993 tests which . included:
- o Cytotoxicity
- O Sensitization
- Irritation or Intracutaneous Reactivity o
- Acute Systemic Toxicity O
- Material Mediated Pyrogenicity but no claim the device is non-pyrogenic o
The test results supported the materials as non-cytotoxic, non-sensitizers, and non-irritants.
## Sterilization and Packaging
- We have performed Single Lot Release validation which includes the following tests:
- Bioburden
- Bioburden Validation ●
- Fractional Cycle – BI show no growth
- Fractional Cycle - Sterility Test
- Sterility Validation
- Full Cycle – BI show no growth
- Full Cycle - ETO Residual
- Bacterial Endotoxin test as per LAL method ●
- Package integrity Test
- Package Stability Test ●
- . The results confirm that the single lot Ethylene Oxide sterilization process capability has achieved SAL of at least 10-6
## Animal
No animal testing was performed
## Clinical
No clinical testing was performed
## Substantial Equivalence Conclusion
Based upon the foregoing performance testing and comparison to the legally marketed predicate device for indications for use, technology, and performance we believe we have demonstrated that the Lung Vision Tool is substantially equivalent in safety and effectiveness to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.