K171291 · Fujifilm Medical Systems U.S.A, Inc. · EOQ · Jan 19, 2018 · Ear, Nose, Throat
Device Facts
Record ID
K171291
Device Name
FUJIFILM Bronchoscope Model EB-530P
Applicant
Fujifilm Medical Systems U.S.A, Inc.
Product Code
EOQ · Ear, Nose, Throat
Decision Date
Jan 19, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4680
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The FUJIFILM Bronchoscope Model EB-530P is intended for the observation, diagnosis, and endoscopic treatment of the trachea and bronchial tree.
Device Story
Flexible video bronchoscope; used for visualization, diagnosis, and treatment of trachea and bronchial tree. Input: optical images captured via distal CCD sensor; illumination provided by internal glass fiber bundles. Device transforms optical input into electronic signals for display on external monitor via video processor. Operated by clinicians in clinical settings. Features control section for distal tip angulation; channels for air/suction and endoscopic accessories (e.g., forceps). Output: real-time endoscopic video feed. Assists clinicians in identifying pathology and performing interventions. Benefits: enables minimally invasive access to airway for diagnostic and therapeutic procedures.
Clinical Evidence
Bench testing only. Evaluated electrical safety (ANSI/AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-18), biocompatibility (ISO 10993-1, -5, -10), and endoscope-specific performance (ISO 8600-1, -3, -4). Validated reprocessing instructions (cleaning, high-level disinfection, EO sterilization). Performance metrics included field of view, bending capability, suction rate, working length, forceps channel diameter, viewing direction, and light guide output. All tests met specifications.
Technological Characteristics
Flexible video bronchoscope; CCD image sensor; glass fiber bundles for illumination. Distal end/flexible portion diameter 3.8mm; working length 600mm. Connectivity: compatible with FUJIFILM video processors (EPX-4440HD). Sterilization: EO sterilization. Biocompatibility per ISO 10993; electrical safety per IEC 60601 series. Mechanical performance per ISO 8600 series.
Indications for Use
Indicated for observation, diagnosis, and endoscopic treatment of the trachea and bronchial tree in patients requiring bronchoscopic procedures.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
Predicate Devices
Fujinon/Fujifilm EB-530 Series Bronchoscopes Model EB-530S (K122535)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 19, 2018
FUJIFILM Medical Systems U.S.A., Inc. Jeffrey Wan Specialist, Regulatory Affairs 10 High Point Drive Wayne, NJ 07470
Re: K171291
Trade/Device Name: FUJIFILM Bronchoscope Model EB-530P Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible or Rigid) and Accessories Regulatory Class: Class II Product Code: EOO Dated: December 18, 2017 Received: December 19, 2017
Dear Jeffrey Wan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{1}------------------------------------------------
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Eric A. Mann -S
for Malvina Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171291
Device Name FUJIFILM Bronchoscope Model EB-530P
Indications for Use (Describe)
The FUJIFILM Bronchoscope Model EB-530P is intended for the observation, diagnosis, and endoscopic treatment of the trachea and bronchial tree.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
# FUJIFILM Medical Systems U.S.A., Inc.'s FUJIFILM Bronchoscope Model EB-530P
#### Date: January 15, 2018
#### Submitter's Information
FUJIFILM Medical Systems U.S.A., Inc., Endoscopy Division 10 High Point Drive Wayne, NJ 07470 USA FDA Establishment Registration Number: 2431293
#### Contact Person:
Jeffrey Wan Specialist, Regulatory Affairs Telephone: (973) 709-2219 Facsimile: (201) 995-2452 E-Mail: jeffrey.wan@fujifilm.com
## Identification of the Proposed Device:
| Proprietary/Trade Name: | FUJIFILM Bronchoscope Model EB-530F |
|-------------------------|-------------------------------------|
| Common Name: | Bronchoscope |
| Device Class: | Class II |
| Review Panel: | Ear, Nose, and Throat |
Classification Information:
| Classification Name | CFR Section | Product Codes |
|--------------------------------------------------|-----------------|---------------|
| Bronchoscope (flexible or rigid) and accessories | 21 CFR 874.4680 | EOQ |
#### Predicate Devices
- Fujinon/Fujifilm EB-530 Series Bronchoscopes Model EB-530S (K122535) ●
#### Intended Use / Indications for Use
The FUJIFILM Bronchoscope Model EB-530P is intended for the observation, diagnosis, and endoscopic treatment of the trachea and bronchial tree.
{4}------------------------------------------------
# Device Description
The FUJIFILM Bronchoscope Model EB-530P is comprised of three general sections: a control/ operating section, an insertion and an umbilicus. The control/operating section controls the angulation (up/down) of the distal end of the endoscope. The insertion portion contains glass fiber bundles, several channels and a complementary charged coupled device (CCD) image sensor in its distal end. The channels in the insertion portion assist in delivering air/suction as well as endoscope accessories, such as forceps. The glass fiber bundles allow light to travel through the endoscope and emit light from the tip of the insertion to illuminate the body cavity. This provides enough light to the CCD image sensor to capture an image and display it on the monitor. The umbilicus consists of electronic components needed to operate the endoscope when plugged in to the video processor and the light source.
The subject device is used in combination with FUJIFILM's video processors, light sources and peripheral devices such as monitor, printer, foot switch, and cart. All of these combinations were previously cleared in K122535.
# Technological Characteristics
A comparison of the technological characteristics between the subject and predicate devices is provided in the Table 7-1 below.
## Performance Data
Electrical safety of the subject device was evaluated using following standards: ANSI/AAMI ES 60601-1:2012, IEC 60601-1-2:2007, IEC 60601-1-6:2013, and IEC 60601-2-18:2009.
Biocompatibility of the subject devices was evaluated using the following consensus standards: ISO 10993-1:2009, ISO 10993-5:2009, and ISO 10993-10:2010.
Cleaning, high-level disinfection, and EO sterilization were performed to validate the reprocessing instructions recommended in the labeling.
Endoscope specific testing was conducted using the following consensus standards: ISO 8600-1:2015, ISO 8600-3:1997, and ISO 8600-4:2014.
Subject devices met performance specifications in the following additional testing:
- . Field of view
- . Bending capability
- Rate of suction
- Working length
- Dianeter of forceps channel .
- Viewing direction
- . LG Output
{5}------------------------------------------------
K171291
# Table 7.1: Comparison of FUJIFILM Bronchoscope Model EB-530P with its predicate device Fujinon/Fujifilm EB-530 Series Bronchoscope Model EB-530S (K122535)
| | | Predicate Device<br>Fujinon/Fujifilm EB-530 Series<br>Bronchoscopes Model EB-530S (K122535) | Proposed Device<br>FUJIFILM Bronchoscope Model EB-530P |
|----------------------------|---------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
| Device Name | | FUJIFILM Video Bronchoscope | FUJIFILM Video Bronchoscope |
| | | EB-530S | EB-530P |
| 510(k) Number | | K122535 | K171291 |
| Indications for Use (IFU) | | The device is intended for the observation,<br>diagnosis, and endoscopic treatment of<br>trachea and bronchial tree. | Same as predicate device |
| Viewing direction | | Forward/ 0 degree | Same as predicate device |
| Observation range | | 3mm - 100mm | Same as predicate device |
| Field of View | | 120 degrees | Same as predicate device |
| Distortion characteristics | | Orthogonal Projection | Same as predicate device |
| Image sensors | | CCD | Same as predicate device |
| Distal end diameter | | 4.9mm | 3.8mm |
| Flexible portion diameter | | 4.9mm | 3.8mm |
| Maximum insertion diameter | | 5.9mm | 4.2mm |
| Bending capability | Up | 180 degrees | Same as predicate device |
| | Down | 130 degrees | Same as predicate device |
| Forceps channel diameter | | 2.0mm | 1.2mm |
| Working length | | 600mm | Same as predicate device |
| Total length | | 870mm | 890mm |
| Video Processor | | EPX-2500<br>EPX-4400<br>EPX-4400HD<br>EPX-4440HD | EPX-4440HD |
| | Predicate Device<br>Fujinon/Fujifilm EB-530 Series<br>Bronchoscopes Model EB-530S (K122535) | Proposed Device<br>FUJIFILM Bronchoscope Model EB-530P | |
| Peripherals | Endoscopic Accessory (i.e. Forceps) | Endoscopic accessory i.e. forceps caliber<br>under φ1.2mm | |
| | Monitor | Same as predicate device | |
| | Printer | Same as predicate device | |
| | Electrosurgical Instruments | N/A | |
| Peripherals | Foot Switch | Same as predicate device | |
| | Cart | Same as predicate device | |
| Standard Accessories | Channel Cleaning brush<br>WB3212FW2<br>WB3503FW | Channel Cleaning Brush<br>WB2212FW2<br>WB3503FW | |
| | Valve Cleaning brush<br>(WB11002FW2) | Same as predicate device | |
| | Cleaning Adapter Kit (CA-500C) | Same as predicate device | |
| | Forceps Valve<br>(FOV-DV7) | Same as predicate device | |
| | Suction button<br>(SB-500B/D) | Same as predicate device | |
| | Ventilation Adapter<br>(AD-7) | Same as predicate device | |
| Optional Accessories | Air leak tester LT-7F | Same as predicate device | |
| | Sterile Suction button<br>(SB-602) | Same as predicate device | |
{6}------------------------------------------------
{7}------------------------------------------------
# Substantial Equivalence
FUJIFILM Bronchoscope Model EB-530P has the same intended use and indications for use, as well as similar technological characteristics and principles of operation as its predicate device Fujinon/Fujifilm EB-530 Series Bronchoscopes Model EB-530S (K122535). The modifications done to the subject device include changes in number of light quide fiber bundles, dimensions, accessories, transport and storage conditions, and materials.
These minor dimensional and material differences between the FUJIFILM Bronchoscope Model EB-530P and its predicate device Fujinon/Fujifilm EB-530 Series Bronchoscopes Model EB-530S were made for the purpose of overall product enhancement and general technological advancement, and raise no new issues of safety or effectiveness. The material changes in the subject device do not raise new concerns regarding biocompatibility as discussed in section XVII. Performance data demonstrated that the FUJIFILM Bronchoscope Model EB-530P has substantially equivalent performance to the predicate device Fujinon/Fujifilm EB-530 Series Bronchoscopes Model EB-530S (K122535). The subject device is successfully validated for the electrical safety and reprocessing instructions, supporting their safety equivalent to the predicate device.
# Conclusions
The subject device, Bronchoscope Model EB-530P is substantially equivalent to its predicate device Fujinon/Fujifilm EB-530 Series Bronchoscopes Model EB-530S (K122535), based on the same intended use/indications for use and similar technological characteristics. The differences in the dimensions and materials between the subject device and its predicate device raise no new issues of safety or effectiveness. Bench testing data demonstrated that the subject device has substantially equivalent performance to the predicate. Thus, the subject device FUJIFILM Bronchoscope Model EB-530P is substantially equivalent to its predicate device Fujinon/Fujifilm EB-530 Series Bronchoscopes Model EB-530S (K122535).
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.