Neo-Bronchoscope -Digital Video Bronchoscopy System(Rigid/Flexible)
K152466 · Prosurg, Inc. · EOQ · May 19, 2016 · Ear, Nose, Throat
Device Facts
Record ID
K152466
Device Name
Neo-Bronchoscope -Digital Video Bronchoscopy System(Rigid/Flexible)
Applicant
Prosurg, Inc.
Product Code
EOQ · Ear, Nose, Throat
Decision Date
May 19, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4680
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Prosurg's Neo-Bronchoscope™ - Digital Video Bronchoscopy System -Rigid & Flexible (which includes Digital Video Bronchoscope, Single view (Volto Sheath) is intended for use for examination, diagnosis and treatment of Airways and Tracheobronchial tree, using surgical devices and accessories. The Neo-Bronchoscope achieves its purpose by providing the user with a visual confirmation of where the tip of the endoscope is in the human anatomy. The NeoFlex – Bronchscope allows user to guide the tip of desired location with ease & safety.
Device Story
Digital video bronchoscopy system consisting of a single-use bronchoscope (rigid or flexible) with distal CMOS sensor and LED/fiber optic illumination. Device connects via USB to a user-supplied laptop/tablet running video processing software. Physician operates the bronchoscope to navigate the tracheobronchial tree; distal tip provides real-time visual feedback for guidance. System includes working channels for suction, irrigation, and delivery of micro-instruments. Output is a live video feed displayed on the computer screen, enabling visual confirmation of scope position and anatomical structures. Facilitates diagnostic and therapeutic procedures in clinical settings; improves procedural ease and safety through direct visualization.
Clinical Evidence
Bench testing only. Testing included dimensional measurements, field of view, image resolution, LED performance, tip deflection, USB compatibility, electrical safety (IEC 601-1), electromagnetic compatibility (IEC 601-1-2), thermal safety, suction capacity, packaging integrity, accelerated aging, sterilization assurance (ISO 11135-1:2007), ETO residuals (ISO 10993-7:2008), and biocompatibility (ISO 10993-1:2009).
Technological Characteristics
System comprises a digital video bronchoscope (CMOS sensor, LED/fiber optic), rigid/flexible outer sheath with suction/irrigation/instrument channels, and USB 2.0 connectivity. Operates via external Windows-based laptop/tablet. Sterilization via ETO (ISO 11135-1:2007). Biocompatibility per ISO 10993-1:2009. Electrical safety per IEC 601-1; EMC per IEC 601-1-2.
Indications for Use
Indicated for endoscopic examination, diagnosis, and treatment of the airways and tracheobronchial tree in patients requiring bronchoscopy, using surgical devices and accessories.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 19, 2016
Prosurg, Inc Ashvin Desai Manager. Regulatory Affairs 2193 Trade Zone Blvd San Jose, California 95131
Re: K152466
Trade/Device Name: Neo-bronchoscope Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: EOQ Dated: April 26, 2016 Received: April 28, 2016
Dear Ashvin Desai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
ਜਾ
#### Indications for Use
310(k) Number (if known)
K152466
Device Name
Neo- Bronchoscope ™ - Digital Video Bronchoscopy System (Rigid / Flexible)
Indications for Use (Describe)
The Prosurg's Neo-Bronchoscope™ - Digital Video Bronchoscopy System -Rigid & Flexible (which includes Digital Video Bronchoscope, Single view (Volto Sheath) is intended for use for use for use for use for use for use for use for use for use for use for use for use for use for examination, diagnosis and treatment of Airways and Tracheobronchial tree, using surgical devices and accessories.
The Neo-Bronchoscope achieves its purpose by providing the user with a visual confirmation of where the tip of the endoscope is in the human anatomy. The NeoFlex – Bronchscope allows user to guide the tip of desired location with ease & safety.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
- Food and Drug Administration
- Office of Chief Information Officer
- Paperwork Reduction Act (PRA) Staff
- PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
-13-
FORM FDA 3881 (1/14)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "ProSurg" in a stylized font. The "Pro" part of the word is in pink, while the "Surg" part is in blue with horizontal lines running through it, giving it a textured appearance. The overall design is simple and clean, with a focus on the text itself.
# 510K Summary # K152466 (Traditional 510K)
## Neo- Bronchoscope™ -Digital Video Bronchoscopy System (Rigid / Flexible)
| Submitter : | Prosurg, Inc<br>2193 Trade Zone Blvd<br>San Jose CA 95131<br>Tel: 408 945 4044 Fax: 408 945 1390 | | |
|------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|--------------------------|
| Contact person: | Ashvin Desai<br>Manager, Regulatory Affairs<br>Prosurg, Inc<br>2193 Trade Zone Blvd San Jose CA 95131<br>Tel: 408 945 4044<br>Fax : 408 945 1390 | | |
| Date Summary prepared : | March 30, 2016 | | |
| Device Trade Name: | Neo- BronchoscopeTM - Digital Video<br>Bronchoscopy System (Rigid / Flexible) | | |
| Device Common Name: | Bronchoscope (Flexible/ Rigid) | | |
| Device Classification: | 21CFR 874.4680 (Class II) | | |
| Product Code: | Endoscope & Accessories - EOQ | | |
| Legally Marketed Devices<br>to which the substantially<br>Equivalency is claimed : | Manufacturer:<br>Ambu A/S | Trade Name:<br>Ambu aScope<br>Ambu Monitor<br>(21 CFR 874.4680) | 510K #<br>K130845<br>EOQ |
### Device Description :
The Single Use Digital Video Bronchoscope design incorporates a CMOS imaging sensor and built-in LEDs (Light Emitting Diodes) or fiber optic mounted at the distal end, connecting wires along the length of the hollow tubular structure and a USB Connector at the proximal end. The distal end of the Bronchoscope containing CMOS sensor and LEDs / Fiber optic is protected by sealed, clear optical window. The NeoFlex™ - Bronchoscope , Flexible Video Bronchoscope
{4}------------------------------------------------
device with an Articulating Tip is designed with built-in fluid suction / irrigation and working channel for Micro Instrument delivery.
## Indications For Use:
The Prosurg's, Neo-Bronchoscope™ -Digital Video Bronchoscopy System -Rigid & Flexible (which includes Digital Video Bronchoscope, Single use Diagnostic / Operating Outer Sheath) is intended for use for endoscopic examination, diagnosis and treatment of Airways and Tracheobronchial tree, using surgical devices & accessories.
The, Neo- Bronchoscope achieves its purpose by providing the user with a visual confirmation of where the tip of the endoscope is in the human anatomy. The Neoflex- Bronchoscope allows user to guide the Bronchoscope in the desired location with ease and safety.
## Technological Characteristics:
The Neo- Bronchoscope ™ Video Bronchoscopy System consists of four main components:
- (a) Digital Video Bronchoscope (Rigid / Flexible) with CMOS Sensor & Built-in LEDs / Fiber optic.
- (b) Single Use. Rigid Diagnostic / Operating Outer Sheath with Suction / Irrigation / Instrument channel and a separate, dedicated channel for video Bronchoscope insertion. Or a flexible Bronchoscope with suction / Irrigation & Instrument channel with Articulating tip deflection.
- (c ) )Laptop /Tablet computer with windows Operating System, Video Module and Video processing Software. (Commercially Available, Not supplied by Prosurg, Inc)
- (d) USB 2.0 / HD / S-Video Connecting Cables. (Commercially Available, Not supplied by Prosurg, Inc)
## Substantial Equivalence:
The Neo -Bronchoscope™ Digital Video Bronchoscopy System -(Rigid & Flexible) and Predicate device Ambu ascope and Ambu Monitor (Ref: K130845 ) are similar in design, Components, including CMOS Imaging Sensor, LED, Outer tube Handle materials & USB 2.0 Connectors. The Manufacturing & Assembly process, Product Packaging, Sterilization method & Recommended for Single use are also identical.
The proposed Indications for Use for Neo-Bronchoscope is based on Predicate device Ambu aScope & aScope Monitor (K130845). The Neo - Bronchoscope - Digital Video System is substantially equivalent to Ambu aScope. For details, please refer to Substantial Equivalence Comparison chart summary outlining Physical, functional characteristics and Indications for use.
## Performance Testing : (Bench Testing)
Prosurg has conducted performance testing (Bench Evaluation- As per FDA Guidance Document) of Neo- Bronchoscope ™ device to demonstrate Safety & effectiveness and its intended use. The
{5}------------------------------------------------
Following Testing was completed to demonstrate safety & effectiveness of the proposed device for its intended use.
- Dimensional Measurements (Outer Diameter, Working Length) .
- Field of View (in air)
- Image Resolution (Number of Pixels) ●
- Direction of View
- LED Brightness, Voltage & Current measurement.
- Tip Deflection Control Mechanism (flex) & Angle of Tip deflection
- . USB Connection compatibility with Laptop / Tablet / Monitor
- Electrical Safety Test (IEC 601-1)
- Electromagnetic Compatibility (IEC 601-1-2)
- Thermal (Heat) Temperature at the Distal Tip (LED)
- Maximum Safe amount of Suction (Vacuum) testing
- Product Labeling & IFU Requirements & Accuracy
- . Packaging Integrity & Sterility Assurance Test
- Accelerated Aging Test (one Year)
- . Sterilization Assurance Level (ISO 11135-1:2007)
- . ETO Residual Levels (ISO 10993-7:2008)
- . Biocompatibility (ISO 10993-1:2009)
### Conclusion:
The results of the performance testing demonstrate that proposed Neo -Bronchoscope™ Digital Video Bronchoscopy System -(Rigid & Flexible), performs as well as predicate device for its intended use and is considered safe and effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.