K150280 · Sanovas, Inc. · EOQ · Jun 11, 2015 · Ear, Nose, Throat
Device Facts
Record ID
K150280
Device Name
PulmoVia Working Channel
Applicant
Sanovas, Inc.
Product Code
EOQ · Ear, Nose, Throat
Decision Date
Jun 11, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4680
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PulmoVia Working Channel is intended to be used as an access channel through which a bronchoscope and other endoscopic tools may be introduced to treat targeted tissue of the airways. The PulmoVia Working Channel can also provide means for bronchoalveolar lavage. The PulmoVia Working Channel is indicated for adult patients.
Device Story
PulmoVia Working Channel is a single-use, sterile, hollow catheter for transtracheal access to lung bronchi. Device features angled, flexible distal tip for anatomical navigation; radiopaque marker for localization via fluoroscopy/CT; printed depth markings. Includes Tuohy-Borst adaptor for compatibility with third-party ultrathin bronchoscopes. Used in clinical settings by physicians to facilitate bronchoalveolar lavage or provide a conduit for endoscopic tools to reach targeted airway tissue. Output is physical access to bronchial tree; enables therapeutic or diagnostic procedures. Benefits include improved access to targeted lung areas and simplified tool introduction.
Clinical Evidence
No clinical studies performed. Usability assessments conducted in a simulated use environment validated that users can operate the device safely and effectively. Non-clinical performance testing (sterility, shelf-life, packaging, biocompatibility, and mechanical bench testing) serves as the primary basis for substantial equivalence.
Technological Characteristics
Single-use, sterile (EO) catheter. Materials: Tecothane and Pebax extrusion, Platinum-Iridium marker band, Polycarbonate hub/Tuohy-Borst connector with silicone gasket. Features angled, flexible distal tip. Dimensions: 485 mm working length, 5.25 mm ID, 6.0 mm OD. Biocompatibility per ISO 10993-1:2009. Mechanical testing per internal specs.
Indications for Use
Indicated for adult patients requiring an access channel for bronchoscopy, introduction of endoscopic tools to treat airway tissue, or bronchoalveolar lavage.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
Predicate Devices
LungPoint™ Tools (LungPoint Sheath and LungPoint Dilation Balloon) (K131234)
KimVent BAL Cath Bronchial Aspirate Sampling Catheter (K112562)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image contains the logo for the U.S. Department of Health and Human Services. The logo consists of two main elements: a circular seal on the left and a stylized symbol on the right. The circular seal contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. The symbol on the right is a graphic representation of a human figure, with three profiles layered on top of each other to create a sense of depth and dimension.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 11, 2015
Sanovas, Inc. c/o Cynthia Nolte, Ph.D., RAC Senior Director of Regulatory Affairs ICON Clinical Research LLC 62 Forest Street, Suite 300 Marlborough, MA 01752
Re: K150280
Trade/Device Name: PulmoVia Working Channel Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible or Rigid) Regulatory Class: Class II Product Code: EOO Dated: April 22, 2015 Received: April 22, 2015
Dear Dr. Nolte:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K150280
Device Name
PulmoVia Working Channel
## Indications for Use (Describe)
The Pulmo Via Working Channel is intended to be used as an access channel through which a bronchoscope and other endoscopic tools may be introduced to treat targeted tissue of the airways. The Pulmo Via Working Chamel can also provide means for bronchoalveolar lavage. The PulmoVia Working Channel is indicated for adult patients.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Publishing Services (301)-443-6740 B
{3}------------------------------------------------
#### ട്. 510(K) SUMMARY
# Sanovas, Inc. PulmoVia Working Channel
(per 21CFR 807.92 and http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/default.htm)
#### 1. SUBMITTER/510(K) HOLDER
| Sanovas, Inc. |
|-----------------------------------------------------|
| 2597 Kerner Blvd |
| San Rafael, CA 94901 |
| Phone: 415-729-9391 |
| Contact: Roy Morgan, VP of Research and Development |
| Contact's Phone: 415-729-9391 |
| Contact's Email: regulatory@sanovas.com |
| Date Prepared: June 10, 2015 |
#### 2. DEVICE NAME
| Trade Name: | PulmoVia Working Channel |
|----------------------------|--------------------------------------------------|
| Classification Name: | Bronchoscope (flexible or rigid) and accessories |
| Classification Panel: | Ear, Nose, and Throat Devices Panel |
| Classification Regulation: | 21 CFR 874.4680 |
| Device Classification: | Class II |
| Product Code: | EOQ |
#### 3. PREDICATE DEVICES
- LungPoint™ Tools (LungPoint Sheath and LungPoint Dilation Balloon), K131234 ●
- KimVent BAL Cath Bronchial Aspirate Sampling Catheter, K112562 ●
- Olympus Guide Sheath, K060243 ●
#### DEVICE DESCRIPTION 4.
The PulmoVia Working Channel is a single-use disposable, terminally-sterilized device for transtracheal access to bronchi of the lungs, and consists of a hollow tube with an angled flexible tip to assist in anatomical navigation. A radiopaque marker is located 4 mm from the distal tip to aid in localization under fluoroscopy/CT. Printed markings are positioned at 370 mm from the distal tip, and every 50 mm proximally from that point. The distal tip of the device has a pliable leading edge.
The device also contains a disposable Tuohy-Borst adaptor to allow use of third-party devices such as ultrathin bronchoscopes. The full assembly will be packaged in a polyethylene and Tyvek pouch, sterilized via ethylene oxide (EO).
{4}------------------------------------------------
#### 5. INDICATION FOR USE/INTENDED USE
The PulmoVia Working Channel is intended to be used as an access channel through which a bronchoscope and other endoscopic tools may be introduced to treat targeted tissue of the airways. The PulmoVia Working Channel can also provide means for bronchoalyeolar lavage. The PulmoVia Working Channel is indicated for adult patients.
### SUMMARY OF TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE PREDICATE 6. DEVICES
Both the proposed PulmoVia Working Channel and the predicate Lung Point Sheath are catheters designed for use with a bronchoscope to reach a targeted area of the bronchial tree. The working length of the PulmoVia Working Channel (485 mm) is less than that of the predicate Lung Point Sheath and reference devices. The inner diameter of the catheter (5.25 mm) is greater than that of the Lung Point and Olympus Guide Sheath. The inner diameter was established in order to be compatible with bronchoscopes with an OD of 4.0 mm.
The predicate Lung Point Sheath is supplied with a stylet. Neither the proposed PulmoVia Working Channel nor the reference device Olympus Guide Sheath are supplied with a stylet.
The proposed PulmoVia Working Channel has an angled, flexible tip to aid in navigation of the PulmoVia Working Channel within the lung. The BAL Cath tip is also designed with a flexible tip.
The materials used for the proposed device are provided in the side-by-side comparison table. Like the predicate devices, the PulmoVia Working Channel sheath and connector are composed of polymers. As will be discussed in Section 7, the PulmoVia Working Channel meets the biocompatibility requirements of ISO 10993-1:2009 for the nature and duration of contact.
### SUMMARY OF NON-CLINICAL PERFORMANCE TESTING AS BASIS FOR SUBSTANTIAL 7. EQUIVALENCE
Non-clinical performance testing consisted of sterility, shelf-life, packaging validation, transport, biocompatibility and mechanical bench testing which demonstrated that the device met all mechanical design and safety requirements.
The biocompatibility testing included:
- . Cytotoxicity (ISO 10993-5:2009, Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity)
{5}------------------------------------------------
- Irritation/ Intracutaneous (ISO 10993-10:2010, Biological Evaluation Of Medical ● Devices - Part 10: Tests For Irritation And Skin Sensitization)
- Sensitization (ISO 10993-10:2010, Biological Evaluation Of Medical Devices Part . 10: Tests For Irritation And Skin Sensitization)
- Acute systemic toxicity (ISO 10993-11:2006, Biological Evaluation Of Medical . Devices - Part 11: Tests For Systemic Toxicity)
- . Pyrogenicity (ISO 10993-11:2006 Biological Evaluation of Medical Devices - Part 11: Tests for Systemic Toxicity)
The mechanical testing included: (Side by side testing indicated by an asterisk *)
- . Pull test*
- Torque test*
- Kink test* ●
- . Bend test*
- . Leak test*
- Vacuum test*
- Operating temperature/humidity test* ●
- Radiopaque test ●
- Gravity fed hydrostatic flow test* ●
#### SUMMARY OF CLINICAL TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE 8.
No clinical studies were performed using the PulmoVia Working Channel. Usability assessments were conducted in a simulated use environment to optimize the safety of the device design and validate the usability of the device. The results demonstrated that users can operate the PulmoVia Working Channel safely and effectively.
#### SUMMARY OF OTHER INFORMATION 9.
No other information is available.
## 10. CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL TESTS
Non-clinical testing confirms that the PulmoVia Working Channel is sterile, has a packaging system adequate to maintain sterility and transportation hazards, and a validated shelf life. The PulmoVia Working Channel is biocompatible and side by side testing confirms that the proposed device has equivalent physical and mechanical characteristics to the identified predicates. Human factors testing confirms that users can operate the PulmoVia Working Channel safely and effectively.
{6}------------------------------------------------
The similarities in intended use, operational characteristics, functional technological characteristics and conclusions drawn from nonclinical testing between the proposed PulmoVia Working Channel and the predicate LungPoint Sheath and KimVent BAL Cath lead to a conclusion of substantial equivalence between the proposed and predicate devices. A side-by-side comparison of the predicate devices and the proposed device is provided in the table at the end of this section.
| | PulmoVia™ | LungPoint™ Tools | Olympus Guide<br>Sheath | BAL Cath* |
|--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | Working Channel<br>(Proposed) | K131234<br>(Predicate) | K060243<br>(Reference) | K112562<br>(Reference) |
| Indication for Use | The PulmoVia<br>Working Channel is<br>intended to be used<br>as an access channel<br>through which a<br>bronchoscope and<br>other endoscopic<br>tools may be<br>introduced to treat<br>targeted tissue of<br>the airways. The<br>PulmoVia Working<br>Channel can also<br>provide means for<br>bronchoalveolar<br>lavage. The<br>PulmoVia Working<br>Channel is indicated<br>for adult patients. | The LungPoint<br>Sheath is intended<br>to be used as a<br>working channel<br>through which<br>endoscopic tools<br>may be introduced<br>to targeted tissue.<br>Not for pediatric<br>use. | The instrument has<br>been designed to be<br>used with Olympus<br>bronchoscopes,<br>endo-therapy<br>accessories or<br>ultrasound probe to<br>the targeted area<br>within the<br>respiratory organs. | The BAL Cath is<br>used in the<br>diagnosis of diffuse<br>lung disease by<br>allowing collection<br>of bronchoalveolar<br>lavage (BAL)<br>specimens from<br>deep within the<br>lung. The use of a<br>bronchoscope is not<br>necessary. This<br>catheter is used in<br>adult intubated<br>patients. |
| Single Use | Single use | Single use | Single use | Single use |
| Sterile | Yes (Ethylene<br>oxide) | Yes (E-beam) | Yes | Yes |
| Catheter Working<br>Length | 485 mm | 900 mm | 1050 mm | Not known |
| Catheter Internal<br>Diameter (ID) | 5.25 mm | 2.0 mm | 2.1 mm | 4 mm |
| Maximum Catheter<br>Outer Diameter<br>(OD) | 6.0 mm | 2.65 mm | 2.7 mm | 5.2 mm |
| Bronchoscope<br>Compatibility | Max OD: 4.0 mm<br>Min Working<br>Length: 600 mm | Not known | Not known | Not used with a<br>bronchoscope |
| Catheter Length | 495 mm | 975 mm | 900 mm | Not known |
| Tuohy-Borst<br>Adaptor | Yes | No | Not known | Not known |
| Stylet | No | Yes | No | No |
| Distal Tip | Angled, flexible | Not known | Not pliable | Flexible, directional |
| Radiopaque | Yes | Yes | Yes | Yes |
| Feature<br>Markers | PulmoViaTM<br>Working Channel<br>(Proposed) | LungPointTM Tools<br>K131234<br>(Predicate) | Olympus Guide<br>Sheath<br>K060243<br>(Reference) | BAL Cath*<br>K112562<br>(Reference) |
| Materials | Extrusion:<br>Tecothane and<br>Pebax<br>Marker Band:<br>Platinum-Iridium<br>Hub: Polycarbonate<br>Tuohy Borst<br>Connector:<br>Polycarbonate with<br>silicone gasket | Extrusion: Not<br>known^<br>Marker band: Not<br>known^ | Extrusion:<br>Polymer^<br>Marker band:<br>Radio-opaque metal<br>Accessory Port:<br>Polymer^<br>Tri-Channel<br>Connector:<br>Polymer^ | Extrusion:<br>Polymer^<br>Marker band: Not<br>known^<br>Collar: Polymer^ |
## Side-by-Side Comparison of the PulmoVia™ Working Channel with the LungPoint™ Tools and Reference Devices
{7}------------------------------------------------
*BAL Cath: Kimberly-Clark KimVent BAL Cath Bronchial Aspirate Sampling Catheter ^Material composition is not known
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.