K121035 · Fujifilm Medical System U.S.A., Inc. · EOQ · Jun 4, 2012 · Ear, Nose, Throat
Device Facts
Record ID
K121035
Device Name
FUJINON/FUJIFILM ULTRASONIC ENDOSCOPE EB-530US
Applicant
Fujifilm Medical System U.S.A., Inc.
Product Code
EOQ · Ear, Nose, Throat
Decision Date
Jun 4, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4680
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Fujinon/Fujifilm Ultrasonic Endoscope EB-530US is intended for the observation, diagnosis and endoscopic treatment of the trachea, bronchial tree and surrounding organs using ultrasonic images. It is used with Fujinon/Fujifilm ultrasonic processor, video processor, light source, other peripheral equipment and endoscopic accessories. It is not intended for use on children and infants.
Device Story
The EB-530US is an ultrasonic bronchoscope used for visualization and ultrasound-guided procedures in the trachea and bronchial tree. It functions as an endoscope integrated with an ultrasonic transducer. The device emits ultrasound waves and scans reflected signals to generate ultrasonic images when connected to a compatible Fujinon/Fujifilm ultrasonic processor, video processor, and light source. A single-patient-use balloon, inflated with sterile water, is used to eliminate air between the transducer and target tissue to improve signal transmission. The device supports ultrasound-guided needle aspiration. It is operated by clinicians in a clinical setting. The output consists of real-time endoscopic and ultrasonic images displayed on a monitor, which assist the physician in diagnosing and treating pulmonary conditions. The device benefits patients by enabling minimally invasive visualization and targeted sampling of tissues within the bronchial tree and surrounding structures.
Clinical Evidence
No clinical testing was conducted. Substantial equivalence is supported by bench testing, including compliance with IEC 60601-1 (general safety), IEC 60601-2-18 (endoscopic equipment), IEC 60601-2-37 (ultrasonic diagnostic equipment), and ISO 10993-1 (biocompatibility). Reprocessing instructions were validated.
Technological Characteristics
Flexible bronchoscope with integrated electrical curved linear array ultrasonic transducer. Operating frequencies: 5, 7.5, 10, 12 MHz. Viewing direction: 10-degree forward oblique. Field of view: 120 degrees. Distal end diameter: 6.7 mm. Forceps channel: 2.0 mm. Connectivity: Compatible with Fujinon/Fujifilm ultrasonic and video processors. Biocompatible patient contact materials. Sterilization: Reprocessing instructions provided.
Indications for Use
Indicated for observation, diagnosis, and endoscopic treatment of trachea, bronchial tree, and surrounding organs in adult patients using ultrasonic imaging. Contraindicated for use on children and infants.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
K121035
# FUJIFILM
JUN - 4 2012
## 510(k) Summary
## Date: June 4, 2012
## Submitter's Information:
FUJIFILM Medical Systems U.S.A., Inc. 10 High Point Drive Wavne, NJ 07470 USA FDA Establishment Registration Number: 2431293
### Contact Person:
| Name: | Gina Walljasper |
|------------|---------------------------------------------|
| Title: | Director, Quality and Regulatory Compliance |
| Telephone: | (973) 633-5600 Ext. 484 |
| Facsimile: | (973) 633-8818 |
| E-Mail: | gwalljasper@fujifilm.com |
## Identification of the Proposed Device:
Proprietary/Trade Name: Common Name: Device Class: Review Panel:
Fuiinon/Fuiifilm Ultrasonic Endoscope EB-530US Ultrasonic Bronchoscope Class 2 Ear Nose & Throat
Classification Information:
| Classification Name | CFR Section | Product Codes |
|--------------------------------------------------|-----------------|---------------|
| Bronchoscope (flexible or rigid) and accessories | 21 CFR 874.4680 | EOQ |
| Diagnostic Ultrasonic Transducer | 21 CFR 892.1570 | ITX |
#### INDICATIONS FOR USE l,
The Fujinon/Fujifilm Ultrasonic Endoscope EB-530US is intended for the observation, diagnosis and endoscopic treatment of the trachea, bronchial tree and surrounding organs using ultrasonic images. It is used with Fujinon/Fujifilm ultrasonic processor, video processor, light source, other peripheral equipment and endoscopic accessories. It is not intended for use on children and infants.
### DEVICE DESCRIPTION II.
Fujinon/Fujifilm Ultrasonic Endoscope EB-530US is intended for the observation, diagnosis and endoscopic treatment of the trachea, bronchial tree and surrounding organs using ultrasonic images. EB-530US is used in combination with the Fujinon/Fujifilm's ultrasonic processor, video endoscope processor, light source, monitor, cart, foot switch, endoscope accessories and other peripheral devices. When used with a Fujinon/Fujifilm's ultrasonic processor, EB-530US emits ultrasound wave and scans the reflected signals to orovide ultrasonic images. For ultrasound procedure, EB-530US can be used with a
{1}------------------------------------------------
K12/035
# FUJKFILM
single-patient-use balloon inflated with sterile water for an ultrasonic endoscope eliminates air between the ultrasonic transducer and target tissue so that the ultrasound wave can travel with little interference. Additionally EB-530US supports ultrasound guided needle aspiration.
#### SUMMARY OF STUDIES lll.
Fujinon/Fujifilm Ultrasonic Endoscope EB-530US conforms to the following internal and international IEC testing requirements in addition to the applicable quality system regulations:
| IEC 60601-1 | Medical electrical equipment - Part 1: General requirements for safety |
|----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC60601-1-1 | Medical electrical equipment - Part 1-1: General requirements for safety - Collateral standard: Safety requirements for medical electrical systems |
| IEC60601-1-2 | Medical electrical equipment - Part 1-2: General Requirements for Safety - Collateral standard: Electromagnetic Compatibility - Requirements and tests |
| IEC60601-2-18 | Medical electrical equipment - Part 2-18: Particular requirements for the safety of endoscopic equipment |
| IEC 60601-2-37 | Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment |
| ISO10993-1 | Biological evaluation of medical devices |
The balloon, B20BU for the Fujinon/Fujifilm Ultrasonic Endoscope EB-530US conform to the applicable internal and international IEC testing requirements for Sterility.
The reprocessing instructions were validated.
No clinical testing was conducted.
#### SUBSTANTIAL EQUIVALENCE IV.
Fujinon/Fujifilm Ultrasonic Endoscope EB-530US is substantially equivalent to the following device:
| Legally Marketed Device(s) | 510(k) # |
|----------------------------------------------|----------|
| EVIS EXERA Ultrasonic Bronchofibervideoscope | K042140 |
| OLYMPUS BF type UC160F-OL8 | |
{2}------------------------------------------------
# FUJIFILM
Comparison is outlined in the table below.
| | Legally Marketed Device<br>K042140 | Proposed Device<br>Model EB-530US |
|--------------------------------|------------------------------------|------------------------------------------------------|
| Viewing direction | 35 degree forward oblique | 10 degree forward oblique |
| Observation range | 2-50 mm | 3-100 mm |
| Field of view | 80 degree | 120 degree |
| Distal end diameter | 6.9 mm | 6.7 mm |
| Flexible portion diameter | 6.2 mm | 6.3 mm |
| Bending capability Up | 120 degree | 130 degree |
| Bending capability Down | 90 degree | 90 degree |
| Forceps channel diameter | 2.0 mm | 2.0 mm |
| Working length | 600 mm | 610 mm |
| Total length | 890 mm | 880 mm |
| Scanning method | Electrical curved linear array | Electrical curved linear array |
| Scanning range | 50 degree | 60 degree (with SU-7000)<br>65 degree (with SU-8000) |
| Ultrasonic operating frequency | 7.5 MHz | 5 MHz, 7.5 MHz, 10 MHz,<br>12 MHz |
All patient contact materials used in the proposed EB-530US remain the same as the previously-cleared Fujinon/Fujifilm endoscopes.
### V. CONCLUSION
Fujinon/Fujifilm Ultrasonic Endoscope EB-530US is substantially equivalent to the legally marketed device and conforms to applicable medical device safety and performance standards.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three wing-like strokes. The logo is surrounded by a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in capital letters.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
FUJIFILM Medical Systems U.S.A., Inc. c/o Mr. Mark Job Regulatory Technology Services, LLC 1394 25th Street NW Buffalo, MN 55313
Re: K121035
Trade/Device Name: Fujinon/Fujifilm Ultrasonic Endoscope EB-530US Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible or Rigid) and Accessories Regulatory Class: Class II Product Code: EOQ, ITX Dated: April 4, 2012 Received: April 5, 2012
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
JUN - 4 2012
{4}------------------------------------------------
Page 2 - Mr. Mark Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K121035
## Indications for Use
510(k) Number (if known): __
Device Name: Fujinon/Fujifilm Ultrasonic Endoscope EB-530US To be used with Fujinon/Fujifilm Ultrasonic Processor (SU-7000/SU-8000)]
Indications for Use:
The Fujinon/Fujifilm Ultrasonic Endoscope EB-530US is intended for the observation, . The Tighton't ajimm Oneonent of the trachea, bronchial tree and surrounding organs using ultrasonic images. It is used with Fujinon/Fujifilm ultrasonic processor, video processor, light source, other peripheral equipment and endoscopic accessories. It is not intended for use on children and infants.
× Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
. Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
121035
510(k) Number
**Prescription Use**
(Per 21 CFR 801.109)
{6}------------------------------------------------
i2 10 35
## Diagnostic Ultrasound Indications For Use
510(k) Number (If Known):
Fujinon/Fujifilm Ultrasonic Processor (SU-7000/SU-8000) System Name: Ultrasonic Endoscope EB-530US Transducer:
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|----------------------|------------------------------------------------|-------------------|---|-----|-----|------------------|-----------|-------|
| General | Specific . | B | M | PWD | CWD | Color<br>Doppler | Combined¹ | Other |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ<br>(Thyroid, Breast, Testes, etc.) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| General | Adult Cephalic | | | | | | | |
| Application | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Tran-esoph. (non-Card.) | N | N | N | | N | N¹ | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | N¹ | |
| | Other (Specify) | N | N | N | | N | N¹ | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Tran-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral | Peripheral vessel | | | | | | | |
| Vascular | Other (Specify) | | | | | | | |
vessel
Combined modes includes B+M, B+CD, B+PWD, B+CD+PWD modes
Other (Specify): Trachea, bronchial tree and surrounding organs
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Ophthalmic, Neurological and Ear,
Nose and Throat Devices
510(k) Number
Prescription Use (Per 21 CFR 801.109)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.