← Product Code [EOQ](/submissions/EN/subpart-e%E2%80%94surgical-devices/EOQ) · K071473

# MODIFICATION TO SUPERDIMENSION/BRONCHUS PREMIUM 2 (K071473)

_Superdimension, Ltd. · EOQ · Jul 12, 2007 · Ear, Nose, Throat · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/EOQ/K071473

## Device Facts

- **Applicant:** Superdimension, Ltd.
- **Product Code:** [EOQ](/submissions/EN/subpart-e%E2%80%94surgical-devices/EOQ.md)
- **Decision Date:** Jul 12, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 874.4680
- **Device Class:** Class 2
- **Review Panel:** Ear, Nose, Throat

## Indications for Use

The superDimension/Bronchus Premium 2 is indicated for displaying images of the tracheobronchial tree to aid the physician in guiding endoscopic tools in the pulmonary tract. It does not make a diagnosis and is not an endoscopic tool. It is not for pediatric use.

## Device Story

The superDimension/Bronchus Premium 2 is an image-guided navigation system for the pulmonary tract. It takes CT scan data as input to generate a visual representation of the tracheobronchial tree. The device transforms these inputs to provide a path for guiding endoscopic tools toward targets in or adjacent to the bronchial tree. It is used in a clinical setting by physicians to assist in navigation; it does not perform diagnoses and is not an endoscopic tool itself. The output is displayed to the physician to facilitate tool guidance, potentially improving the accuracy of reaching pulmonary targets.

## Clinical Evidence

No clinical data. Bench testing and risk analysis were performed to validate design changes.

## Technological Characteristics

Computed tomography x-ray system and bronchoscope accessory. System utilizes CT scan data for image-guided navigation of the tracheobronchial tree. Includes disposable components. Software-based image display.

## Regulatory Identification

A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.

## Predicate Devices

- superDimension/Bronchus Premium 2 ([K062315](/device/K062315.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K071473

# 510(k) Summary

JUL 12 2007

superDimension Ltd. Special 510(k) Changes to superDimension/Bronchus Premium 2

#### Date Prepared:

05/25/2007

#### 510(k) Applicant:

superDimension Ltd. 8 Hamenofim St., P.O. Box 2045 Herzliya 46120 Israel Ph: +972-(0)9-971-3700 Fax: +972-(0)9-971-3701

## 510(k) Application Correspondent:

Clay Anselmo President and COO Reglera 555 Zang St., Suite 100 Lakewood, CO 80228 303-223-4303 Ph: 303-907-2955 Cell: 303-832-6700 Fax: Email: anselmoc@reglera.com

### Name of Device:

superDimension/Bronchus Premium 2 Trade name: Common name: Bronchoscope Computed tomography x-ray system Classification name: 21 CFR 892.1750 Product code JAK

#### Equivalent Legally-Marketed Device:

superDimension/Bronchus Premium 2, K062315

{1}------------------------------------------------

### Description:

The superDimension/Bronchus Premium 2 is a device that guides a bronchoscope and bronchial tool to a target in or adjacent to the bronchial tree on a path indicated by CT scan, and visualizes the target and the interior of the tree.

#### Intended Use:

The superDimension/Bronchus Premium 2 is indicated for displaying images of the tracheobronchial tree to aid the physician in guiding endoscopic tools in the pulmonary tract. It does not make a diagnosis and is not an endoscopic tool. It is not for pediatric use.

#### Summary of Characteristics Compared to Predicate Device:

Changes are being made to the superDimension/Bronchus Premium 2 to include modifications to some of its disposable components and changes to portions of its labeling.

The intended use and indications for use of the modified device, as described in its labeling, are the same as the intended use and indications for use of the 510(k) Applicant's unmodified predicate device.

The modified device has the same technological characteristics as the unmodified predicate device.

#### Performance Data:

The changes being made to the superDimension/Bronchus Premium 2 were subjected to superDimension's design control process. A risk analysis was performed to analyze the hazards associated with the changes. Appropriate verification and validation tests were performed to assure that the design output met the design input requirements and that the modified device continues to meet its user needs and intended uses.

## Clinical Data:

Clinical tests were not required to validate the changes made to the superDimension/Bronchus Premium 2.

#### Conclusion:

The superDimension/Bronchus Premium 2 is safe and effective for its intended use.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird with three wing-like shapes and a stylized body.

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

SuperDimension Ltd. % Mr. Clay Anselmo President and CEO Reglera 555 Zang Street, Suite 100 LAKEWOOD CO 80228

JUL 1 2007

Re: K071473

Trade/Device Name: superDimension/Bronchus Premium 2 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulation Numer: 21 CFR 874.4680 Regulation Name: Bronchoscope (flexible or rigid) and accessories Regulatory Class: II Product Code: JAK and EOQ Dated: May 25, 2007 Received: May 29, 2007

## Dear Mr. Anselmo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/2/Picture/10 description: The image shows a circular logo with the text "1906-2006" at the top. Below that is the acronym "FDA" in a stylized font. The word "Centennial" is written below the acronym. There are three stars below the word "Centennial".

closing and Premoting Public Hoatth

{3}------------------------------------------------

Page 2 -

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and fisting (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology)                      | 240-276-0120 |
| Other          |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150

or at its Internet address http://www.fda.gov/cdrlv/industry/support/index.html.

Sincerely vours.

Nancy C. Hogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known): ___ Ko714 73

Device Name: ___superDimension/Bronchus Premium 2

Indications for Use:

Indicated for displaying images of the tracheobronchial tree to aid the physician in guiding endoscopic tools in the pulmonary tract. It does not make a diagnosis and is not an endoscopic tool.

Not for pediatric use.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Hulut Reiner

(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number __

Page _ of _

000006

---

**Source:** [https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/EOQ/K071473](https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/EOQ/K071473)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
