SHARPSITE AC

K965233 · Xomed, Inc. · EOB · Apr 4, 1997 · Ear, Nose, Throat

Device Facts

Record IDK965233
Device NameSHARPSITE AC
ApplicantXomed, Inc.
Product CodeEOB · Ear, Nose, Throat
Decision DateApr 4, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.4760
Device ClassClass 2

Indications for Use

The Xomed SHARPSITE Ac rigid rod lens endoscopes are intended for use in surgical imaging in otolaryngology and Head and Neck procedures, including rhinology, and endoscopic plastic and reconstructive surgery.

Device Story

SHARPSITE Ac: series of rigid rod lens endoscopes; diameters 2.7mm, 4.0mm, 5.0mm; tip angles 0, 30, 70 degrees; lengths 175mm-188mm. Features removable light port adapter for compatibility with various light cables. Used by surgeons in otolaryngology, head/neck, rhinology, and plastic/reconstructive surgery settings. Provides optical visualization of internal anatomy to assist surgical procedures.

Clinical Evidence

Bench testing only.

Technological Characteristics

Rigid rod lens endoscope; diameters 2.7mm, 4.0mm, 5.0mm; tip angles 0, 30, 70 degrees; lengths 175mm-188mm; removable light port adapter.

Indications for Use

Indicated for surgical imaging in otolaryngology, Head and Neck procedures, rhinology, and endoscopic plastic and reconstructive surgery.

Regulatory Classification

Identification

A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.

Predicate Devices

Submission Summary (Full Text)

{1} APR - 4 1997 K965233 revised 4/9 # 510(k) Summary 1.0 Date Prepared December 23, 1996 2.0 Submitter (Contact) David Timlin Xomed Surgical Products Jacksonville, FL (904) 279-7532 3.0 Device Name Proprietary Name: SHARPSITE Ac Common Name(s): Rigid Rod Lens Endoscopes Classification Name: Endoscope, rigid Nasopharyngoscope (flexible or rigid) 5.0 Device Classification Endoscope, rigid Procode 78GCM Class II; 21 CFR 876.1500 Tier 2 Nasopharyngoscope Procode 77EOB Class II; 21 CFR 874.4760 Tier 2 (flexible or rigid) 6.0 Device Description The SHARPSITE Ac endoscopes are a series of small diameter, rigid rod lens endoscopes; available in diameters of 2.7mm, 4.0mm and 5.0mm with tip angles of 0, 30 and 70 degrees. The length of the scopes range from 175mm to 188 mm. Each scope is fitted with a removable light port adapter for use with light cables from different manufacturers. 7.0 Intended Use The Xomed SHARPSITE Ac rigid rod lens endoscopes are intended for use in surgical imaging in otolaryngology and Head and Neck procedures, including rhinology, and endoscopic plastic and reconstructive surgery. 8.0 Substantial Equivalence Substantial equivalency is based on the fact that the SHARPSITE Ac endoscopes are equivalent to current Xomed endoscopes in their intended use and materials. They are likewise equivalent in design, intended use and performance claims to sinuscopes currently marketed by Smith & Nephew (S&N) ENT.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...