Video Rhinolaryngoscope System (Disposable Video Rhinolaryngoscope SP11A Endoscopic Video Processor M110B)
K253184 · Zhuhai Wesee Meditech Co., Ltd. · EOB · Mar 25, 2026 · Ear, Nose, Throat
Device Facts
| Record ID | K253184 |
| Device Name | Video Rhinolaryngoscope System (Disposable Video Rhinolaryngoscope SP11A Endoscopic Video Processor M110B) |
| Applicant | Zhuhai Wesee Meditech Co., Ltd. |
| Product Code | EOB · Ear, Nose, Throat |
| Decision Date | Mar 25, 2026 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 874.4760 |
| Device Class | Class 2 |
Intended Use
The Disposable Video Rhinolaryngoscope SP11A is a sterile, single-use, flexible endoscope intended for endoscopic procedures and examination within the nasal lumens and upper airway anatomy. The endoscope is used in conjunction with the Endoscopic Video Processor (Models M110B) to provide the images. The endoscope is intended for use in a hospital environment. It is designed for use in adults. The Endoscopic Video Processor M110B is used in conjunction with corresponding Wesee endoscopes to process and display images for endoscopic procedure.
Device Story
System comprises sterile, single-use flexible rhinolaryngoscope (SP11A) and reusable endoscopic video processor (M110B). CMOS sensor at distal end of endoscope captures video signals of body cavity; signals transmitted via cable to processor. LED lamp at distal end provides illumination. Processor receives, processes, and displays real-time video on integrated 10.1-inch monitor. Used in hospital settings by clinicians for visualization of nasal/upper airway anatomy. Output assists in endoscopic examination and procedural guidance. Benefits include sterile, single-use design to mitigate cross-contamination risks.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), photobiological safety (IEC 62471), biocompatibility (ISO 10993), and optical performance testing (resolution, depth of field, noise, dynamic range, color reproduction, intensity uniformity, frame frequency, system delay, geometric distortion).
Technological Characteristics
System includes flexible endoscope (3.2mm distal diameter, 300mm length) and video processor. Illumination via distal LED. Connectivity via HDMI. Power via internal lithium battery or grid. Sterilization of insertion portion via Ethylene Oxide (EO). Complies with ISO 8600-1/3/4, IEC 60601-1, IEC 60601-2-18, and ISO 10993 series.
Indications for Use
Indicated for endoscopic procedures and examination of nasal lumens and upper airway anatomy in adults in a hospital environment.
Regulatory Classification
Identification
A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
Predicate Devices
- Ambu® aScope TM 4 RhinoLaryngo Slim (K191080)
- Ambu® aView™ 2 Advance (K223299)
Related Devices
- K232435 — Rhinolaryngoscope system · Shenzhen HugeMed Medical Technical Development Co., Ltd. · Apr 26, 2024
- K230536 — Single-use Flexible Rhinolaryngoscope; Digital Video Monitor · Hunan Vathin Medical Instrument Co., Ltd. · Jun 14, 2023
- K192090 — CMOS Video Rhino-Laryngoscope SSU · Karl Storz SE & CO. KG · Jan 27, 2020
- K190972 — Ambu aScope 4 RhinoLaryngo Intervention · Ambu A/S · Jul 12, 2019
- K221581 — Single-Use Flexible Rhinolaryngoscope · Hunan Vathin Medical Instrument Co., Ltd. · Nov 18, 2022
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 25, 2026
Zhuhai Wesee Meditech Co., Ltd.
Ruiliang Xie
Management Representative
3-301, 178 Dingxing Road, Tangjiawan Town, Hitech Zone
Zhuahai, Guangdong 519085
China
Re: K253184
Trade/Device Name: Video Rhinolaryngoscope System (Disposable Video Rhinolaryngoscope SP11A Endoscopic Video Processor M110B)
Regulation Number: 21 CFR 874.4760
Regulation Name: Nasopharyngoscope (Flexible Or Rigid) And Accessories
Regulatory Class: Class II
Product Code: EOB
Dated: September 26, 2025
Received: September 26, 2025
Dear Ruiliang Xie:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253184 - Ruiliang Xie
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K253184 - Ruiliang Xie
Page 3
Sincerely,
JOYCE C. LIN -S
for Shu-Chen Peng, Ph.D.
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K253184 | ? |
| Please provide the device trade name(s). | | ? |
| Video Rhinolaryngoscope System (Disposable Video Rhinolaryngoscope SP11A
Endoscopic Video Processor M110B) | | |
| Please provide your Indications for Use below. | | ? |
| The Disposable Video Rhinolaryngoscope SP11A is a sterile, single-use, flexible endoscope intended for
endoscopic procedures and examination within the nasal lumens and upper airway anatomy. The
endoscope is used in conjunction with the Endoscopic Video Processor (Models M110B) to provide the
images. The endoscope is intended for use in a hospital environment. It is designed for use in adults.
The Endoscopic Video Processor M110B is used in conjunction with corresponding Wesee endoscopes to
process and display images for endoscopic procedure. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
1 / 8
# 510(k) Summary
510(k) Number: K253184
Date of Preparation: 03/25/2026
1. Sponsor Identification
**Zhuhai Wesee Meditech Co., Ltd.**
3-301, 178 Dingxing Road, Tangjiawan Town, Hitech Zone, Zhuhai, Guangdong, China
Establishment Registration Number: Not assigned yet
Contact Person: Ruiliang Xie
Position: Management Representative
Tel: +86- 07566286262
Email: xie_rl@weseemed.com
2. Designated Submission Correspondent
Mr. Ruiliang Xie
**Zhuhai Wesee Meditech Co., Ltd.**
P.O. Box 3-301, Zhuhai, 519085, China
Tel: +86- 07566286262
Email: xie_rl@weseemed.com
3. Identification of Subject Device
Trade Name: Video Rhinolaryngoscope System (Disposable Video Rhinolaryngoscope SP11A Endoscopic Video Processor M110B)
Common Name: Video Rhinolaryngoscope System
**Regulatory Information**
Classification Name: Nasopharyngoscope (flexible or rigid) and accessories
Classification: II
Product Code: EOB
Regulation Number: 21 CFR 874.4760
Review Panel: Ear Nose & Throat;
Indication for Use:
The Disposable Video Rhinolaryngoscope SP11A is a sterile, single-use, flexible endoscope intended for endoscopic procedures and examination within the nasal lumens and upper airway anatomy. The endoscope is used in conjunction with the Endoscopic Video Processor (Models M110B) to provide the
{5}
images. The endoscope is intended for use in a hospital environment. It is designed for use in adults.
The Endoscopic Video Processor M110B is used in conjunction with corresponding Wesee endoscopes to process and display images for endoscopic procedure.
## 4. Identification of Predicate Device
Predicate Device 1:
K Number: K191080
Trade/Device Name: Ambu® aScope TM 4 RhinoLaryngo Slim
Indication for Use:
The endoscope is a sterile, single-use, flexible endoscope intended for endoscopic procedures and examination nasal lumens and upper airway anatomy. The endoscope is intended to provide visualization via a monitor.
The endoscope is intended for use in a hospital environment. It is designed for use in adults.
Predicate Device 2:
K Number: K223299
Trade/Device Name: Ambu® aView™ 2 Advance
Indication for Use:
The Ambu® aView™ 2 Advance is intended to display live imaging data from compatible Ambu visualization devices.
## 5. Device Description
The Video Rhinolaryngoscope System consists of the Disposable Video Rhinolaryngoscope SP11A, Video Cable, Endoscopic Video Processor M110B and Accessories. To avoid risk of cross-contamination, the insertion part of the Disposable Video Rhinolaryngoscope SP11A is designed as a sterile, single use device. And other parts of the subject system are reusable device.
The endoscope has a long, thin, flexible tube to enter the patient's body, and the CMOS sensor embedded in the distal end of the endoscope is used to capture video signals of body cavity and real-time transfer the video signals to the endoscopic video processor through the video cable. The endoscopic video processor receives and processes the video signals according to the surgeon's operation on the control panel of the video image processor. Meanwhile, the light emitting diode (LED) lamp of the endoscope provides an illumination for the body cavity and the clear video is displayed on the monitor.
2 / 8
{6}
6. Substantially Equivalent (SE) Comparison
Table 1. General Comparison
| ITEM | Subject Device | Predicate Device 1 K191080 | Predicate Device 2 K223299 | Remark |
| --- | --- | --- | --- | --- |
| Product Code | EOB | EOB | EOQ | Similar |
| Regulation Number | 21 CFR 874.4760 | 21 CFR 874.4760 | 21 CFR 874.4680 | Similar |
| Class | II | II | II | Same |
| Manufacturer | Zhuhai Wesee Meditech Co., Ltd. | Ambu A/S | Ambu A/S | / |
| Product name | Video Rhinolaryngoscope System | Ambu® aScope™ 4 RhinoLaryngo Slim | Ambu® aView™ 2 Advance | / |
| Indication for Use | The Disposable Video Rhinolaryngoscope SP11A is a sterile, single-use, flexible endoscope intended for endoscopic procedures and examination within the nasal lumens and upper airway anatomy. The endoscope is used in conjunction with the Endoscopic Video Processor (Models M110B) to provide the images. The endoscope is intended for use in a hospital environment. It is designed for use in adults. | The endoscope is a sterile, single-use, flexible endoscope intended for endoscopic procedures and examination within the nasal lumens and upper airway anatomy. The endoscope is intended to provide visualization via a monitor. The endoscope is intended for use in a hospital environment. It is designed for use in adults. | / | Same |
| | The Endoscopic Video Processor M110B is used in conjunction with corresponding Wesee endoscopes to process and display images for endoscopic procedure. | / | The Ambu® aView™ 2 Advance is intended to display live imaging data from compatible Ambu visualization devices. | |
{7}
Similar-Product Code and Regulation Number
The predicate device 2's Product Code and Regulation Number refer to the bronchoscope and accessories. However, the Ambu® aView™ 2 Advance was also cleared for use with other compatible Ambu devices including the previously-cleared Ambu® aScope™ 4 RhinoLaryngo Slim. Therefore, the Ambu® aView™ 2 Advance can be used as the predicate device for the subject Endoscopic Video Processor (Models M110B).
Table 2. Specifications Comparison of Disposable Video Rhinolaryngoscope
| ITEM | Subject Device | Predicate Device 1 K191080 | Remark |
| --- | --- | --- | --- |
| Product name | Disposable Video Rhinolaryngoscope | Ambu® aScope™ 4 RhinoLaryngo Slim | / |
| Model | SP11A | Ambu® aScope™ 4 RhinoLaryngo Slim | / |
| Illumination method | LED on the distal end | LED on the distal end | Same |
| Distal end diameter | 3.2mm | 3.5mm | Similar |
| Maximum diameter of insertion portion | 3.6mm | 3.5mm | Similar |
| Working length | 300mm | 300mm | Same |
| Working channel | No working channel | No working channel | Same |
| Field of View | 110° | 85° | Different |
| Depth of Field | 3mm~50mm | 6mm~50mm | Similar |
| Bending of insertion portion | 130° up, 130° down | 130° up, 130° down | Same |
| Method of | EO | EO | Same |
| insertion portion | 130° up, 130° down | 130° up, 130° down | Same |
{8}
Similar-Distal end diameter
The distal end diameter of subject endoscope is similar as that of the predicate endoscope. The slight difference on distal end diameter will not affect the safety and effectiveness of the subject device.
Similar-Maximum diameter of insertion portion
The maximum outer diameter of insertion part of subject endoscope is similar as the predicate endoscope. Meanwhile, based on ISO 8600 Testing, the test results of distal end diameter maximum insertion portion width demonstrate the subject device meets the product design requirements. Therefore, this item will not affect the safety and effectiveness of the subject device.
Different-Field of View
The field of view of subject endoscope is larger than that of the predicate endoscope. Meanwhile, based on ISO 8600 Testing, the test results of field of view demonstrate the subject device meets the product design requirements. Therefore, this item will not affect the safety and effectiveness of the subject device.
Similar-Depth of Field
The depth of field of subject endoscope is similar as that of the predicate endoscope. Meanwhile, based on Depth of Field Testing, the subject device has good resolution in the intended depth of field. Therefore, this item will not affect the safety and effectiveness of the subject device.
Table 3. Specifications Comparison of Endoscopic Video Processor
| ITEM | Subject Device Endoscopic Video Processor | Predicate Device 2 Endoscopic Image Processor K223299 | Remark |
| --- | --- | --- | --- |
| Product name | Endoscopic Video Processor | Ambu® aView™ 2 Advance | / |
| Model | M110B | Ambu® aView™ 2 Advance | / |
| Dimension | 267mm (W) × 47mm (D) × 200mm (H) | 331mm (W) × 52mm (D) × 215mm (H) | Similar |
| Display screen | 10.1” | 12.8” | Different |
| Orientation | Landscape | Landscape | Same |
| Storage medium | SD card | USB drive | Different |
| Video output port | HDMI | HDMI 3G-SDI | Similar |
| Power supply | built-in lithium battery or power grid via the power adaptor. | built-in lithium battery or power grid via the power adaptor. | Same |
{9}
6 / 8
Similar- Dimension
The dimension of subject endoscopic video processor is slightly smaller than that of the predicate device. The differences on dimension will not affect the safety and effectiveness of the subject device.
Different-Display screen
The display screen of the subject device is different from that of the predicate device. Based on the **optical performance testing** (resolution, depth of field, noise and dynamic range, color reproduction, intensity uniformity, image frame frequency and system delay, geometric distortion), the optical performances of the subject system were similar to those of the predicate system. Therefore, the difference on display screen size will not affect the safety and effectiveness of the subject device.
Different-Storage medium
The subject device and predicate device have different storage medium. Considering both SD card and USB drive are common storage mediums, this difference on storage medium will not affect the safety and effectiveness of the subject device.
Similar-Video output
The subject device has HDMI video output port, while the predicate device has HDMI and 3G-SDI video output ports. The optical performance was conducted by using the HDMI video output port, based on the **optical performance testing** (resolution, depth of field, noise and dynamic range, color reproduction, intensity uniformity, image frame frequency and system delay, geometric distortion), the optical performances of the subject system were similar to those of the predicate device by using the HDMI video output port. Therefore, the difference on video output port will not affect the safety and effectiveness of the subject device.
7. Non-Clinical Test Conclusion
Non-clinical tests were conducted to verify that the subject device met all design specifications and was Substantially Equivalent (SE) to the predicate devices.
Electrical Safety
Comply with IEC 60601-1:2005+AMD1:2012+AMD2:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
EMC
Comply with IEC 60601-1-2:2014+AMD1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
Endoscope requirements
Comply with IEC 60601-2-18:2009 Medical electrical equipment - Part 2-18: Particular requirements
{10}
for the basic safety and essential performance of endoscopic equipment
## Shelf life of endoscope
In order to ensure the safety and effectiveness of subject endoscope in the shelf life, the sponsor conducted accelerated aging on the subject endoscope and performed the testing in term of package integrity before aging and after simulated distribution, package integrity after accelerated aging, optical performance testing before aging and after accelerated aging to provide valid data for the endoscope's shelf life per the following standard.
ASTM D4169-23 Standard Practice for Performance Testing of Shipping Containers and Systems
ASTM F88/F88M-23, Standard Test Method for Seal Strength of Flexible Barrier Materials.
ASTM F1929-23 Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration.
ASTM F1886/F1886M-16 Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
## Service life of Endoscopic Video Processor
In order to ensure the safety and effectiveness of Endoscopic Video Processor in the service life, the sponsor conducted accelerated aging on the subject Endoscopic Video Processor and performed the function and performance testing to provide valid data for the Endoscopic Video Processor's service life.
## Endoscope Performance Testing
The subject device was tested per the following standard, to evaluate Surface and Edges, Maximum insertion portion width, Field of View and Direction of View, Bending Angle. The test results demonstrate that the subject device complies with the product design requirements.
ISO 8600-1:2015 Endoscopes - Medical endoscopes and endotherapy devices -Part 1: General requirements
ISO 8600-3:2019 Endoscopes - Medical endoscopes and endotherapy devices -Part 3: Determination of field of view and dire
ISO 8600-4:2023 Endoscopes - Medical endoscopes and endotherapy devices -Part 4: Determination of maximum width of insertion portion
## Photobiological Safety Testing
The subject device is tested per IEC 62471:2006 Photobiological safety of lamps and lamp systems standard, to evaluate photobiological safety. The test result demonstrated that the subject system complies with the standard requirements.
## Sterilization
EO/ECH residual testing was conducted on the subject device per ISO 10993-7:2008 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals.
7 / 8
{11}
Biocompatibility
Comply with ISO 10993-5:2009 Biological evaluation of medical devices-Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2010 Biological evaluation of medical devices-Part 10: Tests for irritation and skin sensitization
ISO 10993-23:2021 Biological evaluation of medical devices - Part 23: Tests for irritation
8. Clinical Test Conclusion
No clinical study is included in this submission.
9. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the subject device is substantially equivalent compared to the legally marketed predicate devices.
8 / 8