Zsquare ENT-Flex endoscope is intended to visualize the internal cavities of the ear, airways, nose and sinus cavities during diagnostic endoscope procedures.
Device Story
Flexible, high-resolution endoscope system; consists of sterile, single-use shell (handle, steering mechanism, steerable shaft with imaging/illumination fibers, distal optics) and reusable imaging core (lens array, camera). Used in hospital/clinic environments by physicians. Imaging core inserts into shell; shell contacts patient and physician hands; shell disposed after procedure; imaging core cleaned with ethanol wipes. Console (illumination unit, CCU, monitor, keyboard, footswitch) processes video data from endoscope. Physician views real-time images on monitor to guide diagnostic procedures; enables visualization of internal cavities; reduces need for high-level disinfection/sterilization by using disposable patient-contact shell.
Clinical Evidence
Bench testing only. Included ISO 8600 image parameter testing, mechanical/dimensional inspection, illumination performance, mating cycle testing, IP testing, latency testing, and color performance. Sterilization validated per ISO 11135; shelf-life per ISO 11607. Biocompatibility per ISO 10993-1. Reprocessing validated per ST-724-2021/P.
Technological Characteristics
Flexible endoscope; 2.3mm insertion diameter; 290mm working length; 93° FOV; 130° articulation. Single-use shell (sterile, ETO) + reusable imaging core (disinfected with 70% isopropyl alcohol). Console-based image processing; LED illumination. Complies with IEC 60601-1, IEC 60601-2-18, IEC 60601-1-2, and IEC 62471.
Indications for Use
Indicated for visualization of internal cavities of ear, airways, nose, and sinus cavities during diagnostic endoscopic procedures.
Regulatory Classification
Identification
A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 14, 2022
Zsquare, Ltd Liat Diamant-Porat CEO, LDP Consulting Ltd. 43 Hasivim Street Petah Tikva, 4959501 Israel
Re: K220004
Trade/Device Name: Zsquare ENT-Flex Regulation Number: 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: EOB Dated: Mav 4, 2022 Received: May 10, 2022
Dear Liat Diamant-Porat:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K220004
Device Name Zsquare ENT-Flex
Indications for Use (Describe)
Zsquare ENT-Flex endoscope is intended to visualize the internal cavities of the ear, airways, nose and sinus cavities during diagnostic endoscope procedures.
**Type of Use (Select one or both, as applicable)**X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
| Owner's Name & Address | Zsquare, Ltd<br>43 Hasivim Street<br>Petah Tikva, Israel 4959501<br>Phone: +972-3-698-7600 |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Liat Diamant Porat, LL.B, B.Sc<br>LDP Consulting Ltd.<br>44 Hey Beiyar Street<br>Tel-Aviv, Israel<br><a href="mailto:liat@ldp-consulting.com">liat@ldp-consulting.com</a><br>+972-50-525-2099 |
| Date | June 09, 2022 |
| Trade Name | Zsquare ENT-Flex |
| Common / Classification Name | Ear & Nose Surgery |
| Product Code | EOB |
| Classification | Class II |
| Predicate Device | 3NT endoscopy system (K162916), Ambu aScope 4<br>RhinoLaryngo Slim (K191080) |
| Device Description | Zsquare ENT-Flex endoscope is a flexible, high-resolution endoscope with a narrow insertion diameter composed of a Single-Use Shell, provided sterile, that fully encompasses a reusable Imaging Core. The Zsquare ENT-Flex endoscope is intended to visualize the internal cavities of the ear, airways, nose, and sinus cavities during diagnostic endoscope procedures. The endoscope is to be used with a Console provided by Zsquare in a hospital or clinic environment. This device is designed to reduce the need for high-level disinfection and/or sterilization between patients by providing a sterile disposable Single-Use Shell that is used in combination with a reusable Imaging Core.<br>The Zsquare ENT-Flex system consists of a/an:<br>• Endoscope comprised of a:<br>• Disposable Single-Use Shell that is comprised of the handle, the steering mechanism, and the steerable shaft (imaging fiber, illumination fiber and distal optics). It is the only part that contacts the patient and physician's soiled hands and is disposed of at the end of the procedure.<br>• Reusable Imaging Core, which is a hand-held multi-use unit containing the optical system that is composed of a lens array and a camera. At the end |
{4}------------------------------------------------
of a procedure, the Imaging Core is disengaged from the Shell and cleaned with ethanol wipes.
- · Console comprised of a/an:
- o Illumination Unit
- o Camera Control Unit (CCU)
- o Monitor/Screen
- Keyboard with touchpad
- o Footswitch
- o Cart
The console enables the image-processing. A data cable connects to the endoscope and transfers the video data to the image-processing unit. The image is transferred and displayed on a graphical user interface (i.e.: monitor).
Power is supplied to the camera contained in the Imaging Core by a cable from the CCU. A second cable is connected to the illumination source to control the illumination.
Indications for use Zsquare ENT-Flex endoscope is intended to visualize the internal cavities of the ear, airways, nose, and sinus cavities during diagnostic endoscope procedures.
## Technological Characteristics
| Optics Performance: | |
|-------------------------------|---------------------------------------------------|
| Field of View | 93° (air) |
| Direction of View | Forward view |
| Depth of Field | 3.5 – 50 mm |
| Optimal working distance | 7 mm |
| Insertion Tube Diameter | 2.3mm |
| Insertion Tube Working Length | 290 mm |
| Articulation | 130° |
| Reusability | Single-use disposable unit, reusable imaging unit |
#### Non-Clinical Performance Data
## PERFORMANCE - BENCH TESTING
Bench testing included Image Parameter Testing (Including ISO 8600), Shaft Dimensional Visual Inspection and Mechanical Properties, Zsquare ENT-Flex Illumination Source (ZEIS) HW & Optical Performance Testing, Reusable to Console Mating Cycles Testing, Reusable to Disposable Mating Cycles Testing, IP Testing of the Imaging Core, Color Performance Test, Glass to Glass Latency Testing.
STERILIZATION, PACKAGING & SHELF-LIFE Sterilization validation of the Single-Use Shell was tested according to Sterilization ISO 11135.
{5}------------------------------------------------
The shelf life of the Zsquare ENT-Flex Shell (use before date) is established through an accelerated aging study according to ISO 11607.
Testing of the the product after transportation/environmental and shelf-life simulation showed that the sterile barrier packaging integrity is maintained, thus assuring maintenance of the product sterility assurance level (SAL) of 10-6.
Core requires Intermediate The lmaging Level Disinfection, and this process kills pathogenic and other microorganisms by physical or chemical means.
The reprocessing validation was conducted, according to ST-724-2021/P Manual Cleaning Intermediate Level Disinfection Validation.
## BIOCOMPATIBILITY
Biocompatibility assessments were performed in accordance with the appropriate risk category requirements, as defined in ISO 10933-1.
## SOFTWARE
Software verification and validation testing was performed to demonstrate that the software in the subject device meets design specifications.
#### ELECTROMAGNETIC COMPATIBILITY AND ELECTRICAL SAFETY
General requirements for basic safety and essential performance IEC 60601-1. Electrical safety for endoscopes in IEC 60601-2-2018. Electromagnetic compatibility testing per IEC 60601-1-2. Photobiological safety per IEC 62471.
Substantial Equivalence The Zsquare ENT-Flex endoscope has similar indications for use and technological features to the predicate devices, the 3NT Endoscopy System and Ambu aScope 4 RhinoLaryngo Slim. These devices are single-use endoscopes, designed to visualize the internal cavities of ear, airways, nose, and sinus cavities. The the endoscopes are supplied sterile and should not be reused. The 3NT Endoscopy System contains an integrated working channel that allows irrigation through a single channel, whereas the Zsquare ENT-Flex endoscope and the Ambu endoscope do not contain a working channel, however, this difference does not introduce hazards and does not raise questions of safety or effectiveness. The Zsquare ENT-Flex endoscope device has the same basic
{6}------------------------------------------------
device design as the 3NT Endoscopy System and Ambu aScope 4 RhinoLaryngo Slim predicate devices. The principle of operation is similar between the devices, as is the expected workflow, and most of the device features.
All devices include a single-use unit connected to a reusable unit, and a CCU. All endoscopes provide a singleuse, sterile distal unit, which is disposed of at the end of the procedure. All single-use distal units are ETO sterilized. All single-use disposable distal units include the insertion tube that is in direct contact with the patient. The 3NT endoscope is composed of a multi-use handle (includes controls for anqulation of the endoscope tip) that engages the single-use insertion tube. The Zsquare ENT-Flex disposable unit completely encompasses the Imaging Core and includes a handle so that the physicians' sterile hands come in contact only with the disposable unit (images are captured with a foot switch) to minimize crosscontamination. The Ambu endoscope is entirely disposable, however is for use with the Ambu aView 2 Advance that requires disinfection by wipes (that are on lsopropyl alcohol 70-80 %, Isopropyl based alcohol/Ammonium Chloride or Isopropyl alcohol/Ethanol.) The Ambu endoscope contains optics and illumination. At the end of the procedure the single-use endoscope unit (or the whole endoscope for Ambu) in all devices is disengaged and disposed of; the Imaging Core of Zsquare is cleaned with a lint-free cloth dampened with 70% isopropyl alcohol, which is considered intermediate level sterilization, as it is not exposed to contamination. It is similar to the reprocessing recommendations for the reusable Ambu monitor. The 3NT multi use unit is reprocessed and autoclaved between patients since the physician holds the handle during the procedure. The Zsquare illumination system includes an external illumination source and illumination fiber in the shaft similar to the 3NT; the only difference is that the Zsquare illumination source is separate from the CCU and the 3NT illumination source is embedded in the CCU. The Ambu illumination system is two LED lights embedded in the tip. The Zsquare endoscope contains an optical system within the Imaging Core that is composed of a lens array and a camera that is connected to a console (control unit), which includes an image/video processor and a white LED light source. Similarly, the 3NT endoscope is composed of a multi-use handle that engages the single-use insertion tube and a Camera control unit (CCU) which includes a video processor and a white LED light source. The 3NT CCU includes an attachment cable which connects to the single-use endoscope, receives video images from the endoscope, and delivers LED light to the endoscope. The Ambu aScope 4 RhinoLaryngo Slim is to be used with the
Ambu® aView™ 2 Advance, the
separately sold
{7}------------------------------------------------
equivalent of the CCU unit. In all devices the video data is processed in a separate image-processing unit. The minor differences in technological characteristics do not raise new types of safety or effectiveness questions. Performance testing and validations confirm that these differences do not adversely impact performance. Therefore, the proposed device is substantially equivalent to the predicate devices.
## Conclusions
Based on the comparison of intended use, indications for use, technological characteristics and performance testing, Zsquare Ltd. believes that the Zsquare ENT-Flex is substantially equivalent to its predicates.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.