The endoscope is a sterile, single-use, flexible endoscope intended for endoscopic procedures and examination within the nasal lumens and upper airway anatomy. The endoscope is intended to provide visualization via a monitor. The endoscope is intended for use in a hospital environment. It is designed for use in adults.
Device Story
Sterile, single-use flexible endoscope for visualization of nasal lumens and upper airway anatomy. Features maneuverable tip controlled by user via handle lever; integrated camera and LED light source at distal tip. Device connects to external monitor for real-time visualization. Used in hospital settings by clinicians to perform endoscopic examinations. Provides direct visual feedback to assist in clinical assessment of airway/nasal structures. Benefits include reduced risk of cross-contamination due to single-use design and portability.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including performance testing (insertion cord dimensions, bending angle, image sharpness), biocompatibility (ISO 10993), electrical safety (IEC 60601-1, IEC 60601-2-18), electromagnetic compatibility (IEC 60601-1-2), and packaging integrity/sterilization validation (ASTM F1980, ASTM D4169, ASTM F88, etc.).
Technological Characteristics
Flexible endoscope with maneuverable tip; distal LED illumination and camera. Insertion cord diameter 3.0 mm; working length 300 mm. Sterilized by Ethylene Oxide. Single-use. Complies with ISO 8600-1, ISO 8600-3, ISO 8600-4, ISO 594-1, IEC 60601-1, IEC 60601-2-18, and IEC 60601-1-2.
Indications for Use
Indicated for endoscopic procedures and examination of nasal lumens and upper airway anatomy in adults in a hospital environment.
Regulatory Classification
Identification
A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
Predicate Devices
Ambu® aScope™ RLS Slim (K18128)
Reference Devices
Olympus ENF-GP (K01186)
Ambu® aScope™ 4 Broncho Slim (K17372)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 22, 2019
Ambu A/S % Sanjay Parikh Director, QA/RA Ambu Inc. 6230 Old Dobbin Lane, Suite 250 Columbia, Maryland 21045
Re: K191080
Trade/Device Name: Ambu aScope 4 RhinoLaryngo Slim Regulation Number: 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (flexible or rigid) and accessories Regulatory Class: Class II Product Code: EOB Dated: April 23, 2019 Received: April 24, 2019
Dear Sanjay Parikh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Michael Ryan Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K191080
Device Name Ambu aScope 4 RhinoLaryngo Slim
Indications for Use (Describe)
The endoscope is a sterile, single endoscope intended for endoscopic procedures and examination within the nasal lumens and upper airway anatomy. The endoscope is intended to provide visualization via a monitor.
The endoscope is intended for use in a hospital environment. It is designed for use in adults.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <div> <input checked="true" type="checkbox"/> </div> |
|----------------------------------------------|------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div> <input type="checkbox"/> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
This Special 510(k) summary has been prepared in accordance with 21 CFR 807.87(h) and the content and format of the 510(k) summery has been prepared in accordance with 21 CFR 807.92.
| Submitter | Ambu A/S<br>Baltorpbakken 13<br>DK-2750 Ballerup<br>Denmark<br>Tel.: +45 7225 2000<br>Fax.: +45 7225 2050 | | | |
|----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|---------------------------------|----------------|
| Contact Person | Name: Gurpreet Kaur Rehal<br>Job Title: Regulatory Affairs Professional<br>Address: Ambu A/S, Baltorpbakken 13, DK-2750 Ballerup<br>Telephone number: +45 7225 2116<br>Fax number: +45 7225 2050 | | | |
| Date Summary<br>Prepared | April 19, 2019 | | | |
| Device Trade<br>Name | Ambu® aScope™ 4 RhinoLaryngo Slim | | | |
| Device Common<br>Name | Rhino-Laryngoscope | | | |
| Device<br>Classification | Nasopharyngoscopes (Flexible or Rigid)<br>Product Codes: EOB<br>21 CFR 874.4760<br>Class II | | | |
| Legally<br>Marketed | | Manufacturer | Trade Name | 510k<br>number |
| devices to which<br>the device is<br>substantially | Predicate<br>C: | Ambu A/S | Ambu® aScope™ RLS<br>Slim | K18128 |
| equivalent | Reference:<br>A: | Olympus<br>Medical<br>Systems<br>Corporation | Olympus ENF-GP | K01186 |
| | Reference<br>B: | Ambu A/S | Ambu® aScope™ 4<br>Broncho Slim | K17372 |
{5}------------------------------------------------
| Description of<br>the Device | The Ambu® aScope™ 4 RhinoLaryngo Slim is a sterile single use flexible<br>endoscope for examination of the nasal lumens and upper airway<br>anatomy. | | |
|----------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|--------------------------------------|
| | Ambu® aScope™ 4 RhinoLaryngo Slim has the following physical and<br>performance characteristics:<br>Maneuverable tip controlled by the user Flexible insertion cord Camera and LED light source at the distal tip Sterilized by Ethylene Oxide For single use | | |
| Indications for<br>Use | The endoscope is a sterile, single-use, flexible endoscope intended for<br>endoscopic procedures and examination within the nasal lumens and<br>upper airway anatomy. The endoscope is intended to provide<br>visualization via a monitor.<br><br>The endoscope is intended for use in a hospital environment. It is<br>designed for use in adults. | | |
| Summary of the<br>technological<br>characteristics<br>in comparison to<br>the predicate<br>devices | Specification | Ambu® aScope™<br>RLS Slim | Ambu® aScope™ 4<br>RhinoLaryngo Slim |
| | Distal end diameter | 4.2 mm (0.16") | 3.5 mm (0.14") |
| | Insertion cord<br>diameter | 3.8 mm (0.15") | 3.0 mm (0.12") |
| | Working length | 600 mm (23.6") | 300 mm (11.8") |
| | Bending section | 180° up, 180° down | 130° up, 130° down |
| | Max diameter of<br>insertion portion | 4.3 mm (0.17") | 3.5 mm (0.14") |
| | Ambu® aScope™ 4 RhinoLaryngo Slim is similar to the predicate device C<br>and reference A and B devices in the following areas:<br>They are all flexible endoscopes with a maneuverable tip. They all have a handle with a control lever giving the operator<br>ability to steer the tip of the scope up and down. They all provide illumination from the distal tip. They all have the same "field of view" and "direction of view" They all have insertion cord diameters within the same range. The Ambu® aScope™ 4 RhinoLaryngo Slim diameter is similar to the<br>predicate and reference device:<br>Predicate C has a larger diameter. Reference A has the same working length. Predicate C<br>and reference B have a longer working length Predicate C and Reference A have same intended use. Predicate C and reference B are single-use device, which are<br>delivered sterile | | |
{6}------------------------------------------------
. They are all portable endoscopes. Validation from non-clinical testing demonstrated that these technological features do not raise any new issues of safety and effectiveness of the applicant device. Overall, the line extension scope is narrower in diameter and shorter in length than the predicate device. Furthermore, the following non-dimensional changes are made: . The color of the control lever and tube connection has been changed from gray to purple, to indicate the product variant. The control lever and tube connector are components on the handle and do not have direct or indirect contact with the patient and do not raise new questions of safety and effectiveness. ● The type of Tyvek primary packaging material has been changed to Tyvek 2FS with EP 1250 film. Packaging integrity testing and sterilization validation report have been updated and do not raise new questions of safety and effectiveness. . Adhesive used for water tightness of the camera is changed. Tests that documents compliance with design requirements are performed, including verification using ISO10993 tests for cytotoxicity, irritation and sensitization, and do not raise new questions of safety and effectiveness. Ambu® aScope™ 4 RhinoLaryngo Slim has been successfully tested for its Performance functions, performance and safety as per FDA recognized consensus Data -Bench standards. The following bench data are described for the product line extension in the premarket notification: Declaration of Conformity with the following product specific standards ISO 8600-1, ISO 8600-3, ISO 8600-4 and ISO 594-1 ● Result: All tests were passed. Performance test has been conducted to document the following properties of Ambu® aScope™ 4 RhinoLaryngo Slim: ● Length and diameters of insertion cord ● Bending angle, endurance and radius . Image Sharpness Result: All tests were passed. Performance test has been conducted to document the shelf life and sterile barrier of Ambu® aScope™ 4 RhinoLaryngo Slim, where the aging and sterile Packaging Integrity testing is carried out in accordance with: ● ASTM F1980: Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices ● ASTM D4169: Standard Practice for Performance Testing of Shipping Containers and Systems
{7}------------------------------------------------
| Performance<br>Data - Clinical | ASTM F1886/F1886M: Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection ASTM F2096: Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test) (Test Method B - Procedure for Porous Packaging.) Seal Strength tests performed based on: EN 868-5: Packaging for terminally sterilized medical devices – Part 5: Sealable pouches and reels of porous materials and plastic film construction - Requirements and test methods and ASTM F88: Standard Test Method for Seal Strength of Flexible Barrier Materials Test reports that verify the Electromagnetic Compatibility and Electricial Safety: Electromagnetic Compatibility in compliance with IEC 60601-1-2. Electrical Safety in compliance with IEC 60601-1 and IEC 60601-2-18. Result: All tests were passed.<br>Similar requirements and test methods are used in the bench test for the predicate device, which supports the determination of substantial equivalence.<br>The line extension Ambu® aScope™ 4 RhinoLaryngo Slim and predicate device have the same intended use fundamental technology and there are no significant differences that raise different questions of safety and effectiveness.<br>Not applicable. |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion | Based on the indication for use, technological characteristics, performance data and comparison to predicate and reference devices it has been concluded that the functionality and intended use of Ambu® aScope™ 4 RhinoLaryngo Slim is substantially equivalent to the predicate device.<br>It is concluded that Ambu® aScope™ 4 RhinoLaryngo Slim is as safe and as effective and perform as well as the predicate device. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.