Sharpsite™ AC Rigid Endoscope: The Sharpsite™ AC Rigid Endoscopes are intended for use in surgical imaging in otolaryngology and Head and Neck procedures. including thinology and endoscopic plastic and reconstructive surgery. Sharpsite™ AC Sterilization Tray: Sharpsite™ AC Endoscope Sterilization Tray is intended to provide a specific configuration to hold, organize and protect Sharpsite™ AC Rigid Endoscope during their transport, storage and sterilization prior to use in surgical procedures.
Device Story
Sharpsite™ AC Rigid Endoscopes are reusable rod lens instruments for visualization during otolaryngology and head and neck procedures. Devices feature diameters from 2.7 mm to 4.0 mm and angles of view of 0, 30, 45, and 70 degrees. Input consists of optical images captured via rigid rod lens; light is provided by external sources connected via fiber optics. Output is direct visualization or camera-captured images for physician use in surgical settings. The sterilization tray is a thermoplastic case designed to organize, protect, and facilitate reprocessing of the endoscopes. Devices are provided non-sterile and require cleaning and sterilization (steam, EtO, STERRAD) between uses. Clinical benefit includes providing access and observation during surgical procedures.
Clinical Evidence
Bench testing only. Reprocessing and sterilization validation (steam, EtO, STERRAD 100S, STERRAD NX) achieved a sterility assurance level (SAL) of 10^-6. Biocompatibility assessment (cytotoxicity, irritation, sensitization, acute systemic toxicity) performed per ISO 10993-1. No clinical or animal testing required.
Technological Characteristics
Rigid rod lens endoscope; fiber optic light transmission. Materials: Phynox Cobalt-Chromium-Nickel Alloy (CoCr20Ni15Mo), 304 Stainless Steel (EN 10088 DIN 1.4301), plane glass, glass fibers, and epoxy resin (EPO TEK OG214). Reusable; autoclavable. Sterilization: Steam, EtO, STERRAD. Thermoplastic sterilization tray included.
Indications for Use
Indicated for surgical imaging in otolaryngology and head and neck procedures, including rhinology and endoscopic plastic and reconstructive surgery.
Regulatory Classification
Identification
A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters FDA in a blue box, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below.
April 20, 2018
Medtronic Xomed, Inc. Shravan Chawla Quality Engineer 6743 Southpoint Drive North Jacksonville, FL 32216
Re: K173855
Trade/Device Name: Sharpsite™ AC Rigid Endoscope Regulation Number: 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: EOB Dated: March 20, 2018 Received: March 22, 2018
Dear Shravan Chawla:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely yours,
Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K173855
Device Name
Medtronic Sharpsite™ AC Rigid Endoscope and Sharpsite™ AC Endoscope Sterilization Tray
Indications for Use (Describe)
Sharpsite™ AC Rigid Endoscope: The Sharpsite™ AC Rigid Endoscopes are intended for use in surgical imaging in otolaryngology and Head and Neck procedures. including thinology and endoscopic plastic and reconstructive surgery.
Sharpsite™ AC Sterilization Tray: Sharpsite™ AC Endoscope Sterilization Tray is intended to provide a specific configuration to hold, organize and protect Sharpsite™ AC Rigid Endoscope during their transport, storage and sterilization prior to use in surgical procedures.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
This summary of safety and effectiveness is submitted in accordance with the requirements of 21 CFR 807.92.
#### l. Company
| Address of Legal Manufacturer: | Medtronic Xomed, Inc.<br>6743 Southpoint Drive North<br>Jacksonville, FL 32216<br>(904) 296-9600 |
|--------------------------------|--------------------------------------------------------------------------------------------------|
| Date: | 18 December 2017 |
| Contact Person: | Shravan Chawla<br>Quality Engineer |
| Alternate Contact: | Gabriela Parana<br>Sr Regulatory Affairs Manager |
| II. Device | |
| Trade or Proprietary Name: | Sharpsite™ AC Rigid Endoscope |
| Common Name: | Sharpsite™ AC Rigid Endoscope |
Class: ll Product Code: EOB Classification Name: 21 CFR 874.4760 – Nasophayngoscope (flexible or rigid) and accessories
#### lll. Description
The Sharpsite™ AC Endoscope Sterilization Tray has been designed to be used with Sharpsite™ AC Rigid Endoscopes for reprocessing. The Sharpsite™ AC Rigid Endoscope is the parent device and Sharpsite™ Tray is considered an accessory as the tray is specifically designed for the endoscopes. This Traditional 510(k) includes the following devices:
{5}------------------------------------------------
- 1. Sharpsite™ AC Rigid Endoscopes
- 2. Sharpsite™ AC Endoscope Sterilization Tray
### Sharpsite™ AC Rigid Endoscopes
The Sharpsite™ AC endoscopes are rigid rod lens endoscopes for imaging and are available in diameters from 2.7 mm to 4.0 mm and in angles of view of 0, 30, 45 and 70 degrees. These endoscopes are reusable, autoclavable instruments and are provided non-sterile.
The Sharpsite™ Rigid Ear, Nose and Throat (ENT) endoscopes provide access and allow observation during otolaryngology and head and neck procedures. The device consists of rigid rod inserted into body orifice. The endoscope includes an optical system to visualize the image by direct view or with use of a camera. The endoscopes are provided non-sterile and can be reused by multiple methods as mentioned in the IFU. Each Sharpsite™ endoscope is fitted with a commercially available removable light port adapter for use with different manufacturers light cables. The Endoscrub sheath, removable light port adapter and light cables are not in the scope of this 510(k).
The Sharpsite™ AC Rigid Endoscopes are manufactured in multiple configurations that differ in insertion tube outer diameter and optical parameters. These endoscopes were previously cleared by FDA (K965233) for the visualization during otolaryngology and head and neck procedures.
# Sharpsite™ AC Endoscope Sterilization Tray
Sharpsite™ AC Endoscopic sterilization tray is a container which is a reusable device that includes a base and lid. It is a single level thermoplastic instrument case designed to hold Sharpsite™ AC Rigid Endoscopes. The Sharpsite™ tray is specifically configured to organize, store, protect, transport and sterilize Sharpsite™ AC Rigid Endoscopes for use in surgical procedures.
#### IV. Indications for Use
The proposed 510(k) includes Sharpsite™ AC Rigid Endoscopes as the parent subject device. The submission also includes the Sharpsite™ Tray as an accessory. The indications for use for the parent device Sharpsite™ AC Rigid Endoscope are identical to the indications for use cleared in K965233. See below for the indications for use of the parent device and proposed indications for use of the tray.
# Sharpsite™ AC Rigid Endoscopes (Parent Device)
The Sharpsite™ AC Rigid Endoscopes are intended for use in surgical imaging in otolaryngology and Head and Neck procedures, including rhinology and endoscopic plastic and reconstructive surgery.
{6}------------------------------------------------
### Sharpsite™ AC Endoscope Sterilization Tray (Accessory)
Sharpsite™ AC Sterilization Tray is intended to provide a specific configuration to hold, organize and protect Sharpsite™ AC Rigid Endoscope during their transport, storage and sterilization prior to use in surgical procedures.
#### V. Predicate Device
| Trade/Proprietary Name: | Sharpsite™ AC Rigid Endoscope |
|--------------------------|--------------------------------------------------------------------|
| Classification Name: | Nasopharyngoscope (flexible or rigid) and<br>accessories |
| Class/Panel: | Class II, EOB, 21 CFR 874.4760 |
| 510(k) Submitter/Holder: | Medtronic Xomed<br>6743 Southpoint Drive<br>Jacksonville, FL 32216 |
| 510(k) Number: | K965233 (S.E. 04/04/1997) |
#### VI. Comparison of Technological Characteristics
The subject and predicate device are based on the same technological characteristics:
- 1. Identical optical system diameter of the rigid rod
- 2. Same working length of the endoscope
- 3. Direction of view (DOV) is within the range of the predicate device
- 4. Field of View (FOV) is within the range of the predicate device
- 5. Same patient contacting materials are used in the predicate and subject device
The complete comparison of the technological characteristics between the subject and predicate device have been described in the table 5-1 below.
{7}------------------------------------------------
| Attributes | Subject Device:<br>Sharpsite™ AC Rigid<br>Endoscopes | Predicate Device:<br>Sharpsite™ AC Rigid<br>Endoscopes (K965233) | Comparison |
|-------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|
| Light Transmission | Fiber optics | Fiber optics | Same |
| Light Source | External, connected via<br>light guide to light guide<br>connector | External, connected via<br>light guide to light guide<br>connector | Same |
| Image Transmission | Rigid Rod Lens | Rigid Rod Lens | Same |
| Direction of view (DOV) | 0; 30; 45; 70 degrees | 0; 30; 45; 70 degrees | Same |
| Field of View (FOV) | 87; 97; 100; 102 degrees | 87; 97; 100; 102 degrees | Same |
| Single Use/<br>Reusable | Reusable | Reusable | Same |
| Reprocessing | Cleaning, sterilization<br>(steam, EtO, STERRAD) | Cleaning, sterilization<br>(steam, EtO, STERRAD) | Same |
| Materials<br>-insertion tube<br>or shaft (outer<br>surface) | Phynox Cobalt-<br>Chromium-Nickel Alloy<br>(CoCr20Ni15Mo) and<br>304 Stainless Steel (EN<br>10088 DIN 1.4301) | Phynox Cobalt-<br>Chromium Nickel Alloy<br>(CoCr20Ni15Mo) and 304<br>Stainless Steel (EN 10088<br>DIN 1.4301) | Same |
| -insertion part<br>(distal surface) | Plane glass | Plane glass | Same |
| -fiber optics | Glass fibers | Glass fibers | Same |
| -lens bonding | Epoxy resin (EPO TEK<br>OG214) | Epoxy resin (EPO TEK<br>OG214) | Same |
| Packaging | Case with foam insert | Case with foam insert | Same |
| Accessories | Sterilization Tray | Not Included | Use of<br>compatible<br>Sharpsite™ AC<br>Endoscope<br>Sterilization Tray |
| Attributes | Subject Device: Sharpsite™ AC Rigid Endoscopes | Predicate Device: Sharpsite™ AC Rigid Endoscopes (K965233) | Comparison |
| | | | to sterilize and clean the endoscope per instructions in the IFU. |
Table 5-1: Technological Characteristics of the Proposed Device and the Predicate Device
{8}------------------------------------------------
#### VII. Testing
Cleaning and Sterilization validations of the endoscopes in the Sterilization tray were completed to ensure the functionality after worst case cleaning and sterilization testing. Reprocessing and sterilization validation testing of the endoscope met all acceptance criteria and has demonstrated a sterility assurance level (SAL) of 10-6 for sterilization via steam, Ethylene Oxide (EO), STERRAD 100S and STERRAD NX sterilization. Biocompatibility assessment was also conducted to establish the safety and effectiveness of the endoscopes. Biocompatibility assessment included cytotoxicity, irritation and sensitization, and acute systemic toxicity in accordance with ISO 10993-1. Additional bench, animal or clinical testing was not required to establish substantial equivalence. All tests were successfully completed with passing results.
#### Conclusion VIII.
The indications for use, technology and performance characteristics of the subject Sharpsite™ AC Rigid Endoscope is substantially equivalent to the predicate Sharpsite™ AC Rigid Endoscope, cleared under K965233. The Sharpsite™ AC Rigid Endoscopes are as safe, as effective, and performs as well as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.