The Schoelly Sinuscope is intended for use in otolaryngology and Head and Neck procedures, including rhinology, and endoscopic plastic and reconstructive surgery.
Device Story
Schoelly Sinuscope is a rigid, reusable endoscope for visualization during otolaryngology and head/neck surgery. Device requires external light source, light guide, video camera, monitor, and printer (not included). Light is transmitted via fiber optics to the distal tip; images are captured via rigid rod lens system and transmitted to an eyepiece for camera/monitor display. Used in clinical settings by physicians. Provides high-resolution visualization of surgical sites to assist in diagnostic and therapeutic procedures. Benefits include minimally invasive access and improved surgical site visibility.
Clinical Evidence
Bench testing only. Performance validated through temperature testing (IEC 60601-2-18, IEC 60601-1), optical parameter testing (ISO 8600 series), biocompatibility (ISO 10993), and bonding strength testing. Reprocessing validated per AAMI TIR12, AAMI TIR30, ANSI/AAMI ST15883-1, ISO 14937, and ISO 17664.
Technological Characteristics
Rigid reusable endoscope; materials: Phynox (Co-Cr-Ni alloy), 304 stainless steel, glass fibers, glass lenses. Bonding: AuSn brazing alloy and epoxy. Illumination: external light source via fiber optics. Imaging: rigid rod lens system. Configurations: 4mm diameter/175mm length or 2.7mm diameter/110mm length. No working channel. Sterilization: steam, STERRAD (100S, 100NX, NX).
Indications for Use
Indicated for patients undergoing otolaryngology, Head and Neck, rhinology, and endoscopic plastic and reconstructive surgical procedures.
Regulatory Classification
Identification
A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of a human figure, with three overlapping profiles facing to the right. The profiles are rendered in black and are connected by flowing lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 27, 2015
Schoelly Fiberoptic Gmbh % Ms. Pamela Papineau, RAC President Delphi Medical Device Consulting, Inc. 5 Whitcomb Ave Ayer, Massachusetts 01432
> K142249 Trade/Device Name: Schoelly Sinuscope Regulation Number: 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (Flexible or Rigid) And Accessories Regulatory Class: Class II Product Code: EOB Dated: December 22, 2014 Received: December 24, 2014
Dear Ms. Papineau:
Re:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
Device Name Schoelly Sinuscope
Indications for Use (Describe)
The Schoelly Sinuscope is intended for use in otolaryngology and Head and Neck procedures, including thinology, and endoscopic plastic and reconstructive surgery.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K142249 - 510(k) Summary
## General Information
| Preparation date: | 11/24/2014 |
|-------------------------|--------------------------------------------------------------------------|
| Owner's Name: | Schoelly Fiberoptic GmbH (Registration: 8043903) |
| Address: | Robert-Bosch-Str. 1 - 3<br>79211 Denzlingen<br>Germany |
| Telephone Number: | +49-7666-980-0 |
| Fax Number: | +49-7666-908-380 |
| Contact Person: | Dr. Sandra Baumann |
| Subject Device Name: | Schoelly Sinuscope |
| Trade Name: | Schoelly Sinuscope |
| Common/Usual Name: | Sinuscope |
| Classification Name: | EOB - Nasopharyngoscope (flexible or rigid)<br>21 CFR 874.4760; Class II |
| Predicate Device Name: | SHARPSITE Ac |
| Trade Name: | SHARPSITE Ac |
| Common/Usual Name: | Rigid Rod Lense Endoscope |
| Classification Name: | EOB - Nasopharyngoscope (flexible or rigid)<br>21 CFR 874.4760; Class II |
| Premarket Notification: | K965233, Medtronic Xomed Inc., SE date April 4, 1997 |
### Device Description
The Schoelly Sinuscope is a rigid reusable endoscope for visualization during Otolaryngology-Head and Neck surgery in conjunction with a commercially available and approved light guide, light source, video camera, monitor, and printer. Light source, video camera, monitor, and printer are not included in the scope of delivery and are further not within the scope of this application.
Schoelly Sinuscopes are manufactured in multiple configurations that differ in insertion tube outer diameter and working length and with respect to optical parameters (direction of view, field of view). Several models of the Schoelly Sinuscope have already been cleared for marketing by FDA (K133682) for the visualization during arthroscopic procedures.
Like other currently marketed sinuscopes, Schoelly Sinuscopes have outer surfaces mainly made from metal (Phynox cobalt-nickel-chromium alloy, 304 stainless steel) and further comprise fiber optics for light transmission and rigid rod-lenses for image transmission.
The Schoelly Sinuscope is delivered in a non-sterile condition and is already CE marked.
{4}------------------------------------------------
## Technological Characteristics
Light that is created by an external light source is transmitted from the Sinuscope light guide connector through the Sinuscope itself to the tip via a fiber optic system. Images are transferred the other way back through a rigid rod lens system. Technical parameters of the Schoelly Sinuscopes that characterize the optical view are the Direction of View (0°-70°) and the Field of View (80°-100°). The image can be displayed by a camera/monitor system which can be connected to the Sinuscope eyepiece. The Schoelly Sinuscopes are available in two different configurations for the insertion tube: 4mm diameter, working length 175mm and 2.7mm diameter, working length 110mm. All configurations of the Schoelly Sinuscope do not have any working channel.
A detailed comparison of the technological characteristics with the predicate device is shown in the table below.
| | Proposed<br>Device<br>(current submission) | Predicate<br>Device (K965233) |
|--------------------------------------|-----------------------------------------------------------------|-----------------------------------------------------------------|
| Attribute | | |
| Light<br>transmission | Fiber optics | Fiber optics |
| Light source | External, connected via light<br>guide to light guide connector | External, connected via light<br>guide to light guide connector |
| Image<br>transmission | Rigid rod lenses | Rigid rod lenses |
| Direction of<br>view | $0°-70°$ | $0°-70°$ |
| Field of view | $80°-100°$ | $80°-100°$ |
| Image display | Camera/monitor connected via<br>the endoscope eyepiece | Camera/monitor connected via<br>the endoscope eyepiece |
| Working<br>channel | none | none |
| Single Use /<br>Reusable | Reusable | Reusable |
| Reprocessing | Cleaning, sterilization (steam,<br>STERRAD®) | Cleaning, sterilization (steam,<br>STERRAD®, EtO) |
| Materials | | |
| - insertion tube<br>(outer surface) | stainless steel, Co-Cr-Ni alloy | stainless steel |
| - insertion part<br>(distal surface) | stainless steel, glass | stainless steel, glass |
| - fiber optics | glass fibers | glass fibers |
| - lens bonding | AuSn brazing alloy, epoxy | epoxy |
Table 5.1 Technological Characteristics of the Proposed Device and the Predicate Device
{5}------------------------------------------------
| Attribute | Proposed<br>Device<br>(current submission) | Predicate<br>Device (K965233) |
|------------------------|--------------------------------------------|-------------------------------|
| Temperature<br>testing | IEC 60601-2-18 compliant | IEC 60601-2-18 compliant |
| Packaging | Case with foam insert | Case with foam insert |
# Indications for Use
The Schoelly Sinuscope is intended for use in otolaryngology and Head and Neck procedures, including rhinology, and endoscopic plastic and reconstructive surgery.
The indications for use of the proposed devices are identical to the indications for use of the Medtronic predicate device.
# Non-clinical Performance Testing
The Schoelly Sinuscope was subjected to temperature, optical parameter, biocompatibility and performance testing with identical acceptance criteria for the Schoelly and the predicate devices. The results of this testing demonstrated that the Schoelly Sinuscope has met pre-determined acceptance criteria and is substantially equivalent to the predicate device. The risks associated with the use of the new device were found acceptable when evaluated in accordance with ISO 14971. Risks and benefits of predicate device are the same as compared to the Schoelly Sinuscope.
Temperature testing: The device was measured for surface temperatures at various locations over time using different light sources and found to meet requirements specified in IEC 60601-2-18 and IEC 60601-1.
Optical parameter testing: The device was tested for all relevant optical parameters and found to meet the minimum requirements defined in internal specifications and as specified in the ISO 8600 series of standards (field of view, direction of view, diopters, image eccentricity, size of view, and vignetting).
Biocompatibility testing: A series of biocompatibility testing according to ISO 10993, including cytotoxicity, sensitization, irritation, and acute systemic toxicity, demonstrated that the device components that are in contact with the patient are biocompatible.
Performance testing: Performance testing consisted of measurement of the bonding strength for the Sinuscope distal glass, with comparison to the distal glass bonding strength of the currently marketed predicate device.
# Reprocessing
The Schoelly Sinuscope is the subject of completed reprocessing validations including manual cleaning, automated cleaning, steam sterilization, as well as STERRAD® 100S, STERRAD® 100NX and STERRAD® NX sterilization.
Cleaning studies have been performed in accordance with AAMI TIR12:2010 (Designing. Testing and Labeling Reusable Medical Devices for Reprocessing in Health Care Facilities: A
{6}------------------------------------------------
Guide for Device Manufacturers), AAMI TIR30:2011 (A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable devices), and ANSI/AAMI ST15883-1: 2009 (Washer-disinfectors - Part 1 - General requirements, terms, and definitions and tests). Devices that have been used for testing had been undergo multiple reprocessing cycles and surface marring artificially created by scratching with metal tools to simulate end of lifetime use and to address the FDA Draft Guidance Processing Medical Devices in Healthcare Settings (dated: May 2, 2011).
Sterilization studies have been performed in accordance with ISO 14937:2009 (Sterilization of health care products - General requirements for characterization of sterilizing agent and the development, validation and routine control of a sterilization process for medical devices), ANSI/AAMI ST81:2004 (Sterilization of medical devices - Information to be provided by the manufacturer for the processing of resterilizable medical devices), ISO 17664:2004 (Sterilization of medical devices - Information to be provided by the manufacturer for the processing of resterilizable medical devices), and ANSI/AAMI/ISO 17665-1:2006 (Sterilization of health care products - Moist heat - Requirements for the development, validation and routine control of sterilization process for medical devices).
# Conclusion
The Schoelly Sinuscope meets all the pre-determined acceptance criteria of the testing performed to demonstrate substantial equivalence to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.