ACCLARENT CYCLOPS MULTI-ANGLE ENDOSCOPE

K110097 · Acclarent, Inc. · EOB · May 27, 2011 · Ear, Nose, Throat

Device Facts

Record IDK110097
Device NameACCLARENT CYCLOPS MULTI-ANGLE ENDOSCOPE
ApplicantAcclarent, Inc.
Product CodeEOB · Ear, Nose, Throat
Decision DateMay 27, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 874.4760
Device ClassClass 2

Indications for Use

The Acclarent Cyclops Multi-Angle Endoscope is intended to provide an endoscopic means to view the nasal cavity and nasopharynx in an operating room environment.

Device Story

Acclarent Cyclops Multi-Angle Endoscope (CYE002) is a 4mm rigid, unchanneled endoscope for visualizing nasal cavity and nasopharynx. Device features a direction-of-view dial (10° to 90°) and a 320° shaft rotation dial, allowing visualization of structures without rotating the scope body. Rare-earth permanent magnets in the proximal control body drive the direction-of-view change. Device includes a standard eyepiece for camera coupling and a light post compatible with ACMI light sources. Used in operating room environments by physicians. Provides 55° field of view and 5mm to 40mm depth of focus. Reusable; requires cleaning and sterilization before each use. Benefits include enhanced visualization capabilities during endoscopic procedures.

Clinical Evidence

Bench testing only. Evaluated distal shaft diameter, working length, field of view, fixed focus, direction of view, rotation of view, illumination, scope resolution, dial actuation forces, temperature, magnetic field strength, electrical safety, EMC, durability, environmental conditioning, compression, vibration, and shock. Temperature testing used thermocouples with 300W Xenon light source. No clinical data required.

Technological Characteristics

Rigid, 4mm diameter endoscope. Sapphire cover lens. Illumination via glass fibers. Rare-earth permanent magnets (≤10 gauss at 2cm) for articulation. Stainless steel adapters. Compatible with standard camera couplers and ACMI light sources. Standards: IEC 60601-2-18, IEC 60601-1-2, ISO 10993-1, ISO 8600-3, ISO 8600-5. Reprocessing: manual cleaning with enzymatic soak and automated washer; pre-vacuum steam sterilization.

Indications for Use

Indicated for visualization of the nasal cavity and nasopharynx in patients undergoing procedures in an operating room setting.

Regulatory Classification

Identification

A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the word "Acclarent" in a bold, sans-serif font. A curved line extends from the top of the "r" to the right, creating a visual element above the word. The text is black against a white background. , K110097 MAY 2 7 2011 ## 510(k) SUMMARY | Sponsor/Submitter: | Acclarent, Inc.<br>1525-B O'Brien Drive<br>Menlo Park, California 94025 | |-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Keri Yen<br>Regulatory Affairs Manager<br>Phone: (650) 687-5874<br>Fax: (650) 687-4449 | | Date of Preparation: | May 20, 2011 | | Device Trade Name/<br>Model Number: | Acclarent Cyclops Multi-Angle Endoscope<br>CYE002 | | Common Name: | Endoscope | | Device Classification: | Class II | | Regulation Number: | 21 CFR 874.4760 | | Classification Name: | Nasopharyngoscope (Flexible or Rigid) | | Product Code: | EOB | | Predicate Devices: | Acclarent Cyclops Multi-Angle Endoscope (K100577) | | Device Description: | The Acclarent Cyclops Multi-Angle Endoscope is a 4mm rigid<br>unchanneled endoscope that has the capability of varying<br>direction of view from 10° to 90°, which is altered by the<br>direction of view dial. The direction of view is indicated by<br>visible markings on the scope body. Cyclops provides a 55°<br>field of view and a depth of focus from 5 mm to 40mm. The<br>device shaft can also rotate 320° to allow for visualization of<br>structures without rotating the device; this is controlled by the<br>shaft rotation dial. Small rare-earth permanent magnets are<br>incorporated into the proximal scope control body (≤10 gauss<br>at 2cm) and drive the change in the direction of view. A<br>standard eyepiece located on the proximal end of the device is<br>compatible with a standard camera coupler. The light post on<br>the subject device is compatible with an ACMI light source.<br>There are two stainless steel adapters that accompany the<br>Acclarent Cyclops Multi-Angle Endoscope to facilitate | {2}------------------------------------------------ Image /page/1/Picture/0 description: The image contains the word "Acclarent" in a bold, sans-serif font. A curved line extends from the top of the "r" to the top of the "t", creating a visual element above the word. The text is black against a white background. The adapters connect to the light post. The Acclarent Cyclops Multi-Angle Endoscope is a reusable device and must be cleaned and sterilized according to the user manual prior to every use. Indications for Use: The Acclarent Cyclops Multi-Angle Endoscope is intended to provide an endoscopic means to view the nasal cavity and nasopharynx in an operating room environment. | Characteristics: | | | |------------------------|-------------------------------------------------------------------|-----------------------------------------------------------------| | Attribute | Predicate Device<br>(Acclarent Cyclops Multi-<br>Angle Endoscope) | Subject Device<br>(Acclarent Cyclops Multi-<br>Angle Endoscope) | | 510(k) number | K100577 | K110097 | | Model Number | CYE001 | CYE002 | | Rigidity | Rigid | Same | | Viewing Optics | Lens (Sapphire cover) | Same | | Depth of View | 5-45 mm | Same | | Field of View | 60° | 55° | | Direction of View | 10° to 100° | 10° to 90° | | Shaft Body<br>Diameter | 4mm | Same | | Working Length | 6.89 inches (175mm) | Same | | Magnetic Strength | ≤10 gauss at 2cm | Same | | Illumination Fibers | Glass Fibers | Same | ## Technological Performance Data: Performance testing of the Acclarent Cyclops Multi-Angle Endoscope consisted of bench testing and a cadaver study. Bench testing met all acceptance criteria for attributes such as distal shaft diameter, working length, field of view, fixed focus, direction of view, rotation of view, illumination, scope resolution, dial actuation forces, temperature testing, field strength testing of magnets, electrical safety, EMC testing, durability testing, environmental conditioning, compression testing, random vibration testing, and shock (free fall drop) testing. Temperature testing consisted of attaching thermocouples at nine locations using adhesive and measuring the temperature at steady state. Test samples were connected to a 300W Xenon light source at 100% light output with both 3.5mm and 5.0mm light cables. The time to heat and time to cool was also evaluated. Clinical data was not necessary for the subject device. The performance data demonstrates that the Acclarent Cyclops Multi-Angle Endoscope performs as {3}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the word "Acclarent" in a bold, sans-serif font. A curved line extends from the top of the "A" to the right, ending above the "t". The font is black, and the background is white. ・ . | | intended according to IEC 60601-2-18, IEC 60601-1-2,<br>ISO10993-1, ISO 8600-3, ISO 8600-5. | |----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Validated Reprocessing<br>Methods: | Full manual cleaning with extended enzymatic soak<br>plus general instrument automated washer Pre-vacuum steam sterilization (wrapped) | | | In validating the above, the following standards were<br>referenced: AAMI/ANSI ST35, AAMI TIR 12, AAMI TIR 30,<br>and ANSI/AAMI ST79 A1/A2, ANSI/AAMI ST8,<br>ANSI/AAMI ST81, ASTM E1766. | | Summary of Substantial<br>Equivalence: | The Acclarent Cyclops Multi-Angle Endoscope is substantially<br>equivalent to the predicate devices as confirmed through<br>relevant performance tests and attributes. | . . . . . {4}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a stylized image of an eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular pattern around the eagle. The eagle is depicted in black, and the text is also in black. ## Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 MAY 2 7 2011 Acclarent, Inc. c/o Ms. Keri Yen. Regulatory Affairs Manager 1525-B O'Brien Drive Menlo Park, CA 94025 Re: K110097 Trade/Device Name: Acclarent Cyclops Multi-Angle Endoscope (Model CYE002) Regulation Number: 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (Flexible or Rigid) Regulatory Class: Class II Product Code: EOB Dated: April 26, 2011 Received: April 27, 2011 Dear Ms. Yen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {5}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. Ekhm Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## Indications for Use 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: _________________________________________________________________________________________________________________________________________________________________ Indications For Use: The Acclarent Cyclops Multi-Angle Endoscope is intended to provide an endoscopic means to view the nasal cavity and nasopharynx in an operating room setting. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Susan Rudy CRNP (Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices Page 1 of ____________________________________________________________________________________________________________________________________________________________________ K110097. 510(k) Number_
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...