CMOS VIDEO RHINO-LARYNGOSCOPE SYSTEM MODEL 11101CM
K103467 · KARL STORZ Endoscopy-America, Inc. · EOB · Jun 28, 2012 · Ear, Nose, Throat
Device Facts
Record ID
K103467
Device Name
CMOS VIDEO RHINO-LARYNGOSCOPE SYSTEM MODEL 11101CM
Applicant
KARL STORZ Endoscopy-America, Inc.
Product Code
EOB · Ear, Nose, Throat
Decision Date
Jun 28, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 874.4760
Device Class
Class 2
Indications for Use
The KARL STORZ CMOS Flexible Video Rhino-Laryngoscope System may only be used by persons with an appropriate medical qualification and who are acquainted with the rhino/laryngoscopic technique. The endoscope is used for endoscopic diagnosis within the nasal lumens and airway anatomy, and is intended to provide visualization via a video monitor.
Device Story
System comprises flexible endoscope with integrated LED light source and distal CMOS imaging sensor; C-MAC monitor for display. Used for endoscopic diagnosis of nasal lumens and airway anatomy. Operated by qualified clinicians in clinical settings. CMOS sensor captures video signal; transmitted to monitor for real-time visualization. Enables direct inspection of airway/nasal structures to assist clinical diagnosis. Benefits include improved visualization compared to traditional fiber-optic systems.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, ISO 10993, and ISO 8600. Reprocessing validated per AAMI TIR 12:2010, AAMI TIR 30:2003, AAMI 7:1999, and AAMI E 1837.
Technological Characteristics
Flexible endoscope with distal CMOS sensor; LED illumination; fiber optic light transmission. Compatible with C-MAC monitor. Reprocessing via manual cleaning and high-level disinfection (CIDEX OPA). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, ISO 10993, ISO 8600.
Indications for Use
Indicated for endoscopic diagnosis within the nasal lumens and airway anatomy in patients requiring rhino/laryngoscopic examination. For use by qualified medical professionals trained in rhino/laryngoscopic techniques.
Regulatory Classification
Identification
A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
Predicate Devices
KARL STORZ VIDEO RHINO-LARYNGSCOPE SYSTEM. MODEL11101SERIES (K072387)
Submission Summary (Full Text)
{0}------------------------------------------------
K103467
JUN 2 8 2012
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge.
Karl Storz Endoscopy-America, Inc. Applicant: 2151 E. Grand Avenue El Segundo. CA 90245
Contact: Shiven Gandhi Regulatory Affairs Associate Telephone: (424) 218-8322 FAX: (424) 218-8519 E-mail:shiven.gandhi@karlstorz.com
Date Prepared: 14th June 2012
## Device Identification:
Common Name: Video Rhino-Laryngoscope System
Trade Name: CMOS Video Rhino-Laryngoscope System
Predicate Device: KARL STORZ VIDEO RHINO-LARYNGSCOPE SYSTEM. MODEL11101SERIES (K072387)
Indication: The KARL STORZ CMOS Flexible Video Rhino-Laryngoscope System may only be used by persons with an appropriate medical qualification and who are acquainted with the rhino/laryngoscopic technique. The endoscope is used for endoscopic diagnosis within the nasal lumens and airway anatomy, and is intended to provide visualization via a video monitor.
### Device Description:
The Karl Storz CMOS Video Rhino-Laryngoscope System includes a flexible endoscope and a monitor. The rhino-laryngoscope consists of an integrated light located in the handle. The light is transmitted thought the fiber optic bundles and illuminates the anatomy under visualization. A new Storz specific CMOS imaging sensor located at the distal tip of the endoscope shaft which is inserted into the body cavities. The captured video signal is then transferred to the C-MAC monitor for display purposes. The Karl Storz CMOS Video Rhino-laryngoscope is a Class II device under 21 CFR874.4760, Rhino-Laryngoscope and accessories.
### Substantial Equivalence:
The Karl Storz CMOS Video Rhino-Laryngoscope System is substantially equivalent to the predicate device (K072387) since the intended uses and the technological characteristics are the same. The minor differences between the Karl Storz CMOS Video Rhino-Laryngoscope System and predicate device raise no issue of safety and effectiveness, as these differences have no effect on the performance, function or general intended use of these devices.
{1}------------------------------------------------
| | Subject Device | Predicate Device (K072387) |
|---------------------|----------------------------------------------------------------------------|----------------------------------------------------------------------------|
| Applicant | Karl Storz Endoscopy America,<br>Inc. | Karl Storz Endoscopy America,<br>Inc. |
| Device Name | Karl Storz CMOS Video Rhino-<br>Laryngoscope | Karl Storz Video Rhino-<br>Laryngoscope |
| Type of Device | Videoscope System | Videoscope System |
| Indication for use | For endoscopic diagnosis<br>within the nasal lumens and<br>airway anatomy. | For endoscopic diagnosis<br>within the nasal lumens and<br>airway anatomy. |
| Imaging Sensor | CMOS | CCD |
| Illumination source | LED | Xenon |
| Working Length | Same | Same |
| Shaft Diameter | Same | Same |
| Data Storage | Yes | Yes |
## Technological Characteristics Comparison Table
## Non-Clinical Performance Data:
The Karl Storz Flexible Video Rhino-Laryngoscope system has been successfully tested for its functions, performance and safety as per FDA recognized standards. Testing has been conducted as per IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, and ISO 10993 to address the safety aspect of the device. The performance of the device is tested by complying with ISO 8600.
## Reprocessing:
The CMOS Video Rhino-Laryngoscope is cleaned using manual cleaning with enzymatic detergent and is subjected to high level disinfection using CIDEX OPA. No methods of sterilization are validated for the CMOS Video Rhino-Laryngoscope. The cleaning and highlevel disinfection studies were perform in accordance with AAMI TIR 12:2010 (Designing, Testing and Labeling Reusable Medical Devices for Reprocessing in Health Care Facilities: A Guide for Device Manufacturers), AAMI TIR 30:2003 (A compendium of processes, materials. test methods, and acceptance criteria for cleaning reusable medical devices), AAMI 7: 1999 (Chemical Sterilants and High-Level Sterilants: A Guide to Selection and Use), AAMI E 1837 (Standard Test Method to Determine Efficacy of Disinfection Processes for Reusable Medical Devices) and FDA draft guidance for Processing Medical devices in Healthcare settings (dated: May 2,2011)
## Conclusion :
Hence the CMOS Video Rhino-Laryngoscope system is as safe and as effective to the predicate device in terms of its intended use and technological characteristics.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a circular seal for the Department of Health & Human Services USA. The seal features the department's emblem, which is a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged around the emblem in a circular fashion.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUN 2 8 2012
Karl Storz Endoscopy America Inc. c/o Mr. Shiven Gandhi, Regulatory Affairs Associate 2151 E. Grand Avenue El Segundo, CA 90245
Re: K103467
Trade/Device Name: Karl Storz CMOS Video Rhino Laryngoscope System Regulation Number: 21 CFR 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (Flexible or Rigid) and Accessories Regulatory Class: Class II Product Code: EOB Dated: June 14, 2012 Received: June 15, 2012
Dear Mr. Gandhi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, attresses, and include requirements for annual registration, listing of general volurely provincitice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability adultieration. Teasoneter ODTE assess that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device to additional controls. Existing major regulations affecting your device can be may be subject to additional vehicles, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I lease of advised that I Dr over and that your device complies with other requirements of the Act that I DA has made a associations administered by other Federal agencies. You must or any I carated and the regirements, including, but not limited to: registration and listing (21 comply with an the rece brequirements on 301); medical device reporting of medical CI'N I all 807), labeling (21 OFF 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Mr. Gandhi
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
R. A. Khan, M.D., f
Malvina B. Eydelman, M Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indication for Use
# 510(k) Number (if known):
Device Name: Karl Storz CMOS Video Rhino-Laryngoscope System
Indication for use: The KARL STORZ CMOS Flexible Video Rhino-Laryngoscope System may only be used by persons with an appropriate medical qualification and who are acquainted with the rhino/laryngoscopic technique. The endoscope is used for endoscopic diagnosis within the nasal lumens and airway anatomy, and is intended to provide visualization via a video monitor.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR ·
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel C Cluff
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K103467
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.