The Instrument Clamp Electromagnetic Sphere and Instrument Clamp Electromagnetic Universal are accessories to the Electromagnetic Navigation Unit and are intended for navigating conventional surgical instruments. The Stryker ENT Navigation System is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy. Example procedures include, but are not limited to the following ENT procedures: -Transsphenoidal access procedures; -Intranasal procedures; -Sinus procedures, such as Maxillary antrostomies. Ethmoidectomies, -Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies; -ENT related anterior skull-based procedures.
Device Story
Instrument Clamps Electromagnetic are accessories for Stryker ENT Navigation System; used to track conventional surgical instruments during ENT procedures. Device incorporates electromagnetic sensor coils directly into clamp body; eliminates need for separate tracker attachment. Input: electromagnetic field generated by system emitter; sensor coils transform field into analog signal. Output: tracking data transmitted to navigation platform. Used in OR by surgeons; enables navigation without line-of-sight. Healthcare providers use output to localize instruments relative to patient anatomy; improves surgical precision. PEEK body; titanium clamps/screws. Reusable; limited to 10 uses.
Clinical Evidence
No clinical data. Bench testing only: accuracy testing per ASTM F2554; reprocessing validation per ISO 17664, ISO 17665-1, and ANSI/AAMI ST79; electrical safety per ANSI/AAMI ES 60601-1.
Technological Characteristics
Materials: PEEK body, titanium clamps/screws. Sensing: Integrated electromagnetic sensor coils. Connectivity: Wired to Electromagnetic Navigation Unit. Form factor: Clamp-on accessory for rigid surgical instruments. Sterilization: Steam (ISO 17665-1, ANSI/AAMI ST79). Standards: ASTM F2554 (accuracy), ANSI/AAMI ES 60601-1 (electrical safety).
Indications for Use
Indicated for patients requiring stereotactic surgery in ENT procedures (e.g., transsphenoidal, intranasal, sinus, anterior skull base) where anatomical structures can be referenced to CT/MR models.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 9, 2020
Stryker Corporation Maya Schiel Staff Regulatory Affairs Specialist Boetzinger Strasse 41 Freiburg, Baden-Wuerttemberg D-79111 Germany
Re: K202609
Trade/Device Name: Instrument Clamps Electromagnetic Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: PGW Dated: September 3, 2020 Received: September 9, 2020
Dear Maya Schiel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Malvina Eydelman, M.D. Director OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K202609
Device Name Instrument Clamps Electromagnetic
#### Indications for Use (Describe)
The Instrument Clamp Electromagnetic Sphere and Instrument Clamp Electromagnetic Universal are accessories to the Electromagnetic Navigation Unit and are intended for navigating conventional surgical instruments.
The Stryker ENT Navigation System is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy.
Example procedures include, but are not limited to the following ENT procedures:
- Transphenoidal access procedures:
- Intranasal procedures;
- Sinus procedures, such as Maxillary antrostomies, Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies;
- ENT related anterior skull based procedures.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <input checked="" type="checkbox"/> |
|----------------------------------------------|-------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <input type="checkbox"/> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
Special 510(k): Instrument Clamps Electromagnetic
## Submitter Information
| 510(k) Submitter | Stryker Leibinger GmbH & Co. KG<br>Boetzinger Strasse 41<br>79111 Freiburg, Germany |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------|
| FDA Establishment Number | 8010177 |
| Contact Person | Maya Schiel<br>Staff Regulatory Affairs Specialist<br>Phone: +49 30 311 911 89<br>Mail: maya.schiel@stryker.com |
| Date Submitted | September 3, 2020 |
| Subject Device | |
| Trade / Proprietary Name | Instrument Clamps Electromagnetic |
| Common Name | Ear, Nose, And Throat Stereotaxic Instrument |
| Classification | Class II |
| Classification Product Code | PGW |
| Classification Name | Stereotaxic Instrument |
| Classification Regulation | 21 CFR 882.4560 |
st
## Legally Marketed Predicate Device
| Predicate Device Name | Scopis Extended Instrument Set EM |
|--------------------------------|-----------------------------------|
| | Instrument Clamps |
| | Universal Tracker Electromagnetic |
| Predicate Device 510(k) Number | K171661 |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The letters are black and the background is white. The word is all lowercase.
The Scopis Extended Instrument Set EM was cleared under K171661 by Scopis GmbH in September 2017. This submission was an extension to the previously cleared Scopis Hybrid Navigation System (K161491).
Scopis GmbH was acquired by Stryker in November 2017. In July 2018, the legal manufacturer for the devices was changed to Stryker Leibinger GmbH & Co. KG. Since October 2019, the devices have also been also physically manufactured in Freiburg under the Stryker Leibinger Quality Management System. The 510(k) ownership was also transferred from Scopis GmbH.
As part of the post-acquisition integration the Scopis name of the system was changed from Scopis Hybrid Navigation System to Stryker ENT Navigation System.
The cleared instrument set contains instrument clamps and an electromagnetic tracker, which needs to be mounted on top of a clamp to enable tracking of conventional surgical instruments. Since the Instrument Clamps Electromagnetic are a combination of these devices, they are viewed as variants of the predicate device. The Instrument Clamps Electromagnetic themselves have the same intended purpose and indications and are therefore considered line extensions.
#### Device Description
The Instrument Clamps Electromagnetic are part of the Stryker ENT Navigation System. The basic principle of this system is the generation of an electromagnetic (EM) field with an EMemitter (Field Generator). The EM devices have sensor coils that transform the detected EM field into an analogue signal is transmitted via cable to the Electromagnetic Navigation Unit. The signal is then evaluated, and the tracking data is transmitted to the platform. This technology allows tracking instruments without a line-of-sight requirement.
The Instrument Clamps Electromagnetic are to be used for localizing surgical instruments in a dedicated working volume. They are intended to be mounted on suitable surgical rigid instruments. The body of the clamp incorporates the required sensor coils, which enables tracking of conventional instruments. For accurate navigation, the instrument with the attached electromagnetic clamp needs to be calibrated prior to use.
Two electromagnetic clamps are in scope of this submission:
- Instrument Clamp Electromagnetic Sphere: Clamping interface optimized for tracking instruments with a circular cross-section between 2.0 mm and 4.6 mm in diameter (inclusive)
- . Instrument Clamp Electromagnetic Universal: Clamping interface optimized for tracking instruments with angular cross-section (e.g. forceps) between 3.0 mm and 6.0 mm in diameter (inclusive)
The subject devices are not intended to have patient contact. Their bodies are made of PEEK and the clamps and screws are manufactured from titanium.
They are reusable devices that need to be reprocessed prior to first use. They are limited to 10 uses.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase and black. The background is white.
## Intended Use and Indications
The Instrument Clamp Electromagnetic Sphere and Instrument Clamp Electromagnetic Universal are accessories to the Electromagnetic Navigation Unit and are intended for navigating conventional surgical instruments.
The Stryker ENT Navigation System is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy.
Example procedures include, but are not limited to the following ENT procedures:
- -Transsphenoidal access procedures;
- -Intranasal procedures;
- Sinus procedures, such as Maxillary antrostomies. Ethmoidectomies, -Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies;
- ENT related anterior skull-based procedures. -
#### Comparison of Technological Characteristics
This submission is about the modification of combining two products (clamp and electromagnetic tracker) into one (electromagnetic clamp).
The combination was achieved by integrating the electromagnetic sensor coil into the body of the clamp, as opposed to an EM tracker unit that must be separately attached, resulting in a smaller and lighter device. The one-piece electromagnetic clamp requires one step less of preparation and post-processing in the OR, since the devices do not need to be assembled or disassembled.
The Instrument Clamps Electromagnetic are attached to conventional surgical instruments the same way as the predicate clamps. The step of attaching a separate electromagnetic tracker (as in predicate) is not required, as it is already incorporated into the subject device.
The subject and predicate devices are identical in the following characteristics:
- . 510(k) holder
- FDA product codes and regulation ●
- Intended Use
- . Indications for Use
- Application Area ●
- Type of Use
- Type of Patient Contact ●
- Tracking Technology
- Usage
- Material ●
- Packaging ●
- Fixation ●
- Calibration ●
- Compatibility
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "Stryker" in a bold, sans-serif font. The letters are black and the background is white. The word is horizontally oriented and appears to be a logo or brand name.
The subject and predicate devices differ in the following ways:
The key requirements of the Instruments Clamps Electromagnetic are to be smaller and lighter, but with the same functionalities as the variants of the predicate. This requires a design that allows the device to be both clamp and tracker in one device. Therefore, the same type of electromagnetic sensor coil used in the predicate is directly incorporated into the body of the subject device in a more compact form.
#### Performance Data
#### Non-Clinical Testing
A risk analysis was completed for subject devices in compliance with ISO 14971 to identify required verification and validation activities in order to evaluate the impact of the device modification and to mitigate potential risks. This submission includes appropriate supporting data to show the safety and effectiveness of the Instrument Clamps Electromagnetic.
The following bench testing was performed to evaluate performance (accuracy), reprocessing, and electrical safety.
Table 1: Performance Data
| Scientific discipline | Test summary |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance - Accuracy | To ensure accuracy, subject devices are tested in accordance<br>with ASTM F2554. The devices meet state-of-the-art<br>acceptance criteria for point accuracy, precision and distance<br>accuracy. |
| Reprocessing | Subject devices are reusable instruments and were therefore<br>validated in compliance with the respective FDA guidance for<br>reprocessing medical devices including respective TIRs.<br>The steam sterilization was also conducted in compliance<br>with ISO 17665-1 and ANSI/AAMI ST79.<br>The respective instructions follow ISO 17664. |
| Electrical Safety | The devices are used as parts of the Stryker ENT Navigation<br>System, which is defined as ME Equipment. Therefore, devices<br>were checked and certified for compliance with ANSI/AAMI<br>ES 60601-1 as accessories to a ME Equipment. |
All cited standards are FDA-recognized consensus standards
#### Clinical Testing
No clinical testing was deemed necessary for this 510(k) submission.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "Stryker" in a bold, sans-serif font. The letters are black and the background is white. The word is the logo for Stryker Corporation, a medical technology company.
### Conclusion
The intended use and fundamental scientific technologies of subject devices and their predicate are identical. The subject devices are line extensions to the variants of the predicate and are therefore part of the same ENT Navigation System, following the same operating principles and conditions. The change of combining two devices into one does not affect the indications for use, tracking technology, calibration or other basic features. The device modifications do not raise different questions of safety and effectiveness than the predicate device.
Risk management was conducted in accordance with ISO 14971 and does not show any unacceptable risk or new risks when compared to the predicate. Verification testing demonstrates that the subject devices are at least as safe and effective as the predicate.
In conclusion, the information provided demonstrates that the Instrument Clamps Electromagnetic are at least as safe and effective as their predicate (K171661) and do not introduce different questions of safety or effectiveness, and therefore supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.