K193502 · Stryker Ent · PGW · Feb 14, 2020 · Neurology
Device Facts
Record ID
K193502
Device Name
TGS Universal Headrest with Mounting Arm
Applicant
Stryker Ent
Product Code
PGW · Neurology
Decision Date
Feb 14, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The TGS Universal Headrest with Mounting Arm is an accessory to the Stryker ENT Navigation System and intended for positioning the Field Generator. It also provides a resting surface for the patient to place their head during procedures requiring navigation.
Device Story
TGS Universal Headrest with Mounting Arm is a reusable accessory for Stryker ENT Navigation System. Device consists of headrest, mounting arm, and chair adapters. Mounting arm fixes navigation system field generator in proximity to patient to create electromagnetic field; headrest supports patient head during surgery. Used in clinical ENT settings by physicians. Device facilitates image-guided surgery by ensuring stable positioning of field generator and patient head. Benefits include improved navigation accuracy and procedural workflow for ENT surgeons.
Clinical Evidence
Bench testing only. Testing included dimensional, functional, simulated use, compatibility, and distribution testing. Design validation performed by ENT physicians confirmed that intended users could perform critical tasks in the expected clinical environment.
Technological Characteristics
Reusable, non-sterile device requiring cleaning/disinfection between uses. Features articulating hinged design with locking mechanisms for positioning. Components include headrest, mounting arm, and chair adapters. No electronic or software components; purely mechanical support and positioning system.
Indications for Use
Indicated for patients undergoing ENT stereotactic surgery where rigid anatomical structures (e.g., paranasal sinuses, mastoid anatomy) can be referenced to CT or MR-based anatomical models. Procedures include transsphenoidal access, intranasal, sinus (maxillary antrostomies, sphenoidotomies/explorations, turbinate resections, frontal sinusotomies), and anterior skull base procedures. No contraindications.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services-USA. To the right, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
February 14, 2020
Stryker ENT Denise Thompson Principal Regulatory Affairs Specialist 3600 Holly Lane North. Suite 40 Plymouth, Minnesota 55447
Re: K193502
Trade/Device Name: TGS Universal Headrest with Mounting Arm Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: PGW Dated: December 17, 2019 Received: December 18, 2019
## Dear Denise Thompson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Michael Ryan Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
Device Name
TGS Universal Headrest with Mounting Arm
| Indications for Use (Describe) | | BIRDER BE CONSULTION OF COLLECTION OF CLASS OF CONSULTER OF CLASS OF CONSULTER OF CLASS OF CONSULTER OF CLASS OF CLASS OF CLASS OF CLASS OF CLASS OF CONSULTER OF CONSULTER OF |
|--------------------------------|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | |
The Stryker ENT Navigation System, of which the TGS Universal Headrest with Mounting Arm is an accessory, is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy.
Example procedures include, but are not limited to the following ENT procedures:
- · Transsphenoidal access procedures:
- Intranasal procedures:
- · Sinus procedures, such as maxillary antrostomies; sphenoidotomies/sphenoid explorations, turbinate resections, and frontal sinusotomies:
- · ENT related anterior skull base procedures.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY
# stryker®
Date Prepared: Submitter Information:
Establishment Registration:
Contact Information:
Device Information: Trade Name: Common Name: Classification Name: Product Code: Classification: Regulation Number: Predicate Devices:
December 17th, 2019 Stryker ENT 3600 Holly Lane North, Suite 40 Plymouth, MN 55447
3006345872
Denise Thompson Principal Regulatory Affairs Specialist (269) 888-0219 denise.thompson(@stryker.com
TGS Universal Headrest with Mounting Arm Image Guided Surgery System Ear, Nose, and Throat Stereotaxic Instrument PGW Class II 21 CFR 882.4560 Navigation Headrest Office, K133573 Field Generator Mounting Arm, K161491
#### Device Description:
The TGS Universal Headrest with Mounting Arm is a reusable device intended to be used with the Stryker ENT Navigation System. It consists of three main components: a headrest, mounting arm and chair adapters. The mounting arm provides fixation for the field generator of the navigation system allowing it to be positioned in proximity to the patient creating an optimal electromagnetic field. The headrest provides support for the patient's head while undergoing navigation guided procedures. The chair adapters are fixed to the headrest to allow the headrest to various ENT procedure chairs
Image /page/3/Picture/12 description: The image shows a chair with a headrest and a field generator attached to it. The field generator is mounted on a mounting arm that is attached to the headrest. A chair adapter is used to connect the mounting arm to the chair. The field generator is provided with the navigation system.
TGS Universal Headrest with Mounting Arm
{4}------------------------------------------------
## Intended Use:
The TGS Universal Headrest with Mounting Arm is an accessory to the Stryker ENT Navigation System and intended for positioning the Field Generator. It also provides a resting surface for the patient to place their head during procedures requiring navigation.
#### Indication for Use:
The Stryker ENT Navigation System, of which the TGS Universal Headrest with Mounting Arm is an accessory, is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy.
Example procedures include, but are not limited to the following ENT procedures:
- Transsphenoidal access procedures; ●
- . Intranasal procedures;
- . Sinus procedures, such as maxillary antrostomies; sphenoidotomies; sphenoidotomies/sphenoid explorations, turbinate resections, and frontal sinusotomies;
- . ENT related anterior skull base procedures.
#### Contraindications:
None
### Technological Characteristics:
The technological characteristics of the TGS Universal Headrest and Mounting Arm are very similar to the predicate devices, the Navigation Headrest Office (an accessory to the Fiagon Navigation System, K133573) and the Field Generator Mounting Arm (an accessory to the Stryker ENT Navigation System, K161491).
The TGS Universal Headrest and Mounting Arm and its predicates all:
- . position the field generator of a navigation system,
- have articulating hinged design features, ●
- . achieve different positions through use of locking mechanisms,
- . are for use by healthcare professionals only (physicians), and
- are provided non-sterile and intended to be cleaned and disinfected between uses.
#### Substantial Equivalence:
The TGS Universal Headrest with Mounting Arm has been shown to be substantially equivalent to the predicate devices, the Navigation Headrest Office and the Field Generator Mounting Arm based on Intended Use, Indications for Use, Principles of Operation and technological characteristics.
The TGS Universal Headrest with Mounting Arm most closely resembles the primary predicate, the Navigation Headrest Office. The main difference between devices relates to the mounting mechanism (arm versus bracket) for the field generator. The subject device's mounting mechanism more closely resembles the mounting mechanism of the secondary predicate, the Field Generator Mounting Arm.
{5}------------------------------------------------
While there are some minor design differences between the subject and predicate devices, these differences do not raise different questions of safety or efficacy.
## Performance Data:
Bench testing was performed to demonstrate that the TGS Universal Headrest with Mounting Arm performs as intended and meets the design specifications. The testing included dimensional, functional, simulated use, compatibility and distribution testing. Design validation involved ENT physicians familiar with image guided sinus surgery and was conducted to demonstrate that the intended users could perform the critical tasks related to use of the TGS Universal Headrest with Mounting Arm in its expected use environment.
## Conclusion:
The TGS Universal Headrest with Mounting Arm has been shown to be substantially equivalent to the predicate devices based on Intended Use, Indications for Use, Principles of Operation and technological characteristics. The minor design differences between the subject and predicate devices do not raise different questions of safety or efficacy and bench performance testing has been conducted to demonstrate the TGS Universal Headrest with Mounting Arm performs as intended and equivalently to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.