← Product Code [PGW](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/PGW) · K173628

# Acclarent ENT Navigation System (K173628)

_Acclarent, Inc. · PGW · Mar 10, 2018 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94neurological-surgical-devices/PGW/K173628

## Device Facts

- **Applicant:** Acclarent, Inc.
- **Product Code:** [PGW](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/PGW.md)
- **Decision Date:** Mar 10, 2018
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The ACCLARENT® ENT Navigation System is intended for use during intranasal and paranasal image-guided navigation procedures for patients who are eligible for sinus procedures.

## Device Story

System provides real-time tracking of navigated surgical instruments during intranasal and paranasal procedures. Inputs: pre-acquired CT images and electromagnetic signals from sensors in instrument tips. Operation: Field Ring generates magnetic field; induces current in instrument sensor to calculate position; system registers CT images to patient anatomy; displays tool tip icon on registered image. Used in surgical settings by ENT physicians. Output: visual display of instrument location relative to patient anatomy. Benefit: assists surgeon in navigating sphenoid, frontal, and maxillary sinuses; improves procedural accuracy.

## Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including electrical safety, EMC, mechanical testing, location accuracy verification against a robotic system, software functional testing, and pre-clinical cadaver studies simulating sinuplasty workflows.

## Technological Characteristics

Electromagnetic tracking system; includes Navigation Console, Field Ring, Instrument Hub, Patient Tracker, and Registration Probe. Uses pre-acquired CT images for navigation. Connectivity: Workstation-based. Software-driven tracking of magnetic sensors in flexible tools.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- CARTO® ENT Navigation System ([K161701](/device/K161701.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA acronym is in a blue square, and the full name "U.S. Food & Drug Administration" is in blue text.

March 10, 2018

Acclarent, Inc. James Patrick Garvey II Associate Director, Regulatory Affairs 1525-B O'Brien Drive Menlo Park, CA 94025

Re: K173628

Trade/Device Name: Acclarent® ENT Navigation System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: PGW Dated: February 8, 2018 Received: February 9, 2018

Dear James Patrick Garvey II:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely yours,

## Srinivas Nandkumar -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Acclarent, a medical technology company. The word "Acclarent" is written in a bold, sans-serif font, with the first four letters in black and the last four in purple. Above the word "ent" is a purple arc. Below the word "Acclarent" is the phrase "PART OF THE Johnson & Johnson FAMILY OF COMPANIES" in a smaller font.

## ACCLARENT® ENT Navigation System Special 510(k) Premarket Notification

| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration |  | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020<br>See PRA Statement below. |
|-------------------------------------------------------------------------|--|---------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------|--|---------------------------------------------------------------------------------------------|

| 510(k) Number (if known)                        | K173628                                                                                                                                                                                                 |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name                                     | ACCLARENT® ENT Navigation System                                                                                                                                                                        |
| Indications for Use (Describe)                  | The ACCLARENT® ENT Navigation System is intended for use during intranasal and paranasal image-guided             navigation procedures for patients who are eligible for sinus procedures.             |
| Type of Use (Select one or both, as applicable) | <div> <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> </div> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services
   
Food and Drug Administration
   
Office of Chief Information Officer
   
Paperwork Reduction Act (PRA) Staff
   
PRAStaff@fda.hhs.gov

*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*

FORM FDA 3881 (7/17)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Acclarent, a medical technology company. The word "Acclarent" is written in a bold, sans-serif font, with the first half in black and the second half in purple. Above the word "Acclarent" is a purple arc. Below the word "Acclarent" is the phrase "PART OF THE Johnson & Johnson FAMILY OF COMPANIES", with "Johnson & Johnson" in red.

## ACCLARENT® ENT Navigation System
 Special 510(k) Premarket Notification

## 510(k) Summary

| [807.92(a)(1)] Submitter Information                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|-------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant:                                                                    | Biosense Webster, Inc.<br>33 Technology Drive<br>Irvine, CA 92618, USA<br>Tel.: (800) 729-9010<br>Fax: (949) 450-6886                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Contact Person:                                                               | James Patrick Garvey II<br>Associate Director, Regulatory Affairs<br>Phone: 949-789-8505<br>Fax: 650-687-5889                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Authored by:                                                                  | Anna Gantman<br>Quality and Regulatory Program Manager<br>Biosense Webster (Israel), Ltd.<br>+972-52-808-9735                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Date Summary Prepared:                                                        | January 19, 2018                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| [807.92(a)(2)] Name of Device                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Device Trade Name:                                                            | ACCLARENT® ENT Navigation System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Device Common Name:                                                           | Image Guided Surgery System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Device Classification:                                                        | Class II, 21 CFR 882.4560                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Classification Name:                                                          | Ear, Nose, and Throat Manual Surgical Instrument (21 CFR 882.4560)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Product Code                                                                  | PGW                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| [807.92(a)(3)] Legally Marketed Devices                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Predicate Device:                                                             | CARTO® ENT Navigation System<br>510(k)#: K161701                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| [807.92(a)(4)] Device Description                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Device Description:                                                           | The ACCLARENT® ENT Navigation System is intended to be used<br>during intranasal and paranasal surgical procedures to help ENT<br>physicians to track and display the real-time location of the tip of<br>navigated instruments relative to pre-acquired reference images, such<br>as CT.<br><br>The ACCLARENT® ENT Navigation System enables ENT physicians<br>to access sphenoid, frontal, and maxillary sinuses by using the system<br>magnetic tracking technology, identical to the predicate device.<br><br>The system incorporates a Navigation Console, Field Ring, Instrument<br>Hub, Patient Tracker, Registration Probe, Field Ring and Holder,<br>Workstation and accessories. A magnetic field generated by the Field<br>Ring induces a current in the magnetic sensor embedded in the tip of<br>the flexible navigated tool, which helps to accurately calculate the tool<br>tip position. A CT image is imported and registered to the patient<br>coordinates and a tool tip icon is displayed on top of the registered |
| [807.92(a)(5)] Intended Use<br>Indications for Use:                           | The ACCLARENT® ENT Navigation System is intended for use<br>during intranasal and paranasal image-guided navigation procedures<br>for patients who are eligible for sinus procedures.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Difference in Indications<br>from Predicate Device:                           | The intended use of the device is unchanged from the predicate.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| [807.92(a)(6)] Technical Characteristics<br>Technological<br>Characteristics: | The substantial equivalence of the ACCLARENT® ENT Navigation<br>System to the predicates is shown by similarity in intended use,<br>indications for use, and performance.<br><br>Like the predicate CARTO® ENT System, the ACCLARENT® ENT<br>Navigation System is an image-guided navigation system intended for<br>use during paranasal surgical procedures. ENT physicians can track<br>and display the real-time location of the tip of navigated instruments<br>relative to pre-acquired CT images.<br><br>Like the predicate, the ACCLARENT® ENT Navigation System<br>utilizes electromagnetic tracking technology for navigation, uses<br>anatomical reference points on the patient's anatomy for intraoperative<br>registration to the image-based model of the anatomy, and uses CT<br>image sets as reference images for the image-based model. The<br>instructions for use and product labels have been updated to reflect the<br>revised device.                                                                           |
| Non-clinical Performance<br>Data:                                             | The ACCLARENT® ENT Navigation System was tested to ensure that<br>it functions in accordance with the system design specifications related<br>to substantial equivalence in terms of device safety and effectiveness.<br><br>The following nonclinical tests were performed:<br><br>1. Proof of Design electrical tests, to verify all hardware modules<br>perform within specifications.<br><br>2. Location Accuracy tests, where the ACCLARENT® ENT<br>Navigation System electromagnetic locations were compared to the<br>locations provided by a, very accurate robot system over the entire<br>navigation volume, to verify the system precision claim.<br><br>3. Software functional tests, covering the complete system<br>functionality, and including error handling, usability and time                                                                                                                                                                                                                                      |
| 4.                                                                            | Safety, EMC, and mechanical tests were performed by a nationally recognized testing laboratory to verify compliance with safety and EMC standards for medical devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| 5.                                                                            | Simulated use accuracy test, in which a complete CT image registration and instrument navigation workflow was performed, to verify the overall accuracy of the system.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| 6.                                                                            | Pre-clinical (cadaver) tests were designed to mimic surgical procedures using the ACCLARENT® ENT Navigation System in a simulated clinical environment, to assess the execution of a complete sinuplasty procedure workflow and to qualitatively estimate the system clinical accuracy.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                                                               | The proposed ACCLARENT® ENT Navigation System passed all tests in accordance with appropriate test criteria and standards, and the modified device did not raise new questions of safety or effectiveness.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Clinical Performance Data:                                                    | Clinical data was not necessary to determine that the ACCLARENT® ENT Navigation System was substantially equivalent to the predicate device. The performance data demonstrated that the device performs as intended.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Conclusion:                                                                   | The modified ACCLARENT® ENT Navigation System is substantially equivalent to the currently cleared CARTO® ENT Navigation System based on the completion of non-clinical bench testing as well as similar principles of design, operation and indications for use.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Acclarent, a medical technology company. The word "Acclarent" is written in a bold, sans-serif font, with the first half in black and the second half in purple. A purple arc is above the second half of the word. Below the word "Acclarent" is the phrase "PART OF THE Johnson & Johnson FAMILY OF COMPANIES" in a smaller font.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Acclarent, a company that is part of the Johnson & Johnson family of companies. The word "Acclarent" is written in a bold, sans-serif font, with the first part in black and the "ent" in purple. Above the purple part of the name is a purple arc. Below the name is the text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES", with "Johnson-Johnson" in red script.

performance (latency).

---

**Source:** [https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94neurological-surgical-devices/PGW/K173628](https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94neurological-surgical-devices/PGW/K173628)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
