Fiagon Navigation System

K162176 · Fiagon GmbH · PGW · Dec 1, 2016 · Neurology

Device Facts

Record IDK162176
Device NameFiagon Navigation System
ApplicantFiagon GmbH
Product CodePGW · Neurology
Decision DateDec 1, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 882.4560
Device ClassClass 2

Indications for Use

The Fiagon Navigation System is intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. The Fiagon Navigation System is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy. Example procedures include, but are not limited to: ENT Procedures: Transphenoidal access procedures. Intranasal procedures. Sinus procedures, such as Maxillary antrostomies, Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies. ENT related anterior skull base procedures.

Device Story

Image-guided surgery system using electromagnetic tracking; tracks instruments with integrated sensors and patient localizers within electromagnetic field generated by field generator. Performs image-to-patient registration via fiducial markers, anatomical landmarks, or surface matching to align intraoperative position with preoperative CT/MRI scans. Displays instrument position superimposed on image data at 15-45 Hz update rate. Used in ENT clinical settings by physicians. Modified version includes WiFi connectivity and iPad remote control with automatic registration. Output assists surgeons in precise anatomical localization during procedures like sinus surgery and skull base access, potentially improving surgical accuracy and patient outcomes.

Clinical Evidence

Bench testing only. Testing included positioning accuracy for target registration, wireless coexistence testing, and software testing. Results confirmed the device functioned as intended and similar to the predicate.

Technological Characteristics

Electromagnetic tracking system; field generator; instrument-integrated sensors; patient localizers. Connectivity: WiFi. Remote control: iPad. Software: Navigation and registration algorithms. Sterilization: Not specified.

Indications for Use

Indicated for patients requiring stereotactic surgery for ENT procedures, including paranasal sinus and mastoid anatomy, where anatomical structures can be identified relative to CT or MR models. Applicable for open or percutaneous procedures.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or head in profile, composed of flowing lines. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 1, 2016 Fiagon Gmbh % Ms. Yarmela Pavlovic Regulatory Counsel Hogan Lovells US LLP 3 Embarcadero Center, Suite 1500 San Francisco, CA 94111 Re: K162176 Trade/Device Name: Fiagon Navigation System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: PGW Dated: November 1, 2016 Received: November 1, 2016 Dear Ms. Pavlovic: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Eric A. Mann -S for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K162176 Device Name Fiagon Navigation System Indications for Use (Describe) The Fiagon Navigation System is intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. The Fiagon Navigation System is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy. Example procedures include, but are not limited to: ENT Procedures: Transphenoidal access procedures. Intranasal procedures. Sinus procedures, such as Maxillary antrostomies, Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies. ENT related anterior skull base procedures. Type of Use (Select one or both, as applicable) 区 Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. ### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## SECTION 6 - 510(k) Summary ### August 3, 2016 #### 1. Submitter Information/ 510(k) Holder | Submitter: | Fiagon GmbH | |------------|----------------------------------------------------| | Address: | Neuendorfstrasse 23b<br>16761 Hennigsdorf, Germany | | | Telephone: | +49 3302 201 21 10 | |----------|--------------------|--------------------| | Telefax: | +49 3302 201 21 15 | | - Mr. Dirk Mucha, CTO Contact: #### Device Information 2. | Trade Name: | Fiagon Navigation System | |-----------------|----------------------------------------------| | Common Name: | Image guided surgery system | | Classification: | Class II per 21 CFR 882.4560 | | Device: | Ear, Nose, and Throat Stereotaxic Instrument | | Product Code: | PGW | #### Purpose of Submission 3. The purpose of this submission is to gain clearance for a modified version of the previously cleared Fiagon Navigation System (K133573). #### Predicate Device Information 4. Fiagon Navigation System (K133573) #### ട. Device Description The Fiagon Navigation System displays position instruments in preoperative scans (e.g., CT, MRI, fluoroscopy) utilizing electromagnetic tracking technology. The position of the instrument with integrated sensor and the patient equipped with localizers are localized within an electromagnetic field generated by a field generator. The principle of navigation is based on electromagnetic spatial measuring of localizer element in a generated electromagnetic field. {5}------------------------------------------------ The display of navigation information requires an image-to-patient registration procedure. During registration procedure, the navigation system determines the coordinate transformation between the intraoperative position of the patient and the position of the preoperative scan by fiducial marker, anatomical landmark or surface matching. Thereafter the spatial position of the instrument is displayed superimposed to the image data. The navigation information is updated with a rate of 15 to 45 Hz. #### 6. Intended Use The Fiagon Navigation System is intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. The Fiagon Navigation System is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy. Example procedures include, but are not limited to: ENT Procedures; Transphenoidal access procedures. Intranasal procedures. Sinus procedures, such as Maxillary antrostomies, Ethmoidectomies, Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies. ENT related anterior skull base procedures #### 7. Comparison of Technological Characteristics The device is a modified version of the previously cleared Fiagon Navigation System (K133573). The reason for this Special 510(k) is to describe the addition of WiFi connectivity to the navigation system and the addition of an iPad remote control with automatic registration feature. The changes do not raise new types of safety or effectiveness questions. Bench testing has been conducted to evaluate the modified system, and results confirm that the device performs as intended and in a similar manner compared to the predicate. #### Performance Data 8. The following nonclinical performance testing was conducted to support the substantial equivalence of the device to the predicate. {6}------------------------------------------------ - Positioning accuracy test for target registration . - Wireless coexisting testing ● - Software testing ● In all instances, the device functioned as intended and similar to the predicate, supporting the substantial equivalence to the predicate device. #### 9. Conclusion Based on the indications for use, technological characteristics, performance testing, and comparison to the predicate, the Fiagon Navigation System has been shown to be substantially equivalent to the cleared predicate and the modified device does not present any new issues of safety or effectiveness.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...