The CARTO® ENT System is intended for use during intranasal image-guided navigation procedures for patients who are eligible for sinus procedures.
Device Story
CARTO ENT System provides real-time navigation for ENT physicians during intranasal/paranasal sinus procedures. System components include navigation console, field ring, instrument hub, patient tracker, and workstation. Magnetic field generated by field ring induces current in magnetic sensor at tip of flexible navigated tool; system calculates tool tip position relative to pre-acquired CT images. Patient tracker fixed to patient forehead compensates for head movement. Physician views tool tip icon overlaid on registered CT images on workstation display. Real-time visualization aids precise location of anatomical structures, potentially improving surgical accuracy and patient outcomes. System supports manual and force-sensing registration probe modes.
Clinical Evidence
No clinical data. Evidence consists of bench testing and pre-clinical cadaver studies. Bench tests included proof of design electrical tests, location accuracy (0.55 mm ± 0.7 mm), software functional tests, and simulated use accuracy (0.63 mm ± 0.2 mm). Pre-clinical cadaver tests assessed sinuplasty workflow and qualitative clinical accuracy.
Technological Characteristics
Electromagnetic tracking system; uses CT image sets for reference. Components: Navigation console, field ring, instrument hub, patient tracker, workstation. Flexible navigated tools with embedded magnetic sensors. Registration via manual or force-sensing probe (1-10g range). Software-controlled navigation. Connectivity: Workstation-based. Sterilization: Not specified.
Indications for Use
Indicated for patients eligible for sinus procedures requiring intranasal and paranasal image-guided navigation.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized image of an eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is in all caps and is written in a simple, sans-serif font.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 24, 2016
Biosense Webster Ms. Melissa C. Schultz Manager, Regulatory Affairs 3333 Diamond Canyon Road Diamond Bar, CA 91765
Re: K161701
Trade/Device Name: CARTO ENT System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: PGW Dated: September 21, 2016 Received: September 22, 2016
Dear Ms. Schultz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"
(21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K161701
Device Name CARTO® ENT System
Indications for Use (Describe)
The CARTO® ENT System is intended for use during intranasal image-guided navigation procedures for patients who are eligible for sinus procedures.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Biosense Webster, a part of the Johnson & Johnson family of companies. The logo features the company name in a teal color, with the words "Biosense" and "Webster" stacked on top of each other. Below the name, in smaller letters, it says "PART OF THE Johnson & Johnson FAMILY OF COMPANIES", with the Johnson & Johnson part in red. Above the company name is a teal-colored abstract design that resembles a heart with intersecting lines.
## [807.92(a)(1)] Submitter Information
| Sponsor/Submitter: | Biosense Webster, Inc.<br>3333 S. Diamond Canyon Road<br>Diamond Bar, CA 91765 USA |
|--------------------|-------------------------------------------------------------------------------------------------------|
| Contact Person: | Melissa C. Schultz<br>Email: MSchult1@its.jnj.com<br>Telephone: 909-839-8787<br>Telefax: 909-839-8804 |
Date Summary Prepared: September 16, 2016
[807.92(a)(2)] Name of Device
| Device Trade | CARTO® ENT Navigation System |
|----------------|--------------------------------------------------------------------|
| Common Name: | Image Guided Surgery System |
| Device | Class II |
| Regulation | 21 CFR 882.4560 |
| Classification | Ear, Nose, and Throat Manual Surgical Instrument (21 CFR 882.4560) |
| Product Code: | PGW |
[807.92(a)(3)] Legally Marketed Devices
Predicate Devices: Fiagon Navigation System (K133573)
CARTO® 3 EP Navigation System (K133916) Reference Devices:
[807.92(a)(4)] Device Description
The CARTO® ENT System is intended to be used during intranasal and Device Description: paranasal surgical procedures to help ENT physicians to track and display the real-time location of the tip of navigated instruments relative to pre-acquired reference images, such as CT. The CARTO® ENT device enables ENT physicians to access sphenoid, frontal, and maxillary sinuses by using the system magnetic tracking technology. The system incorporates a Navigation Console, Field Ring, Instrument Hub, Patient Tracker, Workstation and accessories. A magnetic field generated by the Field Ring induces a current in the magnetic sensor embedded in the tip of the flexible navigated tool, which helps to accurately calculate the tool tip position. A CT image is imported and registered to the patient coordinates and a tool tip icon is displayed on top of the registered image, indicating the position of the tool in
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Biosense Webster, which is in teal. Below the logo, it says "PART OF THE Johnson & Johnson FAMILY OF COMPANIES" with Johnson & Johnson in red. The word "CONFIDENTIAL" is in bold black letters at the bottom right.
reference to the patient anatomy. A Patient Tracker is fixed to the patient forehead to compensate for the head movement during the navigation procedure.
| [807.92(a)(5)] Intended Use | |
|--------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use: | The CARTO® ENT System is intended as an aid for precisely locating<br>anatomical structures during intranasal and paranasal image-guided<br>navigation procedures. |
| Difference in<br>Indications from<br>Predicate Device: | The difference in indications for use between the subject and predicate<br>devices is supported is presented in Table 1 of this summary. |
| [807.92(a)(6)] Technical Characteristics | |
| Technological<br>Characteristics: | The substantial equivalence of the CARTO® ENT Navigation System to<br>the predicates is shown by similarity in intended use, indications for use,<br>and performance. |
| | Like the predicate Fiagon System, the CARTO® ENT Navigation<br>System is an image-guided navigation system intended for use during<br>paranasal surgical procedures. ENT physicians can track and display the<br>real-time location of the tip of navigated instruments relative to pre-<br>acquired CT images. |
| | Like the predicate, the CARTO® ENT System utilizes electromagnetic<br>tracking technology for navigation, uses anatomical reference points on<br>the patient's anatomy for intraoperative registration to the image-based<br>model of the anatomy, and uses CT image sets as reference images for<br>the image-based model. |
{5}------------------------------------------------
Biosense Webster, Inc. 3333 Diamond Canyon Road Diamond Bar, CA 91765
Image /page/5/Picture/1 description: The image shows the logo for Biosense Webster, a company that is part of the Johnson & Johnson family of companies. The logo features the company name in a teal color, with the words "Biosense" and "Webster" on the same line. Above the company name is a teal-colored graphic that resembles a heart with lines going through it. Below the company name is the text "PART OF THE Johnson & Johnson FAMILY OF COMPANIES".
## Table 1: Comparison of Predicate and Subject Device
| Attribute | Fiagon Navigation System | CARTO® ENT Navigation System |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Predicate | Subject Device |
| Intended Use | Intended as an aid for precisely locating<br>anatomical structures in either open or<br>percutaneous neurosurgical procedures. | Intended for use during intranasal and paranasal<br>image-guided navigation procedures |
| Fundamental Scientific Technology | Electromagnetic tracking | Same |
| Control Mechanism | Software controlled | Same |
| Navigation method | Electromagnetic location in reference to<br>registered CT/MR background of patient head | Same, CT only |
| Registration method | Fiducial matching and surface matching | Same |
| Bench test location accuracy | 0.9 mm<br>(Standard deviation 0.34 mm) | 0.55 mm<br>(Standard deviation 0.7 mm) |
| Simulated Use Accuracy | 1.79 mm<br>(Standard deviation 0.4 mm) | 0.63 mm<br>(Standard deviation 0.2 mm) |
| Location update rate | 15 to 45 Hz | 10 Hz |
| Main components | 1. Navigation unit<br>2. Head rest with field generator<br>3. Patient reference localizer<br>4. Navigation instrument | 1. Same (Navigation console)<br>2. Same (Field Ring)<br>3. Same (Patient Tracker)<br>4. Same (Acclarent NavWire, not included with<br>CARTO® ENT System) |
| Supported navigation instruments | Flexible tip instruments with magnetic sensor<br>on instrument tip | Same |
| Registration tools | Registration pointer, manual acquisition of<br>anatomic points and surfaces | Registration probe, manual or force sensing<br>acquisition of anatomic points and surfaces |
| Registration Probe Sensing | No | Force measurement range 1 to 10 gr.<br>Automatic acquisition of registration points or<br>surfaces. |
The primary difference between the Fiagon Navigation System and CARTO® ENT Navigation System is that on top of manual registration supported by the Fiagon system, where the physician manually selects points and surfaces to be registered, the CARTO® ENT System also supports a force sensing registration mode, where points and surfaces are automatic acquired when adequate pressure is applied with the registration probe on the patient face. Since the physician is instructed to verify the quality of the registration, this addition does not raise new issues of safety and effectiveness.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Biosense Webster, a part of the Johnson & Johnson family of companies. The logo features the company name in a teal color, with the words "Biosense" and "Webster" stacked on top of each other. Above the company name is a teal-colored graphic design that resembles a heart with lines going through it. Below the company name, in smaller font, it says "PART OF THE Johnson & Johnson FAMILY OF COMPANIES".
The CARTO® ENT System was tested to ensure that it functions Non-Clinical in accordance with the system design specifications related to Performance Data: substantial equivalence in terms of device safety and effectiveness. The following nonclinical tests were performed: 1. Proof of Design electrical tests, to verify all hardware modules perform within specifications. 2. Location Accuracy tests, where the CARTO® ENT System electromagnetic locations were compared to the locations provided by a, very accurate robot system over the entire navigation volume, to verify the system precision claim. 3. Software functional tests, covering the complete system functionality, and including error handling, usability and time performance (latency). Safety and EMC tests were performed by a nationally 4. recognized testing laboratory to verify compliance with safety and EMC standards for medical devices. 5. Simulated use accuracy test, in which a complete CT image registration and instrument navigation workflow was performed, to verify the overall accuracy of the system. 6. Pre-clinical (cadaver) tests were designed to mimic surgical procedures using the CARTO® ENT System in a simulated clinical environment, to assess the execution of a complete sinuplasty procedure workflow and to qualitatively estimate the system clinical accuracy. The results of the above tests support the safety of the device and demonstrate that the CARTO® ENT System perform as intended in the specified use conditions. The tests also demonstrate that the device performs comparably to the predicate device for the same intended use. [807.92(b) (2)] Determination of Substantial Equivalence Clinical data was not necessary for the CARTO® ENT System. Clinical Performance Data: The performance data demonstrated that the device performs as intended.
## [807.92(b) (1)] Determination of Substantial Equivalence
[807.92(b) (3)] Conclusion
{7}------------------------------------------------
Biosense Webster, Inc. 3333 Diamond Canyon Road Diamond Bar, CA 91765
Image /page/7/Picture/1 description: The image shows the Biosense Webster logo, which is a medical device company. The logo is in blue and features the company name and a graphic of a heart with lines around it. Below the logo, it says "PART OF THE Johnson & Johnson FAMILY OF COMPANIES" and the word "CONFIDENTIAL" is at the bottom right.
Conclusion from Non-Clinical and Clinical Tests
Based on the indications for use, technological characteristics, performance testing, and comparison to the predicate device, the CARTO® ENT Navigation System has been shown to be substantially equivalent to the predicate device identified in this submission, and does not present new issues of safety or effectiveness.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.