← Product Code [PGW](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/PGW) · K161491

# Scopis Hybrid Navigation System EM (K161491)

_Scopis GmbH · PGW · Feb 9, 2017 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94neurological-surgical-devices/PGW/K161491

## Device Facts

- **Applicant:** Scopis GmbH
- **Product Code:** [PGW](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/PGW.md)
- **Decision Date:** Feb 9, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The Scopis Hybrid Navigation System EM is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy. Example procedures include, but are not limited to the following ENT procedures: - Transsphenoidal access procedures; - Intranasal procedures; - Sinus procedures, such as Maxillary antrostomies, Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies; - ENT related anterior skull base procedures.

## Device Story

Image-guided surgery system using electromagnetic (EM) tracking to display instrument position on preoperative CT/MR patient models. System components: navigation unit, EM field generator, patient tracker, navigated instruments with integrated localizers, and Nova Basic navigation software. Used in ENT surgery; operated by surgeons. Registration of patient anatomy to image set achieved via fiducial markers, anatomical landmarks, or surface matching. EM field generator tracks instrument tips relative to patient; real-time position overlaid on imaging. Provides surgeons with visual guidance during procedures; assists in navigating complex anatomy; potentially improves surgical precision.

## Clinical Evidence

Bench testing only. Navigation accuracy measured at 0.49±0.27 mm, compared to predicate accuracy of 0.9±0.34 mm.

## Technological Characteristics

Electromagnetic tracking system; uses CT/MR image sets; registration via fiducials, landmarks, or surface matching. Components include EM field generator, patient tracker, and navigated instruments with integrated localizers. Software: Nova Basic. Standalone navigation unit.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Fiagon Navigation System ([K133573](/device/K133573.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle are three stylized human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 8, 2017

Scopis Gmbh % Mr. Roger N. White President Phiama, Inc. 236 McKinley Park Lane Louisville, CO 80027

Re: K161491

Trade/Device Name: Scopis Hybrid Navigation System EM Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: PGW Dated: December 28, 2016 Received: December 30, 2016

Dear Mr. White:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Eric A. Mann -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K161491

Device Name Scopis Hybrid Navigation System EM

#### Indications for Use (Describe)

The Scopis Hybrid Navigation System EM is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy.

Example procedures include, but are not limited to the following ENT procedures:

- · Transsphenoidal access procedures;
- Intranasal procedures;

· Sinus procedures, such as Maxillary antrostomies, Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies;

· ENT related anterior skull base procedures.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for Scopus medical. The word "SCOPUS" is written in large, red, sans-serif letters. Below "SCOPUS", the word "medical" is written in smaller, gray, sans-serif letters. The logo is simple and modern.

#### SECTION 5 510(k) Summary

# 510(k) Summary Scopis Hybrid Navigation System EM

#### 1. Submitter Information

| Submitter: | Scopis GmbH                                      |
|------------|--------------------------------------------------|
| Address:   | Heinrich Heine Platz 10, D-10179 Berlin, GERMANY |

Telephone: +49 (30) 201 69 38 0

Telefax: +49 (30) 201 69 38 20

| Contact: | Dr. Christopher Özbek   |
|----------|-------------------------|
|          | Chief Technical Officer |

Date Prepared: May 27, 2016

#### 2. Device Information

| Trade Name:          | Scopis Hybrid Navigation System EM           |
|----------------------|----------------------------------------------|
| Common Name:         | Image Guided Surgery System                  |
| Classification:      | Class II per 21 CFR 882.4560                 |
| Classification Name: | Ear, Nose, and Throat Stereotaxic Instrument |
| Product Code:        | PGW                                          |

#### 3. Purpose of Submission

The purpose of this submission is to gain clearance for a new Image Guided Surgery System.

### 4. Predicate Device Information

The Scopis Hybrid Navigation System EM described in this submission is substantially equivalent to the following predicate:

| Predicate Device         | Manufacturer | 510(k) No.: |
|--------------------------|--------------|-------------|
| Fiagon Navigation System | Fiagon GmbH  | K133573     |

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image shows the logo for Scopus medical. The word "SCOPUS" is written in large, bold, red letters. Below the word "SCOPUS" is the word "medical" written in smaller, gray letters. The logo is simple and clean, and the colors are eye-catching.

#### 5. Device Description

The Scopis Hybrid Navigation System EM displays the position of navigated instruments on a model of the patient's anatomy based on preoperative images (CT or MRI) using electromagnetic tracking technology. The position of the instruments and the patient are localized within an electromagnetic field produced by a field generator. The navigation of instruments relative to the patient's anatomy is established via registration of the patient's anatomy to the image set via fiducial markers, anatomical landmarks, or surface matching. The position of navigated instruments is then displayed on the model from the image set.

The Scopis Hybrid Navigation System EM consists of:

- 1. Navigation Unit
- 2. Electromagnetic field generator
- 3. Patient tracker with integrated localizers
- 4. Navigation instruments with integrated localizers
- 5. Navigation software (Nova Basic)

#### 6. Intended Use

The Scopis Hybrid Navigation System EM is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy.

Example procedures include, but are not limited to the following ENT procedures:

- . Transsphenoidal access procedures;
- . Intranasal procedures:
- . Sinus procedures, such as Maxillary antrostomies, Ethmoidectomies, Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies;
- . ENT related anterior skull base procedures.

#### 7. Comparison of Technological Characteristics

The substantial equivalence of the Scopis Hybrid Navigation System EM to the predicate is shown by similarity in intended use, indications for use, materials, and performance. Both the Scopis Hybrid Navigation System EM and the predicate utilize:

- Electromagnetic tracking technology for navigation.
- Fiducial or anatomical reference points for procedure registration to the imagebased model of the patient's anatomy.
- Tracking of the navigation instruments via localizers mounted in the tip of the instruments.

{5}------------------------------------------------

Image /page/5/Picture/2 description: The image shows the logo for Scopus medical. The word "SCOPUS" is written in large, red, sans-serif letters. Below the word "SCOPUS", the word "medical" is written in smaller, gray, sans-serif letters. The logo is simple and modern.

- Use of CT or MR image sets as reference images for the image-based model of the patient's anatomy.
#### 8. Performance Data

Bench testing of the navigation accuracy was performed to establish the substantial equivalence to the predicate device. The measured bench navigation accuracy of the Scopis Hybrid Navigation System was 0.49±0.27 mm, compared to the reported predicate device navigation accuracy of 0.9±0.34 mm.

#### 9. Conclusion

Based on the indications for use, technological characteristics, performance testing, and comparison to the predicates, the Scopis Hybrid Navigation System EM has been shown to be substantially equivalent to the predicate devices identified in this submission, and does not present any new issues of safety or effectiveness.

---

**Source:** [https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94neurological-surgical-devices/PGW/K161491](https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94neurological-surgical-devices/PGW/K161491)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
