EM ENT Instruments are indicated for use in navigated ENT procedures to locate points on a CT-based or MR-based digital model and to remove fluids, semi-fluid substances, tissue, and bone dust. EM ENT Instruments are indicated for use with Medtronic computer-assisted surgery systems. The EM ENT instruments are indicated for use in ENT procedures such as: CSF leak repairs related to ENT procedure Intranasal procedures Orbital decompression procedures Polyposis procedures Endoscopic dacryocystorhinostomy Sinus procedures, such as maxillary antrostomies Ethmoidectomies Sphenoidotomies/sphenoid explorations Turbinate resections and frontal sinusotomies The EM ENT instruments can also be used in the ENT surgical approach for the following procedures: Pituitary tumor removal Skull base procedures Transsphenoidal procedures Optic nerve decompression procedures Encephalocele procedures
Device Story
EM ENT Navigated Suctions are handheld, reusable surgical instruments used with Medtronic computer-assisted surgery systems. A mobile emitter generates a low-energy magnetic field to track the position of an EM ENT Instrument Tracker attached to the suction device. The system displays the instrument tip's location within pre-operative CT or MR image planes and anatomical renderings. Used by surgeons in ENT procedures to navigate anatomy and remove fluids, tissue, and bone dust. Provides real-time visual feedback to the surgeon, aiding in precise localization during complex procedures like skull base or sinus surgery; benefits include improved surgical accuracy and navigation.
Clinical Evidence
Bench testing only. Testing included navigational accuracy (1.54 mm error at 95% confidence in controlled environment; 1.73 mm error in simulated surgical environment), suction functionality, design specification conformance, and electrical safety. No clinical data provided.
Technological Characteristics
Materials: Stainless steel, titanium, and AES Santoprene® 8281-90 (patient-contacting cable). Sensing: Electromagnetic tracking via proximal tracker. Connectivity: Integrates with Medtronic computer-assisted surgery systems. Form factor: Handheld suction with various fixed shaft configurations (straight, 45°, 70°, 90°). Reusable instruments; trackers supplied sterile single-use.
Indications for Use
Indicated for patients undergoing navigated ENT procedures, including CSF leak repair, intranasal, orbital decompression, polyposis, endoscopic dacryocystorhinostomy, sinus procedures (maxillary antrostomies, ethmoidectomies, sphenoidotomies/explorations, turbinate resections, frontal sinusotomies), pituitary tumor removal, skull base, transsphenoidal, optic nerve decompression, and encephalocele procedures.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three abstract human profiles facing right, connected by flowing lines. The symbol is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 7. 2016
Medtronic Navigation, Inc. Carey Brenner Regulatory Affairs Specialist 826 Coal Creek Circle Louisville, CO 80027
Re: K153555
Trade/Device Name: EM ENT Navigated Suctions Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: PGW Dated: March 7, 2016 Received: March 8, 2016
Dear Carey Brenner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Denise L. Hampton -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
## 510(k) Number (if known) K153555
## Device Name
ENT EM Fusion and Supplemental Navigated Instruments
## Indications for Use (Describe)
EM ENT Instruments are indicated for use in navigated ENT procedures to locate points on a CT-based or MR-based digital model and to remove fluids, semi-fluid substances, tissue, and bone dust. EM ENT Instruments are indicated for use with Medtronic computer-assisted surgery systems. The EM ENT instruments are indicated for use in ENT procedures such as:
·CSF leak repairs related to ENT procedure
- •Intranasal procedures
- ·Orbital decompression procedures
- •Polyposis procedures
- ·Endoscopic dacryocystorhinostomy
- ·Sinus procedures, such as maxillary antrostomies
- · Ethmoidectomies
- ·Sphenoidotomies/sphenoid explorations
- ·Turbinate resections and frontal sinusotomies
The EM ENT instruments can also be used in the ENT surgical approach for the following procedures:
•Pituitary tumor removal
·Skull base procedures
·Transsphenoidal procedures
·Optic nerve decompression procedures
•Encephalocele procedures
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
| |
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
Electromagnetic (EM) ENT Navigated Suctions 510(k) Summary December 11, 2015
- I. Company: Medtronic Navigation, Inc. 826 Coal Creek Circle Louisville, Colorado 80027 USA Telephone Number: 720-890-3200
- Carey Brenner Contact: Regulatory Affairs Specialist
- II. Proprietary Trade Name:
EM ENT Navigated Suctions
- III. Classification Name:
Stereotaxic Instrument
- IV. Classification:
21 CFR 882.4560 Stereotaxic Instrument
- V. Product Codes:
PGW
- VI. Product Description
The Electromagnetic Ear Nose & Throat (EM ENT) Navigated Suctions are electromagnetically tracked, handheld, re-usable instruments intended to be used with EM ENT Instrument Trackers and ENT software on a Medtronic computer-assisted surgery system. The EM ENT Instrument Tracker is attached to the EM ENT Navigated Suction to track its position. The navigation system's mobile emitter generates a lowenergy magnetic field to locate the tracker mounted on the instrument. Then, the software displays the location of the instrument's tip within multiple patient image planes and other anatomical renderings.
{4}------------------------------------------------
#### VII. Indications for Use
EM ENT Instruments are indicated for use in navigated ENT procedures to locate points on patient anatomy relative to a CT-based or MR-based digital model and to remove fluids, semi-fluid substances, tissue, and bone dust. EM ENT Instruments are indicated for use with Medtronic computer-assisted surgery systems. The ENT instruments are indicated for use in ENT procedures such as:
- . CSF leak repairs related to ENT procedure
- Intranasal procedures ●
- Orbital decompression procedures
- Polyposis procedures
- Endoscopic dacryocystorhinostomy
- Sinus procedures, such as maxillary antrostomies ●
- Ethmoidectomies ●
- Sphenoidotomies/sphenoid explorations
- Turbinate resections and frontal sinusotomies
The EM ENT instruments can also be used in the ENT surgical approach for the following procedures:
- Pituitary tumor removal
- Skull base procedures
- Transsphenoidal procedures
- Optic nerve decompression procedures .
- . Encephalocele procedures
### VIII. Summary of the Technological Characteristics
| | Subject Device:<br>EM ENT Navigated Suction Devices | | Predicate Devices: |
|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | EM ENT Navigated Suctions<br>and Trackers<br>EM ENT Instruments are<br>indicated for use in navigated<br>ENT procedures to locate points<br>on patient anatomy relative to a<br>CT-based or MR-based digital<br>model and to remove fluids,<br>semi-fluid substances, tissue,<br>and bone dust. EM ENT<br>Instruments are indicated for use<br>with Medtronic computer-<br>assisted surgery systems. The<br>EM ENT instruments are<br>indicated for use in ENT<br>procedures such as:<br>•CSF leak repairs related to ENT | The EM ENT instruments can<br>also be used in the ENT surgical<br>approach for the following<br>procedures:<br>• Pituitary tumor removal<br>• Skull base procedures<br>• Transsphenoidal procedures<br>• Optic nerve decompression<br>procedures<br>• Encephalocele procedures | Medtronic Manual Suctions (non-<br>navigated) Class I, 510k Exempt<br>Same intended use<br>StealthStation® System GoldenEye™<br>Micro-Magnetic Tracking System<br>Option<br>Same Indications for use<br>StealthStation® System Update<br>Same Indications for use |
| | procedure | | |
| | •Intranasal procedures | | |
| | •Orbital decompression<br>procedures | | |
| | •Polyposis procedures | | |
| | •Endoscopic<br>dacryocystorhinostomy | | |
| | •Sinus procedures, such as<br>maxillary antrostomies | | |
| | •Ethmoidectomies | | |
| | •Sphenoidotomies/sphenoid<br>explorations | | |
| | •Turbinate resections and frontal<br>sinusotomies | | |
| | Localization<br>Technology | Electromagnetic (proximal tracker attached) | |
| | | | |
| | | | |
| | | | |
| | | StealthStation® System Update<br>Same | |
| | | | |
| System or<br>Instrument<br>Accuracy<br>Requirements | EM ENT Navigated Suctions | StealthStation® System GoldenEye™<br>Micro-Magnetic Tracking System | |
| | Within a standard controlled environment: navigated peg errors of<br>1.54 mm at 95% confidence and 99% reliability | Option<br>Equivalent | |
| | Within a simulated surgical environment: navigated peg errors of<br>1.73 mm at 5% confidence and 99% reliability | StealthStation® System Update<br>Equivalent | |
| | | | |
| Suction<br>Functionality | Yes | Medtronic Manual Suctions (non-<br>navigated) Class I, 510k Exempt<br>Substantially Equivalent | |
| Instrument Tip<br>Configurations | Standard (straight) | Medtronic Manual Suctions (non- | |
| | Angle<br>Olive<br>Ball/Angle | navigated) Class I, 510k Exempt<br>Substantially Equivalent | |
| Materials | EM ENT Navigated Suctions | Medtronic Manual Suctions (non- | |
| | Stainless Steel, titanium | navigated) Class I, 510k Exempt<br>Same Material as Suction | |
| | Patient and Instrument Trackers | | |
| | Patient contacting cable: AES Santoprene® 8281-90 material with<br>Colorant Pantone 301C | StealthStation® System Update<br>Same Material as Tracker Cable | |
| Instrument<br>Shaft<br>Configurations | Fixed- Straight, Small Straight, 45° frontal, 70° curve, 90° curve, 90°<br>frontal | Medtronic Manual Suctions (non-<br>navigated) Class I, 510k Exempt<br>Substantially equivalent | |
| | | | |
| Supplied as<br>"Reusable Use" | EM ENT Navigated Suctions | Medtronic Manual Suctions (non- | |
| | Yes | navigated) Class I, 510k Exempt<br>reusable | |
| | Patient and Instrument Trackers<br>Supplied Sterile Single Use | | |
{5}------------------------------------------------
{6}------------------------------------------------
### IX. Identification of Legally Marketing Devices (Predicate Devices)
| Description<br>(Substantial Equivalence Characteristic) | 510(k) Number | Clearance<br>Date |
|--------------------------------------------------------------------------------------------------------|---------------|---------------------|
| Manual Surgical Instruments (non-navigated), Class I<br>(product code: LRC Regulation: 21CFR 874.4420) | Exempt | Not<br>Applicable a |
| StealthStation® GoldenEye™ Micro-Magnetic Tracking<br>System Option | K001284 | 06/12/2000 |
| StealthStation® System Update | K050438 | 06/02/2005 |
a The Class I Medtronic ENT suction devices were released to market on June 6, 1997. They were recently sold to another company in 2014.
### X. Discussion of the Performance Testing
Testing was completed to ensure the functionality and compatibility with the identified Medtronic products. Specifically, testing with the StealthStation® System and EM ENT Suction instruments was conducted to ensure acceptable navigational accuracy, suction capability, conformance to design specifications and electrical safety testing. Test samples were subjected to simulated real-life use conditions including repeated functional testing.
### XI. Conclusions
The EM ENT Navigated Suction Instruments have been shown through comparative and bench testing to be substantially equivalent to the identified predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.